K161177 · Anthogyr Sas · DZE · Jun 15, 2017 · Dental
Device Facts
Record ID
K161177
Device Name
Axiom PX
Applicant
Anthogyr Sas
Product Code
DZE · Dental
Decision Date
Jun 15, 2017
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Anthogyr dental implants are intended for use as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. Anthogyr dental systems are indicated for one-stage or two-stage surgery. It is up to the practitioner to decide whether immediate or delayed loading is most appropriate, based on clinical factors like good primary stability and appropriate occlusal loading.
Device Story
Axiom® PX is an endosseous dental implant system designed for replacement of missing teeth; functions as artificial root structure. Implants are surgically placed into upper or lower jaw; used for stabilization of removable prostheses or fixation of single tooth/partial dentures. System includes titanium implants (Ti6Al4V-ELI) with BCP® surface treatment and DLC-coated abutment screws. Operated by dental practitioners in clinical settings. Device provides mechanical foundation for prosthetic restorations; aids in functional and aesthetic integration. Performance validated via static/fatigue testing (ISO 14801), biocompatibility (ISO 10993), and sterilization validation (ISO 11137/17665).
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes static and fatigue biomechanical testing (ISO 14801), biocompatibility testing (ISO 10993 series: cytotoxicity, intracutaneous, sensitization, chemical characterization), surface analysis (SEM), and shelf-life validation (ASTM F1980, F1929-15, F88-15). Sterilization validated per ISO 11137 and ISO 17665.
Indicated for patients requiring replacement of missing teeth via artificial root structures; supports stabilization of removable prostheses or fixation of single tooth restorations/partial dentures; suitable for one-stage or two-stage surgery; immediate or delayed loading determined by practitioner based on clinical factors.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a family of three people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 15, 2017
ANTHOGYR SAS Therese Candau Regulatory Affairs Engineer 2237 Avenue Andre Lasquin Sallanches, 74700 FRANCE
Re: K161177
Trade/Device Name: Axiom® PX Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: June 12, 2017 Received: June 13, 2017
Dear Therese Candau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mary S. Runner -A
Lori A. Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161177
Device Name Axiom® PX
Indications for Use (Describe)
Anthogyr dental implants are intended for use as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures. Anthogyr dental systems are indicated for one-stage or two-stage surgery. It is up to the practitioner to decide whether immediate or delayed loading is most appropriate, based on clinical factors like good primary stability and appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 5 510(K) SUMMARY
Section 5 – page 1 / 7 510(k) summary
{4}------------------------------------------------
# 510(k) SUMMARY
#### SUBMITTER'S CONTACT INFORMATION 5.1
# ANTHOGYR SAS
2237, avenue André Lasquin SALLANCHES, FRANCE 74700 Registration Number: 8020776 Phone number: (33) (0)4 50 58 02 37 Fax number: (33) (0)4 50 93 78 60 Contact person: Thérèse CANDAU (m.candau@anthogyr.com) Preparation date: 06/14/2017
#### 5.2 DEVICE
Trade name: AXIOM® PX Classification name: Endosseous Dental Implant Class II Product Code: DZE / NHA CFR Section: 21CFR 872.3640 Device panel: DENTAL 510(k) Number: K161177
#### 5.3 PREDICATE DEVICE
# Primary Predicate Device
K131066, AXIOM® REG manufactured by Anthogyr for implants
# Reference Predicate Device
K101913, Anthogyr Dental Implant System: Axiom manufactured by Anthogyr
K992538, Amorphous diamond coated screw manufactured by NobelBiocare for attachment screws
{5}------------------------------------------------
The AXIOM® PX implant system has been designed in order to enhance the functional and aesthetic integration of implant supported restorations.
The file concerns the implants and abutment screws. The prosthetic components and surgical instrumentation are the same as for the AXIOM® REG (see 510(k) K101913 and K131066) and are compatible with Axiom® PX implants.
Implants:
Replacement of a missing root for placement of a dental restoration. The implant is screwed in the upper or lower jaw.
Material: Ti6AI4V-ELI
Surface treatment: BCP®
Dimensions, Ø3.4, length 8, 10, 12, 14, 16, 18 mm Ø 4.0, length 8, 10, 12, 14, 16, 18 mm Ø4.6, length 6.5, 8, 10, 12, 14 mm
Ø5.2, length 6.5, 8, 10, 12 mm
Screws:
Screwing the abutment into the implant and the secondary parts into the abutment. Material: Ti6AI4V-ELI
Surface treatment: DLC
Dimensions: M1.4, length 5, 7.9 mm
#### INDICATIONS FOR USE 5.5
Anthogyr dental implants are intended for use as artificial root structures for replacement of missing teeth. They can be used for stabilization of removable prostheses or fixation of single tooth restorations or partial dentures.
