Cardinal Health Sterile Polyisoprene Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating
K160875 · Cardinal Health, Inc. · KGO · Jul 11, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160875
Device Name
Cardinal Health Sterile Polyisoprene Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating
Applicant
Cardinal Health, Inc.
Product Code
KGO · General, Plastic Surgery
Decision Date
Jul 11, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
A powder-free sterile surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Device Story
Disposable, sterile, powder-free orthopedic surgical glove; manufactured from synthetic polyisoprene rubber; features hydrogel polymer coating, anti-slip finish, independent thumb, and mechanically locking cuffs to prevent roll down. Used by operating room personnel (surgeons, nurses, technicians) during orthopedic, trauma, and reconstructive procedures. Provides a physical barrier to protect surgical wounds from contamination. Single-use only.
Clinical Evidence
Bench testing only. No clinical data required. Testing included biocompatibility (ISO 10993-10), physical properties, dimensions, freedom from holes (ASTM D5151/21 CFR 800.20), and powder residual (ASTM D6124), all meeting ASTM D3577 standards.
Technological Characteristics
Synthetic polyisoprene rubber; hydrogel polymer coating; hand-specific design with independent thumb and beaded cuff. Meets ASTM D3577 (physical properties, dimensions, powder residual) and 21 CFR 800.20 (freedom from holes). Biocompatibility per ISO 10993-10. Sterile, single-use.
Indications for Use
Indicated for use by operating room personnel as a protective barrier for surgical wounds during orthopedic, trauma, and reconstructive surgical procedures.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Sterile Latex Powder-Free Surgical Orthopedic Gloves with Hydrogel Coating (K151778)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 11, 2016
Cardinal Health, Inc. Ms. Megan Middaugh Manager, Regulatory Affairs 1500 Waukegan Rd Waukegan, IL 60085
Re: K160875
Trade/Device Name: Cardinal Health™ Sterile Polyisoprene Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: June 9, 2016 Received: June 13, 2016
Dear Ms. Middaugh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160875
Device Name
Cardinal Health™ Sterile Polyisoprene Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating
Indications for Use (Describe)
A powder-free sterile surgeon's glove is a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo for Cardinal Health. The logo consists of a stylized red graphic above the company name. The graphic is composed of several curved lines that converge to a point, resembling a bird in flight. The text "CardinalHealth" is written in a bold, sans-serif font.
Cardinal Health 1500 Waukegan Road Waukegan, IL 60085 cardinalhealth.com
Page _1_ of _5_
### 510(k) SUMMARY
### Cardinal Health™ Sterile Polyisoprene Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating
| Manufacturer: | Cardinal Health 200, LLC<br>1500 Waukegan Road<br>Waukegan, IL 60085 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Affairs Contact: | Megan Middaugh<br>1500 Waukegan Road<br>Waukegan, IL 60085 |
| Telephone Number: | (847) 887-6812 |
| Fax Number: | (847) 887-2461 |
| Date Summary Prepared: | June 9, 2016 |
| Product Trade Name: | Cardinal Health™ Sterile Polyisoprene Powder-Free Orthopedic<br>Surgical Gloves with Hydrogel Coating |
| Common Name: | Orthopedic Surgeon's Gloves |
| Classification Name: | Surgeon's Gloves |
| Classification Panel: | General and Plastic Surgery |
| Regulation: | 21 CFR 878.4460 |
| Product Code: | KGO |
| Predicate Devices: | 1) Esteem Ortho Polyisoprene Powder-Free Surgical Sterile<br>Gloves previously cleared under K092304 on September 15,<br>2009<br><br>2) Sterile Latex Powder-Free Surgical Orthopedic Gloves with<br>Hydrogel Coating cleared under K151778 on March 24, 2016 |
