K160981 · Edan Instruments, Inc. · MHX · Dec 22, 2016 · Cardiovascular
Device Facts
Record ID
K160981
Device Name
Patient Monitor
Applicant
Edan Instruments, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Dec 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The monitors are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments. The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanics (RM) and impedance cardiography (ICG). BIS is intended for use on adult and pediatric patients. ICG monitoring is intended for use on adults only. The arrhythmia detection and ST Segment analysis are intended for adult patients. The monitors are additionally intended for use during patient transport inside hospitals. The monitors are not intended for MRI environments.
Device Story
V series patient monitors (elite V5, V6, V8) provide continuous monitoring of physiological parameters including ECG, RESP, TEMP, SpO2, PR, NIBP, IBP, CO2, C.O., AG, BIS, RM, and ICG. Input data is acquired via various parameter-specific modules. The device processes these signals to display waveforms, store data, and trigger alarms for abnormal conditions. Used in hospital environments (including transport) by trained healthcare professionals. Output is presented on a color TFT touch screen (12.1, 15, or 17 inches). Clinicians use the visual and audible alarms to identify patient status changes, facilitating timely intervention. Benefits include continuous patient surveillance and early detection of physiological abnormalities.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance testing against various recognized standards (e.g., IEC 60601-2-27 for ECG, ISO 80601-2-55 for gas monitors).
Technological Characteristics
Modular patient monitoring system with 12.1-17 inch TFT touch screens. Powered by AC or internal Lithium-ion battery. Connectivity includes networking and online software upgrades. Biocompatibility per ISO 10993-1. Safety standards: IEC 60601-1, IEC 60601-1-8, IEC 60601-2-49, ISO 80601-2-55, ISO 80601-2-56, ISO 80601-2-61, ISO 81060-2.
Indications for Use
Indicated for monitoring, storing, and alarming of physiological parameters in adult, pediatric, and neonatal patients in hospital environments. Arrhythmia detection, ST-segment analysis, and ICG are restricted to adult patients. BIS is indicated for adult and pediatric patients. Not for MRI environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2016
Edan Instruments, Inc. % Doug Worth Edan Medical 1200 Crossman Ave. Suite 200 Sunnyvale, California 94089
Re: K160981 Trade/Device Name: Patient Monitor, models elite V5, elite V6 and elite V8 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, DSI, DRT, CBQ, DXN, DSK, CBR, DQA, NHO, CBS, NHQ, NHP, CCK, DSB, CCL, BZQ, BZK, DPS, FLL, DRG, MLD Dated: November 30, 2016 Received: November 30, 2016
Dear Doug Worth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mude Jellm
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160981
Device Name
Patient Monitor, models elite V5, elite V6 and elite V8
Indications for Use (Describe)
The monitors are intended to be used for monitoring, storing, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended healthcare professionals in hospital environments.
The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanics (RM) and impedance cardiography (ICG).
BIS is intended for use on adult and pediatric patients.
ICG monitoring is intended for use on adults only.
The arrhythmia detection and ST Segment analysis are intended for adult patients.
The monitors are additionally intended for use during patient transport inside hospitals.
The monitors are not intended for MRI environments.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (K) Summary
## Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92
| 1. Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>Shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: +86(0755) 26858736<br>Fax: +1 (408) 418-4059 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Queena Chen |
| Preparing date: | March 30, 2016 |
