The HIOSSEN Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures. The OSSTEM Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
Device Story
The ET US SS Prosthetic system consists of various dental abutments, components, and abutment screws used to support prosthetic restorations (crowns, bridges, over-dentures) on dental implants. The system includes temporary abutments, multi-abutments, solid abutments, ComOcta abutments, cover screws, healing abutments, and retraction caps. These components are connected to dental fixtures (implants) via internal or external hex/octa connections. The device is used by dental professionals in a clinical setting to facilitate the restoration of edentulous or partially edentulous patients. The components are made of titanium alloy, titanium, or polymers (PEEK, polycarbonate, POM). The system provides the mechanical interface between the implant fixture and the final or temporary prosthesis, enabling functional and aesthetic restoration. Some components allow for path adjustment (angled abutments) or gingival shaping (healing abutments).
Clinical Evidence
No clinical studies were submitted. Substantial equivalence is supported by non-clinical testing, including biocompatibility assessments, sterilization validation (ISO 17665-1:2006), and fatigue testing conducted according to FDA guidance for root-form endosseous dental implants and abutments.
Indicated for use with dental implants to support prosthetic restorations including crowns, bridges, or over-dentures in patients requiring dental implant-supported prosthetics.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 25, 2017
OSSTEM Implant Co., Ltd. % David Kim Manager HiOSSEN Inc. 85 Ben Fairless Dr. Fairless Hills, Pennsylvania 19030
Re: K160670
Trade/Device Name: ET US SS Prosthetic System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: April 12, 2017 Received: April 12, 2017
Dear David Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo and contact information for OSSTEM Implant Co., Ltd. The logo includes the company name in bold, black font, with the word "OSSTEM" in orange to the left of the company name. Below the company name is the address: 66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea. The phone number is +82 51 850-2500, the fax number is +82 51 850-4341, and the website is www.osstem.com.
## Indications for Use Statement
_K160670 510(k) Number
Device Name: ET US SS Prosthetic system
Indication for use:
#### ET System
The HIOSSEN Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
#### US/SS System
The OSSTEM Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
Prescription Use X (Per 21CFR801 Subpart D) OR
Over-The-Counter Use (Per 21CFR807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
#### 510(k) Summary
Date: April 25, 2017
1. Company and Correspondent making the submission:
| - Submitter's Name: | OSSTEM Implant Co., Ltd. |
|-------------------------|-------------------------------------------------------------------------|
| - Address: | 66-16, Bansong-ro 513beon-gil, Haeundae-gu,<br>Busan, Republic of Korea |
| - Contact: | Mr. Hee Kwon Son |
| - Phone: | +82 51 850 2575 |
| - Correspondent's Name: | HIOSSEN Inc. |
| - Address: | 85 Ben Fairless Dr. Fairless Hills, PA 19030 |
| - Contact: | DAVID KIM |
| - Phone: | 267 759 7031 |
- 2. Device:
| Trade or (Proprietary) Name: | ET US SS Prosthetic system |
|------------------------------|------------------------------------|
| Common or Usual Name: | Dental Abutment |
| Classification Name: | Endosseous dental implant abutment |
| Regulation Number: | 21CFR872.3630 |
| Device Classification: | Class II |
| Product Code: | NHA |
- 3. Predicate Device:
Primary Predicate K062030 US System, OSSTEM IMPLANT CO.,LTD
Reference Predicates
K121585 TS Implant System, OSSTEM IMPLANT CO.,LTD K100245 HS, HG Prosthetic System, OSSTEM IMPLANT CO.,LTD K130662 ET Prosthetic System, OSSTEM IMPLANT CO.,LTD K062051 SS System, OSSTEM IMPLANT CO.,LTD K073247 US.SS.GS System, OSSTEM IMPLANT CO.,LTD K121843 NP Cast Abutment System, OSSTEM IMPLANT CO.,LTD K081575 HU.HS.HG Prosthetic System, OSSTEM IMPLANT CO.,LTD K 132067 Multi Angled Abutment with carrier, OSSTEM IMPLANT CO.,LTD
{4}------------------------------------------------
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
#### 4. Description:
The ET US SS Prosthetic system is intended for use as an aid in prosthetic restoration. It is consisted of abutments, components and abutment screws.
