K160972 · Gexfix International Corp. · KTT · Dec 28, 2016 · Orthopedic
Device Facts
Record ID
K160972
Device Name
Gexfix External Fixation
Applicant
Gexfix International Corp.
Product Code
KTT · Orthopedic
Decision Date
Dec 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1) Bone fracture fixation 2) Osteotomy 3) Arthrodesis 4) Correction of deformity 5) Revision procedures where other devices have been unsuccessful 6) Bone reconstruction procedures
Device Story
Gexfix External Fixation system provides mechanical bone stabilization during healing; assists in fracture management and reconstructive surgery. System comprises invasive components (implanted temporarily), external components (structural support), and ancillary components. Surgeons assemble components to hold bones in position; physical mechanism of action provides stability. Used in clinical settings for orthopedic applications including fingers, feet, and shoulder. Supplied non-sterile; requires sterilization by user. Benefits include bone healing support and stabilization of fractures/deformities. Does not replace normal bone structure.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F1541-02(2011) including bending, traction, compression, sliding, torsion, and pin push-out. All components met predetermined acceptance criteria.
Technological Characteristics
Materials: Stainless steel, titanium alloy (Ti6Al4V), and PEEK. Mechanical fixation via invasive pins/wires and external structural bars/connectors. Sterilization: User-validated cycle to SAL 10^-6. Compliance: ASTM F1541-02(2011) for mechanical testing; ASTM F2503-13 for MR environment labeling.
Indications for Use
Indicated for patients requiring bone fracture fixation, osteotomy, arthrodesis, deformity correction, revision procedures, or bone reconstruction. Applicable to long bones and various orthopedic applications including fingers, feet, and shoulder.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 28, 2016
Gexfix International Corp. % Roger Gray VP, Quality and Regulatory Donawa Lifescience Consulting S.R.L. Piazza Albania, 10 Rome, 00153 Italy
Re: K160972 Trade/Device Name: Gexfix External Fixation Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: November 23, 2016 Received: November 28, 2016
Dear Roger Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Caroline Rhim -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K160972
Device Name
Gexfix External Fixation
Indications for Use (Describe)
1) Bone fracture fixation
2) Osteotomy
3) Arthrodesis
4) Correction of deformity
5) Revision procedures where other devices have been unsuccessful
6) Bone reconstruction procedures
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for a company called "Gexfix". The logo consists of a blue circle with a white "G" inside. The "G" is stylized with a horizontal line extending from the left side of the "G" to the right side, creating a unique design. Below the circle, the word "Gexfix" is written in black, sans-serif font.
# APPENDIX E
# 510(k) Summary in accordance with 21 CFR 807.92(c)
| Type of Submission: | Traditional 510(k) |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date of Submission: | 21 November 2016 |
| Device Proprietary Name: | Gexfix External Fixation |
| FDA Product Code: | KTT |
| FDA Regulation Number: | 21 CFR 888.3030 |
| FDA Classification Name: | Single/multiple component metallic bone fixation appliances and<br>accessories |
| FDA Identification: | Single/multiple component metallic bone fixation appliances and<br>accessories are devices intended to be implanted consisting of one or<br>more metallic components and their metallic fasteners. The devices<br>contain a plate, a nail/plate combination, or a blade/plate combination<br>that are made of alloys, such as cobalt-chromium-molybdenum, stainless<br>steel, and titanium, that are intended to be held in position with<br>fasteners, such as screws and nails, or bolts, nuts, and washers. These<br>devices are used for fixation of fractures of the proximal or distal end of<br>long bones, such as intracapsular, intertrochanteric, intercervical,<br>supracondylar, or condylar fractures of the femur; for fusion of a joint; or<br>for surgical procedures that involve cutting a bone. The devices may be<br>implanted or attached through the skin so that a pulling force (traction)<br>may be applied to the skeletal system. |
| FDA Panel: | Orthopedic |
| FDA Classification: | Class II |
| 510(k) Owner/Submitter: | Gexfix International Corporation<br>(previously registered as Gexfix USA Inc.)<br>55 NE Avenue, Suite 501<br>Boca Raton, FL 33432<br>USA |
| FDA Registration Number: | 3005112659 |
| Device Manufacturer: | Gexfix S.A.<br>Avenue de la Praille 50<br>CH-1227 Carouge<br>Switzerland |
| | Tel: +41 22 308 16 70<br>Fax: +41 22 308 16 79 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Gexfix. The logo consists of a blue circle with a white "G" shape inside. Below the circle, the word "Gexfix" is written in black, bold letters. The logo appears to be for a company or organization named Gexfix.
| FDA Registration Number: | 3005157180 |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Correspondent: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy |
| | Tel: +39 06 578 2665<br>Fax: +39 06 574 3786<br>Email: rgray@donawa.com |
| Intended Use / Indications for Use: | 1) Bone fracture fixation<br>2) Osteotomy<br>3) Arthrodesis<br>4) Correction of deformity<br>5) Revision procedures where other devices have been unsuccessful<br>6) Bone reconstruction procedures |
#### Device Description:
The use of External Fixation devices provides surgeons with a means of bone fixation for the management of fractures and reconstructive surgery.
