EndoVive One Step Button Low Profile Initial Placement PEG Kit with ENFit Connector, EndoVive Low Profile Button Replacement Gastrostomy Tube Kit with ENFit Connector, EndoVive Button Right Angle Feeding Set with ENFit Connector, EndoVive Button Bolus Feeding Set with ENFit Connector, EndoVive Button Decompression Tube with ENFit Connector
K161003 · Boston Scientific Corporation · PIF · Jun 17, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K161003
Device Name
EndoVive One Step Button Low Profile Initial Placement PEG Kit with ENFit Connector, EndoVive Low Profile Button Replacement Gastrostomy Tube Kit with ENFit Connector, EndoVive Button Right Angle Feeding Set with ENFit Connector, EndoVive Button Bolus Feeding Set with ENFit Connector, EndoVive Button Decompression Tube with ENFit Connector
Applicant
Boston Scientific Corporation
Product Code
PIF · Gastroenterology, Urology
Decision Date
Jun 17, 2016
Decision
SEKD
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The One Step Button device is intended to provide nutrition to a patient directly into the stoma. It is indicated for use on patients who are unable to consume nutrition by conventional means. The EndoVive Low Profile Replacement Button Gastrostomy device is intended to provide nutrition to a patient directly into the stomach through a stoma. It is indicated for use on patients who are unable to consume nutrition by conventional means. The Right Angle Feeding Set is intended to provide nutrition to a patient directly into the stomach through a gastrostomy tube. It is indicated for use on patients who are unable to consume nutrition by conventional means. The Bolus Feeding Set is intended to provide nutrition to a patient directly into the stomach through a gastrostomy tube. It is indicated for use on patients who are unable to consume nutrition by conventional means. The Right Angle Decompression Tube is intended to provide stomach decompression through a gastrostomy tube. It is indicated for use on patients who require enteral feeding as a means of nutritional support
Device Story
EndoVive Button kits provide enteral nutrition or stomach decompression via gastrostomy tubes. Kits include feeding tubes, Y-ports, and bolus/decompression accessories. Devices feature ENFit threaded connectors to prevent misconnection with non-enteral devices. Used in clinical settings by healthcare providers to establish or maintain gastric access. Input is nutritional formula or gastric content; output is delivery of nutrition or decompression of stomach. Device facilitates long-term nutritional support for patients unable to feed orally. Modification involves updated connector design per AAMI/CN3 ISO/DIS 80369-3.2 standards to improve safety and reduce misconnection risks.
Clinical Evidence
Bench testing only. No clinical data. Testing included AAMI CN3:2014/ISO/DIS 80369-3.2 and AAMI CN20:2014/ISO 80369-20 compliance. Assessments included risk analysis, enteral connector misconnection assessment, human factors validation study, FMEA, tensile strength, flow rate, and liquid leakage testing.
Technological Characteristics
Gastrointestinal feeding/decompression tubes and accessories. Features ENFit threaded connectors compliant with AAMI/CN3 ISO/DIS 80369-3.2. Connectors utilize rigid materials. Design includes Y-port and bolus/decompression configurations. Non-sterile components (kits may contain antiseptic/iodine).
Indications for Use
Indicated for patients unable to consume nutrition by conventional means or requiring enteral feeding for nutritional support; includes patients requiring stomach decompression via gastrostomy tube.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
One Step Button, Low Profile Initial Placement PEG Kit (K910584)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three figures representing people. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2016
Boston Scientific Corporation Virginia Garcia, MPP, RAC Principal, Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K161003
Trade/Device Name: EndoVive™ One Step Button Low Profile Initial Placement PEG Kit with ENFit Connector, Endovive™ Low Profile Button Replacement Gastrostomy Tube Kit with ENFit Connector, Endovive™ Button Right Angle Feeding Set with ENFit Connector, Endovive Button Bolus Feeding Set with ENFit Connector, Endovive Button Decompression Tube with ENFit Connector Regulation Number: 21 CFR& 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF Dated: April 8, 2016 Received: April 11, 2016
Dear Virginia Garcia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the Endovive One Step Button Low Profile Initial Placement Peg Kit with ENFit Connector and Endovive Low Profile Button Replacement Gastrostomy Tube Kit with ENFit Connector have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate
{1}------------------------------------------------
information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kits contain an antiseptic ointment packet and iodine swabs, which are subject to regulation as drugs.
Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:
Center for Drug Evaluation and Research Office of Compliance 10903 New Hampshire Avenue Bldg. 51, Rm 5271 Silver Spring, MD 20993-0002
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
{2}------------------------------------------------
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K161003
#### Device Name
EndoVive™ One Step Button Low Profile Initial Placement PEG Kit with ENFit Connector, Endovive™ Low Profile Button Replacement Gastrostomy Tube Kit with Connector, Endovive™ Button Right Angle Feeding Set with ENFit Connector, Endovive™ Button Bolus Feeding Set with ENFit Connector, Endovive™ Button Decompression Tube with ENFit Connector
#### Indications for Use (Describe)
The One Step Button device is intended to provide nutrition to a patient directly into the stoma. It is indicated for use on patients who are unable to consume nutrition by conventional means.
The EndoVive Low Profile Replacement Button Gastrostomy device is intended to provide nutrition to a patient directly into the stomach through a stoma. It is indicated for use on patients who are unable to consume nutrition by conventional means.
The Right Angle Feeding Set is intended to provide nutrition to a patient directly into the stomach through a gastrostomy tube. It is indicated for use on patients who are unable to consume nutrition by conventional means.
The Bolus Feeding Set is intended to provide nutrition to a patient directly into the stomach through a gastrostomy tube. It is indicated for use on patients who are unable to consume nutrition by conventional means.
The Right Angle Decompression Tube is intended to provide stomach decompression through a gastrostomy tube. It is indicated for use on patients who require enteral feeding as a means of nutritional support
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.8536 3.85355C14.0488 3.65829 14.0488 3.34171 13.8536 3.14645C13.6583 2.95118 13.3417 2.95118 13.1464 3.14645L6.29289 9.99999L2.85355 6.56065C2.65829 6.36539 2.34171 6.36539 2.14645 6.56065C1.95118 6.75592 1.95118 7.0725 2.14645 7.26776L5.58579 10.7071C5.78105 10.9024 6.09763 10.9024 6.29289 10.7071L13.8536 3.14645C14.0488 2.95118 14.3654 2.95118 14.5607 3.14645C14.7559 3.34171 14.7559 3.65829 14.5607 3.85355L13.8536 3.85355Z" fill="black"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> </svg> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Button
Tube
Boston Scientific Corporation K920894
KNT
Decompression
## 1. Submitter
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4430 Fax: 508-683-5939
Contact: Virginia Garcia, MPP, RAC Principal, Regulatory Affairs Specialist
Date Prepared: April 8, 2016
## 2. Devices
Trade Name:
Endo Vive™ One Step Button Low Profile Initial Placement PEG Kit with ENFit Connector Endovive™ Low Profile Button Replacement Gastrostomy Tube Kit with ENFit Connector Endovive™M Button Right Angle Feeding Set with ENFit Connector Endovive™ Button Bolus Feeding Set with ENFit Connector Endovive™M Button Decompression Tube with ENFit Connector Common Name: Gastrointestinal tube and accessories Product Code: PIF Device Class and panel: Class II, Gastroenterology and Urology 21 CFR 876.5980 Classification Regulation:
| 3. Predicate Devices | | | | |
|------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------|------------------------------------------------|------------------------------------------------|
| Trade Name: | One Step<br>Button, Low<br>Profile Initial<br>Placement PEG<br>Kit | Low Profile<br>Button<br>Replacement<br>Gastrostomy<br>Tube Kit | Button Right<br>Angle Feeding<br>Set | Button Bolus<br>Feeding Set |
| Manufacturer<br>and Clearance<br>Number: | Boston<br>Scientific<br>Corporation<br>K910584 | Boston<br>Scientific<br>Corporation<br>K014297 | Boston<br>Scientific<br>Corporation<br>K014297 | Boston<br>Scientific<br>Corporation<br>K014297 |
| KNT
### 3
Product Code:
Classification Name:
Device Class and panel:
Classification Regulation
KGC
| KNT
21 CFR 876.5980
KNT
Gastrointestinal tube and accessories
Class II, Gastroenterology/Urology
{5}------------------------------------------------
## 4. Device Description
The purpose of this Traditional 510(k) is to demonstrate that the proposed changes to the Y-Port and Bolus/Decompression accessories that are included in the devices identified above do not raise new questions of safety or effectiveness and that the devices are substantially equivalent to the currently marketed EndoVive™ Button kit and Accessory products.
The Y-Port and Bolus/Decompression accessories that are included in the devices identified above have two connections. One is located the distal end and the other is at the proximal end. The connector on the proximal end is identical in design to the ENFit threaded connector of the Y-Port accessory which was cleared via K150679. The Y-Port and Bolus/Decompression accessories are also identical in materials to the Y-Port device which was cleared via K150679. All of the performance and biocompatibility testing that was presented in K150679 to support the ENFit threaded connector is applicable to the Y-Port and Bolus/Decompression accessories of the proposed device. The scope of this submission is to demonstrate that the changes to the distal end of the Y-Port and Bolus/Decompression accessories are substantially equivalent to the predicate device. Neither the Button feeding tube nor the accessories used for tube placement within these kits is changing. The Button feeding tube and placement accessories in the kits are identical to the predicates.
