K160693 · Stryker Corporation · GEI · Apr 13, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160693
Device Name
Neptune E-SEP Smoke Evacuation Pencil
Applicant
Stryker Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 13, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Neptune E-SEP is an Integrated Smoke Evacuation Pencil (pencil) designed for general electrosurgical applications including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit (generator) to the operative site for the desired surgical effect.
Device Story
Single-use monopolar electrosurgical pencil; integrates smoke evacuation with electrosurgical cutting/coagulation. Input: electrical current from user-supplied electrosurgical generator; surgical smoke via suction. Operation: operator uses rocker or push-button switches to deliver current to operative site; smoke is evacuated through integrated tubing to external vacuum/filtration system. Output: electrosurgical effect on tissue; smoke removal. Used in surgical settings by clinicians. Benefits: minimizes surgical smoke plume exposure to personnel; provides integrated surgical tool. Modifications include additional blade electrode lengths (70mm, 125mm, 165mm) and PTFE heat-shrink insulation at electrode tips to focus high-density current.
Clinical Evidence
No clinical testing was deemed necessary. Evidence consists of bench testing, including electrical safety (ISO 60601-1, ISO 60601-2-2), biocompatibility (ISO 10993-1), and sterilization validation (ISO 11135).
Indicated for general electrosurgical applications including cutting and coagulation and smoke removal in conjunction with an evacuation system. Contraindicated for use on small appendages, such as in circumcision or finger surgery.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol consisting of three stylized human profiles facing to the right, stacked on top of each other.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2016
Stryker Corporation Mairead Twomey Associate Regulatory Affairs Manager 4100 E. Milham Avenue Kalamazoo, Michigan 49001
Re: K160693
Trade/Device Name: Neptune E-SEP™ Smoke Evacuation Pencil Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: March 10, 2016 Received: March 14, 2016
Dear Mairead Twomey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Jennifer R. Stevenson**
A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K160693
Device Name
Neptune® E-SEP™ Smoke Evacuation Pencil
Type of Use (Select one or both, as applicable)
Indications for Use (Describe)
The Neptune E-SEP is an Integrated Smoke Evacuation Pencil) designed for general electrosurgical applications including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit (generator) to the operative site for the desired surgical effect.
| <div> <span> <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# 510(k) Summary
| Contact Details | |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, Michigan 49001 USA<br>Ph: +353-21-4532988<br>Fax: +1-269-324-5412 |
| FDA Establishment<br>Registration No. | 1811755 |
| Contact Person | Mairead Twomey<br>Associate Regulatory Affairs Manager<br>Ph: +353-21-4532988<br>Fax: +1-269-389-5412<br>Mairead.Twomey@stryker.com |
| Date Submitted | March 10th, 2015 |
| Device Name | |
| Trade Name | Stryker® Neptune® E-SEP™ Smoke Evacuation Pencil |
| Common Name | Electrosurgical cutting and coagulation device and accessories |
| Classification | Class II |
| Primary Classification<br>Name | Electrosurgical, cutting & coagulation & accessories<br>(21 CFR 878.4400, Product code GEI) |
| Reason for 510(k)<br>Submission | Special 510(k) – Device modifications and line extension with no change<br>to fundamental scientific technology or intended use |
| Device Modification and<br>Line Extension | The following is a list of modifications to the Stryker Neptune E-SEP™<br>Smoke Evacuation Pencil. |
| | Addition of three electrodes (70 mm, 125 mm and 165 mm Blade<br>Electrodes)- this is a line extension to enable marketing of a more<br>competitive offering |
| | The proposed devices have an additional layer of insulation (PTFE) at<br>the tip of the electrodes:PTFE (Polytetraflouroethylene) has been added via heat<br>shrinking to the proposed electrodes to focus the delivery of the<br>high density current to the distal end of the electrode tip |
| | This line extension does not change the intended use, indications for use<br>or the fundamental scientific technology of the system. |
{4}------------------------------------------------
| Legally Marketed Predicate Device(s) | | | |
|-------------------------------------------------------------------------------|--------------|-----------------------------------------------------|--------------|
| 510(k) Number | Product Code | Trade Name | Manufacturer |
| K153679 | GEI | Stryker® Neptune® E-SEP™<br>Smoke Evacuation Pencil | Stryker |
| These predicate devices have not been the subject of a design related recall. | | | |
| Indications for Use | The Neptune E-SEP is an Integrated Smoke Evacuation Pencil<br>(pencil) designed for general electrosurgical applications<br>including cutting and coagulation, and for removing smoke<br>generated by electrosurgery when used in conjunction with an<br>effective smoke evacuation system. The pencil enables the<br>operator to remotely conduct an electrosurgical current from<br>the output connector of an electrosurgical unit (generator) to<br>the operative site for the desired surgical effect. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description | The Stryker® Neptune® E-SEP™ Smoke Evacuation Pencil is a<br>single use monopolar electrosurgical pencil that includes a<br>handpiece with a smoke evacuation function and a range of<br>electrodes (both uncoated and coated). The electrosurgical<br>pencil is compact in size (190 mm long and 17 mm<br>diameter) with 3 m of plastic smoke evacuation tubing and<br>associated electrical cable housed within the tubing. The<br>device is externally powered via an external power<br>generator (supplied by user) connected to a monopolar<br>receptacle using a conventional 3-pin connector. This device<br>remains unchanged from the cleared predicate pencil.<br><br>The Stryker® Neptune® E-SEP™ Smoke Evacuation Pencil is<br>available in two activation switch configurations, a rocker<br>style and push-button style, which activates monopolar cut or<br>coagulate functions operated by pressing the respective<br>button. The pencil is connected to smoke evacuation tubing<br>which features a dual connector (8 and 22 mm) to allow the<br>user to connect to a variety of smoke evacuation systems<br>including filtration or central vacuum systems, thus<br>minimizing exposure of personnel to surgical smoke plume. |
Blade electrodes are available coated and insulated and are listed in the below table.