| 2. Device name and<br>classification: | Device Name: Patient Monitor<br>Model: elite V5, elite V6, elite V8<br>Classification Name/ Product code:<br>870.1025 monitor, physiological, patient(with arrhythmia<br>detection or alarms)/ MHX<br>870.1025 Detector and Alarm, Arrhythmia/ DSI<br>870.1025 Monitor, ST Segment with Alarm/ MLD<br>870.2300 Cardiac monitor (including cardiotachometer and rate<br>alarm)/ DRT<br>870.1130 Non-Invasive blood pressure/ DXN<br>870.1110 Blood pressure computer/ DSK<br>880.2910 Clinical Electronic Thermometers-Temperature<br>Monitor with Probe/ FLL<br>870.2700 Oximeter, Pulse/ DQA<br>868.1400 Carbon Dioxide Gas Analyzer/ CCK<br>868.1500 Analyzer, Gas, Enflurane, Gaseous-Phase (Anesthetic<br>Concentration)/ CBQ<br>868.1500 Analyzer, Gas, Desflurane, Gaseous-Phase<br>(Anesthetic Concentration)/NHO<br>868.1500 Analyzer, Gas, Isoflurane, Gaseous-Phase (Anesthetic<br>Concentration)/NHQ<br>868.1500 Analyzer, Gas, Sevoflurane, Gaseous-Phase<br>(Anesthetic Concentration)/NHP<br>868.1620 Halothane gas analyzer/ CBS<br>868.1700 Nitrous Oxide gas analyzer/ CBR<br>868.1720 Oxygen gas analyzer/ CCL |
{4}------------------------------------------------
| <b>3.Premarket Notification Class III Certification and Summary</b> | 870.2770 Impedance plethysmograph/ DSB<br>868.1850 Monitoring spirometer/ BZK<br>868.2375 Monitor, Breathing Frequency/BZQ<br>870.2340 Electrocardiograph/DPS<br>870.2910 Radiofrequency physiological signal transmitter and receiver/ DRG<br>Regulatory Class: Class II<br>Not applicable, the subject device is Class II. |
|---------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>4. Predicate Device(s):</b> | 1) Edan Instruments, Inc. / Patient Monitor Model elite V8/ K120173<br>2) Shenzhen Mindray Bio-Medical Electronics Co., Ltd/ BeneView T8 /K092449.<br>3) Edan Instruments, Inc. / iM70/ K131971 |
| <b>5. Reason for Submission</b> | Modification for previous product elite V8 and introduce two new models elite V5 and elite V6. |
| <b>6. Pre-Submission, IDE</b> | Not applicable, there is no prior submission. |
| <b>7. Device Description:</b> | V series Patient Monitor including elite V5, elite V6 and elite V8 which can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.<br>The V series Patient Monitor realize the monitoring of physiological parameters by configuration with different parameter modules which include SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure, mean pressure and pulse rate), TEMP, ECG, RESP(respiration), CO2, IBP, C.O. and AG(anesthetic gas), RM(respiratory mechanics), BIS(bispectral index) and ICG(impedance cardiography).<br>The above is the maximum configuration for V series Patient Monitor, the user may select different monitoring parameters in according with their requirements.<br>elite V5 configures with 12.1-inch color TFT touch screen , elite V6 and elite V8 with same screen except different sizes |
{5}------------------------------------------------
15-inch and 17-inch separately. Three models are all build-in Lithium-ion battery, support software upgrade online and networking.
8. Intended Use: The monitors are intended to be used for monitoring, storing, and reviewing of, and to generate alarms for, multiple physiological parameters of adults, pediatrics and neonates. The monitors are intended for use by trained healthcare professionals in hospital environments.
> The monitored physiological parameters include: ECG, respiration (RESP), temperature (TEMP), oxygen saturation of arterial blood (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), cardiac output (C.O.), anesthetic gas (AG), bispectral index (BIS), respiration mechanics (RM) and impedance cardiography (ICG).
BIS is intended for use on adult and pediatric patients.
ICG monitoring is intended for use on adults only.
The arrhythmia detection and ST Segment analysis are intended for adult patients.
The monitors are additionally intended for use during patient transport inside hospitals.
The monitors are not intended for MRI environments.
### 9. Predicate Device Comparison
The subject devices share the same characteristics in all items with the predicate device, concluding from using the same technology and principle. All the technological differences existed between the subject and predicate devices are only some performance parameters improvement, for example accuracy for RESP Module is better, minor differences on NIBP module, different measurement range, accuracy and Sample flow rate for CO2 module and different Warm-up time, Sample flow rate, measurement range, accuracy for AG module. Please refer to the following tables for details.
{6}------------------------------------------------
# Comparison to Primary Predicate elite V8
| Item | Proposed device<br>(elite V5, elite V6, elite V8) | Predicate device<br>(elite V8) | Justification | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------|
| K# | Current Submission | K120173 | — | |
| Intended use | | | | |