Anthogyr dental systems are indicated for one-stage or two-stage surgery. It is up to the practitioner to decide whether immediate or delayed loading is most appropriate, based on clinical factors like good primary stability and appropriate occlusal loading.
{6}------------------------------------------------
#### PERFORMANCE TESTING 5.6
ANTHOGYR AXIOM® PX conforms to Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments -Guidance for Industry and FDA Staff Document issued on May 12, 2004.
# Sterilization validation
Gamma irradiation sterilization was conducted on a representative worst case implant in order to support sterility of the implants, in accordance with ISO 11137 series standards. Sterilization was performed on the predicate device (K131066).
Both implants and DLC screws are sold sterile.
Validation of the recommended cleaning and sterilization parameters for end-user sterilized products was conducted according to ISO 17665-1 and ISO 17665-2 to validate DLC screws sterilization when sold non sterile.
# Shelf-life validation
The shelf-life has been validated in accordance with ASTM F1980 under accelerated and real-time ageing and supports that the product remains sterile throughout the shelf life. Dye penetration test (ASTM F1929-15), peeling testing, seal strength testing (ASTM F88-15) as well as sterility test were performed.
# Biomechanical testing
Static and fatigue testing has been conducted on a representative worst case scenario in air in accordance with ISO 14801 (2007) and Special Controls Guidance Document : Root-form Endosseous Dental Implant and Endosseous Dental Abutments - Guidance for Industry and FDA Staff Document issued on May 12, 2004. The fatigue limit has been compared to the predicate Axiom REG implants fatique limit (K131066). Surface area analysis, insertion torque testing, and pullout strength testing has been included in this submission to demonstrate substantial equivalence.
# Biocompatibility testing
Biocompatibility assessment has been performed according to ISO 10993 series standards for implants and abutment screws.
Implants:
Chemical characterization (ISO 10993-18:2005 – to determine the type and levels of extractable chemicals present).
{7}------------------------------------------------
Cytotoxicity test (ISO 10993-5:2009 - the cytotoxicity test is used to determine the biological response of mammalian cells in vitro by exposing mouse fibroblast cells after extraction).
# Abutment screws:
Chemical characterization (ISO 10993-18:2005)
Cytotoxicity test (ISO 10993-5: 2009)
Intracutaneous study (ISO 10993-10:2010 - Intradermal injection of test extracts in the rabbit to evaluate the potential of the material to produce irritation).
Sensitization study (ISO 10993-10:2010 - to assess the contact sensitization potential of the device in the albino guinea pig after a polar and non-polar extraction).
# Surface analysis
SEM for the implants as required by FDA guidance class II special controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments.
#### TECHNOLOGICAL CHARACTERISTICS 5.7
# 5.7.1 Implants
| | Candidate<br>Axiom PX | Predicate<br>Axiom REG | Comparison |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Picture | Image: Axiom PX | Image: Axiom REG | The thread of the PX<br>implant is different from<br>the REG implant in order<br>to easier the post-<br>extractional placement<br>and low-density bone |
| Material | Titanium: Ti6Al4V-ELI | Titanium: Ti6Al4V-ELI | Same material for both<br>implants |
| Indications<br>for use | Anthogyr dental implants<br>are intended for use as<br>artificial root structures<br>for replacement of missing<br>teeth. They can be used<br>for stabilization of<br>removable prostheses or<br>fixation of single tooth<br>restorations or partial<br>dentures.<br>Anthogyr dental systems<br>are indicated for one-<br>stage or two-stage | Anthogyr dental implants<br>are intended for use as<br>artificial root structures for<br>replacement of missing<br>teeth. They can be used for<br>stabilization of removable<br>prostheses or fixation of<br>single tooth restorations or<br>partial dentures.<br>The Axiom dental system is<br>indicated for one-stage or<br>two-stage surgery.<br>It is up to the practitioner | The indications for use<br>are the same for both<br>implants. The indications<br>for Axiom REG include<br>both implants and<br>prosthetic parts<br>indications.<br>The Axiom PX indications<br>include only the implants<br>indications as the<br>prosthetic parts are not<br>presented here. |