| Reason for 510(k)<br>Submission: | New device |
| Device Description: | The proposed device is a disposable device. It is not made with<br>natural rubber latex. Instead, the gloves are formulated using<br>polyisoprene, which is a synthetic rubber latex, and is brown in<br>color. The glove is coated with hydrogel polymer coating.<br><br>The gloves are manufactured using molds that feature anti-slip<br>finish, independent thumb and mechanically locking cuffs to help<br>prevent cuff roll down. They are offered powder-free and sterile.<br>This glove is for single use only.<br><br>This glove is suitable for use as a specialty surgical glove intended<br>for orthopedic procedures or other procedures where a thicker<br>barrier may be desired such as trauma and reconstructive surgery |
| Intended Use: | A powder-free sterile surgeon's glove is a disposable device<br>made of synthetic rubber intended to be worn by operating room<br>personnel to protect a surgical wound from contamination. |
| Substantial Equivalence: | Cardinal Health™ Sterile Polyisoprene Powder-Free Orthopedic<br>Surgical Gloves with Hydrogel Coating are substantially<br>equivalent to Esteem Ortho Polyisoprene Powder-Free Surgical<br>Sterile Gloves and the Sterile Latex Powder-Free Surgical<br>Orthopedic Gloves with Hydrogel Coating in regards to intended<br>use, sizes, physical characteristics, design and product features. All<br>three gloves are intended for use as specialty orthopedic surgeon's<br>gloves |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cardinal Health. The logo consists of a stylized red bird-like symbol above the words "CardinalHealth" in a bold, black font. The bird symbol is made up of several curved lines that resemble feathers or wings.
Page _2_ of __5_
## 510(K) SUMMARY (CONT'D)
{5}------------------------------------------------
# Page _3_ of __5_
## Table 1: Summary of Technological Characteristics
| Summary of the technological characteristics of the device compared to the predicate devices | | | |
|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | New Device<br>Cardinal Health™ Sterile Polyisoprene Powder-Free<br>Orthopedic Surgical Gloves with Hydrogel Coating | Predicate Device<br>Esteem Ortho Polyisoprene Powder-Free<br>Surgical Sterile Gloves<br>(K092304) | Predicate Device<br>Sterile Latex Powder-Free Surgical<br>Orthopedic Gloves with Hydrogel Coating<br>(K151778) |
| Material Composition | Synthetic Polyisoprene coated with Hydrogel polymer | Synthetic Polyisoprene | Natural Rubber Latex coated with Hydrogel |
| Design | Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated<br>Polymer Coating | Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated | Hand Specific<br>Independent Thumb<br>Beaded Cuff<br>Lubricated<br>Polymer Coating |
| Intended Use /<br>Indication for Use | Powder-Free Orthopedic Surgeon's Glove | Powder-Free Orthopedic Surgeon's Glove | Powder-Free Orthopedic Surgeon's Glove |
| Label Claims | Sterile<br>Powder-free<br>Polyisoprene Surgical Glove with Hydrogel Coating<br>Not made with natural rubber latex<br>Single Use Only<br>Orthopedic<br>AQL 0.65 (Freedom from Holes per ASTM D5151) | Sterile<br>Powder-free<br>Not made with natural rubber latex<br>For Single Use Only<br>Orthopedic | Sterile<br>Powder-free<br>Latex Surgical Glove with Hydrogel<br>Coating<br>Single Use Only<br>Orthopedic<br>AQL 0.65 (Freedom from Holes per ASTM<br>D5151) |
| Dimensions &<br>Physical Properties | Meets ASTM D3577 | Meets ASTM D3577 | Meets ASTM D3577 |
| Freedom from Holes | Meets 21CFR 800.20 & ASTM D3577 requirements<br>of AQL 1.5 | Meets 21CFR 800.20 & ASTM D3577<br>requirements of AQL 1.5 | Meets 21CFR 800.20 & ASTM D3577<br>requirements of AQL 1.5 |
| Powder Residual | Meets requirements of ≤2.0 mg/glove for Powder-<br>Free designation per ASTM D3577 | Meets requirements of ≤2.0 mg/glove for<br>Powder-Free designation per ASTM D3577 | Meets requirements of ≤2.0 mg/glove for<br>Powder-Free designation per ASTM D3577 |
