K161199 · Endymed Medical, Ltd. · PBX · Aug 1, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161199
Device Name
EndyMed Contour Handpiece
Applicant
Endymed Medical, Ltd.
Product Code
PBX · General, Plastic Surgery
Decision Date
Aug 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage: - Relief of minor muscle aches and pain, relief of muscle spasm - Temporary improvement of local blood circulation - Temporary reduction in the appearance of cellulite.
Device Story
EndyMed Contour Handpiece is an accessory for the EndyMed Imagine TC Skin Treatment System. It utilizes bipolar RF energy and vacuum-assisted mechanical massage. The vacuum module provides suction for deep tissue massage, promoting sub-dermal and adipose tissue shrinkage and improving electrode-to-tissue contact. The device is operated by a clinician in a clinical setting. The system console manages power supply, RF generation, and suction. The output is thermal heating of tissue combined with mechanical massage. The device aims to alleviate muscle-related discomfort, improve circulation, and reduce cellulite appearance. Clinical benefit is achieved through the combination of RF-induced thermal effects and vacuum-based massage.
Clinical Evidence
No clinical data. Safety and efficacy were established through bench performance testing, design validation, and software verification. Testing confirmed power output up to 85W, tissue temperature maintenance between 40°C and 45°C, and vacuum pressure up to 500mBar.
Technological Characteristics
System includes console (power supply, RF generator, suction module, CPU, display) and applicators (cable, connector, vacuum pump). Employs bipolar RF energy and vacuum mechanical manipulation. Complies with IEC 60601-1 (safety), IEC/EN 60601-1-2 (EMC), and IEC 60601-2-2 (high-frequency surgical equipment).
Indications for Use
Indicated for patients requiring relief of minor muscle aches, pain, and muscle spasms; temporary improvement of local blood circulation; and temporary reduction in the appearance of cellulite.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2016
Endymed Medical Ltd. % Mr. Yoram Levy QSite 31 Haavoda St. Binyamina, Israel 30500 Israel
Re: K161199
Trade/Device Name: Endymed Contour Handpiece Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX Dated: Julv 6, 2016 Received: July 12, 2016
Dear Mr. Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161199
Device Name EndyMed Contour Handpiece
Indications for Use (Describe)
The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour applicator for delivery of non-thermal RF combined with massage:
- · Relief of minor muscle aches and pain, relief of muscle spasm
- · Temporary improvement of local blood circulation
- · Temporary reduction in the appearance of cellulite.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for EndyMed, a company specializing in 3DEEP technology. The logo features a grid of dots on the left, followed by the company name "EndyMed" in a stylized font. Below the name, the tagline "THE 3DEEP COMPANY" is displayed in smaller letters. The overall design is clean and professional, reflecting the company's focus on advanced technology.
#### K161199 510(K) SUMMARY
## EndyMed Contour Handpiece 510(k) Number K161199
- Applicant's Name: EndyMed Medical Ltd. 12 Leshem Street, North Industrial Park, Caesarea, 3088900 Israel Tel: (972)4-630-9100 Fax: (972)4-630-9101
- Contact Person: Yoram Levy, Qsite 31 Haavoda Street Binyamina, Israel 30500 Tel (972)4-638-8837; Fax (972)4-638-0510 Yoram@qsitemed.com
Trade Name: EndyMed Contour Handpiece
Preparation Date: April 14. 2016
Classification: Name: Electrosurgical cutting and coagulation device and accessories Product Code: PBX Regulation No: 21 CFR 878.4400 Regulation Description: Electrosurgical cutting and coagulation device and accessories Class: II
- Classification Panel: General & Plastic Surgery
#### Device Description:
EndyMed's EndyMed Contour Handpiece is a treatment handpiece to be attached to the FDA cleared EndyMed Imagine TC Skin Treatment System (Imagine K08346). The EndyMed Contour Handpiece combines suction module that creates vacuum energy and emits bipolar RF energy that flows between the electrodes to create thermal heating of the tissue.
The Vacuum is used for the massaging of deep tissues by creating mild to deep suction. The vacuum massage contributes to the sub-dermal and adipose tissues shrinkage and improves the contact surface between electrodes and tissue.
