The Insulin Algorithms System (IA System) is a glycemic management software support program designed to provide insulin dosing recommendations to the clinician in order to adjust and maintain blood glucose levels in insulin-dependent diabetic patients within a clinician determined target range. The IA System evaluates a set of current and cumulative patient blood glucose values together with biometric information (including age, weight, and gender) and suggests changes (if necessary). The system is designed for use by healthcare professionals. The IA System is indicated for use in adults and pediatric patients ages 7 and above who weigh more than 36 lbs. The IA System is not intended to be used to provide insulin dosing recommendations to the clinician in order to adjust insulin treatment for patients hospitalized in an acute care setting. Also, the IA System is not intended to be used to provide insulin dosing recommendations for patients on insulin pump therapy or IV insulin therapy. Nor is IA System intended to provide recommendations for mealtime injections of insulin that are based on carb counting. The IA System is not a substitute for, but rather an adjunct to clinical reasoning. No medical decision should be based solely on the recommended guidance provided by this software program.
Device Story
IA System is a glycemic management software; inputs include patient biometrics (age, height, weight, gender), insulin dosing history, meal information, and glucometer readings. Custom algorithm processes data based on established insulin dosing conventions to generate regimen change recommendations. Used by healthcare professionals in clinical settings; software operates as a server-client architecture. Clinicians review output as an adjunct to clinical reasoning to adjust patient insulin therapy. Benefits include optimized glycemic control for insulin-dependent patients.
Clinical Evidence
No clinical data. Evidence consists of software verification and validation testing per IEC 62304:2006 and FDA software guidance, plus usability testing confirming medical professionals understand system scope and limits.
Technological Characteristics
Software-only solution; server-client architecture. Server application performs data storage and algorithmic analysis. Client application serves as user interface. HIPAA-compliant security controls (individual ID/password). No hardware components.
Indications for Use
Indicated for insulin-dependent diabetic patients, adults and pediatric patients ages 7+ weighing >36 lbs, to assist clinicians in adjusting insulin regimens to maintain blood glucose within target ranges. Contraindicated for patients in acute care settings, patients on insulin pump or IV insulin therapy, and for mealtime insulin dosing based on carb counting.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2017
INSULIN ALGORITHMS C/O SIGI CARON PRESIDENT MEDTECH CONSULTANTS, INC. 20370 SKYHAWK LANE TOPANGA CA 90290
Re: K160673
Trade/Device Name: Insulin Algorithms System Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive pulmonary-function value calculator Regulatory Class: II Product Code: NDC Dated: February 2, 2016 Received: February 3, 2016
Dear Sigi Caron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney H. Lias -S
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) k160673
Device Name Insulin Algorithms System
#### Indications for Use (Describe)
The Insulin Algorithms System (IA System) is a glycemic management software support program designed to provide insulin dosing recommendations to the clinician in order to adjust and maintain blood glucose levels in insulin-dependent diabetic patients within a clinician determined target range. The IA System evaluates a set of current and cumulative patient blood glucose values together with biometric information (including age, weight, and gender) and suggests changes (if necessary). The system is designed for use by healthcare professionals. The IA System is indicated for use in adults and pediatric patients ages 7 and above who weigh more than 36 lbs.
The IA System is not intended to be used to provide insulin dosing recommendations to the clinician in order to adjust insulin treatment for patients hospitalized in an acute care setting. Also, the IA System is not intended to be used to provide insulin dosing recommendations for patients on insulin pump therapy or IV insulin therapy. Nor is IA System intended to provide recommendations for mealtime injections of insulin that are based on carb counting.
