K161185 · Viax Dental Lab · GEX · Sep 26, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161185
Device Name
LUCERNA Dental Laser
Applicant
Viax Dental Lab
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The VIAX Dental Lab LUCERNA VDL980-1 is intended for use in dental intraoral soft tissue, general, oral maxilla-facial and cosmetic surgery. It is intended for ablating, incising, excising, vaporizing and coagulation of soft tissues using a fiber optic delivery system. The following are the indications for which the device will be marketed: - Excision and incision biopsies - Hemostatic assistance - Treatment of apthous ulcers - Frenectomy - Frenotomy - Gingival incision and excision - Gingivectomy - Gingivoplasty - Incising and draining abscesses - Operculectomy - Oral papillectomy - Removal of fibromas - Soft tissue crown lengthening - Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) - Tissue retraction for impression - Photo initiation of gingival barriers and dams - Laser-assisted bleaching/whitening of teeth
Device Story
Portable diode laser system for dental soft tissue surgery and photo-initiation. Inputs: user-selected modes via LCD touch screen (procedural, curing, whitening). Transformation: microprocessor controls solid-state diode laser to emit 980nm (procedural) or 450nm (curing/whitening) energy; fiber optic delivery system directs beam to target tissue. Output: laser energy for ablation, incision, vaporization, coagulation, or photo-activation. Used in dental clinics by trained clinicians. Features include PIN-based access, adjustable power settings, and optional wireless footswitch. Benefits: precise soft tissue management and efficient dental material curing/whitening. Healthcare providers use visual status indicators on the display to monitor and adjust parameters during procedures.
Clinical Evidence
Bench testing only. Comparative performance testing against predicate devices measured laser output accuracy in continuous wave and pulse modes. Subject device showed 1.4% variance (CW) and 0.5% variance (Pulse), outperforming predicate variance of 2.2% (CW) and 2.7% (Pulse). All results met IEC 60601-2-22 requirements (±20% tolerance). Compliance with electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and photobiological safety (IEC 60825-1/IEC 60601-2-22) standards confirmed.
Technological Characteristics
Portable diode laser; 980nm (procedural) and 450nm (curing/whitening) wavelengths. Power: 0.1-3.0W (CW), 0.1-5.0W (Pulse). Convection cooled. Fiber optic delivery (200µm tips). User interface: LCD touch screen. Power: 12VDC (AC/DC supply or 7.2VDC Li-ion battery). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22, ISO 14971. Software validation per IEC 61508.
Indications for Use
Indicated for dental patients requiring intraoral soft tissue, general, oral maxillofacial, or cosmetic surgery, including biopsies, frenectomy, gingivectomy, abscess drainage, fibroma removal, crown lengthening, sulcular debridement, tissue retraction, gingival barrier initiation, and laser-assisted teeth whitening.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged in a way that suggests movement or progress. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 26, 2016
Viax Dental Lab Mr. Edwin First Regulatory Affairs Consultant Calle Potrerillos, Flexipark Bodegas D2 Alajuela, San Rafael Cost Rica 20108
Re: K161185 Trade/Device Name: VIAX Dental Lab LUCERNA VDL 980-1 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: August 24, 2016 Received: August 24, 2016
Dear Mr. First:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K161185
#### Device Name
VIAX Dental Lab LUCERNA VDL980-1
#### Indications for Use (Describe)
The VIAX Dental Lab LUCERNA VDL980-1 is intended for use in dental intraoral soft tissue, general, oral maxillafacial and cosmetic surgery. It is intended for ablating, incising, vaporizing and coagulation of soft tissues using a fiber optic delivery system. The following are the indications for which the device will be marketed:
- . Excision and incision biopsies
- . Hemostatic assistance
- .Treatment of apthous ulcers
- . Frenectomy
- . Frenotomy
- . Gingival incision and excision
- . Gingivectomy
- . Gingivoplasty
- . Incising and draining abscesses
- . Operculectomy
- . Oral papillectomy
- . Removal of fibromas
- . Soft tissue crown lengthening
- . Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
- . Tissue retraction for impression
- . Photo initiation of gingival barriers and dams
- . Laser-assisted bleaching/whitening of teeth
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness
## VIAX Dental Lab VIAX DENTAL LAB LUCERNA VDL980-1
Submitted for: Viax Dental Lab,
| Phone: | 949-633-3554 |
|-----------------|----------------|
| Facsimile: | (888) 557-7207 |
| Contact Person: | Edwin First |
| Date Prepared: | 20 April 2016 |
Device Proprietary Name(s): VIAX DENTAL LAB LUCERNA VDL980-1 Dental Laser Common or Usual Name: General Dental Use Soft tissue and Curing Laser Product Classification: Powered Laser Surgical Instrument Product Code: GEX Predicate Device(s): (1) Biolase EPIC soft tissue Laser (2) ILT Systems ACL 5500 Argon Curing Laser(3) Denmat Sapphire ST Portable
## Rationale for Substantial Equivalence
Both the subject and predicate devices share similar intended uses and indications for use, technical characteristics, features, and specifications. The Laser characteristics of the VIAX Dental Lab LUCERNA VDL980-1. including working wavelengths and outputs, Light delivery methods, safety features, and performance specifications are similar to those of the cleared Biolase Epic, ILT ACL5500, and Denmat Sapphire ST Portable Dental Lasers. The VIAX Dental Lab LUCERNA VDL980-1 operating system and controls of the subject device are similar to those used by the previouslycleared predicate devices that have proven safety and effectiveness records in the treatment of the claimed indications. Safety and performance test results have been shown to satisfy applicable international standards recognized by the Agency.
