AmorChem Titanium Porous Fixation Device

K160791 · Amorchem Holdings, Inc. · HWC · Nov 15, 2016 · Orthopedic

Device Facts

Record IDK160791
Device NameAmorChem Titanium Porous Fixation Device
ApplicantAmorchem Holdings, Inc.
Product CodeHWC · Orthopedic
Decision DateNov 15, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AmorChem 3.8 mm Headless Porous Titanium Compression Screw is used for fixation of bone fractures, bone reconstruction, and fixation of osteotomies.

Device Story

AmorChem Porous Titanium Fixation Device is 3.8mm diameter, 15mm long, variable pitch, fully threaded, cannulated, headless compression screw; fabricated from 40% porous titanium. Device inserted into pre-drilled bone hole, guided by k-wire; headless design allows flush seating. Intended for use by orthopedic surgeons in clinical settings for bone fracture fixation, reconstruction, and osteotomies. Porous structure facilitates biological fixation via bone ingrowth. Output is mechanical stabilization of bone segments; clinical decision-making relies on surgeon assessment of fracture stability and bone healing. Benefits include secure fixation and potential for enhanced biological integration compared to solid metallic screws.

Clinical Evidence

No human clinical trials. Evidence includes bench testing and animal studies. Biomechanical testing (torsion, torque-in/out, holding strength, 3-point static/fatigue bend) per ASTM F543, F2502, F1264, F1541 showed superior holding power and strength vs. k-wire predicate. Biocompatibility assessed via ISO 10993-1, -5, -6, -12, -17, -18, and USP tests; chemical analysis showed leachables below detectable levels. Ovine model (6 and 12 weeks) confirmed biological fixation via interconnected porous structure.

Technological Characteristics

Material: 40% porous titanium. Design: 3.8mm diameter, 15mm length, variable pitch, fully threaded, cannulated, headless compression screw. Drive: Hexagonal drive along entire length. Standards: ASTM F543 (torsion), ASTM F2502, ASTM F1264, ASTM F1541 (bend testing), ISO 10993 (biocompatibility). Single-use, provided non-sterile.