| Description | The monitors are intended to be used for<br>monitoring, storing, and reviewing of,<br>and to generate alarms for, multiple<br>physiological parameters of adults,<br>pediatrics and neonates. The monitors<br>are intended for use by trained<br>healthcare professionals in hospital<br>environments.<br>The monitored physiological parameters<br>include: ECG, respiration (RESP),<br>temperature (TEMP), oxygen saturation<br>of arterial blood (SpO2), pulse rate (PR),<br>non-invasive blood pressure (NIBP),<br>invasive blood pressure (IBP), carbon<br>dioxide (CO2), cardiac output (C.O.),<br>anesthetic gas (AG), bispectral index<br>(BIS), respiration mechanics (RM) and<br>impedance cardiography (ICG).<br>BIS is intended for use on adult and<br>pediatric patients.<br>ICG monitoring is intended for use on<br>adults only.<br>The arrhythmia detection and ST<br>Segment analysis are intended for adult<br>patients.<br>The monitors are additionally intended<br>for use during patient transport inside<br>hospitals.<br>The monitors are not intended for MRI<br>environments. | The monitor monitors parameters<br>such as ECG (3-lead, 5-lead<br>selectable), Respiration (RESP),<br>Functional arterial oxygen saturation<br>(SpO2), Invasive or noninvasive<br>blood pressure (dual-IBP, NIBP),<br>Temperature (dual-TEMP), Expired<br>CO2.<br>The monitor is equipped with alarms<br>that indicate system faults (such as<br>loose or defective electrodes).<br>physiologic parameters that have<br>exceeded the limits set by the<br>operator, or both.<br>The arrhythmia detection and ST<br>Segment analysis are not intended for<br>neonatal patients. | Similar | |
| Intended patient<br>population | adult, pediatric and neonatal patients | adult, pediatric and neonatal patients | Same | |
| Intended<br>application<br>environment | Hospital environment. | Hospital environment. | Similar | |
| ECG monitor | | | | |
| Lead Mode | 3-Lead: I, II, III<br>5-Leads: I, II, III, aVR, aVL, aVF, V<br>12-leads: I, II, III, aVR, aVL, aVF, V1 to<br>V6 | 3-Lead: I, II, III<br>5-Leads: I, II, III, aVR, aVL, aVF, V<br>12-leads: I, II, III, aVR, aVL, aVF,<br>V1 to V6 | Same | |
| Lead Naming<br>Style | AHA, IEC | AHA, IEC | Same | |
| Display<br>Sensitivity | 1.25mm/mV (×0.125), 2.5mm/mV<br>(×0.25), 5mm/mV (×0.5), 10mm/mV<br>(×1), 20mm/mV (×2), 40mm/mV (×4),<br>AUTO gain | 1.25 mm/mV (X0.125), 2.5 mm/mV<br>(X0.25), 5 mm/mV (X0.5),<br>10 mm/mV (X1), 20 mm/mV (X2),<br>40 mm/mV (X4), Auto | Same | |
| Sweep | 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s | 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50<br>mm/s | Same | |
| Measurement<br>Range | Neonate: 15 to 350 bpm<br>Pediatric: 15 to 350 bpm<br>Adult: 15 to 300 bpm | Neonate: 15 to 350 bpm<br>Pediatric: 15 to 350 bpm<br>Adult: 15 to 300 bpm | Same | |
| CMRR<br>(Common Mode<br>Rejection Ratio) | Diagnosis: >95dB<br>Monitor: >105dB<br>Surgery: >105dB | Diagnosis: >95dB (the Notch filter is<br>off)<br>Monitor: >105dB (the Notch filter is<br>on)<br>Surgery: >105dB (the Notch filter is<br>on) | Similar | |
| Accuracy | ±1 bpm or ±1%, whichever is<br>greater | ±1 bpm or ±1%, whichever is<br>greater | Same | |
| Resolution | 1 bpm | 1 bpm | Same | |
| Sensibility | 200μV | 200µV (lead II) | Same | |
| Differential Input<br>Impedance | >5MΩ | ≥5 MΩ | Same | |
| Leakage Current<br>of Patient | <10μΑ | <10 μΑ | Same | |
| ST value | | | | |
| Measurement<br>Range | -2.0 mV ~ +2.0 mV | -2.0 mV ~ +2.0 mV | Same | |
| Accuracy | -0.8 mV to +0.8 mV: ±0.02 mV or 10%,<br>whichever is greater.<br>Beyond this range: not specified. | -0.8 to 0.8 mV: Beyond this<br>range:±0.02 mV or ±10%, whichever<br>is greater. Not specified. | Same | |
| Pace | | | | |
| Pulse Indicator | Amplitude: ±2 mV to ±700 mV<br>Width: 0.1 ms to2.0 ms<br>Ascending time: 10 µs to 100 µs | Amplitude: ±2 to ±700 mV<br>Width: 0.1 to 2 ms<br>Rise time: 10 to 100 µs | Same | |
| Pulse Rejection | Amplitude: ±2 mV to ±700 mV<br>Width: 0.1 ms to 2.0 ms<br>Ascending time: 10 µs to 100 µs | Amplitude: ±2 to ±700 mV<br>Width: 0.1 to 2 ms<br>Rise time: 10 to 100 µs | Same | |
| HR | | | | |
| Measurement range | Neonate: 15 to 350 bpm<br>Pediatric: 15 to 350 bpm<br>Adult: 15 to 300 bpm | | Neonate: 15 to 350 bpm<br>Pediatric: 15 to 350 bpm<br>Adult: 15 to 300 bpm | Same |
| Accuracy | ±1% or 1 bpm, whichever is greater | | ±1 bpm or ±1%, whichever is greater | Same |
| Resolution | 1 bpm | | 1 bpm | Same |
| RESP monitor | | | | |
| Principle of Operation | Thoracic impedance | | Trans-thoracic impedance | Same |
| Measurement Range | Adult: 0 to 120 rpm<br>Pediatric/neonate: 0 rpm to 150rpm |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.