## Intended Uses and Indications for Use
The VIAX Dental Lab LUCERNA VDL980-1 is intended for use in dental intraoral soft tissue, general, oral maxilla-facial and cosmetic surgery. It is intended for ablating, incising, excising, vaporizing and coagulation of soft tissues using a fiber optic delivery system. The following are the indications for which the device will be marketed:
- Excision and incision biopsies
- Hemostatic assistance ●
- Treatment of apthous ulcers ●
- Frenectomy ●
{4}------------------------------------------------
- Frenotomy ●
- Gingival incision and excision ●
- Gingivectomy
- Gingivoplasty ●
- Incising and draining abscesses
- Operculectomy
- Oral papillectomy
- . Removal of fibromas
- Soft tissue crown lengthening ●
- Sulcular debridement (removal of diseased or inflamed soft tissue in the ● periodontal pocket)
- Tissue retraction for impression
- Photo initiation of gingival barriers and dams
- Laser-assisted bleaching/whitening of teeth ●
## Device Description
The VIAX Dental Lab LUCERNA VDL980-1 Portable Diode Laser is comprised of three basic parts: The control box ("control module") with microprocessor and user interface for selection of modes: (a) Procedural (continuous wave or pulsed) 980mm laser output (b) Curing (dental photo initiated materials) at 450mm (+/-10mm) with timed outputs of 1.3, and 5 second(s) of output duration; and (c) Whitening procedures at 450nm (+/-10nm) timed to provide 15 minute durational output(s); A touch screen LCD display to provide visual indications of power settings and of the unit"s status, and accept inputs via the touch screen to select modes and operational parameters and adjust / set and features the units power ON/OFF functions, Application of user PIN number for operation and initialization, Selection of Language (English, Spanish, Portugese) and System settings (PIN/Password setting); the Second component is the laser handpiece that houses the laser working beam and aiming beam fiber optic component(s) and the disposable fiberoptic tip assemblies (Procedural, Curing, and Whitening). Two optional footswitches are available that can be used to activate the laser working beam: (a) Corded that plugs into the back of the system unit and (optional) wireless footswitch. Electrical power is taken from the AC/DC 12 vdc Power Supply that plugs into a standard electrical outlet and charges the L-ion 7.2VDC battery at 24W, a 7,2 VDC lithium-ion battery that generates 4 amps used to drive the diode laser(s) and system overhead.
## Conformity to International Standards
The VIAX Dental Lab LUCERNA VDL980-1 Portable Diode Laser complies with the performance requirements listed in 21 CFR 1040.10 and 1040.11, with permissible deviations pursuant to Laser Notice 50, dated July 26, 2001. Additionally, the subject device has been shown to conform to the same international electrical safety standards for electrical medical devices in general, and lasers in particular, as the predicate devices: IEC 60601-1, IEC 60601-2-2, IEC 60825-1, and IEC 60601-2-22.
## Comparative Performance Data
{5}------------------------------------------------
The VIAX Dental Lab LUCERNA VDL980-1 Portable Diode Laser has been tested side-by-side against one of the predicate devices. Measurements of the output of the subject device's working beam ranging from 0.1 to 3.0W output in Continuous Wave and Pulse modes were shown to vary from the VIAX Dental Lab LUCERNA VDL980-1 settings by an average of only 1.4% in CW and 0.5% in P(ulse) compared to the predicate's variance of 2.2% in CW and 2.7% in P(ulse). The intended performance of these devices, based on IEC 60601-2-22, is that laser output should vary from the device"s setting by less than ± 20% of the setting. Both the subject and predicate devices have been shown to satisfy this standard.
## Comparison of Features and Characteristics
Table 1, following, lists key Features and Characteristics of the subject and three predicate devices.
