INTRAFIX(TM) Advance Tibial Fastener System

K160804 · Medos International SARL · MAI · Jul 19, 2016 · Orthopedic

Device Facts

Record IDK160804
Device NameINTRAFIX(TM) Advance Tibial Fastener System
ApplicantMedos International SARL
Product CodeMAI · Orthopedic
Decision DateJul 19, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DePuy Mitek INTRAFIX™ Advance Tibial Sheath and Screw System is indicated for fixation of tissue including ligament, or tendon to bone during cruciate ligament reconstruction.

Device Story

Implant system for cruciate ligament reconstruction; secures soft tissue (ligament/tendon) to bone. System comprises tibial sheaths and screws. Sheaths available in absorbable BR or non-absorbable polypropylene; screws available in absorbable BR or non-absorbable PEEK. Supplied sterile for single-use. Used by surgeons in clinical settings. Provides mechanical fixation to facilitate healing. Performance verified via pull-out testing and insertion/failure torque assessments.

Clinical Evidence

Bench testing only. Verification activities included pull-out testing, insertion/failure torque testing, and bacterial endotoxin testing. No clinical data provided.

Technological Characteristics

Implant system consisting of tibial sheaths and screws. Materials: absorbable BR, non-absorbable PEEK, and non-absorbable Polypropylene. Supplied sterile. Mechanical fixation principle.

Indications for Use

Indicated for fixation of tissue, including ligament or tendon, to bone during cruciate ligament reconstruction.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, representing the department's focus on people and health. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 19, 2016 Medos International SARL % Kristine Christo Director. Regulatory Affairs DePuy Mitek, Inc. a Johnson & Johnson Company 325 Paramount Drive Raynham, Massachusetts 02767 Re: K160804 Trade/Device Name: INTRAFIX™ Advance Tibial Fastener System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI, MBI Dated: June 13, 2016 Received: June 14, 2016 Dear Ms. Christo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K160804 Device Name INTRAFIXTM Advance Tibial Fastener System Indications for Use (Describe) The DePuy Mitek INTRAFIX™ Advance Tibial Sheath and Screw System is indicated for fixation of tissue including ligament, or tendon to bone during cruciate ligament reconstruction. Type of Use (Select one or both, as applicable) | <div> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |--------------------------------------------------------------------------------| | <div> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## SECTION 2 - 510(k) SUMMARY INTRAFIX™ Advance Tibial Fastener System | Submitter's<br>Name and<br>Address | Medos International SARL<br>Chemin-Blanc 38, Le Locle Neuchatel<br>CH 2400, Switzerland<br>Date Prepared: March 21, 2016 | |------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Kristine Christo<br>Director, Regulatory Affairs<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA<br>Telephone: 508-828-3359<br>Facsimile: 508-977-6911<br>e-mail: kchristo@its.jnj.com | | Name of<br>Medical Device | Proprietary Name: INTRAFIXTM Advance Tibial Fastener System | | | Classification Name: a) Single/multiple component metallic bone fixation<br>appliances and accessories<br>b) Smooth or threaded metallic bone fixation fasteners | | | Common Name: Fastener, fixation, soft tissue | | Substantial<br>Equivalence | The INTRAFIXTM Advance Tibial Fastener System is substantially equivalent to:<br>■ K123362 MILAGRO® ADVANCE Interference Screw<br><br>Reference devices:<br>■ K130539 Healix AdvanceTM Knotless PEEK Anchor<br>■ K102443 Intrafix Tibial Sheath | | Device<br>Classification | INTRAFIXTM Advance BR Screw and Sheath:<br>■ Single/multiple component metallic bone fixation appliances and<br>accessories, classified as Class II, product code MAI, regulated under<br>21 CFR 888.3030.<br><br>INTRAFIXTM Advance PEEK Screw and INTRAFIXTM Advance PP Sheath:<br>■ Smooth or threaded metallic bone fixation fastener, classified as Class<br>II, product code MBI, regulated under 21 CFR 888.3040. | | Device<br>Description | The INTRAFIX™ Advance Tibial Fastener System is an implant system used for<br>fixation of tissue including ligament, or tendon to bone during cruciate ligament<br>reconstruction. The INTRAFIX™ Advance Sheaths are available in absorbable BR<br>and non-absorbable Polypropylene materials. The INTRAFIX™ Advance Screws are<br>available in absorbable BR and non-absorbable PEEK materials. The INTRAFIXTM<br>Advance Tibial Sheaths and Screws are supplied sterile ready to use. | | Technological<br>Characteristics | The proposed INTRAFIX™ Advance Sheaths and Screws are similar to the predicate<br>MILAGRO® ADVANCE Interference Screw (K123362) in that they share the same<br>intended for use, BR material, screw design, sterilization method, and shelf life. The<br>proposed INTRAFIX™ Advance Screws are similar to the Healix Advance™<br>Knotless PEEK Anchor (K130539) in that they share the same PEEK material. The<br>proposed INTRAFIX™ Advance Sheaths are similar to the Intrafix Tibial Sheath<br>(K102443) as they share the same Polypropylene material and have a similar design<br>and function. | | Indications for<br>Use | The DePuy Mitek INTRAFIX™ ADVANCE Tibial Sheath and Screw System is<br>indicated for fixation of tissue including ligament, or tendon to bone during cruciate<br>ligament reconstruction. | | Non clinical<br>Testing | Verification activities were performed on the implant and / or its predicate. Testing<br>assessments include pull out testing, and insertion / failure torque. | | Safety and<br>Performance | Results of performance testing have demonstrated that the proposed devices are<br>suitable for their intended use. | | | The proposed device also met requirement of bacterial endotoxin testing. | | | Based on similarities in the indications for use, technological characteristics, and<br>performance in comparison to the predicate devices, the proposed INTRAFIX™<br>Advance Tibial Fastener System has shown to be substantially equivalent to the<br>predicate device under the Federal Food, Drug and Cosmetic Act. | {4}------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...