The ET US SS Prosthetic system is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics.
The ET US SS Prosthetic system is substantially equivalent in design, function and intended use to the Predicate Devices as above.
| Item | Content | |
|-----------------------------|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Quick Temporary<br>Abutment | Description | Used temporary until final prosthesis is made to maintain<br>esthetic appearance.<br>Quick Temporary Abutment Set is consisted of Quick<br>Temporary Abutment and Ti Screw. |
| | Material | Titanium Alloy + Plastic (Polyetherethereketone) |
| | Diameter (mm) | 4, 4.5 |
| | Height (mm) | 13.9, 14.3, 14.5, 14.6, 17.4, 17.8, 18.0, 18.1 |
| Multi Abutment | Description | Used for edentulous mandible or maxilla.<br>Usually use to make full denture.<br>Screw Retained Restoration.<br>Used in screw type prosthetics in multiple cases.<br>Multi Abutment Set is consisted of Multi Abutment and<br>Multi Abutment Carrier. |
| | Material | Titanium Alloy (TiN Coating) |
| | Diameter (mm) | 4.8 |
| | Height (mm) | 8.3, 8.7, 9.3 9.7, 10.3, 10.7, 11.3, 11.7, 12.3, 12.7 |
#### ET System
#### SS System
| Item | Content |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Solid Abutment | Description<br>Used in producing ordinary cement-retained prosthetics.<br>Solid Abutment Set is consisted of Solid Abutment and<br>Solid Protect Cap. |
| | Material<br>Titanium Alloy (ASTM F 136) |
| | Diameter (mm)<br>3.5, 4.3 |
| | Height (mm)<br>9.5, 9.8, 11, 11.3, 12.5, 12.8 |
| Solid Protect Cap | Description<br>Used to protect Solid abutments in the oral cavity. |
| | Material<br>Polymer (Polycarbonate) |
| | Diameter (mm)<br>5.3, 6.02 |
| Excellent Solid Abutment | Description<br>Used in producing cement-retained prosthetics.<br>Excellent Solid Abutment Set is consisted of Excellent<br>Solid Abutment and Solid Protect Cap. |
| | Material<br>Titanium Alloy (ASTM F 136) |
| | Diameter (mm)<br>4.3, 5.29 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Osstem Implant. The word "OSSTEM" is written in a bold, sans-serif font in orange color. Below the word "OSSTEM", the word "IMPLANT" is written in a smaller, sans-serif font in gray color. The logo is simple and modern.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| | Height (mm) | 9.5, 9.7, 11, 11.2, 12.5, 12.7 |
|--------------------------------|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Excellent Solid<br>Protect Cap | Description | Used to protect Excellent Solid Abutments in the oral cavity. |
| | Material | Polymer (Polycarbonate) |
| | Diameter (mm) | 5.3, 6.6 |
| ComOcta Abutment | Description | Used for making general cement-type prosthesis.<br>ComOcta Abutment is consisted of ComOcta Abutment and Ti Screw. |
| | Material | Titanium Alloy (ASTM F 136) |
| | Diameter (mm) | 4.3, 5.26 |
| | Height (mm) | 6.53, 6.55, 6.88, 6.90, 8.03, 8.05, 8.38, 8.40, 9.53, 9.55,<br>9.88, 9.90 |
| ComOcta Plus Abutment | Description | Used for making general cement-type prosthesis and used when there is deep gingiva or a fixture is to be deeply inserted.Gold coloring on gingiva region is for aesthetics.<br>ComOcta Plus Abutment Set is consisted of ComOcta Plus Abutment and Ti Screw. |
| | Material | Titanium Alloy (ASTM F 136) |
| | Diameter (mm) | 4.8, 6.0 |
| | Height (mm) | 6.5, 7, 7.5, 8, 8.5, 9, 9.5, 10 |
| ComOcta Angled<br>Abutment | Description | Used for prosthetic restoration and path adjustment of prosthesis in case of 15° and 20° axial angle.<br>ComOcta Angled Abutment Set is consisted of ComOcta Angled Abutment and Ti Screw. |
| | Material | Titanium Alloy (ASTM F 136) |
| | Diameter (mm) | 3.75, 4.3 |
| | Height (mm) | 5.5, 6.7 |
| ComOcta Retraction Cap | Description | Used to protect ComOcta abutments in the oral cavity |
