K161156 · Health & Life Co., Ltd. · DXN · Jun 14, 2016 · Cardiovascular
Device Facts
Record ID
K161156
Device Name
Full Automatic (NIBP) Blood Pressure Monitor
Applicant
Health & Life Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jun 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
HL858TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use. HL858TA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.
Device Story
Oscillometric blood pressure monitor; measures systolic/diastolic pressure and heart rate during cuff inflation. Input: pressure signals from upper arm cuff. Processing: oscillometric algorithm; detects irregular heartbeats (rhythm variation >25% from average); calculates risk category based on readings. Output: LCD display of BP, heart rate, irregular heartbeat symbol, and risk category. Used in home environment by adult patients. Healthcare providers use displayed values for tracking BP levels. Benefits: enables self-monitoring of hypertension and heart rhythm irregularities.
Clinical Evidence
Clinical validation performed per ISO 81060-2:2013. Study included 105 subjects. Results demonstrated that mean difference and standard deviation met required limits for automated non-invasive sphygmomanometers.
Technological Characteristics
Oscillometric measurement during inflation; pressure range 0-300mmHg; pulse 40-199 bpm. Materials: ABS housing/keys. Power: 4x AAA alkaline batteries. Connectivity: None. Software: Embedded firmware. Standards: ISO 81060-2, IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 10993-1/5/10, IEC 80601-2-30, ISO 14971, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for adults aged 18+ with arm circumference 9-17 inches (23-43 cm) for home-use blood pressure and heart rate monitoring.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858DI (K153214)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The seal is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 14, 2016
Health & Life Co., Ltd. Ms. Sarah Su Director 9f, No. 186, Jian Yi Road Zhonghe District, New Taipei City, 23553 CN
Re: K161156
Trade/Device Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858TA Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 21, 2016 Received: April 25, 2016
Dear Ms. Sarah Su.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161156
#### Device Name
Full Automatic (NIBP) Blood Pressure Monitor, Model HL858TA
#### Indications for Use (Describe)
HL858TA automatically measures human's Systolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
HL858TA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K161156
# Page 1 of 6
# 510(k) SUMMARY
(As Required By 21 CFR 807.92)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is: _ K161156 __ Date: ____________________________________________________________________________________________________________________________
#### 1. Submitter:
Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301
Contact person: Sarah Su/ QA & RA Dept. E-mail: sarah.su@hlmt.com.tw Tel: 886-2-8227-1300 ext.1201 Fax: 886-2-8227-1301
#### 2. Name of the Device:
Trade Name: Full Automatic (NIBP) Blood Pressure Monitor, Model HL858TA Common Name: Blood Pressure Monitor Classification Name: Non-invasive Blood Pressure Measurement System Classification: Class II, 21 CFR 870.1130 Classification Panel: 74 Cardiovascular Product Code: DXN
{4}------------------------------------------------
### 3. Information for the 510(k) Cleared Device (Predicate Device):
Full Automatic (NIBP) Blood Pressure Monitor, Model: HL858DI (K153214)
#### 4. Device Description:
HL858TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
The device will display a symbol ( heartbeat rhythm as defined as a rhythm is more than or less than 25% from the average heartbeat intervals during the measurement. Additionally, after measurement, the Risk Category Indicator function will show the information with the readings on the screen for the user tracking their blood pressure level.
#### 5. Intended Use
HL858TA automatically measures human's Systolic, Diastolic blood pressure and heart rate by using the oscillometric method during inflation. All values can be read out in one LCD panel. Measurement position is at human being's upper arm. The intended use of this over-the-counter device is for adults aged 18 years and older with arm circumference ranging from 9 inches to 17 inches (approx.23 cm to 43 cm) and for home use.
HL858TA detects the appearance of irregular heartbeats during measurement; an indicated symbol will appear with measuring reading. And the Risk Category Indicator will show the information with the readings on the screen for the user tracking their blood pressure level.
