ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator
Applicant
Surgitools Pty , Ltd.
Product Code
HEW · Obstetrics/Gynecology
Decision Date
Oct 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1640
Device Class
Class 2
Indications for Use
The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is indicated for use by a surgeon in laparoscopic procedures where uterine manipulation and visualization of the position of the vaginal fornices for colpotomy incisions is required, and for maintaining pneumoperitoneum during vaginal vault closure.
Device Story
Device consists of CerviGrip Uterine Manipulator, ColpoWave Colpotomizer, and optional ColpoWave Balloon; used by surgeons in laparoscopic procedures. Manipulator features integrated cervical screw, sliding uterine tip, and tail screw for locking; colpotomizer is double-ended (30mm/40mm) with raised lips for distance gauging and green color for tissue contrast. Balloon accessory provides vaginal occlusion. Device is single-use, sterile, and non-latex. Surgeon operates device to manipulate uterus and visualize vaginal fornices for colpotomy incisions; balloon maintains pneumoperitoneum during vaginal vault closure. Benefits include improved visualization and maintenance of surgical field integrity during laparoscopic surgery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including mechanical performance (tensile, manipulation, balloon stability), biocompatibility (ISO 10993), and sterilization validation (ISO 11135-1).
Technological Characteristics
Materials: ABS polymer, 304 stainless steel. Components: Uterine manipulator with cervical screw, sliding tip, colpotomizer (30mm/40mm), pneumatic balloon. Single-use. Sterilization: Ethylene Oxide (EtO) per ISO 11135-1. Biocompatibility: ISO 10993-1, -10, -11. Shelf life: 3 years.
Indications for Use
Indicated for use by surgeons in laparoscopic procedures requiring uterine manipulation, visualization of vaginal fornices for colpotomy incisions, and maintenance of pneumoperitoneum during vaginal vault closure.
Regulatory Classification
Identification
A culdoscope is a device designed to permit direct viewing of the organs within the peritoneum by a telescopic system introduced into the pelvic cavity through the posterior vaginal fornix. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include trocar and cannula, instruments used through an operating channel, scope preheaters, light source and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2016
Surgitools Pty Ltd. % Elizabeth Pugh Regulatory Affairs Consultant Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, Texas 78701
Re: K161065
> Trade/Device Name: ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator Regulation Number: 21 CFR 884.1640 Regulation Name: Culdoscope and accessories Regulatory Class: Class II Product Code: HEW Dated: September 28, 2016 Received: September 28, 2016
Dear Elizabeth Pugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
For Division
Douglas Silverstein -S 2016.10.27 15:26:25 -04'00' Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161065
#### Device Name
ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator
Indications for Use (Describe)
The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is indicated for use by a surgeon in laparoscopic procedures where uterine manipulation of the position of the vaginal fornices for colpotomy incisions is required, and for maintaining pneumoperitoneum during vaginal vault closure.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K161065
# 1. Submission Sponsor
Surgitools Pty Ltd. 231 Timberlane Dr. Woodvale Western Australia 6026, Australia Phone number: 61-08-93093222 Contact: Jai Singh Title: Director: Global Sales & Marketing
# 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Rd, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Elizabeth Pugh, Consultant, RA Email: project.management@emergogroup.com
#### 3. Date Prepared
October 14, 2016
# 4. Device Identification
| Trade/Proprietary Name: | ColpoWave™ Colpotomizer and CerviGrip™ Uterine Manipulator |
|-------------------------|--------------------------------------------------------------------------------|
| Common/Usual Name: | Uterine Manipulator and Colpotomizer |
| Classification Name: | Culdoscope and accessories |
| Regulation Number: | 884.1640 |
| Product Code: | HEW, Culdoscope and accessories<br>LKF, Cannula, manipulator/injector, uterine |
| Device Class: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
# 5. Legally Marketed Predicate Device(s)
K110819, Singh Colpotomizer System, Surgitools Pty Ltd.
#### 6. Device Description
The Surgitools Pty Ltd. ColpoWave Colpotomizer and CerviGrip Uterine Manipulator consists of the ColpoWave™ Colpotomizer, CerviGrip™ Uterine Manipulator, and the ColpoWave™ Balloon
{4}------------------------------------------------
accessory. All parts are made without natural latex rubber and supplied sterile to be disposed of after use.
The single-use CerviGrip™ Uterine Manipulator is designed with an integrated cervical screw, a sliding uterine tip, tail screw that can lock the uterine tip in place, and a screw to hold the colpotomizer in place. The CerviGrip™ Uterine Manipulator is manufactured with ABS polymer and 304 Stainless Steel.
The ColpoWave™ Colpotomizer is non-fenestrated and has incorporated two different sizes due to the double-ended colpotomizer design (30mm or 40mm diameter). Cup shape with one partial circumferential raised lip at each end and a second partial raised circumferential lip 20mm from each end providing the surgeon with a known distance to help gauge pelvic distances during dissection. The ColpoWave™ Colpotomizer is manufactured with green ABS polymer to provide better visualization (greater contrast than white against tissue).
The ColpoWave™ Balloon is an optional accessory, which is inflated to fit various vaginas.
# 7. Indications for Use Statement
The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is indicated for use by a surgeon in laparoscopic procedures where uterine manipulation and visualization of the position of the vaginal fornices for colpotomy incisions is required, and for maintaining pneumoperitoneum during vaginal vault closure.