{5}------------------------------------------------
| Model Numbers | Model Descriptions |
|---------------|-------------------------------------------------------------|
| 0703-070-002 | Neptune ® E-SEP TM 70mm Blade Electrode Coated & Insulated |
| 0703-125-002 | Neptune ® E-SEP TM 125mm Blade Electrode Coated & Insulated |
| 0703-165-002 | Neptune ® E-SEP TM 165mm Blade Electrode Coated & Insulated |
#### Table 6-1: Model Numbers and Descriptions of proposed Electrodes
| Performance Data<br>(Non-Clinical Tests) | The results of the performance testing demonstrate that the<br>functionality, integrity, and safety and effectiveness of the<br>electrodes for use with the Neptune E-SEP pencil is sufficient<br>for their intended use and support a determination of<br>substantial equivalence to the predicate device. |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Summary of Performance<br>Testing | Biocompatibility testing was performed on the subject devices<br>in accordance with ANSI/AAMI/ISO 10993-1:2009: Biological<br>Evaluation of Medical Devices - Part 1: Evaluation and Testing.<br>Results of testing validation the subject device is biocompatible<br>as intended for use.<br><br>The electrodes for use with the Neptune E-SEP pencil are only<br>available in sterile packaged form. The sterile products will be<br>terminally sterilized using ethylene oxide (EO). The sterilization<br>method was validated and performed in accordance with<br>ANSI/AAMI/ISO 11135:2014, Sterilization of health-care<br>products Ethylene oxide. A sterility assurance of 10-6 has been<br>validated for these products.<br><br>Performance testing was conducted on the subject devices as<br>determined by the risk analysis for the products. The following<br>areas were evaluated:<br>Electrical Safety Testing Functional / Performance Testing Thermal spread testing Biocompatibility testing Sterilization and packaging testing |
{6}------------------------------------------------
| | Results of these tests demonstrate that the functionality,<br>integrity, and safety and effectiveness of the subject devices<br>are sufficient for their intended use and support a<br>determination of substantial equivalence. |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Tests | No clinical testing was deemed necessary for this 510(k). |
{7}------------------------------------------------
| | Description | STRYKER E-SEP Pencil [Predicate] | STRYKER® Neptune® E-SEP™ Smoke Evacuation Pencil | Explanation of<br>Difference |
|-------------------------------------------|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| | | | [Proposed] | |
| Regulatory Information | 510(k) | K153679 | K160693 | N/A |
| | Product Code | GEI | GEI | Same |
| | Indication for Use | The Neptune E-SEP is an Integrated Smoke Evacuation Pencil (pencil) designed for general electrosurgical applications including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit (generator) to the operative site for the desired surgical effect. | The Neptune E-SEP is an Integrated Smoke Evacuation Pencil (pencil) designed for general electrosurgical applications including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit (generator) to the operative site for the desired surgical effect. | Same |
| | Classification of Device | Class II | Class II | Same |
| | Regulation Number | 21 CFR 878.4400 | 21 CFR 878.4400 | Same |
| | Regulation Name | Electrosurgical cutting and coagulation device and accessories | Electrosurgical cutting and coagulation device and accessories | Same |
| | Condition of Use | Single Use | Single Use | Same |
| | Type of Use | Prescription Use Only | Prescription Use Only | Same |
| | Patient Population | General | General | Same |
| | Contra -<br>indications | Do not use monopolar electrosurgery on small appendages, as in circumcision or finger surgery. | Do not use monopolar electrosurgery on small appendages, as in circumcision or finger surgery. | Same |
| Description | STRYKER E-SEP Pencil [Predicate] | STRYKER® Neptune® E-SEP™ Smoke Evacuation Pencil [Proposed] | Explanation of Difference | |
| Overall Design Concept | Overall Design | Device designed to integrate smoke evacuation into electrosurgery by combining both features into a single handpiece. | Device designed to integrate smoke evacuation into electrosurgery by combining both features into a single handpiece. | Same |
