K160723 · Carestream Health, Inc. · JAK · Aug 30, 2016 · Radiology
Device Facts
Record ID
K160723
Device Name
OnSight 3D Extremity System
Applicant
Carestream Health, Inc.
Product Code
JAK · Radiology
Decision Date
Aug 30, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 or over.
Device Story
OnSight 3D Extremity System is a cone beam computed tomography (CBCT) device for 3D volumetric imaging and 2D projection radiography of extremities (feet, ankles, legs, knees, hands, wrists, arms, elbows). Device acquires 2D projection x-rays via amorphous silicon flat panel detector while x-ray source and detector rotate 216.5 degrees around anatomy; x-ray tube is pulsed during 20-25 second scan. System supports weight-bearing and non-loaded imaging. Used in office-based practices, hospitals, and imaging facilities; operated by healthcare professionals. Software processes projection images to generate 3D anatomical reconstructions. Output used by clinicians for diagnostic procedures. Benefits include diagnostic-quality 3D imaging of extremities in a standalone, point-of-care format.
Clinical Evidence
Clinical reader study performed to evaluate diagnostic capability. Due to unavailability of predicate images, Philips Brilliance CT (K060937) used as reference for 3D images and DRX-1C Detector (K120062) as reference for 2D images. Results indicated diagnostic capability of OnSight 3D Extremity System is statistically equivalent to or better than reference devices.
Technological Characteristics
Cone beam computed tomography (CBCT) system. Amorphous silicon (a-Si:H) flat panel detector. Pulsed x-ray source. Standalone unit. Supports weight-bearing and non-loaded imaging. 216.5-degree rotation scan. 20-25 second scan duration. Software-based image reconstruction and processing.
Indications for Use
Indicated for x-ray computed tomography and projection x-ray imaging of upper and lower extremities in adults and pediatric patients aged 12 and older.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 30, 2016
Carestream Health, Inc. Carolyn Wagner Sr. Supervising Manager, Regulatory Affairs & Quality Systems 150 Verona Street Rochester, New York 14608
Re: K160723
Trade/Device Name: OnSight 3D Extremity System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: July 22, 2016 Received: August 1, 2016
Dear Carolyn Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name
OnSight 3D Extremity System
Indications for Use (Describe)
The device is intended to be used for x-ray computed tomography and projection x-ray imaging of upper and lower extremities of adult patients and pediatric patients aged 12 and over.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <div> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Carestream
# "510(k) Summary"
| 510(k) Owner Name: | Carestream Health, Inc. |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner Address: | 150 Verona Street<br>Rochester, New York 14608 |
| 510(k) Owner Phone: | 585 627-6588 |
| 510(k) Owner Fax: | 585 323-7643 |
| Contact Person & Info: | Carolyn Wagner<br>Sr. Regulatory Affairs Manager, X-ray Solution<br>carolyn.wagner@carestream.com<br>585-627-6588 |
| Date Summary Prepared: | March 15, 2016 |
| Device Trade Name: | OnSight 3D Extremity System |
| Device Common Name: | Computed Tomography X-ray System |
| Classification Name: | Computed tomography x-ray system |
| Device Class: | Class 2 |
| Device Code: | JAK |
| Regulation Number: | 21 CFR 892.1750 |
| Predicate DeviceTrade Name: | Planmed Verity<br>Manufactured by: Planmed Oy<br>510(k) No.: K121418 (February 1, 2013) |
| Device Common Name: | Computed Tomography X-ray System |
| Classification Name: | Computed tomography x-ray system |
| Device Class: | Class 2 |
| Device Code: | JAK |
| Regulation Number: | 21 CFR 892.1750 |
### Device Description:
The Carestream Health Onsight 3D Extremity System is a medical x-ray imaging device designed to acquire three-dimensional, volumetric CT data of patient extremities (feet, ankles, lower leg, knees, hands, wrists, arms and elbows). The device is configured as a "cone beam computed tomography" system (CBCT) in that the x-ray field covers the whole anatomy of interest (~25cm in length) and the data is acquired with a single rotation of the detector and x-ray source (actually a short scan of
{4}------------------------------------------------
180 degrees plus "fan angle" for a total of 216.5 degrees) with no patient motion through the irradiation field. The device is intended for use in a range of different locations including officebased medical practice, hospital departments and other imaging facilities. The system is also capable of acquiring standard two-dimensional projection radiographs of the same body parts.
#### Indications for Use / Intended Use:
The Indications for Use for the device, as described in its labeling, are:
"The device is intended to be used for x-rav computed tomography and projection x-rav imaging of upper and lower extremities of adult patients and pediatric patients aged 12 or over."
