K161130 · Calgary Scientific, Inc. · LLZ · Aug 24, 2016 · Radiology
Device Facts
Record ID
K161130
Device Name
ResolutionMD
Applicant
Calgary Scientific, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
ResolutionMD® software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar reformatting and 3D visualization of medical image data and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports. ResolutionMD is intended for use as a diagnostic, review, and analysis tool by trained healthcare professionals to drive clinical management. When interpreted by a trained physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations. ResolutionMD is not to be used for primary mammography diagnoses.
Device Story
ResolutionMD is an enterprise medical image viewer; operates on off-the-shelf servers, web browsers, and mobile devices. Inputs: DICOM and non-DICOM medical image data and reports. Transforms inputs via 2D display, multi-planar reformatting (MPR), and 3D visualization; Vessel Analysis module performs semi-automatic lumen boundary detection and stenosis measurement. Outputs: Visualized images, analysis measurements, and reports. Used by trained healthcare professionals (radiologists/clinicians) in clinical settings for diagnostic review and analysis. Facilitates collaboration via integrated secure audio-video communication. Output aids physicians in clinical decision-making and management of vascular anatomy. Benefits include remote access to diagnostic-quality images and collaborative review capabilities.
Clinical Evidence
Bench testing only. Verification testing included >2000 functional, smoke, and regression tests. Validation testing performed by trained radiology personnel focused on usability assessment and consistency across client platforms. Compliance with ISO 14971 and IEC 62304 standards documented.
Technological Characteristics
Enterprise software viewer; supports 2D, MPR, and 3D visualization. Connectivity via web browsers and mobile devices. Compliant with ISO 14971 (risk management) and IEC 62304 (software life cycle). Vessel Analysis module uses semi-automatic algorithms for vascular measurement.
Indications for Use
Indicated for trained healthcare professionals to view, analyze, and report on medical images (DICOM/non-DICOM) for clinical management. Includes a Vessel Analysis module for semi-automatic determination of contrasted lumen boundaries, stenosis measurements, and evaluation of lumen diameters and lengths in vascular anatomy. Not for primary mammography diagnosis. Mobile use is not a replacement for full radiology workstations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in black and have a flowing, wave-like quality. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profiles.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 24, 2016
Calgary Scientific, Inc. Kyle Peterson Director, Regulatory & Corporate Affairs Suite 208, 1210- 20th Avenue SE Calgary, Alberta, T2G 1M8 CANADA
Re: K161130
Trade/Device Name: ResolutionMD Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications System Regulatory Class: Class II Product Code: LLZ Dated: July 22, 2016 Received: July 26, 2016
Dear Kyle Peterson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Jeff Rodges
For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161130
Device Name ResolutionMD
Indications for Use (Describe) Indications for Use Statement for ResolutionMD:
ResolutionMD® software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar reformatting and 3D visualization of medical image data and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports.
ResolutionMD is intended for use as a diagnostic, review, and analysis tool by trained healthcare professionals to drive clinical management. When interpreted by a trained physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.
ResolutionMD is not to be used for primary mammography diagnoses.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161130
Device Name ResolutionMD
#### Indications for Use (Describe)
Indications for Use Statement for ResolutionMD with Vessel Analysis module:
ResolutionMD® software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar reformatting and 3D visualization of medical image data and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports.
ResolutionMD incorporates a Vessel Analysis module which is used as a post-processing diagnostic review and analysis application for images viewed from ResolutionMD. It is a tool for use by trained healthcare professionals to review, edit, analyze and report findings of vascular anatomy. Clinicians can semi-automatically determine contrasted lumen boundaries and stenosis measurements, and evaluate maximum lumen diameters and length measurements.
ResolutionMD is intended for use as a diagnostic, review, and analysis tool by trained healthcare professionals to drive clinical management. When interpreted by a trained mages may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.
ResolutionMD is not to be used for primary mammography diagnoses.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☑</span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <span style="font-size: 10pt;"> <span style="font-family: Symbol;">☐</span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Calgary Scientific. The logo consists of the word "Calgary" in a stylized font, followed by a blue integral symbol. The word "Scientific" is written in a similar stylized font to the right of the integral symbol. The overall design is clean and modern.