| | Material | Polymer (PolyOxy Methylene) |
| | Diameter (mm) | 6.4, 7.7 |
#### US System
| Item | Content | |
|-------------------|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Cover Screw | Description<br>Material<br>Diameter (mm)<br>Height (mm) | Used to protect other substances from ingressing into the fixture after the sugical operation.<br>Titanium (ASTM F 67)<br>3.7, 4.3, 5.2, 5.3<br>5.9 |
| Healing Abutment | Description<br>Material<br>Diameter (mm)<br>Height (mm) | Used to form the appropriate gingival shape after the second surgery.<br>Titanium (ASTM F 67)<br>4, 5, 6, 7<br>7.9, 8.9, 9.9, 11.4, 12.9 |
| Cemented Abutment | Description<br>Material<br>Diameter (mm) | Used for making general cement-type prosthesis.<br>Titanium Alloy (ASTM F 136)<br>4.0, 4.1, 5.0, 5.1, 6.0, 7.0 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Osstem Implant. The logo is in orange and consists of the word "OSSTEM" in a bold, sans-serif font, with the word "IMPLANT" in a smaller, lighter font underneath. The logo is simple and modern, and the orange color is eye-catching.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
| | Height (mm) | 5.0, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0, 9.5, 10.0, 11.0 |
|------------------------------------|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Ti Screw | Description | Used to connect abutment with fixture. |
| | Material | Titanium Alloy (ASTM F 136) |
| | Diameter (mm) | 2.3, 2.9, 3.2 |
| | Height (mm) | 8.0, 8.2 |
| US Multi Angled<br>Abutment | Description | If a few number of fixture were implanted in mandibular<br>bone for making full denture, some of fixture path should<br>be leaned. In that case, US Multi Angled Abutment is used<br>to adjust path of prosthesis in case of 17 and 30° axial<br>angle.<br>US Multi Angled Abutment Set is consisted of US Multi<br>Angled Abutment, Ti Screw, and Abutment Carrier. |
| | Material | Titanium Alloy (ASTM F 136) |
| | Diameter (mm) | 3.4, 4.0, 4.1 |
| | Height (mm) | 3.97, 3.98, 4.28, 4.77, 4.78, 5.28, 5.47, 6.28 |
| Ti Screw | Description | Used to connect abutment with fixture. |
| | Material | Titanium Alloy (ASTM F 136) |
| | Diameter (mm) | 2.2, 2.3 |
| | Height (mm) | 4.9, 5.55 |
| Esthetic-low Abutment | Description | Used in producing screw-retained aesthetic prosthetics.<br>Structure producing prosthetics in cylinder after attaching<br>abutment in the oral cavity.<br>Esthetic-low Abutment Set is consisted of Esthetic-low<br>Abutment and Esthetic-low Abutment Screw. |
| | Material | Titanium (ASTM F67) |
| | Diameter (mm) | 4.8, 5.5 |
| | Height (mm) | 2.1, 2.2, 3.0, 3.1, 4.0, 4.1, 5.0, 5.1 |
| Esthetic-low Abutment<br>Screw | Description | Used to connect Esthetic-low abutment with fixture. |
| | Material | Titanium (ASTM F67) |
| | Diameter (mm) | 3.1, 3.6 |
| | Height (mm) | 7.8, 7.9, 8.8, 8.9, 9.8, 9.9, 10.8, 10.9 |
| Esthetic-low Temporary<br>Cylinder | Description | Used in producing temporary prosthetics.<br>Esthetic-low Temporary Cylinder is consisted of Esthetic-<br>low Temporary Cylinder and Ti Screw. |
| | Material | Titanium (ASTM F 67) |
| | Diameter (mm) | 4.8, 5.5 |
| | Height (mm) | 12.0 |
| O-ring Abutment | Description | Used in creating stud type overdenture prosthetics. |
| | Material | Titanium Alloy |
| | Diameter (mm) | 5.0, 5.6 |
| | Height (mm) | 2.0, 3.0, 4.0, 5.0, 6.0 |
#### 5. Indications for Use:
#### ET System
The HIOSSEN Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
#### US/ SS System
The OSSTEM Prosthetic system is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or over-dentures.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Osstem Implant. The logo is in orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray and is located below the word "OSSTEM".