#### 8. Predicate Comparison
The following table compares the Surgitools Colpotomizer and CerviGrip Uterine Manipulator to the predicate device with respect to indications for use, technological characteristics, and materials:
| Trade Name | Singh Colpotomizer System<br>K110819 (Surgitools Pty Ltd.) | ColpoWave Colpotomizer and CerviGrip<br>Uterine Manipulator<br>K161065 (Surgitools Pty Ltd.) |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The Singh Colpotomizer System is<br>indicated for use by a surgeon in<br>laparoscopic procedures requiring<br>uterine manipulation and the<br>visualization of the position of the<br>vaginal fornices for colpotomy incisions<br>into the vaginal vault. The surgeon<br>makes the colpotomy incisions to access<br>or remove intraperitoneal tissue from<br>the pelvic cavity. | The ColpoWave Colpotomizer and<br>CerviGrip Uterine Manipulator is<br>indicated for use by a surgeon in<br>laparoscopic procedures where uterine<br>manipulation and visualization of the<br>position of the vaginal fornices for<br>colpotomy incisions is required, and for<br>maintaining pneumoperitoneum during<br>vaginal vault closure. |
{5}------------------------------------------------
| System<br>Components | Uterine manipulator with cervical<br>screw attachment; Sliding uterine tip (inner rod); Sliding and rotating funnels<br>(available in two sizes) with a lip<br>acting as a colpotomizer; O-ring and screw to hold the<br>funnel in place; and Vaginal occluder (plastic<br>vaginal plug) | Uterine manipulator with cervical<br>screw attachment Sliding uterine tip Sliding and rotating colpotomizer<br>(incorporating two sizes) ColpoLock and screw to hold the<br>colpotomizer in place; Pneumatic plug (PneumoBloc); and Vaginal occluder (ColpoWave Balloon |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Uterine Tips | Single length inner rod | Single length inner rod |
| Vaginal Plug | OD 43mm<br>Length 50mm | ColpoWave Balloon is inflated to fit<br>various vaginas |
| Design<br>Colpotomizer | Two colpotomizers of different<br>sizes. 35mm and 40mm Cup shape with partial<br>circumferential raised lip at its end. White. Non-fenestrated. | One colpotomizer incorporating two<br>different sizes. One side diameter<br>30mm and one of diameter 40mm Cup shape with one partial<br>circumferential raised lip at each end<br>and a second partial raised<br>circumferential lip 20mm from each<br>end. Green. Non-fenestrated. |
| Uterine<br>manipulator | Uterine manipulator with cervical<br>screw attachment; Sliding uterine tip (inner rod); Tail screw with hydrotubation port;<br>and O-ring and screw to hold the funnel<br>in place | Uterine manipulator with integrated<br>cervical screw and hydrotubation<br>port; Sliding uterine tip; Tail screw that can lock the uterine<br>tip in place; and ColpoLock and screw to hold<br>colpotomizer in place. |
| Materials | Stainless steel and polymer | Stainless steel and polymer |
| Single use or reusable | Reusable | Single use |
| Biocompatible materials | Yes | Yes |
The Surgitools ColpoWave Colpotomizer and CerviGrip Uterine Manipulator has the same intended use but different technological characteristics compared to the predicate device. The differences in technological characteristics do not raise different questions of safety and effectiveness.
# 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the ColpoWave Colpotomizer and CerviGrip Uterine Manipulator and in showing substantial equivalence to the predicate device, Surgitools Pty Ltd. completed a number of tests and validations. The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator meets all the requirements for overall design, sterilization, and biocompatibility confirming that the design output meets the design inputs and specifications for the device.
{6}------------------------------------------------
The ColpoWave Colpotomizer and CerviGrip Uterine Manipulator passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
- Performance Testing
The device successfully passed bench testing to evaluate mechanical and functional properties to ensure the device is robust to withstand forces that exceed the expected forces during use, including tensile tests, manipulation, and balloon strength and stability.
- Biocompatibility testing per ISO 10993-1, ISO 10993-10 and ISO 10993-10 and ISO 10993-11
The subject device is an external communicating device that contacts tissue for less than 24 hours. Cytotoxicity, Intracutaneous Reactivity, Sensitization, and Acute Systemic Toxicity Testing were performed. All tests were conducted on the finished device after manufacturing, packaging, and sterilization had been completed to reflect the actual state of the device when exposed to the patient. All patient-contacting device components passed biocompatibility testing required by ISO 10993-1 and FDA General Program Memorandum G95-1.
- . Sterilization Validation
The sterility of the Surgitools Uterine Manipulator and Colpotomizer system is assured by using a validated sterilization method qualified in accordance with ISO 11135-1. The validation results met the acceptance criteria defined in the standard and demonstrates the ability of the process to deliver SAL of 10 °. Examination of the EtO residuals has been conducted in accordance with ISO 10993-7, which has validated a minimum release period of 7 days.
- Shelf Life Testing
After the sterilization cycle was complete, the products were accelerated aged and tested to demonstrate no degradation in strength or functionality over the anticipated shelf life period of three (3) years.
#### 10. Clinical Performance Data
There was no human clinical testing required to support the substantial equivalence of the subject device.
#### 11. Conclusion
Based on the comparison and analysis above, the Surgitools ColpoWave Colpotomizer and CerviGrip Uterine Manipulator is substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.