{8}------------------------------------------------
| | surgery.<br>It is up to the practitioner to decide whether immediate or delayed loading is most appropriate, based on clinical factors like good primary stability and appropriate occlusal loading. | to decide whether immediate or delayed loading is most appropriate, based on clinical factors like good primary stability and appropriate occlusal loading.<br>Prefabricated components are intended for use as accessories to dental implants to support implant-supported restorations. | |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| Compatible<br>abutments | Axiom REG Aesthetic<br>abutments<br>Axiom REG Standard<br>abutments<br>Axiom REG conical<br>abutments<br>Axiom REG Healing screws | Axiom REG Aesthetic<br>abutments<br>Axiom REG Standard<br>abutments<br>Axiom REG conical<br>abutments<br>Axiom REG Healing screws | Same compatible<br>abutments |
| Dimensions | Ø3.4 mm;<br>Length 8-18 mm<br>Ø4.0 mm;<br>Length 8-18 mm<br>Ø4.6 mm;<br>Length 6.5-14 mm<br>Ø5.2 mm;<br>Length 6.5-12 mm | Ø3.4 mm;<br>Length: 8-18 mm<br>Ø4.0 mm;<br>Length: 6.5-18 mm<br>Ø4.6 mm;<br>Length: 6.5-14 mm<br>Ø5.2 mm;<br>Length: 6.5-14 mm | The PX implants<br>dimensions available are<br>included in the range of<br>REG implants |
| Surgical<br>approach | One stage surgery<br>Two stages surgery<br>Immediate loading | One stage surgery<br>Two stages surgery<br>Immediate loading | The surgical approach is<br>the same |
| Surface<br>treatment | sand blasting BCP +<br>passivation | sand blasting BCP +<br>passivation | Same treatment surface<br>for both implants |
| Sterilization<br>method | Gamma radiation | Gamma radiation | Same sterilization<br>process for both<br>implants |
# 5.7.2 Abutment screws
| | Candidate<br>Axiom PX screw | Predicate<br>NobelBiocare | Comparison |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Titanium: Ti6Al4V-ELI | Titanium: Ti6Al4V-ELI | Both made from same<br>material |
| Indications<br>for use | Anthogyr dental implants<br>are intended for use as<br>artificial root structures for<br>replacement of missing<br>teeth. They can be used for<br>stabilization of removable<br>prostheses or fixation of<br>single tooth restorations or<br>partial dentures.<br>Anthogyr dental systems<br>are indicated for one-stage | The Amorphous Diamond<br>Coated Screw is used to<br>retain prosthetic<br>components to dental<br>implants or to other<br>prosthetic components. The<br>amorphous diamond<br>coating will add a greater<br>pre-load to the screw,<br>which in turn help prevent<br>the screw and prosthetic | The indication for the<br>predicate is specific to<br>the Diamond Coated<br>Screw, while the<br>indication for the<br>subject device<br>describes the entire<br>implant system,<br>including the Axiom PX<br>screw. |
{9}------------------------------------------------
| | or two-stage surgery.<br>It is up to the practitioner<br>to decide whether<br>immediate or delayed<br>loading is most appropriate,<br>based on clinical factors like<br>good primary stability and<br>appropriate occlusal<br>loading. | components from<br>loosening. | |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| Compatible<br>abutments | Axiom REG Aesthetic<br>abutments<br>Axiom REG Standard<br>abutments<br>Axiom REG conical<br>abutments | Pre-fabricated titanium<br>abutments compatible with<br>NobelActive implants | Each system has its<br>own compatible range |
| Connection<br>type | Internal conical connection | Internal conical connection | Both internal conical<br>connection |
| Surface<br>treatment | Diamond Like Carbon | Diamond Like Carbon | Both DLC surface<br>treatment |
| Sterilization<br>method | Delivered not sterile or<br>sterile with abutment | Delivered not sterile | Anthogyr screw can be<br>delivered sterile. The<br>sterilization method<br>has been validated |
# 5.8 CONCLUSION
ANTHOGYR AXIOM® PX is substantially equivalent to its predicate devices in terms of intended use, material, design, mechanical properties and function. Non clinical performance testing according to special control demonstrate that ANTHOGYR AXIOM® PX is substantially equivalent to its predicate device.
The mechanical testing shows that the performance of the Axiom PX implants is substantially equivalent to the predicate device.
Based on technological characteristics included in this submission, the Axiom PX implants have been shown to be substantially equivalent to the Axiom REG implants.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.