The IA System is not a substitute for, but rather an adjunct to clinical reasoning. No medical decision should be based solely on the recommended guidance provided by this software program.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 1. 510(K) SUMMARY - K160673
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| DATE PREPARED | June 7, 2017 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Insulin Algorithms<br>269 S. Beverly Dr. #613<br>Beverly Hills, CA 90212<br>www.insulinalgorithms.com<br>Tel: (310) 606-2055 |
| OFFICIAL<br>CORRESPONDENT | Sigi Caron<br>MedTech Consultants, Inc.<br>20370 Skyhawk Lane<br>Topanga, CA 90290<br>sigi@medtechconsultants.com<br>Tel: (310) 455-3473<br>Fax: (888) 295-1535 |
| TRADE NAME | IA System |
| COMMON NAME | Drug Dose Calculator |
| DEVICE CLASSIFICATION | Name: Predictive pulmonary-function value calculator<br>Regulation No: §868.1890<br>Product Code: NDC<br>Class: II<br>Panel: Anesthesiology |
| PREDICATE DEVICE | • Glucommander System – K113853<br>• EndoTool Glucose Management System – K132547 |
### SUBSTANTIALLY EQUIVALENT TO:
The IA System device is substantially equivalent in intended use and technological features to the Glucommander System - K113853 and to the EndoTool Glucose Management System -K132547.
### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The Insulin Algorithms System (IA System) takes basic biometric information (age, height, weight, gender), insulin dosing information (prescribed insulins, injection times, and amounts), meal information (meal times, before meal glucose targets, after meal glucose targets), along with a collection of glucometer readings and processes this information though a custom algorithm to create a recommendation as to what changes to the patient's insulin regimen should be made. The algorithm was developed using the insulin dosing conventions taught and used in
{4}------------------------------------------------
### INDICATIONS FOR USE:
The IA System is a glycemic management software support program designed to provide insulin dosing recommendations to the clinician in order to adjust and maintain blood glucose levels in insulin-dependent diabetic patients within a clinician determined target range. The IA System evaluates a set of current and cumulative patient blood glucose values together with biometric information (including age, weight, height, and gender) and suggests changes (if necessary). The system is designed for use by healthcare professionals. The IA System is indicated for use in adults and pediatric patients ages 7 and above who weigh more than 36 lbs.
The IA System is not intended to be used to provide insulin dosing recommendations to the clinician in order to adjust insulin treatment for patients hospitalized in an acute care setting. Also, the IA System is not intended to be used to provide insulin dosing recommendations for patients on insulin pump therapy or IV insulin therapy. Nor is IA System intended to provide recommendations for mealtime injections of insulin that are based on carb counting.
The IA System is not a substitute for, but rather an adjunct to clinical reasoning. No medical decision should be based solely on the recommended guidance provided by this software program.
## TECHNICAL CHARACTERISTICS:
The IA System is a software solution. It is comprised of two separate software systems: a server application containing all algorithms and mitigations and a client application that serves as an interface to the server. All data are stored and analyzed on the server application. Each user has an individual identification and password to access the application. The IA System is designed to safeguard the confidentiality, integrity, and availability of electronic health information in conformance with the Health Insurance Portability and Accountability Act (HIPAA).
### PERFORMANCE STANDARDS:
No applicable performance standards have been issued under 514 of the Food, Drug and Cosmetic Act for a predictive Pulmonary Function Value Calculator §868.1890.
## SUMMARY OF NONCLINICAL TESTING:
Software verification & validation testing was conducted in accordance with IEC 62304:2006: Medical device software -- Software Life Cycle Processes and with FDA guidances for software development and testing (Guidance for the Content of Premarket Submission for Medical Device Containing Software, FDA-CDRH-CBER, Rockville, MD, 05/2005; Guidance for Off-the-Shelf Software Use in Medical Devices, FDA- CDRH, Rockville, MD, 06/1997; and General Principles of Software Validation; Final Guidance for Industry and FDA Staff, FDA-CDRH, Rockville, MD, 01/2002). Usability testing indicates that the target group of medical
{5}------------------------------------------------
professionals understood the scope and limit of the application and were able to use the program with ease and confidence. All features provided for the IA System have been verified to operate as specified. Testing confirms that the IA System can be used according to its intended use and in an equivalent manner to the predicate devices.
### BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
The IA System is substantially equivalent to the listed predicate devices with respect to their indications for use (intended use) and technical characteristics. The information and data provided in this 510(k) submission identifies no new safety or effectiveness issues.
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.