Powder Free Nitrile Examination Gloves Flock-Lined, Black Color
K161074 · Shen Wei (Usa), Inc. · LZA · Nov 17, 2016 · General Hospital
Device Facts
Record ID
K161074
Device Name
Powder Free Nitrile Examination Gloves Flock-Lined, Black Color
Applicant
Shen Wei (Usa), Inc.
Product Code
LZA · General Hospital
Decision Date
Nov 17, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
Powder-Free Nitrile Examination Glove Flock-Lined, Black Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.
Device Story
Disposable, non-sterile, ambidextrous nitrile examination glove; black color; interior cotton flock-lined to facilitate donning; beaded cuff. Used by healthcare professionals in clinical settings to prevent cross-contamination between patient and examiner. Device acts as a physical barrier. Performance verified against ASTM D6319-10 standards for dimensions, physical properties, and freedom from pinholes. Biocompatibility confirmed via ISO 10993-10:2010 testing (primary skin irritation and dermal sensitization).
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D6319-10 (dimensions, physical properties, powder residue, freedom from pinholes) and ISO 10993-10:2010 (biocompatibility).
Technological Characteristics
Synthetic nitrile rubber; cotton flock interior lining; ambidextrous; beaded cuff; non-sterile. Complies with ASTM D6319-10 (dimensions, physical properties, powder residue) and ASTM D5151-11 (freedom from pinholes). Biocompatibility per ISO 10993-10:2010.
Indications for Use
Indicated for use as a disposable, single-use medical glove worn on the examiner's hand to prevent contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Non Sterile, Powder Free Nitrile Examination Glove, Blue with Polymer Coating (K070639)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2016
Shen Wei (USA), Inc. % Ms. Cheryl Bailey-Kroll Compliance Manager 33278 Central Ave. Suite 102 Union City, California 94587
Re: K161074
> Trade/Device Name: Powder Free Nitrile Examination Gloves Flock-Lined, Black Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: October 12, 2016 Received: October 18, 2016
Dear Cheryl Bailey-Kroll,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K161074
Device Name
Powder-Free Nitrile Examination Glove Flock-Lined, Black Color
Indications for Use (Describe)
Powder-Free Nitrile Examination Glove Flock-Lined, Black Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is labeled "Attachment One" and features a logo for "Shen Wei usa inc." The logo includes a green hand graphic above the text. Below the company name is the tagline "PROMOTING YOUR SUCCESS!" in a smaller font size.
## 510(k) Summary
#### 1.0 Submitter:
| Name: | Shen Wei USA Inc. |
|---------------|-------------------------------------------------------|
| Address: | 33278 Central Ave. Suite 102<br>Union City, CA. 94587 |
| FDA Est Reg#: | 2939388 |
| Phone No: | 510-429-8692 |
| Fax No: | 510-487-5347 |
### Manufacturer:
| Name:<br>Address: | ZHANGJIAGANG DAYU RUBBER PRODUCTS CO., LTD<br>XIZHANG TOWN<br>ZHANGJIAGANG CITY, CHINA 215614 |
|-------------------|-----------------------------------------------------------------------------------------------|
| FDA Est Reg#: | 1000254914 |
| Phone No: | 520-845-0023 |
| Fax No: | 520-845-0311 |
Date 510k Summary was prepared: October 10, 2016
#### 2.0 Contact Person:
| Name: | Cheryl Reep |
|-----------|----------------------|
| Phone No: | 510-429-8692 |
| Fax No: | 510-487-5347 |
| Email: | Creep@shenweiusa.com |
#### 3.0 Name of the device:
| Trade Name: | Powder Free Nitrile Examination Gloves Flock-Lined, Black Color |
|----------------------|-----------------------------------------------------------------|
| Device Name: | Powder Free Nitrile Examination Gloves Flock-Lined, Black Color |
| Common Name: | Patient Examination Gloves |
| Classification Name: | Patient examination glove |
| Device Class: | Class 1 |
| Regulation Number: | 21 CFR 880.6250 |
| Product Code: | 80 LZA |
#### 4.0 Predicate Device Information:
Device Name: Non Sterile, Powder Free Nitrile Examination Glove, Blue with Polymer Coating Company Name: Siam Sempermed Corp. Ltd. 510K Number: K070639
#### 5.0 Description of the Device:
Powder Free Nitrile Examination Gloves Flock-Lined, Black Color Non-Sterile, is a disposable device intended for
{4}------------------------------------------------
medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only. The gloves are lined with cotton flock, ambidextrous with beaded cuff, black colored, single use disposable devices that comes is six sizes (XS, S, M, L, and XL) supplied in non-sterile state.
The gloves are made of synthetic rubber, designed and manufactured in accordance with ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical and performance characteristics meet all requirements of ASTM D6319-10. Powder was not used for this glove's manufacturing process.
