← Product Code [FAJ](/productcode/FAJ) · K160766

# Ambu USR, Ambu M (K160766)

_Ambu A/S · FAJ · Aug 17, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160766

## Device Facts

- **Applicant:** Ambu A/S
- **Product Code:** [FAJ](/productcode/FAJ.md)
- **Decision Date:** Aug 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Ambu® USR is a sterile, single use flexible cystoscope designed for removal of double loop ureteral stents accessible in the bladder via a urethral insertion in adults. Ambu® USR has been designed to be used with the reusable Ambu® M to visualize the observations obtained by Ambu® USR.

## Device Story

Ambu® USR is a sterile, single-use flexible cystoscope with an integrated rat-tooth grasper; used for removal of double loop ureteral stents in adults. Device features a maneuverable tip, flexible insertion cord, and distal camera/LED light source. Operated by a clinician, the cystoscope connects to the reusable Ambu® M monitor. The monitor displays live video, captures snapshots, and records video. The system facilitates visualization of the bladder and stent during the retrieval procedure. The integrated grasper is controlled by the user to engage and remove the stent. The device is intended for use in clinical settings where cystoscopy is performed.

## Clinical Evidence

No clinical data. Bench testing only. Testing included optical performance (ISO 8600-1, 3, 4), mechanical properties (insertion cord, bending section, grasper functionality/pull force), shelf life, packaging integrity, biocompatibility (ISO 10993-1, 5, 10), cleaning/disinfection validation for the monitor, and electrical/EMC safety (IEC 60601-1, 60601-2-18, 60601-1-2). All tests passed.

## Technological Characteristics

Flexible cystoscope with maneuverable tip, integrated stainless steel rat-tooth grasper, and distal LED/camera. Luer lock connector for irrigation/aspiration. Sterilized by Ethylene Oxide. Reusable monitor with TFT-screen, battery or AC power. Standards: ISO 8600-1/3/4 (optics), ISO 594-1 (Luer), ISO 10993-1/5/10 (biocompatibility), IEC 60601-1/2-18/1-2 (electrical/EMC).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- VISERA Cystovideoscope CYF-V/VA and CYF-V2/VA2 ([K021074](/device/K021074.md) and [K133538](/device/K133538.md))
- CST-5000 Flexible Video Cystoscope ([K072180](/device/K072180.md))
- Ambu® aScope 3™ 5.0/2.2 and Ambu® aView ([K130845](/device/K130845.md))
- Olympus FG-4L-1 ([K955066](/device/K955066.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus-like symbol with three figures in profile, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17, 2016

Ambu A/S % Sanjay Parikh Director, QA/RA Ambu, Inc. 6230 Old Dobbin Lane, Suite 250 Columbia, MD 21045

Re: K160766 Trade/Device Name: Ambu® USR, Ambu® M Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FAJ, OCZ Dated: July 11, 2016 Received: July 12, 2016

Dear Sanjay Parikh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K160766

Device Name Ambu® USR Ambu® M

Ambu® USR is a sterile, single use flexible cystoscope designed for removal of double loop ureteral stents accessible in the bladder via a urethral insertion in adults.

Ambu® USR has been designed to be used with the reusable Ambu® M to visualize the observations obtained by Ambu® USR.

| Type of Use (Select one or both, as applicable)                                                |                                                                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Section 5 510(k) Summary

This 510(k) summary has been prepared in accordance with 21 CFR 807.87(h) and the content and format of the 510(k) summery has been prepared in accordance with 21 CFR 807.92.

| Submitter                                                                | Ambu A/S<br>Baltorpbakken 13<br>DK-2750 Ballerup<br>Denmark<br>Tel.: +45 7225 2000<br>Fax.: +45 7225 2050                                                                                          |                                                      |                     |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------|---------------------|
| Contact Person                                                           | Name: Camilla Wismar<br>Job Title: Senior Regulatory Affairs Professional<br>Address: Ambu A/S, Baltorpbakken 13, DK-2750 Ballerup<br>Telephone number: +45 7225 2208<br>Fax number: +45 7225 2050 |                                                      |                     |
| Date Summary<br>Prepared                                                 | August 17, 2016                                                                                                                                                                                    |                                                      |                     |
| Device Trade<br>Name                                                     | Ambu® USR<br>Ambu® M                                                                                                                                                                               |                                                      |                     |
| Device Common<br>Name                                                    | Flexible Cystoscope                                                                                                                                                                                |                                                      |                     |
| Device<br>Classification                                                 | Cystoscope and Accessories, Flexible/Rigid<br>Product Codes: FAJ<br>Endoscopic grasping/cutting instrument, non-powered<br>Product Codes: OCZ<br>21 CFR 876.1500<br>Class II                       |                                                      |                     |
| Legally Marketed devices to which the device is substantially equivalent | Manufacturer                                                                                                                                                                                       | Trade Name                                           | 510k number         |
| A:                                                                       | Olympus Corporation                                                                                                                                                                                | VISERA Cystovideoscope<br>CYF-V/VA<br>and CYF-V2/VA2 | K021074 and K133538 |
| B:                                                                       | Vision Sciences                                                                                                                                                                                    | CST-5000 Flexible Video Cystoscope                   | K072180             |
| C+E:                                                                     | Ambu A/S                                                                                                                                                                                           | Ambu® aScope 3™ 5.0/2.2<br>and Ambu® aView           | K130845             |

