K160761 · Dongwon Medical Co., Ltd. · NEW · Nov 22, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160761
Device Name
DW-1C
Applicant
Dongwon Medical Co., Ltd.
Product Code
NEW · General, Plastic Surgery
Decision Date
Nov 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DW-1 C comprised of PDO is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
Device Story
DW-1C is a synthetic, absorbable, monofilament surgical suture composed of polydioxanone (PDO). The device features bi-directional barbs along the long axis of the monofilament, allowing for tissue approximation without the need for surgical knots. The suture is dyed with D&C Violet No. 2 and is provided sterile via ethylene oxide (EO) gas. It is intended for use by surgeons for soft tissue approximation. The barbs embed into the tissue upon placement, providing holding force. The device degrades or dissolves over time in the body. It is a mechanical implant device used in clinical settings.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993) included cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity (Ames, chromosomal aberration, micronucleus), and 12-week muscle implantation. Performance testing included USP 32/37-NF 32 monographs for diameter, tensile strength, and absorption. In vivo absorption observed at 10-220 days. Holding forces, biodegradation rates, and mechanical properties were evaluated, showing no significant differences from reference samples and meeting all acceptance criteria.
Technological Characteristics
Synthetic absorbable monofilament suture; material: polydioxanone (PDO); bi-directional cog-shaped barbs; EO gas sterilization; USP size 1; implantable; duration of contact >30 days.
Indications for Use
Indicated for soft tissue approximation in patients where absorbable sutures are appropriate.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
Dongwon Medical Co., Ltd. % Mr. Peter Chung Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K160761
Trade/Device Name: Dw-1c Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable Polydioxanone Surgical Suture Regulatory Class: Class II Product Code: NEW Dated: October 17, 2016 Received: October 24, 2016
Dear Mr. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160761
Device Name DW-IC
Indications for Use (Describe)
The DW-1 C comprised of PDO is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
[as required by 807.92(c)]
#### 1. Applicant
- 1) Company : Dongwon Medical Co.,Ltd.
- 2) Address : 6F-601, 79-1, Mongnyeon-ro 153beon-gil, Gwangsan-gu, Gwangju, Korea
- 3) Tel : 82-62-430-6521
- 4) Fax : 82-62-430-6520
- 5) Contact person : Peter Chung, 412-687-3976
- 6) Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
- 7) Submission date : Oct. 17, 2016
#### 2. Device Information
- 1) Trade Name : DW-1C
- 2) Common Name : Absorbable polydioxanone surgical suture
- 3) Classification Name : Suture, Surgical, Absorbable, Polydioxanone
- 4) Product Code : NEW
- 5) Regulation Number : 21 CFR 878.4840
- 6) Class of device : Class II
- 7) Panel : General & Plastic Surgery
## 3. The legally marketed device to which we are claiming equivalence : K130191 MINT
#### 4. Device description :
DW-1C synthetic absorbable PDO suture is made of polydioxanone. The pigment for the violet dye is D&C Violet No.2. The DW-1C is available sterile after ethylene oxide (EO) gas sterilization and degrades or dissolves over time in tissue.
Each dyed (violet) suture has bi-directional barbs along the long axis of the suture monofilament without needle attachment. The DW-1C Synthetic Absorbable PDO suture approximate tissue, without the need to tie surgical knots, by using the opposing barbs on the surface to embed in the tissues after the surgeon precisely places the suture within the tissues.
While the formation of barbs in the DW-1C reduces the tensile strength relative to non-barbed suture material of the same size, tying of knots in non-barbed suture materials also reduce their effective strength. For this reason, the strength of the DW-1C can be compared with USP knot strength of nonbarbed suture and the USP size of DW-1C is 1
#### 5. Indication for Use :
DW-1C comprised of PDO is indicated for use in soft tissue approximation where use of absorbable sutures is appropriate.
#### 6. Technological characteristics
DW-1C, absorbable polydioxanone surgical suture has the same fundamental scientific technology as the MINT, absorbable polydioxanone surgical suture, K130191.