#### 6.0 Labeling and Intended Use of the Device:
The Powder Free Nitrile Examination Gloves Flock-Lined, Black Color is a disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. This device is single use only.
#### 7.0 Shelf Life or Expiration Data: Not Applicable
#### 8.0 Summary of the Technological Characteristics of the Device:
The Powder Free Nitrile Examination Gloves Flock-Lined, Black Color is summarized with the following technological characteristics compared to ASTM or equivalent standards.
| Characteristics | Standards | Device Performance |
|---------------------------------------------|----------------------------------------------------------------|--------------------------------|
| Dimensions | ASTM D6319-10 | Meets |
| Physical Properties<br>(Before/After Aging) | ASTM D6319-10 | Meets |
| Freedom from pinholes | ASTM D6319-10 / ASTM<br>D5151-11 | Meets |
| Powder Residual | ASTM D6319-10 | Meets <2mg/glove |
| Biocompatibility | ISO 10993-10:2010:<br>i. Primary Skin Irritation<br>in Rabbits | i. Not a primary skin irritant |
| | ii. Dermal Sensitization | ii. Not a contact sensitizer |
#### 9.0 Device Interior Lining and Manufacturing Process:
The Powder Free Nitrile Examination Gloves Flock-Lined, Black Color is sprayed with a white cotton flock on the interior of the glove to help facilitate donning.
#### 10.0 Substantial Equivalence Based On Assessment of Non-Clinical Performance Data
Comparison to Predicate- The subject device Powder Free Nitrile Examination Gloves Flock-Lined, Black Color compares favorably to the predicate device, Non Sterile, Powder Free Nitrile Examination Glove, Blue with Polymer Coating, as indicated in the tabulated summary below:
| Characteristics | Predicate Device | Proposed New Device | Comments |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| Name | Non Sterile, Powder Free<br>Nitrile Examination Glove,<br>Blue with Polymer Coating | Powder Free Nitrile<br>Examination Gloves Flock-<br>Lined, Black Color | N/A |
| 510K Number | K070639 | K161074 | N/A |
| Company | Siam Sempermed Corp. Ltd. | Shen Wei USA Inc. | N/A |
| Device Description/<br>Regulation Number | Patient examination glove /<br>21 CFR 880.6250 | Patient examination glove /<br>21 CFR 880.6250 | Substantially Equivalent |
| Product Code | LZA | LZA | Substantially Equivalent |
| Intended use | Intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br> | Intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. This<br>device is single use only. | Substantially Equivalent |
| Instruction for Use | A disposable device<br>intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. This<br>device is single use only. | A disposable device<br>intended for medical<br>purpose that is worn on the<br>examiner's hand to prevent<br>contamination between<br>patient and examiner. This<br>device is single use only. | Substantially Equivalent |
| Non-Sterile | Non-Sterile | Non-Sterile | Substantially Equivalent |
| Materials | Nitrile | Nitrile | Substantially Equivalent |
| Color | Blue | Black | Difference |
| Design | Ambidextrous in different<br>sizes per ASTM 6319-10-<br>dimension requirement | Ambidextrous in different<br>sizes per ASTM 6319-10-<br>dimension requirement | Substantially Equivalent |
| Lining / Coating | Polymer Coating | Flock Lining | Difference |
| Single Patient Use | Single Patient Use | Single Patient Use | Substantially Equivalent |
| Performance: | | | |
| I. Freedom from Holes | | | Substantially Equivalent |
| II. Dimension | Meets ASTM D6319-10 | Meets ASTM D6319-10 | Substantially Equivalent |
| III. Physical Properties | | | Substantially Equivalent |
| IV. Powder Residue | | | Substantially Equivalent |
| Single Use | Yes | Yes | Substantially Equivalent |
| Biocompatibility Test | Passes:<br>i. Primary Skin Irritation<br>Test ii. Dermal Sensitization<br>Test | Passes:<br>i. Not a primary skin irritant<br>ii. Not a contact sensitizer | Substantially Equivalent |
# Substantial Equivalence Comparison Table
{5}------------------------------------------------
# Summary of Difference and Comparison of Safety and Effectiveness- The subject device differs from the predicate in that:
- 1) The subject device is color Black while the predicate device is color Blue. Difference in color does not affect subject device's safety and effectiveness; subject device was tested and met requirements for Biocompatibility and ASTM standards.
- 2) The subject device is manufactured with the inner surface coated or lined with flock to facilitate donning. The predicate device is manufactured with the inner surface coated with polymer to assist in donning. Though the materials of construction differ, the subject device's material is functionally equivalent to those of the cited predicate.
- Summary of Clinical Testing: Not applicable. No clinical testing was performed. 11.
12. Conclusion: The conclusion drawn from the nonclinical test demonstrate that the subject device, Powder Free Nitrile Examination Gloves Flock-Lined, Black Color, is as safe, as effective, and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.