NexSite™ HD Hemodialysis Step Tip Catheter For Long Term Use (24 cm, 28 cm, 32 cm, 36 cm, 40 cm, 55 cm)
K161026 · Marvao Medical Devices, Ltd. · MSD · Aug 9, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K161026
Device Name
NexSite™ HD Hemodialysis Step Tip Catheter For Long Term Use (24 cm, 28 cm, 32 cm, 36 cm, 40 cm, 55 cm)
Applicant
Marvao Medical Devices, Ltd.
Product Code
MSD · Gastroenterology, Urology
Decision Date
Aug 9, 2016
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5540
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NexSite HD Hemodialysis Step Tip Catheter for long term use is indicated for use in attaining long term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily inserted in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein. Catheters greater than 40cm are indicated for femoral vein insertion.
Device Story
Long-term vascular access catheter for hemodialysis and apheresis; percutaneously inserted into internal jugular, subclavian, or femoral veins (for >40cm lengths). Device includes Dacron cuff for tissue integration and Dermal Ingrowth Support Collar (DISC) for catheter direction. Supplied with accessories (tunneler, guidewire, introducer/dilator) to facilitate placement. Single-use, sterile, ethylene oxide-sterilized. Used by clinicians in clinical settings to provide blood access for dialysis/apheresis procedures. Benefits include reliable, long-term vascular access for patients with end-stage renal disease or other conditions requiring blood purification.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via in vitro bench testing (visual/dimensional, priming volume, leakage, joint strength, pressure, flow rate, recirculation, clamp fatigue, hemolysis, chemical tolerance, radiopacity, corrosion resistance) and in vivo implantation studies.
Technological Characteristics
Polyurethane and Dacron construction. Step-tip design. Includes DISC (polyurethane/Dacron). Sterile, ethylene oxide-sterilized. Single-use. Complies with ISO 10555-1:2013. Mechanical device; no software or energy source.
Indications for Use
Indicated for adult patients requiring long-term vascular access for hemodialysis and apheresis. Primary insertion site is the internal jugular vein; alternate sites include the subclavian vein. Catheters >40cm are indicated for femoral vein insertion.
Regulatory Classification
Identification
A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
Special Controls
*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Marvao NexSite HD Hemodialysis Catheter for long term use (K121933, K133796, K140492)
Marvao NexSite HD Hemodialysis Split Tip Catheter for long term use (K143567)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus-like design with three human profiles facing right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 8, 2016
Marvao Medical Devices, Ltd. % Joseph DePaolo Regulatory Manager Boston Biomedical Associates 100 Crowley Drive, Suite 216 Marlborough, MA 01752
K161026 Re:
> Trade/Device Name: NexSite™ HD Hemodialysis Step Tip Catheter for long term use (24 cm, 28 cm, 32 cm, 36 cm, 40 cm, 55 cm) Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood Access Device and Accessories Regulatory Class: II Product Code: MSD Dated: July 5, 2016 Received: July 6, 2016
Dear Joseph DePaolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Douglas Silverstein -S 2016.08.09 10:56:34 -04'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161026
#### Device Name
NexSite™ HD Hemodialysis Step Tip Catheter for long term use (24 cm, 32 cm, 36 cm, 40 cm, 55 cm)
#### Indications for Use (Describe)
The NexSite HD Hemodialysis Step Tip Catheter for long term use in attaining long term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily inserted in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein. Catheters greater than 40cm are indicated for femoral vein insertion.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------|
| <span style="font-size:100%;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> |
| <span style="font-size:100%;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21CFR Part 807.92.
## I. SUBMITTER
| Name: | Marvao Medical Devices, Ltd. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | GMIT Innovation in Business Centre, Dublin Road<br>Galway, Ireland |
| Phone: | +353 (0)91 759 301 |
| Fax: | +353 (0)91 762 055 |
| Contact Person: | Joseph DePaolo<br>Boston Biomedical Associates<br>386 West Main Street, Suite 7<br>Northborough, MA 01532<br>jdepaolo@boston-biomedical.com<br>Phone: (508) 691-7026<br>Fax: (508) 351-8637 |
| II. DEVICE | |
| Device/Trade Name:<br>Date of Preparation:<br>Common Name:<br>Classification Name:<br>Classification Number:<br>Product Code/ Classification Panel: | NexSite™ HD Hemodialysis Step Tip Catheter for long<br>term use (24 cm, 28 cm, 32c m, 36 cm, 40 cm, 55 cm)<br>August 4, 2016<br>Blood access devices and accessories<br>catheter, hemodialysis, implanted<br>876.5540<br>MSD/Gastroenterology /Urology |
| III.PREDICATE DEVICE(S): | |
| K121933 / K133796 /K140492 | Marvao NexSite HD Hemodialysis Catheter for long<br>term use, Manufactured by Marvao Medical<br>Devices, Ltd. |
| K143567 | Marvao NexSite HD Hemodialysis Split Tip |
510(k) Summary Page 1 of 3
Catheter for long term use, Manufactured by
Marvao Medical Devices, Ltd.