{4}------------------------------------------------

D: Olympus FG-4L-1 K955066 Olympus Corporation Grasping Forceps The system consists of Ambu® USR and Ambu® M Monitor. Ambu® USR is Description of the Device a single use flexible cystoscope with an integrated grasper designed for removal of double loop ureteral stents accessible in the bladder via a urethral insertion. Ambu® USR is connected to the reusable Ambu® M Monitor that displays the video image from Ambu® USR. Ambu® USR has the following physical and performance characteristics: Maneuverable tip controlled by the user ● . Integrated grasper controlled by the user . Flexible insertion cord . Camera and LED light source at the distal tip ● Luer lock connector for irrigation and aspiration ● Sterilized by Ethylene Oxide sterilization . For single use Ambu® M Monitor has the following physical and performance characteristics: . Displays the image from Ambu® USR on the screen. Records snapshots or video of image from Ambu® USR. . Indications for Ambu® USR is a sterile, single use flexible cystoscope designed for removal of double loop ureteral stents accessible in the bladder via a Use urethral insertion in adults. Ambu® USR has been designed to be used with the reusable Ambu® M to visualize the observations obtained by Ambu® USR. Ambu® USR is similar to the predicate devices in the following areas: Summary of the technological They are all flexible endoscopes with a maneuverable tip ● characteristics . They all have a handle with a control button giving the operator in comparison to ability to steer the tip of the scope up and down the predicate . They are all video scopes with a camera located in the distal tip devices to provide an image on a separate monitor ● They all provide illumination from the distal tip ● They are all connected to a monitor by a cable Ambu® USR has an integrated grasper. Predicate A, B and C . have a working channel that can be used for graspers or other instrumentation. . Grasper in Ambu® USR and Predicate D are rat tooth designs with two jaws in a V-shape and teeth at the end.

{5}------------------------------------------------

| Grasper in Ambu® USR and Predicate D are made of stainless steel components They all allow for irrigation                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ambu® M Monitor is similar to Predicate E in the following areas:  Both display a live image on a TFT-screen Both supply power to the endoscope via a connection to the monitor Both are powered by a rechargeable battery or by a power supply Both systems are operated either on battery or standard wall outlet Both monitors are portable |                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Performance<br>Data -Bench                                                                                                                                                                                                                                                                                                                     | The following data has been submitted in the premarket notification:<br>Declaration of conformity to the following recognized consensus standards applicable for Ambu® USR:  ISO 8600-1, ISO 8600-3 and ISO 8600-4 Optics and optical instruments – Medical endoscopes and certain accessories. ISO 594-1 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment.              Result: All tests were passed. |
|                                                                                                                                                                                                                                                                                                                                                | Performance test reports to document the following properties of the Ambu® USR:  Length and diameters of insertion cord Bending angle and durability of bending section Image quality during 1 hour operation Irrigation system functionality Grasper functionality and image quality Grasper stent pull force              Result: All tests were passed.                                                                                               |
|                                                                                                                                                                                                                                                                                                                                                | Performance test report to document shelf life of Ambu® USR. Tests were performed on finished, sterilized, shipped and aged products:  Performance test of Ambu® USR Sterile Packaging Integrity              Result: All tests were passed.                                                                                                                                                                                                             |
|                                                                                                                                                                                                                                                                                                                                                | Biocompatibility tests reports to document that Ambu® USR complies with the requirements of ISO 10993-1:  Cytotoxicity (ISO 10993-5) Sensitization (ISO 10993-10) Intracutaneous reactivity test (ISO 10993-10)              Result: All tests were passed.                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                | Performance test report to document the following properties of the Ambu® M monitor (reusable):  Cleaning and Low Level Disinfection validation              Result: The test was passed.                                                                                                                                                                                                                                                                |
|                                                                                                                                                                                                                                                                                                                                                | Test reports that verify the Electromagnetic Compatibility and Electrical Safety:  Electromagnetic Compatibility in compliance with IEC 60601-1-2. Electrical Safety in compliance with IEC 60601-1 and IEC 60601-2-18.                                                                                                                                                                                                                                  |
|                                                                                                                                                                                                                                                                                                                                                | Result: All tests were passed.                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Performance<br>Data – Clinical                                                                                                                                                                                                                                                                                                                 | Not applicable.                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Conclusion                                                                                                                                                                                                                                                                                                                                     | Based on the indication for use, technological characteristics,<br>performance data and comparison to predicate devices it has been<br>concluded that the functionality and intended use of Ambu® USR and<br>Ambu® M is equivalent to the predicate devices.<br><br>It is concluded that Ambu® USR and Ambu® M are as safe and as<br>effective and perform as well as or better than the chosen legally<br>marketed predicate devices.                   |

{6}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K160766](https://fda-staging.innolitics.com/device/K160766)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).