The Volume Viewing software is a visualization package for PACS workstations. It is intended to support the medical professional in the reading, analysis and diagnosis of DICOM compliant volumetric medical datasets. The software is intended as a general purpose digital image processing tool, with optional functionality to facilitate visualization and measurement of vessel features. Other optional functionality is intended for the registration of anatomical (CT) on a second CT dataset or on functional volumetric image data (MR) to facilitate the comparison of various lesions. Volume and distance measurements are intended for evaluation and quantification of tumour measurements, and other analysis and evaluation of both hard and soft tissues. The software also supports interactive segmentation of a region of interest (ROI), has a dedicated tool set for lung lesion segmentation, quantification and follow-up of lesions selected by the user and provides tools to define and edit paths such as centerlines through structures, which may be used to analyze cross-sections of structures, or to provide flythrough visualizations rendered along such centerline. Caution: Web-browser access is available for review purposes. Images accessed through a web-browser (via a mobile device or by other means) should not be used to create a diagnosis, treatment plan, or other decision that may affect patient care.
Device Story
IMPAX Volume Viewing 4.0 is a PACS workstation software accessory for 3D medical image processing. It inputs DICOM-compliant volumetric datasets (CT, MR). The software performs multi-planar reformatting (MPR), curved planar reformatting (CPR), volume rendering, and image registration. It provides automated tools for table/bone removal, vessel/airway segmentation, and lung lesion quantification. Users (radiologists/clinicians) operate the software on PACS workstations to view, measure, and analyze anatomical structures. Output includes 2D/3D visualizations, cross-sectional views, and quantitative measurements (distance, volume). A web-browser variant allows remote review, though it is restricted from diagnostic use. The device assists clinicians in lesion management and follow-up, potentially improving diagnostic efficiency and patient monitoring.
Clinical Evidence
No clinical trials were performed. Evidence consists of bench-top verification and validation. Verification included regression testing of measurement algorithms against reference values (meeting +/- scanner resolution criteria) and crosshair position checks (meeting half-voxel criteria). Validation involved three radiologists evaluating new functionality (lung nodule segmentation, lesion management, fly-through) in a lab environment using clinical datasets; results met established acceptance criteria.
Technological Characteristics
Software-based PACS accessory. Features 3D volume rendering, MPR, CPR, MIP, MinIP, AvgIP, and multi-modality registration. Includes automated segmentation (vessels, airways, lung lesions) and lesion management tools. Connectivity via PACS network or web-browser. Developed per ISO 13485, ISO 14971, ISO 27001, ISO 62366, and IEC 62304 standards.
Indications for Use
Indicated for medical professionals to read, analyze, and diagnose DICOM-compliant volumetric medical datasets. Supports visualization/measurement of vessel features, registration of CT/MR datasets for lesion comparison, quantification of tumors, and analysis of hard/soft tissues. Includes tools for ROI segmentation, lung lesion segmentation/follow-up, and flythrough visualization of vessels/airways. Not for diagnostic use via web-browser/mobile access.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with medicine and healthcare.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 22, 2016
AGFA Healthcare N.V. % ShaeAnn Cavanagh, RAC Regulatory Affairs Manager AGFA HealthCare Corporation 10 South Academy Street GREENVILLE SC 29601
Re: K161061
Trade/Device Name: IMPAX Volume Viewing 4.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 8, 2016 Received: June 8, 2016
Dear Ms. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161061
Device Name IMPAX Volume Viewing 4.0
### Indications for Use (Describe)
The Volume Viewing software is a visualization package for PACS workstations. It is intended to support the medical professional in the reading, analysis and diagnosis of DICOM compliant volumetric medical datasets. The software is intended as a general purpose digital mage processing tool, with optional functionality to facilitate visualization and measurement of vessel features.