{4}------------------------------------------------
## IV. DEVICE DESCRIPTION
The proposed device, the NexSite HD Hemodialysis Step Tip Catheter for long term use is a long term catheter intended for use in attaining long term vascular access for hemodialysis and apheresis. The proposed device is available in the following models below and has a Dacron cuff distal to the bifurcation hub.
Models:
NexSite HD Hemodialysis Step Tip Catheter for long term use (24cm) NexSite HD Hemodialysis Step Tip Catheter for long term use (28cm) NexSite HD Hemodialysis Step Tip Catheter for long term use (32cm) NexSite HD Hemodialysis Step Tip Catheter for long term use (36cm) NexSite HD Hemodialysis Step Tip Catheter for long term use (40cm) NexSite HD Hemodialysis Step Tip Catheter for long term use (55cm)
A Dermal Ingrowth Support Collar (DISC), manufactured from polyurethane and Dacron®. and supplied with the Catheter, is implanted subcutaneously, and is intended to assist with the direction of the catheter. The Catheter and DISC are packaged with accessories (stainless steel Tunneler and Sleeve, 0.038" Guidewire, 16Fr Introducer/Dilator, and Luer Caps) that are used to facilitate catheter placement.
The NexSite HD Hemodialysis Step Tip Catheter for long term use is provided as a sterile, singleuse device, and is sterilized using a validated ethylene oxide process. It is a blood contact device with greater than 30 days of exposure.
# V. INTENDED USE
The NexSite HD Hemodialysis Step Tip Catheter for long term use is indicated for use in attaining long term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily inserted in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein. Catheters greater than 40cm are indicated for femoral vein insertion.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The descriptive characteristics of the proposed device and the identified predicates were compared and determined to be substantially equivalent in terms of intended use, dimensions, materials, and basic design. Comparative performance testing using finished proposed NexSite™ HD Step Tip catheters and identified predicate devices was completed. The test results support the substantial equivalence of the proposed NexSite™ HD Step Tip catheter to the predicate devices.
#### VII. PERFORMANCE DATA
In vitro testing was performed on the NexSite HD Hemodialysis Step Tip Catheter to assure reliable design and performance in accordance with ISO 10555-1:2013.. The non-clinical tests performed include visual and dimensional, priming volume, catheter leakage, catheter joint strength, catheter pressure, catheter flow rate testing, catheter recirculation testing, catheter clamp
{5}------------------------------------------------
fatigue testing, mechanical hemolysis, chemical tolerance to disinfectant agents, radiopacity and corrosion resistance. The test results demonstrate that the NexSite™ HD Hemodialysis Step Tip Catheter meets the requirements in the applicable standards and specifications, and is substantially equivalent to legally marketed predicate devices.
In vivo implantation studies were also performed to demonstrate that the device would perform as intended. Clinical studies were not deemed necessary since in vivo and in vitro testing were sufficient to demonstrate safety and effectiveness by way of comparison to a legally marketed predicate device.
#### Guidance
The FDA Guidance on Premarket Notification [(510(k)] Submission for Short-Term and Long-Term Intravascular Catheters, dated 3/16/95, was utilized in order to meet the FDA requirements for content and organization of this submission, as well as the t Guidance Industry and Food and Drug Administration Staff Implanted Blood Access Devices for Hemodialysis, January 21, 2016.
## VIII. CONCLUSIONS
Marvao Medical Devices, Ltd. believes the proposed NexSite HD Hemodialysis Step Tip Catheter for long term use is substantially equivalent to legally marketed predicate devices. The indications for use, methods of operation, design and materials used are either identical or substantially equivalent to existing legally marketed predicate products. In addition, performance testing supports substantial equivalence of the proposed and predicate devices.
510(k) Summary Page 3 of 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.