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
#### 6. Substantial Equivalence Matrix:
| | Proposed Device | Primary Predicate Device | Reference Predicate |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510K | - | K062030 | K130662 |
| Indications for<br>Use | ET System<br>The HIOSSEN<br>Prosthetic system is<br>intended for use with a<br>dental implant to<br>provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or over-dentures.<br><br>US/ SS System<br>The OSSTEM Prosthetic<br>system is intended for use<br>with a dental implant to<br>provide support for<br>prosthetic restorations<br>such as crowns, bridges,<br>or over-dentures. | Us system and ssii mini are<br>intended for use in partially<br>or fully edentulous mandibles<br>and maxillae, in support of<br>single or multiple-unit<br>restorations including:<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal or<br>intermediate abutment<br>support for fixed bridgework.<br><br>Us system is for two stage<br>surgical procedures. It is not<br>for one stage or immediate<br>load. Ssii mini is for one and<br>two stage surgical procedures.<br>It is not for immediate load. | ET Prosthetic<br>Systems is intended<br>for use with a dental<br>implant to provide<br>support for prosthetic<br>restorations such as<br>crowns, bridges, or<br>overdentures. |
The Indications for Use of the submission device differs from the primary predicate as in that the submission Indications are specific to the sub-set of devices, that is abutments, which have been included in the current submission. The primary predicate contains devices for a complete implant system and the Indications reflects this. Since the current submission is a sub-set of devices included in the primary predicate, the modification of the Indications to a sub-set of the primary predicate Indications is also appropriate. The Indications for the submission device are also identical to a reference predicate which only contains a similar abutment sub-set of devices.
Subtantial Equivalence Matrixes for each specific device are provided as shown below.
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|------------------------------------------------------------|------------------------------------------------------------|
| 510K | - | K121843 |
| Material | Titanium Alloy<br>PEEK | Titanium Alloy<br>PEEK |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to make temporary<br>prosthesis | Used to make temporary<br>prosthesis |
| Principles of<br>Operation | Quick Temporary Abutment is<br>used in producing temporary | Quick Temporary Abutment is<br>used in producing temporary |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Osstem Implant. The logo is orange and features the word "OSSTEM" in a stylized font. Below the word "OSSTEM" is the word "IMPLANT" in a smaller, gray font.
| | prosthetics for immediate loading<br>Upper part is made with plastic<br>(PEEK) for easy altering/removing<br>when make temporary prosthetics | prosthetics for immediate loading<br>Upper part is made with plastic<br>(PEEK) for easy altering/removing<br>when make temporary prosthetics |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Dimension<br>(Diameter) | 4mm, 4.5mm | 4.5mm, 5.5mm |
| Connection | Internal Hex Connection | Internal Hex Connection |
| Characteristic | Cement retained restoration.<br>Capable of easily altering/removing<br>shape of plastic material.<br>Abutment + Screw | Cement retained restoration.<br>Capable of easily<br>altering/removing shape of plastic<br>material.<br>Abutment + Screw |
| Design | Image: Abutment with screw | Image: Abutment with screw |
| SE | Difference:<br>Shape of upper part (Plastic part) is different from predicate, Quick<br>Temporary Abutment<br>It is designed to be used when limitations in space (narrow space )<br>The Quick Temporary Abutment had been cleared with K121585 but it<br>is being sumitted to add revised shape therefore Quick Temporary<br>Abutmet and Predicate devices have same principles of operation,<br>function, material, Connection structure, charateristic and intended use | |
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| 510K | - | K062051 |
| Material | Titanium Alloy | Titanium Alloy |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to make Cement-retained<br>prosthesis | Used to make Cement-retained<br>prosthesis |
| Principles of<br>operation | As general one body cement<br>retained restoration, it is connected<br>with fixture and cemented crown<br>on the abutment | As general one body cement<br>retained restoration, it is connected<br>with fixture and cemented crown<br>on the abutment |
| Dimension<br>(Height) | 9.5mm, 9.8mm, 11mm, 11.3mm,<br>12.5mm, 12.8mm | 9.5mm, 9.8mm, 11mm, 11.3mm,<br>12.5mm |
| Connection | Internal Octa Connection | Internal Octa Connection |
| Characteristic | Cement retained restoration<br>Abutment + Protect Cap | Cement retained restoration<br>Abutment + Protect Cap |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is written in orange, and the word "IMPLANT" is written in gray below it. The font is sans-serif and appears to be bolded.