Other optional functionality is intended for the registration of anatomical (CT) on a second CT dataset or on functional volumetric image data (MR) to facilitate the comparison of various lesions. Volume and distance measurements are intended for evaluation and quantification of tumour measurements, and evaluation of both hard and soft tissues. The software also supports interactive segmentation of a region of interest (ROI), has a dedicated tool set for lung lesion segmentation, quantification and follow-up of lesions selected by the user and provides tools to define and edit paths such as centerlines through structures, which may be used to analyze cross-sections of structures, or to provide flythrough visualizations rendered along such centerline.
Caution: Web-browser access is available for review purposes. Images accessed through a web-browser (via a mobile device or by other means) should not be used to create a diagnosis, treatment plan, or other decision that may affect patient care.
Type of Use (Select one or both, as applicable)
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY:
### Agfa IMPAX Volume Viewing 4.0
#### SUBMITTER I.
Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Koen Cobbaert, Prepared: April 14, 2016 Telephone: +32 3 444 7539
#### II. DEVICE
Name of Device: IMPAX Volume Viewing 4.0 Common Name: System, Image Processing, Radiological Classification Name: Picture Archiving and Communications System (PACS) Regulatory Classification: Class II, 21 CFR 892.2050 Product Code: LLZ
#### III. PREDICATE DEVICES
This is a 510(k) for Agfa's IMPAX Volume Viewing 4.0, which is a picture archiving and communications system. It is substantially equivalent to Agfa's IMPAX Volume Viewing 3.0 (K133135) and TeraRecon's iNtuition 4.4.11 (K121916).
These predicates have not been subject to a design-related recall. No reference devices were used in this submission.
#### DEVICE DESCRIPTION IV.
IMPAX Volume Viewing is a general purpose medical image processing tool for the reading and analysis of 3D image datasets. It is also intended for the registration of anatomical (CT) image data onto functional (MR) data to facilitate the comparison of various lesions. Volume and distance measurements facilitate the quantification of lesions and the analysis of both soft and hard tissue.
A variant of the software also provides web-browser access for review purposes. Images accessed through a web-browser (via a mobile device or by other means) should not be used to create a diagnosis, treatment plan, or other decision that may affect patient care.
The new device is similar to the predicate devices. All are PACS system accessories that allow the user to view and manipulate 3D image data sets. This new version adds a dedicated tool set for lesion management and flythrough visualizations rendered along a centerline for endoscopic view of vessels and airways.
Principles of operation and technological characteristics of the new and predicate devices are the same.
{4}------------------------------------------------
# INTENDED USE
The Volume Viewing software is a visualization package for PACS workstations. It is intended to support the medical professional in the reading, analysis and diagnosis of DICOM compliant volumetric medical datasets. The software is intended as a general purpose digital image processing tool, with optional functionality to facilitate visualization and measurement of vessel features.
Other optional functionality is intended for the registration of anatomical (CT) on a second CT dataset or on functional volumetric image data (MR) to facilitate the comparison of various lesions. Volume and distance measurements are intended for evaluation and quantification of tumour measurements, and other analysis and evaluation of both hard and soft tissues. The software also supports interactive segmentation of a region of interest (ROI), has a dedicated tool set for lung lesion segmentation, quantification and follow-up of lesions selected by the user and provides tools to define and edit paths such as centerlines through structures, which may be used to analyze cross-sections of structures, or to provide flythrough visualizations rendered along such centerline.
Caution: Web-browser access is available for review purposes. Images accessed through a web-browser (via a mobile device or by other means) should not be used to create a diagnosis, treatment plan, or other decision that may affect patient care.
Intended use has not changed as a result of any labeling modification(s).
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH V. PREDICATE DEVICES
Agfa's IMPAX Volume Viewing 4.0 has an Indication for Use statement largely similar to the statements for the predicate devices (K133135, K121916). Intended uses are the same. As software accessories to PACS systems, the predicates and the new device have the same technological characteristics. The software is used to identify characteristic patterns within 3D image data which the user can then view and manipulate.
IMPAX Volume Viewing 4.0 comprises the same functionality to manipulate and analyze 3D image data as its primary predicate IMPAX Volume Viewing 3.0. The device subject of this submission however differs on the following technological characteristics:
- 1. The visualization algorithm for multi-planar reformatting (MPR) now employs a different interpolation. MPR visualization is present in the primary predicate IMPAX Volume Viewing 3.0, but uses a different algorithm.