| | Power Supply | Monopolar Generator supplied by user | Monopolar Generator supplied by user | Same |
| | Voltage Rating | Maximum 5.5 kV peak | Maximum 5.5 kV peak | Same |
| | Electrical Connector | US-3-Pin | US-3-Pin | Same |
| | Electrical Safety Testing | ISO 60601-1<br>ISO 60601-2-2 | ISO 60601-1<br>ISO 60601-2-2 | Same |
| | Sterility | Sterile Single Use Only, EtO sterilized<br>SAL = 10-6 | Sterile Single Use Only, EtO sterilized<br>SAL = 10-6 | Same |
| | Packaging | Electrodes will be individually packaged in a Tyvek sealed pouch, sold 10 per box. | Electrodes will be individually packaged in a Tyvek sealed pouch, sold 10 per box. | Same |
| Electrode Technology and Materials | Electrode Rod Material | 303 Series Stainless Steel. | 303 Series Stainless Steel. | Same |
| | Electrode Rod Diameter | 2.36 mm | 2.36 mm | Same |
| | Electrode Connector Feature | Pentagon Connector<br>Image: Pentagon Connector | Pentagon Connector<br>Image: Pentagon Connector | Same |
| | Description | STRYKER E-SEP Pencil [Predicate] | STRYKER® Neptune® E-SEP™ Smoke Evacuation Pencil [Proposed] | Explanation of Difference |
| Electrode Technology and Materials | Electrode Connector Material | ABS Polylac PA-757 (ABS Plastic)<br>Image: Electrode Connector | ABS Polylac PA-757 (ABS Plastic)<br>Image: Electrode Connector | Same |
| | Electrode Over Mould Insulation | ABS Polylac PA-757 (For 125 mm and 165mm blade electrodes)<br>Image: Electrode Over Mould Insulation | ABS Polylac PA-757 (For 125 mm and 165mm blade electrodes)<br>Image: Electrode Over Mould Insulation | Same |
| | Electrode tip material | 303 Series Stainless Steel | 303 Series Stainless Steel | Same |
| | Electrode tip coating | PTFE (Polytetrafluoroethylene) (non-stick coating) | PTFE (Polytetrafluoroethylene) (non-stick coating) | Same |
| | Description | STRYKER E-SEP Pencil [Predicate] | STRYKER® Neptune® E-SEP™ Smoke Evacuation Pencil<br>[Proposed] | Explanation of<br>Difference |
| Electrode Technology and Materials | Electrode tip<br>working length | 70 mm - 165 mm | 70 mm - 165 mm | Same |
| | Electrode tip<br>geometry | Blade<br>Image: Electrode tip geometry | Blade<br>Image: Electrode tip geometry | Same |
| | Electrode Tip<br>Insulation | None<br>Image: Electrode tip insulation | PTFE (Polytetrafluroethylene) has been added via heat<br>shrinking to the electrodes.<br>Image: Tip insulation | Different |
| Smoke Evacuation Technology and Materials | Adjustable<br>Suction Sleeve<br>Material | Styrene butadiene copolymer<br>Acrylonitrile Butadiene Styrene with barium sulfate | Styrene butadiene copolymer<br>Acrylonitrile Butadiene Styrene with barium sulfate | Same |
| | Suction Sleeve<br>Lengths | 70mm, 125mm, 165 mm | 70mm, 125mm, 165 mm | Same |
| | Evacuation<br>Tubing<br>Dimension | 10 mm diameter x 3 m length | 10 mm diameter x 3 m length | Same |
| | Smoke<br>Evacuation<br>System<br>Connector | 8mm, 22mm | 8mm, 22mm | Same |
| Description | STRYKER E-SEP Pencil [Predicate] | STRYKER® Neptune® E-SEP™ Smoke Evacuation Pencil [Proposed] | Explanation of Difference | |
| Pencil Technology and Materials | Handpiece Housing Material | Acrylonitrile Butadiene Styrene with Thermoplastic Elastomer | Acrylonitrile Butadiene Styrene with Thermoplastic Elastomer | Same |
| | Handpiece Dimension | 15mm Diameter x 190 mm Length | 15mm Diameter x 190 mm Length | Same |
| | Operation Function Switches | CUT button labeled yellow and proximal to electrode<br>COAG button labeled blue and distal to electrode | CUT button labeled yellow and proximal to electrode<br>COAG button labeled blue and distal to electrode | Same |
Table 6-2: Summary of Predicate Comparison
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
## Conclusion/Substantial Equivalence Rationale
The Stryker E-SEP Pencil and Accessories are either identical or similar in intended use, indications for use, technological characteristics, safety and effectiveness to the previously cleared Pencil. The products have the same fundamental scientific technology, basic design, functional characteristics and applications. The modifications and line extension of electrodes introduced raise no new issues of safety and effectiveness.
Therefore, the subject device is at least as safe and effective as the predicate and evidence supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.