| Design | Image: dental abutment | Image: dental abutment |
|--------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| | Difference:<br>New dimension is added. | |
| S E | The Solid Abutment had been cleared with K062051, but it is being<br>sumitted to add a new dimension. Therefore, Solid Abutment and<br>Predicate devices have same principles of operation, function, material,<br>connection structure, charateristic and intended use. | |
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| | Solid Protect Cap | Solid Protect Cap |
| 510K | | K062051 |
| Material | Polymer (Polycarbonate) | Polymer (Polycarbonate) |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to protect Solid Abutment in<br>the oral cavity | Used to protect Solid Abutment in<br>the oral cavity |
| Principles of<br>operation | Solid Protect Cap is connected<br>with Solid Abutment to protect in<br>the oral cavity and minimize<br>foreign body sensation by the<br>patient | Solid Protect Cap is connected<br>with Solid Abutment to protect in<br>the oral cavity and minimize<br>foreign body sensation by the<br>patient |
| Dimension<br>(diameter) | 5.3mm, 6.02mm | 5.5mm, 6.7mm |
| Characteristic | Protect abutment in the oral cavity | Protect abutment in the oral cavity |
| Design | Image: Proposed Device Design | Image: Predicate Device Design |
| SE | Difference:<br>New dimension is added and shape is changed.<br>The Solid Protect Cap had been cleared with K062051, but it is being<br>sumitted to add a new dimension added and its shape changed.<br>Therefore, Solid Protect Cap and Predicate devices have same principles<br>of operation, function, material, charateristic and intended use. | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Osstem Implant. The word "OSSTEM" is in large, bold, orange letters. Below it, the word "IMPLANT" is in smaller, gray letters. The logo is simple and modern.
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| 510K | - | K062051 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to make Cement-retained<br>prosthesis | Used to make Cement-retained<br>prosthesis |
| Principles of<br>operation | As general one body cement<br>retained restoration, it is connected<br>with fixture and cemented crown<br>on the abutment | As general one body cement<br>retained restoration, it is connected<br>with fixture and cemented crown<br>on the abutment |
| Dimension<br>(Height) | 9.5mm, 9.7mm, 11mm, 11.2mm,<br>12.5mm, 12.7mm | 9.5mm, 9.7mm, 11mm, 11.2mm,<br>12.5mm |
| Connection | Internal Octa Connection | Internal Octa Connection |
| Characteristic | Cement retained restoration<br>Abutment + Protect Cap | Cement retained restoration<br>Abutment + Protect Cap |
| Design | Image: Proposed Device Design | Image: Predicate Device Design |
| SE | Difference:<br>New dimension is added.<br>The Excellent Solid Abutment had been cleared with K062051, but it is<br>being sumitted to add a new dimension. Therefore, Solid Abutment and<br>Predicate devices have same principles of operation, function, material,<br>connection structure, charateristic and intended use. | |
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Excellent Solid Protect Cap | Excellent Solid Protect Cap |
| 510K | - | K062051 |
| Material | Polymer (Polycarbonate) | Polymer (Polycarbonate) |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to protect Excellent Solid<br>Abutment in the oral cavity | Used to protect Excellent Solid<br>Abutment in the oral cavity |
| Principles of<br>operation | Excellent Solid Protect Cap is<br>connected with Solid Abutment to<br>protect in the oral cavity and<br>minimize foreign body sensation<br>by the patient | Excellent Solid Protect Cap is<br>connected with Solid Abutment to<br>protect in the oral cavity and<br>minimize foreign body sensation<br>by the patient |
| Dimension<br>(Height) | 6.35mm, 7.85mm, 9.35mm,<br>9.65mm | 6.35mm, 7.85mm, 9.35mm |
{11}------------------------------------------------
I M P L A N 1
| Characteristic | Protect abutment in the oral cavity | Protect abutment in the oral cavity |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Design | Image: Protect abutment in the oral cavity | Image: Protect abutment in the oral cavity |