The Gexfix devices, in common with similar devices from other manufacturers, are intended only to assist during the healing process, and so are not intended to replace normal bone structures.
The Gexfix External Fixation devices are intended use a physical mechanism of action to hold bones in positions that aid stability during healing, thus facilitating the normal healing process. Some elements of the system are implantable during the healing process, with the main structural support being provided by external mechanical elements.
the predicate device, as cleared under K052605.
The Gexfix range includes external fixation devices for use in a variety of orthopedic applications. They are supplied non-sterile for single use, although some accessory items can be reused. Users are instructed to sterilize the components before use, using a sterilization cycle validated to achieve a sterilization Assurance Level (SAL) of at least 10-6.
The Gexfix range of external fixation devices comprises three main elements, these being:
- Invasive components
- . External components
- Ancillary components
This submission covers new components of the Gexfix system that are being added to the range of all three of the above types of component when compared with the predicate device. Some new components have an extended range of available dimensions, while other new components are
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image features the logo of Gexfix. The logo consists of a blue circle with a stylized letter 'G' inside. The 'G' is also blue and has a break in the upper left quadrant of the circle. Below the circle is the word 'Gexfix' in a bold, black font.
additions to the range, to fulfil user needs for greater variety in the range of surgical applications, such as the introductions of components to allow external fixation of bones in the fingers, feet and shoulder.
Biocompatibility of patient-contacting materials is assured by using materials that meet applicable standards (e.g. for stainless steel and titanium alloy components) or by using materials which have received previous 510(k) clearance (e.g. PEEK).
#### Nonclinical testing:
Testing of the new components has been conducted on the basis of compliance with ASTM F1541-02(2011), 'Standard Specification and Test Methods for External Skeletal Fixation Devices'.
Testing has included, as appropriate to the constructs concerned, mechanical testing in:
- . Bending
- Traction
- Compression
- Sliding
- Torsion
- . Pin push-out
Tests were carried out on the following components/constructs:
- Digifix and mini-pins
- . K-wires
- Lateral entry connections
- Shoulder osteosynthesis device (DOS) ●
- External fixation
- Minibone
- Gex-foot
- Caslau bar
- Pins
- Implantable elements
All tests resulted in the components being tested meeting predetermined acceptance criteria.
#### Labeling
MR Environment labeling has been introduced in accordance with ASTM F2503–13, 'Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment'.
Unique Device Identification labeling has also been introduced.
#### Substantial equivalence:
The predicate device for comparison with Gexfix External Fixation is the original version of the device, as cleared by FDA in K052605:
Predicate Device:.......................Gexfix External Fixation Sponsor:...................................................................................................................................................................... 510(k) Number: ......................K052605 Clearance Date: .....................................10 November 2005 FDA Product Code: .......................KTT Classification Name: .........Single/multiple component metallic bone fixation appliances and accessories
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Gexfix. The logo features a stylized letter 'G' in a blue circle. Below the circle, the word 'Gexfix' is written in a bold, sans-serif font.
#### Regulation No:...............................21 CFR 888.3030
The subject device and the predicate devices have many identical or similar properties or features and share identical intended uses and indications for use.
The fundamental technology of the original and modified device is unchanged. Both the original and extended range of components in the Gexfix External Fixation comprise various elements that can be assembled by the orthopedic surgeon to help treat a wide variety of bone fractures and associated conditions. The range includes elements which are invasive, including those for temporary implantation; elements which are attached to the invasive components but which themselves are external to the patient; and elements that are required in order to either assemble the invasive and external components together, or to aid in the temporary implantation process.
The technical specifications are identical between the two devices except for the following:
Invasive components:
- The introduction of titanium (Ti6Al4V) versions of components alongside the original stainless steel versions
- . An extension of the range of available dimensions of certain components
### External components:
- The introduction of Lateral Entry Connection components ●
- . The introduction of the Shoulder Osteosynthesis Device (DOS)
- The introduction of Digifix
- The introduction of Adynex
- The introduction of Gex-finger
- The introduction of Caslau Bar
- The introduction of Gex-foot
- The introduction of Minibone
- The introduction of titanium (Ti6Al4V) versions of certain components
- An extension of the range of available dimensions of certain components
#### Ancillary components:
- The introduction of two new components: Allen Key, Universal Key, K-wire Finger Implantation ● Guide
- An extension of the range of Trays available ●
The original device and the subject device interact with the patient in exactly the same ways.
The original device and the subject device interact with other devices in exactly the same way.
These differences have no significant effect on the safety or effectiveness of the subject device.
#### Summary:
All of the new range additions detailed in this submission have been reviewed and tested, and none have been identified as having any significant effect on the safety and effectiveness of the device when compared with the predicate.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Gexfix. The logo consists of a blue circle with a white horizontal line through the middle, and a white "G" inside the circle. The text "Gexfix" is written in black below the circle.
### Substantial Equivalence Conclusion:
Based on the information contained within this submission, it is concluded that the components that make up the extension to the range of components in the Gexfix External Fixation device are substantially equivalent to the identified predicate device components cleared under K052605, which are already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.