#### Table 1.
| | VIAX Dental Lab<br>VIAX Dental Lab<br>LUCERNA<br>VDL980-1 | Biolase<br>EPIC Portable<br>Dental Laser | ILT<br>ACL5500<br>Dental Curing<br>Laser | Denmat Holdings<br>Sapphire ST<br>Portable Laser |
|-----------------------|----------------------------------------------------------------------------------------------------|-------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------------------------------|
| Application | Dental Laser<br>(Multi Function) | Dental Laser<br>(Multi Function) | Curing Laser | Dental Laser |
| Wavelength | $980nm \pm 10nm$ | $980 \pm 5 nm$ | $450 \pm 20 nm$ | $808 \pm 5 nm$ |
| Power | 0.1 - 3.0 W (CW)<br>0.1 - 5.0 W (Pulse) | 0.1 - 3.0 W (CW)<br>0.1 - 5.0 W (Pulse) | 500mW | 0.1 - 3.0 W (CW)<br>0.1 - 5.0 W (Pulse) |
| Aiming<br>Beam | 640 nm<br>(±10nm),<br>maximum<br>2mW<br>(adjustable) | 640 nm (±10 nm),<br>maximum 2mW<br>(adjustable) | N/A | 640 nm (±10 nm),<br>maximum 2mW<br>(adjustable) |
| Cooling<br>System | Convection cooled | Convection cooled | FAN forced air | Convection Cooled |
| Pulse<br>Control | Digital<br>emission<br>Contr<br>ol | Digital emission<br>control | N/A | Digital emission<br>control |
| Laser<br>Source | Solid-state diode | Solid-state diode | Argon Ion | Solid State Diode |
| Power<br>Requirements | 24W 12VDC<br>supplied from 110 -<br>120 VAC @ 60 Hz<br>or 220 - 240 VAC<br>@ 50 Hz<br>(switchable) | | 100-240VAC @ 50-<br>60 Hz, (switchable) | 24W 12VDC<br>supplied from 110 -<br>120 VAC @ 60 Hz or<br>220 - 240 VAC<br>@ 50 Hz<br>(switchable) |
| User<br>Interface | LCD Touch Screen<br>Interface | LCD Touch Screen<br>Interface | Discreet switches | Membrane touch<br>pad,LED indicators |
{6}------------------------------------------------
| Fiberoptic<br>Tip | Disposable, 200 µm unit<br>dose | Disposable, 200µm<br>unit dose | Optic/mirror via<br>400um fiber<br>optic | Disposable, 400 µm unit dose |
|-------------------|---------------------------------|--------------------------------|------------------------------------------|-------------------------------|
| Sterility | Disposable, Barrier<br>sleeve | Disposable, Barrier<br>sleeve | Disposable, Barrier<br>sleeve | Disposable, Barrier<br>sleeve |
| 510(k)<br>Number | Pending this<br>application | K140120 | K930210 | K103667 |
#### Conclusion
The subject device shares the same principle of operation as the three predicate devices. All are diode lasers that emit radiant energy at approximately 808nm to 980mm and 450nm with outputs that range from 0.1 to 5.0W. All deliver collimated laser energy to subject target tissue via 200 to 400 um fiberoptic tips controlled by trained, experienced clinicians. All share the same indications for use in dental intraoral soft tissue general, oral maxilla-facial and cosmetic surgery. All have photo initiated dental teeth whitening, and one other has dental materials curing. All have been found to satisfy international safety standards relating to electrical medical devices in general and medical lasers in particular. All have similar safety labeling, device interlocks, and associated safety features. Both the subject and a predicate devices output were measured and compared to their settings to determine the accuracy of the devices" controls. Both met international standards pertaining to accuracy of output of the working beam, but the difference between the subject devices output and its setting was much less than the predicate's, demonstrating not only conformance to the standard, but also superior control over laser emissions. Performance testing has been accomplished for the Viax Dental Lab Lucerna VDL980-1that compares it to the predicate devices and that it meets international standards pertinent to validating its performance and compliance. The Viax Dental Lab Lucerna VDL980-1 has been tested and is in compliance with IEC60601-1 (Medical Electrical Equipment Safety); IEC60601-1-2 (EMC): IEC60825-1/IEC60601-2-22 (Photobiological Safety); Software validation in compliance with IEC61508; and ISO 14971 (Risk Management).
## The VIAX Dental Lab LUCERNA VDL980-1
Viax Dental Lab Lucerna VDL980-1 device shares intended uses, principle of operation, technical attributes, functional capabilities, and performance characteristics with the listed predicate devices. Both the subject and predicate devices have been shown to comply with applicable Federal and international safety and performance standards. The Sapphire ST Portable Laser is substantially equivalent to the listed predicate laser surgical devices and does not raise any issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.