| S E | Difference:<br>New dimension is added.<br>The Excellent Solid Protect Cap had been cleared with K062051, but it<br>is being sumitted to add a new dimension. Therefore, Solid Protect Cap<br>and Predicate devices have same principles of operation, function,<br>material, charateristic and intended use. | |
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| 510K | ComOcta Abutment | K073247 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to make Cement-retained<br>prosthesis | Used to make Cement-retained<br>prosthesis |
| Principles of<br>operation | As general cement retained<br>restoration, it is connected with<br>fixture and cemented crown on the<br>abutment | As general cement retained<br>restoration, it is connected with<br>fixture and cemented crown on the<br>abutment |
| Dimension<br>(Height) | 6.53mm, 6.55mm, 6.88mm,<br>6.90mm, 8.03mm, 8.05mm,<br>8.38mm, 8.40mm, 9.53mm,<br>9.55mm, 9.88mm, 9.90mm | 6.53mm, 6.55mm, 6.88mm,<br>6.90mm, 8.03mm, 8.05mm,<br>8.38mm, 8.40mm, 9.53mm,<br>9.55mm |
| Connection | Internal Octa Connection | Internal Octa Connection |
| Characteristic | Cement retained restoration<br>Abutment + Screw | Cement retained restoration<br>Abutment + Screw |
| Design | Image: ComOcta Abutment design | Image: ComOcta Abutment design |
| SE | Difference:<br>New dimensions are added.<br>The ComOcta Abutment had been cleared with K062051, but it is being<br>sumitted to add new dimension. Therefore, ComOcta Abutment and | |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Osstem Implant. The logo is in orange and gray. The word "OSSTEM" is in orange, and the word "IMPLANT" is in gray.
66-16, Bansong-ro 513beon-gil, Haeundae-gu, Busan, Republic of Korea Tel: +82 51 850-2500 Fax: +82 51 850-4341 www.osstem.com
> Predicate devices have same principles of operation, function, material, connection structure, charateristic and intended use.
| Part Name | Proposed Device | Predicate Device |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| ComOcta Plus Abutment | ComOcta Plus Abutment | |
| 510K | - | K073247 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description<br>(Intended for use) | Used to make Cement-retained<br>prosthesis<br>Used when there is deep gingiva or<br>a fixture is to be deeply inserted | Used to make Cement-retained<br>prosthesis<br>Used when there is deep gingiva or<br>a fixture is to be deeply inserted |
| Principles of<br>operation | As general cement retained<br>restoration, it is connected with<br>fixture and cemented crown on the<br>abutment | As general cement retained<br>restoration, it is connected with<br>fixture and cemented crown on the<br>abutment |
| Dimension<br>(Height) | 6.5mm, 7mm, 7.5mm, 8mm,<br>8.5mm, 9mm, 9.5mm, 10mm | 7.5mm, 8mm, 9.5mm, 10mm |
| Connection | Internal Octa Connection | Internal Octa Connection |
| Characteristic | Cement retained restoration<br>Abutment + Screw<br>Gold coloring on gingiva region<br>for asthetics | Cement retained restoration<br>Abutment + Screw<br>Gold coloring on gingiva region<br>for asthetics |
| Design | Image: abutment | Image: abutment |
| SE | Difference:<br>New dimensions and Non-Octa type abutments are added.<br>The ComOcta Plus Abutment had been cleared with K073247, but it is<br>being sumitted to add new dimensions and Non-Octa type abutments.<br>Therefore, ComOcta Plus Abutment and Predicate devices have same<br>principles of operation, function, material, charateristic and intended use. | |
| Part Name | Proposed Device | Predicate Device |
|--------------|---------------------------------|---------------------------------|
| | ComOcta Angled Abutment | ComOcta Angled Abutment |
| 510K | - | K073247 |
| Material | Titanium Alloy (ASTM F 136) | Titanium Alloy (ASTM F 136) |
| Manufacturer | OSSTEM Implant Co., Ltd. | OSSTEM Implant Co., Ltd. |
| Description | Used for prosthetic restoration | Used for prosthetic restoration |
{13}------------------------------------------------
MPLAN7
| (Intended for use) | Used for the path adjustment of<br>prosthesis in case of 15° and 20°<br>axial angle…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.