Dextera MicroCutter 5/80 Blue Reload, Dextera MicroCutter 5/80 Blue Curved Tip Reload
K161137 · Cardica, Inc. · GDW · Jul 20, 2016 · General, Plastic Surgery
Device Facts
Record ID
K161137
Device Name
Dextera MicroCutter 5/80 Blue Reload, Dextera MicroCutter 5/80 Blue Curved Tip Reload
Applicant
Cardica, Inc.
Product Code
GDW · General, Plastic Surgery
Decision Date
Jul 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The MicroCutter 5/80™ is intended for transection in multiple open or minimally invasive urologic, thoracic, and pediatric surgical procedures, as well as application for transection, resection, and/or creation of anastomoses in the small and large intestine, and the transection of the appendix.
Device Story
The Dextera MicroCutter 5/80™ is a single-patient-use surgical stapler system consisting of a stapler and reload cartridges (Blue for 1.5mm tissue compression; White for 1.0mm tissue compression). The device is operated by surgeons in open or minimally invasive surgical settings. It functions by delivering two double-staggered rows of 316L stainless steel staples while simultaneously transecting tissue between the rows. The device provides a 28mm cut length and 30mm staple line. It is used to facilitate tissue transection, resection, and anastomosis in urologic, thoracic, and pediatric procedures. The device benefits patients by enabling precise tissue management during complex surgeries.
Clinical Evidence
No clinical data. Bench testing only, including staple forming and release, cartridge loading/unloading forces, cartridge boss strength, cartridge retention, dimensional inspection, leak/burst pressure testing, staple form height, and deployment force.
Technological Characteristics
Materials: 316L stainless steel staples. Form factor: Single-patient-use stapler with reload cartridges. Blue reload (1.5mm tissue compression), White reload (1.0mm tissue compression). Staple line length: 30mm; Cut length: 28mm. Mechanical actuation.
Indications for Use
Indicated for patients undergoing open or minimally invasive urologic, thoracic, or pediatric surgical procedures requiring tissue transection, resection, or anastomosis of the small and large intestine, or appendiceal transection.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 20, 2016
Cardica, Inc. Mr. Greg Watson Vice President of Operations 900 Saginaw Drive Redwood City, California 94063
Re: K161137
Trade/Device Name: Dextera Microcutter 5/80™ Blue Reload, Dextera Microcutter 5/80™ Blue Curved Tip Reload, Dextera Microcutter 5/80TM White Reload, Dextera Microcutter 5/80TM White Curved Tip Reload Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: April 21, 2016 Received: April 22, 2016
Dear Mr. Watson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
# 510(k) Number (if known)
K161137
### Device Name
Dextera MicroCutter 5/80™ Blue Reload; Dextera MicroCutter 5/80™ Curved Tip Blue Reload; Dextera MicroCutter 5/80™ White Reload; Dextera MicroCutter 5/80™ Curved Tip White Reload
### Indications for Use (Describe)
The MicroCutter 5/80 is intended for transection in multiple open or minimally invasive urologic, thoracic, and pediatric surgical procedures, as well as application for transection, and/or creation of anastomoses in the small and large intestine, and the transection of the appendix.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - Dextera MicroCutter 5/80™
#### A. Date Prepared
April 21, 2016
#### B. 510(k) Owner
Cardica, Inc. 900 Saginaw Drive Redwood City, California 94063 Phone: 650-364-9975 Fax: 650-364-3134
#### C. Contact Person
Greg Watson Vice President, Operations (650) 331-7163 (office) (949) 412-9808 (cell) watson@cardica.com
#### D. Device Information
Trade name: Dextera MicroCutter 5/80™ Blue Reload, Dextera MicroCutter 5/80™ Blue Curved Tip Reload, Dextera MicroCutter 5/80™ White Reload and Dextera MicroCutter 5/80TM White Curved Tip Reload
Common name: Staple
Classification Name: Implantable staple
Regulation Number: 21 CFR §878.4750
Product Code: GDW
#### E. Marketed Predicate Devices
MicroCutter XCHANGE® 30 Blue Cartridge MicroCutter XCHANGE® 30 Blue Curved Tip Cartridge MicroCutter XCHANGE® 30 White Cartridge MicroCutter XCHANGE® 30 White Curved Tip Cartridge (K132581, K140118, K140170 and K151081)
#### F. Device Description
The size of the Dextera MicroCutter 5/80TM Blue Reload and Dextera MicroCutter 5/80™ Blue Curved Tip Reload staple is compatible with tissue that can be easily compressed to 1.5mm. The size of the Dextera MicroCutter 5/80™ White Reload and Dextera MicroCutter 5/80TM White Curved Tip Reload staple is compatible with tissue that can be easily compressed to 1.0mm. The cut length of the staple line is 28.0mm (1.10in) and the staple line length is 30mm long. The Dextera MicroCutter 5/80™
{4}------------------------------------------------
Reloads are deployed using the Dextera MicroCutter 5/80™ Stapler. The MicroCutter 5/80 Stapler is a single patient use stapler that delivers two, double staggered rows of 316L stainless steel staples while simultaneously transecting tissue between staple rows.
#### G. Intended Use
The MicroCutter 5/80™ is intended for transection in multiple open or minimally invasive urologic, thoracic, and pediatric surgical procedures, as well as application for transection, resection, and/or creation of anastomoses in the small and large intestine, and the transection of the appendix.
#### H. Comparison to Predicate Device
The Dextera MicroCutter 5/80™ Blue Reload, Dextera MicroCutter 5/80TM Blue Curved Tip Reload, Dextera MicroCutter 5/80™ White Reload and Dextera MicroCutter 5/80™ White Curved Tip Reload have minor modifications in comparison to the legally marketed predicate devices. The primary design changes include two minor modifications to the Reloads. These modifications were done to minimize interference and provide a better fit with other Reload components, and to help ensure the Reload is secured within the Stapler. The modifications do not alter the fundamental scientific technology and or intended use of the devices.
#### I. Non-Clinical Performance Data
Bench testing, including staple forming and release, cartridge loading force, cartridge unloading force, cartridge boss strength, cartridge retention during and outside of deployment, dimensional inspection, leak/burst pressure testing, staple form height, and deployment force, was performed.
#### J. Conclusion
Bench testing was conducted to verify the performance of the staple. The results demonstrated that the subject devices perform as intended and are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.