# 5. Indication for Use:
Endo Vive™ One Step Button Low Profile Initial Placement PEG Kit with ENFit Connector The One Step Button Device is intended to provide nutrition to a patient directly into the stomach through a stoma. It is indicated for use on patients who are unable to consume nutrition by conventional means.
Endovive™ Low Profile Button Replacement Gastrostomy Tube Kit with ENFit Connector The EndoVive Low Profile Replacement Button Gastrostomy Device is intended to provide nutrition to a patient directly into the stomach through a stoma. It is indicated for use on patients who are unable to consume nutrition by conventional means.
Endovive™ Button Right Angle Feeding Set with ENFit Connector The Right Angle Feeding Set is intended to provide nutrition to a patient directly into the stomach through a gastrostomy tube. It is indicated for use on patients who are unable to consume nutrition by conventional means.
Endovive™ Button Bolus Feeding Set with ENFit Connector
The Bolus Feeding Set is intended to provide nutrition to a patient directly into the stomach through a gastrostomy tube. It is indicated for use on patients who are unable to consume nutrition by conventional means.
Endovive™M Button Decompression Tube with ENFit Connector The Right Angle Decompression Tube is intended to provide stomach decompression through a gastrostomy tube. It is indicated for use on patients who require enteral feeding as a means of nutritional support
{6}------------------------------------------------
# 6. Technological Characteristics:
The Y-Port and Bolus/Decompression accessories included in the initial placement kits, the replacement kits, and replacement accessory sets are the subject of this submission.
The initial placement kit contains a feeding tube that is placed via a procedure called Percutaneous Endoscopic Gastrostomy (PEG). It also contains accessories assembled for the preference of the physician placing the Button feeding set accessories included in the kit are for use in connecting the feeding tube to nutrition.
The replacement kit contains a feeding tube that is used to replace an existing feeding tube using the accessories assembled in the kit. The feeding set accessories included in the kit are for use in connecting the feeding tube to nutrition.
The feeding set and decompression tube accessories that are included in the kits are also sold separately. The Y-Port and Bolus/Decompression accessory within the feeding and decompression tube sets is being modified to comply with the new AAMI/CN3 ISO/DIS 80369-3.2 standard requirements to reduce the risk of misconnection with non-enteral feeding devices. The AAMI/CN3 ISO/DIS 80369-3.2 ENFit thread design will be added to the connector and the connector will be made of a more rigid material. Other than the connector changes, the Button feeding tube and placement accessories in the kits are identical to the predicates.
## 7. Performance Data:
The Endo Vive™ Button Kit with ENFit and Accessories have been tested according to the following standards:
- AAMI CN3:2014 (PS)/ ISO/DIS 80369-3.2 Small bore connectors for liquids and . gases in healthcare applications - Part 3: Connectors for enteral applications
- . AAMI CN20:2014 (PS)/ ISO 80369-20 gases in healthcare applications - Part 20: Common test methods
The information presented in this submission to support substantial equivalence of the Endo Vive Button Kit with ENFit and Accessory with ENFit Connector to the legally marketed predicate devices include: device description, indications for use, device comparison, material information, and labeling. Bench testing of the connector ISO/DIS 80369-3.2 ENFit thread design supports compliance to the material, and non-interconnectability, principles of the ISO 80369-1: 2010 standard. This performance assessment includes Risk analysis, Enteral Connector Misconnection Assessment, Enteral Connector Risk Management Report, Human Factors Validation Study, and ENFit Misconnection data with FMEA, tensile strength, flow rate, and liquid leakage.
Boston Scientific has assessed the similarities between the proposed Endo Vive Button Kit with ENFit and Accessories with ENFit Connector and its predicates in terms of intended use and technological characteristics. The differences in the technological characteristics are
{7}------------------------------------------------
minor and do not present any new issues of safety or effectiveness. This evidence supports a finding of substantial equivalence between the products.
# 8. Conclusion:
Boston Scientific Corporation has demonstrated that the proposed Endo Vive Button Kit with ENFit Connector and Accessories with ENFit Connector are substantially equivalent to the currently marketed EndoVive Initial Placement Button Kit (K910584), EndoVive Replacement Button Kit (K014297) and Decompression Tube (K920894).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.