{5}------------------------------------------------
# 6. Comparison of device to predicate device:
Product Specification Comparison Table of Subject Device HL858TA, and Predicate Device HL858DI (K153214)
| Item | Predicate Device<br>HL858DI (K153214) | Subject Device<br>HL858TA |
|-------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Method of<br>measurement | Oscillimetric | Same as left |
| Measurement<br>Type | During inflation | Same as left |
| Range of<br>measurement | Pressure 0- 300mmHg,<br>Pulse 40-199 Beats/minute | Same as left |
| Accuracy | Pressure ± 3mmHg<br>Pulse ± 5% | Same as left |
| Pressure<br>Changed Rate | 2~5mmHg/sec. | Same as left |
| Exhaust | Automatic exhaust valve | Same as left |
| Display | Liquid Crystal Digital | Same as left |
| Power Supply | 6V 1A, 4 × AA/1.5V (LR6)<br>Alkaline batteries, or<br>AC adapter (optional) | 6V 1A, 4 × AAA/1.5V (LR03)<br>Alkaline batteries |
| Storage/<br>Transportation<br>Environment | - 25°C ~ + 70°C (- 13°F ~<br>+158°F),<br>≤ 93% R.H. | Same as left |
| Operating<br>Environment | 5℃~40℃ (41°F~104°F),<br>15%~93% R.H. | Same as left |
| Material | ABS housing and<br>ABS keys | Same as left |
| Sets of<br>memory | 2*60,total 120 | 1*60 |
| Number of | 5 keys | 2 keys |
| Push Button | | |
| Storage pouch | Yes | No |
| Cuff size | Arm circumference approx.<br>23~43cm / 9~17 inch<br>(Universal cuff) | 1. Arm circumference approx. 23~33cm / 9~13 inch (Standard cuff)<br>2. Arm circumference approx. 23~43cm / 9~17 inch (Universal cuff; optional) |
| Unit Weight | Approx. 310 ± 10g<br>(Excluding cuff and Batteries ) | Approx. 221 ± 5g<br>(Excluding cuff and Batteries ) |
| Risk Category<br>Indicator | Yes | Same as left |
| Irregular<br>Heartbeat<br>Detector | Yes | Same as left |
{6}------------------------------------------------
#### Changes from the predicate devices HL858DI (K153214):
*Changing the battery type form four AA (LR6) alkaline batteries and AC adapter to four AAA (LR03) alkaline batteries only.
- * Changing the memory of 120 (2*60, total 120) to 60 (1*60).
* Changing the 5 keys to the 2 keys.
- * Adding the Standard cuff (23~33cm / 9~13 inch).
These features have been verified and validated and do not affect the safety and effectiveness of subject device HL858TA.Please refer to Section 12. Substantial Equivalence Discussion for detail information.
#### 7. Discussion of Clinical Tests Performed:
HL858TA is compliant to the standard of ISO 81060-2: Second Edition 2013-05-01 Non-invasive sphygmomanometers- Part 2: Clinical validation of automated measurement type. The results of this clinical investigation show that the required limits for mean difference and standard deviation are fulfilled by the subject device HL858TA in the group of 105 subjects with qualified distribution. Thus, all the
{7}------------------------------------------------
relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
# 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:
The subject device was tested to evaluate its safety and effectiveness, including the followings:
- a. EMC Test: IEC 60601-1-2 Edition 3:2007-03, Medical Electrical Equipment -Part 1-2: General requirements for safety - collateral standard: Electromagnetic compatibility - Requirements and Tests
- b. Safety Test:
-IEC 60601-1:2005, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
-IEC 60601-1-11:2010, Medical electrical equipment-Part 1-11: General Requirement for basic safety and essential performance- Collateral Standard: Requirements for medical electrical systems used in the home healthcare environment
- c. FCC Test:
FCC 47 CFR Part 15, Subpart B
- d. Biocompatibility Test:
-ISO 10993-1:2009, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
-ISO 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for In Vitro cvtotoxicity
-ISO 10993-10:2010, Third Edition Biological evaluation of medical devices-Part 10: Tests for irritation and skin sensitization
- e. Reliability Test:
IEC 80601-2-30 Edition1.1 2013-07 Medical electrical equipment-Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers.
- f. Risk Assessment: ISO 14971:2007 Second Edition, Medical devices -
{8}------------------------------------------------
Application of risk management to medical devices
## g. Software Verification and Validation:
-IEC 62304 Ed.1.0 (2006), Medical device software - Software life cycle processes,
-IEC 60601-1-4 Medical electrical equipment - Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1
### h. Usability Validation:
-IEC 62366-1:2015 Medical devices - Application of usability engineering to medical devices
-IEC 60601-1-6 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
## 9. Conclusions:
The subject device was tested and fulfilled the requirements of those standards mentioned above, and it's concluded that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.