| Biocompatibility<br>(Irritation, ISO<br>10993-0:2010;<br>Sensitization, ISO<br>10993-10: 2010) | Non-Irritating, under the conditions of the study<br>Non-sensitizing, under the conditions of the study | Non-Irritating, under the conditions of the study<br>Non-sensitizing, under the conditions of the<br>study | Non-Irritating, under the conditions of the<br>study<br>Non-sensitizing, under the conditions of the<br>study |
{6}------------------------------------------------
### Table 2: Summary of Non-Clinical Tests
| PERFORMANCE DATA | | |
|-----------------------------------------------------------------------------------------|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION<br>OF SUBSTANTIAL EQUIVALENCE | | |
| Performance Test Summary-New Device | | |
| Characteristic | Standard/Test/FDA<br>Guidance | Results Summary |
| Biocompatibility: | | |
| Primary Skin<br>Irritation | ISO 10993-10 | Non-Irritating, under the conditions of the study |
| Guinea Pig<br>Maximization | ISO 10993-10 | Non-sensitizing, under the conditions of the study |
| Physical Characteristics: | | |
| Dimensions | ASTM D3577 | Meets requirements |
| Physical Properties | ASTM D3577 | Meet requirements for synthetic surgical<br>gloves |
| Freedom from Holes | 21 CFR 800.20 &<br>ASTM D3577 | Tested in accordance with ASTM D 5151<br>and meets 21CFR 800.20 & ASTM<br>D3577 requirements of AQL 1.5 |
| Powder Residual | ASTM D3577 tested<br>using ASTM<br>standard D6124 | Gloves meet powder level requirements for<br>"Powder-Free" designation per ASTM<br>D3577. Results generated values < 2mg of<br>residual powder per glove. |
{7}------------------------------------------------
| Comparative Performance Information Summary | | | | |
|---------------------------------------------|----------------------------|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Requirement | New Device<br>Cardinal HealthTM<br>Sterile Polyisoprene<br>Powder-Free<br>Orthopedic Surgical<br>Gloves with<br>Hydrogel Coating | Predicate Device<br>Esteem Ortho<br>Polyisoprene<br>Powder-Free<br>Surgical Sterile<br>Gloves (K092304) | Predicate Device<br>Sterile Latex<br>Powder-Free<br>Surgical Orthopedic<br>Gloves with<br>Hydrogel Coating<br>(K151778) |
| Biocompatibility: | ISO 10993-1 | Meets<br>requirements | Meets<br>requirements | Meets<br>requirements |
| Primary Skin<br>Irritation | ISO 10993-10 | Non-Irritating,<br>under the<br>conditions of the<br>study | Non-Irritating,<br>under the<br>conditions of the<br>study | Non-Irritating,<br>under the<br>conditions of the<br>study |
| Guinea Pig<br>Maximization | ISO 10993-10 | Non-sensitizing,<br>under the<br>conditions of the<br>study | Non-sensitizing,<br>under the<br>conditions of the<br>study | Non-sensitizing,<br>under the<br>conditions of the<br>study |
| Dimensions | ASTM D3577 | Meets<br>requirements | Meets<br>requirements | Meets<br>requirements |
| Physical<br>Properties | ASTM D3577 | Meets<br>requirements | Meets<br>requirements | Meets<br>requirements |
| Freedom from<br>Holes | 21CFR800.20,<br>ASTM D3577 | Meets<br>requirements | Meets<br>requirements | Meets<br>requirements |
| Powder Residual | ASTM D3577 | Meets<br>requirements | Meets<br>requirements | Meets<br>requirements |
### Table 3: Summary of Comparative Performance
### SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE AND/OR OF CLINICAL INFORMATION
Clinical data is not required.
### CONCLUSIONS DRAWN FROM NON-CLINICAL DATA
Non-clinical data demonstrates Cardinal Health™ Sterile Polyisoprene Powder-Free Orthopedic Surgical Gloves with Hydrogel Coating meet the technological characteristics of ASTM D3577 standard, and are as safe, as effective, and performs as well as the legally marketed devices identified in this summary.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.