#### Intended Use Statement:
The EndyMed Contour Handpiece is intended for the treatment of the following medical conditions; using the Contour
EndyMed Contour Applicator 510k Notification
Page 1 of 3
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for EndyMed, a company that specializes in 3DEEP technology. The logo features a grid of blue dots on the left, followed by the company name "EndyMed" in a stylized blue font. Below the company name is the tagline "THE 3DEEP COMPANY" in a smaller, lighter font.
K161199
applicator for delivery of non-thermal RF combined with massage:
- Relief of minor muscle aches and pain, relief of muscle spasm
- Temporary improvement of local blood circulation
- . Temporary reduction in the appearance of cellulite.
Predicate Devices: Substantial equivalence to the following predicate devices is claimed:
| Device Name | 510(k) No. | Date of Clearance |
|------------------|------------|-------------------|
| Venus Legacy CX | K143554 | August 4, 2015 |
| Viora V10 System | K150035 | May 1, 2015 |
- Venus Legacy CX (K143554) - in respect to intended use and technological characteristics;
- Viora V10 System (K150035) in respect to intended ● use and technological characteristics;
#### Reference Devices:
- EndyMed Medical Ltd. Imagine (K083461) - in respect to technological characteristics;
#### Performance Standards:
EndyMed Contour Handpiece complies with the following standards:
- IEC 60601-1:2005 / EN 60601-1:2006 Medical electrical equipment - part 1 General requirements for basic safety and essential performance, 3td edition
- · IEC/EN 60601-1-2:2014 Medical Electrical Equipment -Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard; Electromagnetic Compatibility - Requirements and Tests 4th edition
- · IEC 60601-2-2:2009 Medical Electrical Equipment Part 2-2: Particular Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories
#### Summary of Technologies:
EndyMed Contour Handpiece implements RF energy and vacuum mechanical manipulation to achieve its intended use. The components of the EndyMed Pro, includes the console (with power supply, RF generator, suction module, CPU and display panel) and the applicators (with cable, connector to console, and vacuum pump).
EndyMed Contour Applicator 510k Notification
Page 2 of 3
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for EndyMed. The logo consists of a grid of dots on the left, followed by the word "EndyMed" in a stylized font. Below the word "EndyMed" is the text "THE 3DEEP COMPANY" in a smaller font. The color of the logo is a muted blue.
K161199
#### Performance Data:
The safety and efficacy of the EndyMed Contour Handpiece were established by a series of performance tests. Lab performance tests, design validation and software verification and validation.
The EndyMed Contour Handpiece was validated for its power increase to 85W while maintaining a temperature between 40℃ and 45℃ and vacuum pressure of up to 500mBar.
Verification and testing have shown that the EndyMed Contour Handpiece device performs according to its specifications.
#### Summary of Clinical performance data:
safety and efficacy of the EndyMed Contour The -Handpiece was evaluated by performance testing.
The EndyMed Contour Handpiece has the same intended use, clinical indication and technology and no clinical studies were necessary to show substantial equivalency with its predicate devices.
#### Substantial Equivalence:
EndyMed Contour Handpiece device has the same intended use, clinical indication and basic technology as its predicate devices.
The envelope of power, frequency and vacuum parameters of the submitted EndyMed Contour Handpiece is covered by the envelopes of its predicate devices.
The EndyMed Contour Handpiece, the Venus Legacy CX device and the K150035 Viora V10 System are based on the same core technology of RF heating and vacuum massaging for the same indications for use. The design and components in the subject device including the console and the applicators are the same. The minor technological differences do not alter device core technology or performance. The performance of the EndyMed Contour Handpiece is substantially equivalent to that of the Venus Legacy CX device and the K150035 Viora
V10 System, as demonstrated in testing. Any minor differences in the human interface and accessories design do not raise any new types of safety and effectiveness issues, as verified by performance testing. Therefore the EndyMed Contour Handpiece is substantially equivalent to its predicate devices.
#### Conclusions:
The EndyMed Contour Handpiece is based on the same intended use and indications for use, technological characteristics, and principles of operation. As a result the EndyMed Contour Handpiece device is substantially equivalent to the Venus Legacy CX and the K150035 Viora V10 System predicate devices
> EndyMed Contour Applicator 510k Notification
Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.