The intended use for this device, as determined by descriptions and the proposed labeling contained in this submission, is similar to the Indications for Use statement provided above. The OnSight 3D Extremity Syster is a computed tomography x-ray system (cone beam CT) used to generate 2D and 3D digital images of upper and lower extremities for diagnostic procedures. We believe that the OnSight 3D Extremity System and the predicate device (Planmed Verity) have the same intended use.
The Indications for Use for the subject device is the same as for the predicate device and the intended use remains unchanged. Any variation in features or technical specifications have been identified and addressed through testing (described below) to support a substantial equivalence determination.
### Comparison of Technological Characteristics:
Based upon information provided within this submission. we believe that the OnSight 3D Extremity System is substantially equivalent to the legally marketed Planmed Verity (predicate device). Both the Planmed Verity and OnSight 3D systems utilize Cone Beam Computed Tomography (CBCT) techniques to create a 3 dimensional reconstruction of a patient's upper or lower extremity. Both devices are designed as standalone units which enable the complete patient imaging procedure from acquisition of the x-ray images through patient information management, image processing and archiving of to be performed. Both systems image a single extremity at a time and support weight bearing imaging of the foot/ankle, leg or knee, in addition to the "non-loaded" imaging of these and the upper extremities (hand, wrist, arm and elbow).
Both systems utilize an amorphous silicon flat panel detector (a-Si:H) to acquire a multitude of 2-D projection x-rays sequentially acquired as the detector and x-ray source rotate around the patient's anatomy. During the acquisition of the multiple exposures the x-ray tube is pulsed (rather than continuous radiation). The duration of the complete scan is 20-25 seconds for the OnSight 3D device.
{5}------------------------------------------------
#### Discussion of Testing
The performance characteristics and operation / usability of the OnSight 3D Extremity System were evaluated in non-clinical (bench) testing. These studies demonstrated the intended workflow, related performance, overall function, verification and validation of requirements for intended use, and reliability of the system including both software and hardware requirements. Non-clinical test results have demonstrated that the device conforms to its specifications. Predefined acceptance criteria were met and demonstrated that the device is as safe, as effective, and performs as well as or better than the predicate device.
A clinical study was performed to demonstrate the diagnostic capability of the OnSight 3D Extremity System. Since it was not possible to obtain clinical images from the Planmed Verity system, two reference devices were used for comparison. The Philips Brilliance CT X-ray system (K060937) was used to evaluate 3D images. The DRX-1C Detector (K120062) as a component of the DRX-Evolution x-ray system (K091889) was used as a comparison to evaluate the 2D images.
Results of the Reader Study indicated that the diagnostic capability of the OnSight 3D Extremity System is statistically equivalent to or better than that of the reference devices.
#### Conclusions
Both the Planmed Verity and OnSight 3D Extremity System utilize Cone Beam Computed Tomography (CBCT) techniques to create a 3 dimensional reconstruction of a patient's upper or lower extremity. Both devices are designed as standalone units which enable the complete patient imaging procedure from acquisition of the x-ray images through patient information management, image processing and archiving to be performed. Both systems image a single extremity at a time and support weight bearing imaging of the foot/ankle, leg or knee, in addition to the "non-loaded" imaging of these and the upper extremities (hand, wrist, arm and elbow).
Both systems utilize an amorphous silicon flat panel detector (a-Si:H) to acquire a multitude of 2-D projection x-rays sequentially acquired as the detector and x-ray source rotate around the patient's anatomy. During the acquisition of the multiple exposures the x-ray tube is pulsed (rather than continuous radiation). This multitude of projection images is then processed by advanced software algorithms to generate a 3-dimensional representation of the patient's anatomy, in much the same way as a "slice based" computed tomography system generates a traditional CT scan of a patient. Both systems can also produce standard 2D projection radiographs of the patient's anatomy.
The differences between the OnSight 3D Extremity System and the predicate device (Planmed Verity) are minimal and do not affect the intended use of the devices or alter their fundamental scientific technology.
{6}------------------------------------------------
Results of bench testing described above in the "Discussion of Testing" section demonstrate that the OnSight 3D Extremity System conforms to its specifications. Results of clinical studies (also described in the "Discussion of Testing" section above) demonstrate that the OnSight 3D Extremity System produces both 2D and 3D images of diagnostic quality that are equivalent or superior to the reference devices used in the clinical study. We believe that these results support a claim of substantial equivalence for the OnSight 3D Extremity System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.