{4}------------------------------------------------
## 7. Comparison of technological characteristics with the predicate device
The comparison of features and operation principles between DW-1C, absorbable polydioxanone surgical suture from Dongwon Medical Co., Ltd., and MINT, absorbable polydioxanone surgical suture from HansBiomed Corporation is listed as follows:
| Proprietary | Submission Device | Predicate Device | Substantially<br>Equivalent or Not<br>Substantially<br>Equivalent |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| 510(k) Number | N/A | K130191 | N/A |
| Common Name | Absorbable polydioxanone<br>surgical suture | Absorbable polydioxanone<br>surgical suture | Substantially<br>Equivalent |
| Trade name | DW-1C | MINT™ | N/A |
| Manufacturer | Dongwon Medical Co., Ltd. | HansBiomed Corporation | N/A |
| Product Classification | Absorbable polydioxanone<br>surgical | Absorbable polydioxanone<br>surgical | Identical |
| Indication for use | DW-1C comprised of PDO is<br>indicated for use in soft tissue<br>approximation where use of<br>absorbable sutures is<br>appropriate. | MINT comprised of PDO is<br>indicated for use in soft<br>tissue approximation<br>where use of absorbable<br>sutures is appropriate. | Identical |
| Raw Material | Polydioxanone suture,<br>Samyang Corporation | Polydioxanone suture,<br>Samyang Corporation | Identical |
| Suture Characteristic | Synthetic Absorbable<br>Monofilament | Synthetic Absorbable<br>Monofilament | Identical |
| Technique of<br>Deployment | Needles are not attached | Needles are not attached | Identical |
| Technological<br>Characteristic | Bi-directional barbs along the<br>long axis of the suture<br>monofilament | Bi-directional barbs along the<br>long axis of the suture<br>monofilament | Identical |
| Material | Polydioxanone | Polydioxanone | Identical |
| Sterilization | EO gas | EO gas | Identical |
| Size (USP) | 1 | 1 | Identical |
| Absorbable | Absorbable | Absorbable | Identical |
| Patient contact | Implant | Implant | Identical |
| Duration of contact | Over 30 Days | Over 30 Days | Identical |
| Braid/Monofilament | Monofilament | Monofilament | Identical |
| Number of barbs per<br>linear length of suture | 85 | 82 | Similar |
| Barb shape | Cog shape | Cog shape | Identical |
| Barb size | 0.567mm | 0.601mm | Similar |
| Barb direction | A section and B section are<br>opposite direction | A section and B section are<br>opposite direction | Identical |
| Pattern of the barbs | bi-directional barbs along the<br>long axis of the suture | bi-directional barbs along the<br>long axis of the suture | Identical |
#### Substantial equivalence summary
The DW-1C has a substantially equivalent intended use as the identified predicate, MINT comprised of PDO (Polydioxanone) manufactured by HansBiomed corporation and is made of polydioxanone intended for soft tissue approximation. The subject and predicate are composed of the same material (PDO) and
{5}------------------------------------------------
have bi-directional barbs, and have the same indications for use.
#### 8. Performance data:
Laboratory testing regarding characteristics was performed on the DW-1C to verify its safety and performance. A biocompatibility assessment was performed on the patient contact materials of DW-1C.
| Test Identification | Method | Criteria | Result PASS/FAIL | Report |
|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Cytotoxicity MEM<br>Elution Method<br>(Cytotoxicity) | ISO10993-5 Tests<br>for Cytotoxicity, in<br>vitro Methods | No evidence<br>of causing cell lysis<br>or toxicity to<br>fibroblast cells | Not Cytotoxic<br>(PASS) | R-C-R-8400 |
| ISO Guinea Pig<br>Maximization<br>Sensitization Test<br>(Sensitization) | ISO10993-10<br>Test for Irritation<br>and sensitization<br>6.2 Maximization<br>sensitization test | No delay in dermal<br>contact sensitization | No evidence SC and<br>CSO extracts causing<br>delayed dermal<br>contact sensitization<br>(PASS) | R-C-R-8400 |
| ISO Acute<br>Intracutaneous<br>Reactivity Study in<br>the Rabbit<br>(Intracutaneous Test) | ISO10993-10<br>Test for Irritation<br>and sensitization<br>5.4 Intracutaneous<br>reactivity test | No irritation | Primary Irritation<br>Index<br>Characterization<br>SC extract = 0.0<br>CSO extract = 0.0<br>(PASS) | R-C-R-8400 |
| ISO Systemic Toxicity<br>Study in Mice<br>(Acute systemic<br>toxicity) | ISO10993-11 Tests<br>for systemic toxicity | No reaction<br>No Mortality during<br>this study or<br>evidence<br>of systemic toxicity | No Mortality or<br>Evidence or<br>Significant Systemic<br>toxicity<br>(PASS) | R-C-R-8400 |
