Latera Absorbable Nasal Implant

K161191 · Spirox, Inc. · NHB · Jun 23, 2016 · Ear, Nose, Throat

Device Facts

Record IDK161191
Device NameLatera Absorbable Nasal Implant
ApplicantSpirox, Inc.
Product CodeNHB · Ear, Nose, Throat
Decision DateJun 23, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 874.3620
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Spirox Latera Absorbable Nasal Implant is indicated for supporting nasal upper and lower lateral cartilage.

Device Story

Latera Absorbable Nasal Implant system supports nasal upper and lower lateral cartilage; consists of absorbable PLLA-PDLA copolymer implant (1mm diameter, 24mm length) and stainless steel delivery cannula. Implant features forked distal end for anchoring and flexible proximal end. Used by physicians in clinical settings to provide minimally invasive nasal wall support. Delivery device includes silicone-coated cannula and positioning guide for pre-procedure planning. Implant provides structural support to nasal cartilage, potentially improving nasal airflow. System is single-use and sterile.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including verification testing, sterilization validation, packaging/shelf-life testing, transit testing, biocompatibility testing, and human factors/usability testing.

Technological Characteristics

Implant: PLLA-PDLA copolymer. Delivery device: 16-gauge medical grade stainless steel cannula with silicone coating. Dimensions: 1mm diameter, 24mm length. Sterilization: Gamma irradiation (implant), e-beam irradiation (delivery device). Single-use, sterile, non-electronic.

Indications for Use

Indicated for supporting nasal upper and lower lateral cartilage in patients requiring nasal wall support.

Regulatory Classification

Identification

Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a human profile. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 23, 2016 Spirox, Inc. Mr. Mike Rosenthal Chief Operating Officer 3475-O Edison Wav Menlo Park, CA 94025 Re: K161191 Trade/Device Name: Latera Absorbable Nasal Implant Regulation Number: 21 CFR 874.3620 Regulation Name: Ear, Nose, And Throat Synthetic Polymer Material Regulatory Class: Class II Product Code: NHB Dated: May 27, 2016 Received: May 31, 2016 Dear Mr. Rosenthal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Eric A. Mann -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161191 Device Name Latera Absorbable Nasal Implant Indications for Use (Describe) The Spirox Latera Absorbable Nasal Implant is indicated for supporting nasal upper and lower lateral cartilage. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K161191: Special 510(k) Summay #### I: SUBMITTER INFORMATION | Submitter: | Spirox, Inc.<br>3475-0 Edison Way<br>Menlo Park, CA 94025 | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Mike Rosenthal<br>Chief Operating Officer, Spirox, Inc.<br>Phone: 650.503.3329<br>Fax: 650.618.1440<br>Email: mrosenthal@spiroxmed.com | | Date Summary Prepared: | 26 April 2016 | #### II: SUBJECT DEVICE INFORMATION | Device Trade Name: | Latera Absorbable Nasal Implant | |----------------------|----------------------------------------------------------------------------| | Common Name: | Ear, nose and throat synthetic polymer material | | Classification Name: | Polymer, Ear, Nose and Throat, Synthetic, Absorbable<br>(21 CFR §874.3620) | | Product Code: | NHB | #### III: PREDICATE DEVICE INFORMATION Predicate Device: INEX Absorbable Nasal Implant (K152958) No recalls, market withdrawals or safety alerts were identified in FDA's database for the above referenced predicate device. No reference devices were used in this submission. #### IV: DEVICE DESCRIPTION The Spirox Latera Absorbable Nasal Implant System is intended to support cartilage in the nasal lateral wall. The System consists of the Latera Absorbable Nasal Implant) and Accessory Delivery Device (Delivery Device). The Implant is composed of a PLLA-PDLA copolymer that is predominantly cylindrical in shape with an approximate diameter of 1mm and overall length of 24mm. The distal end of the Implant is forked to facilitate anchoring during implantation and the proximal end is narrower for increased flexibility. The disposable Delivery Device is comprised of a non-patient contacting handle assembly and a medical grade stainless steel 16 gauge delivery cannula. The Delivery Device enables placement of the Implant in a minimally invasive manner. The Latera Absorbable Nasal Implant and Accessory Delivery Device are provided sterile and are intended for single-use only. {4}------------------------------------------------ ### V. INDICATIONS FOR USE The Spirox Latera Absorbable Nasal Implant is indicated for supporting nasal upper and lower lateral cartilage. # VI. MODIFICATIONS OF SUBJECT DEVICE The INEX Absorbable Nasal Implant, which includes the INEX Absorbable Nasal Implant) and Accessory Delivery Tool (Delivery Tool), was cleared on December 4, 2015 (K152958). The cleared device has been modified to include changes to device packaging and enhance usability aspects of the system. Importantly, the modified device, the Latera Absorbable Nasal Implant and Accessory Delivery Device, has the same fundamental scientific technology and intended use/indications for use as the cleared predicate device. The following characteristics remain unchanged between the modified and predicate device: - . No modifications are being made to the Implant device itself; the Implant retains the same dimensional and material attributes previously described under K152958. - . No changes are being made to the sterilization methods; the Implant is sterilized via gamma irradiation and the Delivery Device is sterilized via e-beam irradiation pursuant to the applicable ISO standards previously described under K152958. - . No modifications are being made to the shelf-life; both the Implant and the Delivery Device will retain the same shelf-life previously described under K152598. The proposed modifications to the subject device include: - The addition of a silicone coating to the cannula portion of the Delivery Device. ● - The addition of a positioning guide "Implant Positioning Guide" to facilitate pre-procedure planning including external visualization of the internal target implant site. - . The inclusion of more prominent Implant orientation features on the Delivery Device handle distal end. - . Changes to the Implant packaging to accommodate an implant tray to facilitate implant handling and further protect the implant during shipping. - . Changes to the Delivery Device packaging to accommodate modifications to handle shape, inclusion of the Implant Positioning Guide. # VII. DESIGN CONTROL ACTIVITIES The risk analysis method used to assess the impact of the modifications was a Failure Mode and Effects Analysis (FMEA). The verification and validations tests that were performed as a result of this risk analysis assessment included: 1) Verification Testing; 2) Sterilization Validation; 3) Packaging and Shelf-Life Testing; 4) Transit Testing; 5) Biocompatibility Testing; and 6) Human Factors/Usability Testing. The methods used for the verification and validation tests for the modified device are the same as those submitted in the original 510(k) application for the predicate device. # VIII CONCLUSION Based on the same intended use / indications for use and fundamental scientific technology, the modified Latera Absorbable Nasal Implant Device has a safety and effectiveness profile that is substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...