- 2. It includes additional visualization tools for fly-through along a centerline intended for endoscopic viewing of vessels and airways. This functionality is present in the secondary predicate TeraRecon's inNuition, but uses proprietary algorithms,
- 3. New tools were added for the automated segmentation and management of lung lesions. This functionality is present in the secondary predicate TeraRecon's inNuition, but uses proprietary algorithms.
Descriptive characteristics and performance data are adequate to ensure equivalence.
Differences in devices do not alter the intended therapeutic/diagnostic effect.
IMPAX Volume Viewing 4.0 is an accessory to Agfa's IMPAX PACS systems. It is a general
{5}------------------------------------------------
purpose medical image processing tool for the reading and analysis of 3D image datasets. It is the successor to Agfa's Volume Viewing 3.0 (K133135) and adds the following new functionality: a dedicated tool set for lung lesion segmentation, quantification and follow-up of lesions selected by the user and tools to define and edit paths such as centerlines through structures, which may be used to analyze cross-sections of structures, or to provide flythrough visualizations rendered along such centerline.
It is a tool for conveniently viewing and manipulating cross-sectional image series for display in any orientation and slice thickness. A second series can be registered or fused to the first automated, manually or with user defined landmarks. Automated table removal, removal of bonelike structures and segmentation of blood vessels and air-filled structures facilitate the visualization of vessels and airways. The software includes measurement tools for stenosis and lesions. A web-browser variant with a limited tool set provides remote access for review purposes.
In addition the software provides color maps, subtraction views, multiple screen layouts and anaglyphic images. Anaglyphic images are used to provide a stereoscopic 3D effect, when viewed with glasses where the two lenses are different (usually chromatically opposite) colors, such as red and cyan. Images are made up of two color lavers, superimposed, but offset with respect to each other to produce a depth effect. The stereoscopic 3D view is available next to the "regular" 3D view. A user can toggle the image from the regular 3D view to stereo 3D. Tests by medical professionals showed that there are no specific medical or clinical benefits to using the stereoscopic 3D view. It is perceived as useful to train medical students, but has no clinical benefits in comparison to the "regular" 3D view.
The table on the next page summarizes the similarities and differences between the new device and the predicates.
{6}------------------------------------------------
| | IMPAX<br>Volume Viewing<br>4.0<br>NEW DEVICE | IMPAX<br>Volume Viewing 3.0<br>(K133135)<br>PRIMARY<br>PREDICATE | TeraRecon<br>iNtuition 4.4.11<br>(K121916)<br>PREDICATE |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3D Volume rendering | SAME AS PREDICATES | √ | √ |
| MPR, CPR | SAME AS PREDICATES | √ | √ |
| MIP, MinIP, AvgIP | SAME AS PREDICATES | √ | √ |
| Fusion and subtraction<br>views | SAME AS PREDICATES | √ | √ |
| Can load and register two<br>data sets for comparison<br>(landmark based,<br>automated, manual) | SAME AS PREDICATES | √ | √ |
| Multi-modality (CT and<br>MR) image registration | SAME AS PREDICATES | √ | √ |
| Anaglyphic stereo 3D<br>Viewing | SAME AS PREDICATES | √ | √ |
| Segmentation tools | SAME AS PREDICATES | √ | √ |
| Automated removal of the<br>CT-table | SAME AS PREDICATES | √ | √ |
| Automated removal of<br>bone-like structures | SAME AS PREDICATES | √ | √ |
| Volume measurements | SAME AS PREDICATES | √ | √ |
| Semi-automatied region<br>growing | SAME AS PREDICATES | √ | √ |
| Vessel analysis | SAME AS PREDICATES | √ | √ |
| Stenosis measurements | SAME AS PREDICATES | √ | √ |
| AVI movie generation | SAME AS PREDICATES | √ | √ |
| Reformat to a new dataset | SAME AS PREDICATES | √ | √…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.