#### 510(k) Summary 5
As required by 21 CFR Part 807.87(h)
# Submitter
Calgary Scientific Inc. Suite 208, 1210 - 20th Ave. SE, Calgary, Alberta, T2G 1M8 CANADA
Phone: (403) 767-7945 Fax: (403) 270-2771
Contact person: Kyle Peterson, Director - Regulatory & Corporate Affairs
Date of Submission: April 19, 2016
### Identification of the Device
| Device Proprietary Name: | ResolutionMD® |
|--------------------------|-------------------------------------------------------------------|
| Common Name: | ResolutionMD |
| Classification Name: | Picture Archiving and Communication System per<br>21 CFR 892.2050 |
| Product Code: | LLZ |
| Device Class: | Class II |
#### Marketed Device to which Equivalence is claimed
| Device | Manufacturer | 510(k) Number |
|---------------------|-------------------------|---------------|
| ResolutionMD Mobile | Calgary Scientific Inc. | K133508 |
| ResolutionMD Web | Calgary Scientific Inc. | K120076 |
#### Device Description
ResolutionMD® software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar reformatting and 3D visualization of medical image data and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports.
ResolutionMD incorporates a Vessel Analysis module which is used as a post-processing diagnostic review and analysis application for images viewed from ResolutionMD. It is a tool for use by trained healthcare professionals to review, edit, analyze and report findings of vascular anatomy.
# Indications for Use
Indications for Use for ResolutionMD:
ResolutionMD® software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar
{5}------------------------------------------------
# calgary /ട്cIenTIFIC
reformatting and 3D visualization of medical image data and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports.
ResolutionMD is intended for use as a diagnostic, review, and analysis tool by trained healthcare professionals to drive clinical management. When interpreted by a trained physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.
ResolutionMD is not to be used for primary mammography diagnoses.
The Indications for Use Statement is a unification of the statements of the two previously cleared devices, along with the addition of some suggested wording from the IMDRF (International Medical Device Regulators Forum) guidance on SaMD (Software as a Medical Device) Risk Categorization. This unification is reflective of the similar architecture and device use case of the two cleared products. The intended use is the same between the three products and there are no new diagnostic claims involved in the wording change.
# Indications for Use for ResolutionMD with Vessel Analysis module:
ResolutionMD® software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar reformatting and 3D visualization of medical image data and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports.
ResolutionMD incorporates a Vessel Analysis module which is used as a post-processing diagnostic review and analysis application for images viewed from ResolutionMD. It is a tool for use by trained healthcare professionals to review, edit, analyze and report findings of vascular anatomy. Clinicians can semi-automatically determine contrasted lumen boundaries and stenosis measurements, and evaluate maximum and minimum lumen diameters and length measurements.
ResolutionMD is intended for use as a diagnostic, review, and analysis tool by trained healthcare professionals to drive clinical management. When interpreted by a trained physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.
ResolutionMD is not to be used for primary mammography diagnoses.
For the Indications for Use Statement for the Vessel Analysis module, the first and third paragraph is exactly the same as the Statement for the ResolutionMD device. The second paragraph, which contains the information specific to the Vessel Analysis module, is exactly the same as the previously cleared ResolutionMD Web with Vessel Analysis module.
# Technological Characteristics
The ResolutionMD® software has the same technological characteristics as the predicate ResolutionMD devices and has the same uses and applications as the predicate devices. Both
{6}------------------------------------------------
# carga
the device and predicates are used by the clinician as a diagnostic, review, and analysis tool for radiological images.
# Software Verification and Validation Testing
Verification testing consisting of more 2000 separate tests, each executed multiple times by different testers, was performed for this device. Testing included functional, smoke and regression tests and was also complemented by beta tests performed by Calgary Scientific's distribution partners. The vast majority of tests passed our testing criteria. Any defects found or reported were either fixed or logged in the Unresolved Anomalies report included with this submission and annotated as to any impact on safety or effectiveness including applicable workarounds.
Validation testing based on typical clinical workflows was performed by trained radiology personnel. Validation includes usability assessment and consistency across all client platforms.
### Safety and Effectiveness
The device is designed and manufactured under Quality System Regulations as outlined in 21 CFR 820. All requirements of Picture Archiving and Communications System (21 CFR 892.2050) are met, and software is in compliance with ISO 14971 "Medical devices – Application of risk management to medical devices" standard and IEC 62304 "Medical device software – Software life cycle processes" standard.
# Conclusion
Based on the above considerations, Calgary Scientific Inc. believes that the ResolutionMD software is substantially equivalent to the predicate device and the predicates are post-processing and provide the same or similar essential features of visualization of radiological data on web and mobile devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.