| Bacterial Reverse<br>Mutation study<br>(Ames test) | ISO 10993-3 Test for<br>Genotoxicity,<br>Carcinogenicity and<br>Reproductive<br>Toxicity. | Tester strain char<br>must exhibit<br>sensitivity to<br>crystal violet, UV, no<br>growth biotin plates<br>and growth<br>histatine-biotin<br>plates...number of<br>mutation revertants<br>(including<br>spontaneous<br>reversions)<br>statistically<br>less than negative<br>control results | The Saline and dimethyl<br>sulfoxide test article<br>extracts were considered<br>to be non-mutagenic to<br>Salmonella typhimurium<br>and Escherichia coli<br>tester strains.<br>(PASS) | R-C-R-8400 |
| Genotoxicity : In<br>Vitro Chromosomal<br>Aberration. | EN ISO 10993-3 Test<br>for Genotoxicity,<br>Carcinogenicity and<br>Reproductive<br>Toxicity. | Whether the extract<br>would cause<br>genotoxicity in<br>Chinese Hamster<br>Ovary cells | Not considered genotoxic<br>(Pass) | R-C-R-8400 |
| Mouse Bone Marrow<br>Micronucleus | EN ISO 10993-3 Test<br>for Genotoxicity,<br>Carcinogenicity and<br>Reproductive<br>Toxicity. | No cellular toxicity | Not considered to be<br>genotoxic (Pass) | R-C-R-8400 |
| 4 Week Subchronic<br>Toxicity | ISO10993-11 Tests<br>for systemic toxicity | No signs of<br>behavioral change<br>or toxicity in rats | No significant evidence of<br>systemic toxicity from the<br>subcutaneous | R-C-R-8400 |
| | | | article into rats.(Pass) | |
| ISO Muscle<br>Implantation<br>Study in Rabbits, 12<br>Weeks<br>(Implantation) | ISO 10993-6 Test for<br>Local Effect After<br>Implantation. | No significant<br>macroscopic and<br>microscopic<br>reactions nonirritant<br>compared to USP<br>Negative Control<br>Implant Material | Macroscopic<br>reaction for the test<br>article device was<br>not significant as<br>compared to the<br>USP negative<br>control implant material<br>(PASS) | R-C-R-8400 |
## 1) Biocompatibility test
{6}------------------------------------------------
Based on the above testing results, the DW-1C is biocompatible.
#### 2) Bench (performance) testing
In the following table, the summary of all the bench tests is provided.
| No. | Title | Result summary | File (Protocol/<br>Report) |
|-----|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| 1 | USP 32:2009 Absorbable<br>Surgical Suture | In vivo absorption of absorbable suture were observed<br>at 10, 80, 120, 220days after implantation. Absorption<br>time of MONOSORB was 180-220days | R-C-R-83e |
| 2 | Dimension Test<br>USP 37-NF 32:2014 <861><br>Sutures - Diameter | Acceptance Criteria for this testing ±5%<br>No failed results were performed in any DW-1C during<br>the performing period. | BTR 15-11 |
| 3 | Tensile Strength<br>USP 37-NF 32: 2014<br><881>Tensile Strength | Acceptance Criteria for this testing Not less than 300gf.<br>No failed results were performed in any DW-1C during<br>the performing period. | BTR 15-11 |
| 4 | Holding forces | There was no significant difference between test<br>samples and reference samples | 16-KE-126 |
| 5 | Weights | The weights of test samples in 1st (G1, Week 2), 2nd (G2 ,<br>Week 4), 3rd (G3, Week 6), 4th (G4 , Week 8), 5th (G5,<br>Week 10), 6th (G6, Week 12) and 7th (G7, Week 14)<br>necropsy groups were statistically significantly lower<br>than that of the test samples in before implantation<br>(p<0.001). | 16-KE-126 |
| 6 | Biodegradation<br>(absorption rate) | The biodegradation levels of test samples in 1st, 2nd, 3rd<br>4th, 5th, 6th and 7th necropsy groups were significantly<br>lower than that of the test samples in before<br>implantation (p<0.001). | 16-KE-126 |
| 7 | Mechanical<br>property (Tensile<br>strengths) | The tensile strengths of test samples in 3rd, 4th, 5th and<br>6th necropsy groups were significantly lower than that of<br>test samples in 1st necropsy group (p<0.05, p<0.05,<br>p<0.05 and p<0.01). In case of 7th necropsy groups, the<br>test samples were degraded into small particles, so it<br>was not possible to measure the tensile strengths. | 16-KE-126 |
| 8 | Endotoxin test | Endotoxin concentration of the test sample is less than<br>0.1EU/device. | CT16-074111 |
| 9 | Pyrogen test | There was non-pyrogenicity to the extraction solution. | CT16074112 |
## Safety and performance
Testing was performed per FDA's Class II Special Controls Guidance Document : Surgical Sutures, including testing in accordance with the USP monograph for absorbable sutures, in vitro and in vivo resorption testing, biocompatibility testing in accordance with ISO 10993, and a barb holding strength evaluation.
Holding forces, Weights, Biodegradation (absorption rate), Mechanical property (Tensile strengths), Endotoxin test and Pyrogen test were performed.
No failed results were noted in any product evaluations. The test results satisfied all the acceptance
{7}------------------------------------------------
criteria.
### 9. Conclusion:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification MINT™ comprised of PDO(Polydioxanone) (K130191) concludes that the DW-1C is substantially equivalent to predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.