Indications for Use

Indicated for fixation of bone fractures, bone reconstruction, and fixation of osteotomies in patients requiring orthopedic bone fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 15, 2016 Amorchem Holdings, Inc. % Robert Poggie, Ph.D. President, BioVera, Inc. 65 Promenade Saint Louis Notre Dame De Lile Perrot, Québec, J7V7P2 Canada Re: K160791 Trade/Device Name: Amorchem Porous Titanium Fixation Device Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC, HTY Dated: October 12, 2016 Received: October 13, 2016 Dear Dr. Poggie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K160791 Device Name AmorChem Porous Titanium Fixation Device Indications for Use (Describe) The AmorChem 3.8 mm Headless Porous Titanium Compression Screw is used for fixation of bone fractures, bone reconstruction, and fixation of osteotomies. Type of Use (Select one or both, as applicable) | <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |--------------------------------------------------------------------------------------| | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | | X Prescription Use (Part 21 CFR 801 Subpart D) __ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY # AmorChem Porous Titanium Fixation Device In accordance with 21 CFR 807.92, the following information is a summary of safety and effectiveness for the AmorChem Porous Titanium Fixation Device. ### A. SUBMITTERS INFORMATION | Submitter Name: | BioVera, Inc. | |---------------------|----------------------------------------------------------------------------------| | Submitter Address: | 65 Promenade Saint-Louis, Notre-Dame-De-L'Ile-Perrot, Québec,<br>J7V 7P2, CANADA | | Contact Person: | Robert A Poggie, PhD | | Title: | President | | Phone Number: | (514) 901-0796; (514) 349-7226 | | Fax Number: | (514) 901-0796 | | Date of Submission: | November 14, 2016 | ### B. DEVICE IDENTIFICATION & MANUFACTURER | Manufacturer Name: | AmorChem Holdings, Inc. | |-----------------------|--------------------------------------------------------| | Manufacturer Address: | 1, Westmont Square<br>Montreal, Quebec, H3Z 2P9 Canada | | Registration Number: | Not registered | | Contact Name: | Kevin McBride, MsC, PhD | | Title: | Vice President, Research | | Device Trade Name: | AmorChem Porous Titanium Fixation Device | | Device Common Name: | Metallic bone screw | | Classification Name: | Smooth or threaded metallic bone fixation fastener | | Classification Code: | HWC, HTY ">– Class II | | Classification Panel: | Orthopedic | | Regulation Number: | 21 CFR section 888.3040 | #### C1. PREDICATE DEVICES | K143618 | Zimmer Kirschner Wires and Steinmann pins | |----------------------|-------------------------------------------| | Primary predicate | | | K930834 | Acutract compression screw | | Additional predicate | | | K102528 | N-Force Fixation System | | Additional predicate | | {4}------------------------------------------------ ### C2. REFERENCE DEVICES | K071616 | The ARC Surgical BIOTRAK Pin System | |---------|--------------------------------------------------------| | K041189 | The Arthrex Trimlt family | | K140879 | Smith & Nephew BIOSURE HEALICOIL PK Interference Screw | ### D. DEVICE DESCRIPTION The AmorChem Porous Titanium Fixation Device is a 3.8mm, 15mm long variable pitch, fully threaded, cannulated, and headless compression screw. The AmorChem screw is designed to be inserted into a pre-drilled hole that can be guided with a standard k-wire or Allen-drive k-wire. The headless design is intended to be seated flush with the surface of bone. The device is 40% titanium with interconnected porosity. The AmorChem device is provided clean, not sterile, and is single-use only. Materials: Porous titanium. ## E. INDICATIONS FOR USE The AmorChem 3.8 mm Headless Porous Titanium Compression Screw is used for fixation of bone fractures, bone reconstruction, and fixation of osteotomies. # F. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE The subject AmorChem Porous Titanium Fixation Device is a 3.8mm, 15mm long variable pitch, fully threaded, cannulated, and headless compression screw made entirely of 40% porous titanium. The characteristics of variable pitch thread, tapered design, through-cannulation, and size of the AmorChem device are the same as the predicate Acumed device. The main technological differences are the porous titanium and hexagonal drive along the entire length of the AmorChem device; the Acumed device is fabricated from titanium alloy with the hex drive confined to the proximal end of the screw. The porous titanium material is intended to allow biological fixation. In vitro testing per ASTM F543, 3-point static and fatigue bend testing, ISO 10993 evaluation of cytotoxicity, chemistry and associated risk analysis, and evaluation of biological fixation and biocompatibility in ovine models of gap healing, osteonecrosis, and cortical bone showed the AmorChem device to be substantially equivalent to the predicate devices. ## G. PERFORMANCE DATA Characterization of the AmorChem Porous Titanium Fixation Device was performed per in-vitro and in-vivo laboratory testing, and assessment of biocompatibility per the chemical characterization and risk analysis route described in ISO 10993-1 and recommended by FDA in pre-submission Q150543. More specifically: - . Biomechanical evaluation of the AmorChem device was evaluated for torsional properties, torque-in, torque-out, and holding strength per consensus testing standards ASTM F 543 and F 2502. The results of these tests showed the AmorChem device to {5}------------------------------------------------ possess similar torque-in and -out properties as the Acumed predicate device, superior holding power to the k-wire primary-predicate device. - Strength of the subject AmorChem device and the k-wire primary-predicate device was . evaluated in static and fatigue 3-point bend testing using ASTM F1264 and ASTM F1541 as guides. The AmorChem device was found to be substantially stronger in 3-point bending than the primary-predicate k-wire device. - Biocompatibility was assessed by quantifying the presence of leachables / extractables and toxicological risk assessment by NAMSA per ISO 10993-1, -5, -6, -12, -17, and -18, ISO 14971: 2012, FDA G-95-1 Bluebook quidance, and USP Physiochemical Tests. The chemical analysis showed the levels of residual compounds to be below detectable levels, or presenting no risk of safety. ISO 10993-6 evaluation of local tissue effects by AccelLAB of the subject device showed no adverse effects. The porous titanium material and subject device were found to be biocompatible for permanent implantation. - Ovine modeling of biological fixation and biocompatibility at 6- and 12-weeks post-op . showed the subject AmorChem Porous Titanium Fixation Device to be biologically fixed via the interconnected porous structure and cannula of the device. The results of the performance testing indicate the AmorChem Porous Titanium Fixation Device to be substantially equivalent to the identified predicate devices. ## H. CONCLUSION The AmorChem Porous Titanium Fixation Device is substantially equivalent to the predicate devices cited in this 510(k) application.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...