The Aqueduct 100 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
Device Story
Aqueduct 100 is a triple-balloon cervical dilation catheter; features 3-lumen shaft with anchoring balloon, two cylindrical dilation balloons, and infusion lumen with fixed stiffening stylet. Physician inserts catheter through vagina/cervix into uterus; inflates distal anchoring balloon; withdraws to position against internal os. Two cylindrical balloons align at internal/external cervical os; inflated with saline for mechanical dilation. Infusion lumen allows saline injection into cervical canal to confirm dilation completion. Procedure duration 6-7 minutes; achieves 8-9mm dilation in 5 minutes. Used in clinical settings by physicians to facilitate gynecological procedures. Benefits include rapid, controlled mechanical dilation of cervical canal.
Clinical Evidence
Bench testing only. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation). In vitro performance testing per ISO 10555-1 including dimensional verification, balloon deployment/retraction, burst pressure, fatigue, inflation/deflation time, bond strength, and stylet penetration resistance.
Technological Characteristics
Triple-balloon catheter; 10 French size; 25cm usable length. Materials: polymer shaft and balloons. Energy: manual saline inflation. Connectivity: none. Sterilization: EO (SAL 10-6). Standards: ISO 10993-1 (biocompatibility), ISO 10555-1 (catheter requirements).
Indications for Use
Indicated for patients requiring cervical softening and dilation for procedures including cervical stenosis treatment, IUD placement/removal, Radium placement, uterine cavity drainage, endometrial biopsy, uterine curettage, suction cannula aspiration, or operative hysteroscopy.
Regulatory Classification
Identification
A hygroscopic Laminaria cervical dilator is a device designed to dilate (stretch open) the cervical os by cervical insertion of a conical and expansible material made from the root of a seaweed (Laminaria digitata or Laminaria japonica). The device is used to induce abortion.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2017
GTIMD LLC Richard DeMello Quality Manager 6 Columbia Drive Amherst. NH 03031
Re: K160664 Trade/Device Name: Aqueduct 100 Cervical Dilation Balloon Catheter Regulation Number: 21 CFR 884.4260 Regulation Name: Hygroscopic Laminaria Cervical Dilator Regulatory Class: II Product Code: PON Dated: June 27, 2016 Received: July 1, 2016
Dear Richard DeMello.
This letter corrects our substantially equivalent letter of August 1, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Joyce M. Whang -S" in a large, sans-serif font. The text is black and appears to be the focus of the image. The background is a light blue color, which may be a watermark or part of the image's design.
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160664
Device Name
Aqueduct 100 Cervical Dilation Balloon Catheter
Indications for Use (Describe)
The Aqueduct 100 Cervical Dilation Balloon Catheter is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| <span style="display:inline-block; margin-right:5px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="display:inline-block; margin-left:20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "GTIMD" in a large, sans-serif font. The letters are a bright, light blue color. The letters are evenly spaced and appear to be a logo or brand name.
## 510(K) SUMMARY
| I. Submitter | |
|----------------------|---------------------------------------|
| Submitter's Name: | GTIMD LLC |
| Address: | 6 Columbia Drive<br>Amherst, NH 03031 |
| Phone: | (603) 880-0277 |
| Fax: | (603) 386-6422 |
| Contact Person: | Richard DeMello |
| Date of Preparation: | January 23, 2017 |
- II. Device
| Name of Device: | Aqueduct 100 Cervical Dilation Balloon Catheter |
|----------------------|-------------------------------------------------------------|
| Common Name: | Catheter, Balloon, Dilation of the Cervical Canal |
| Classification Name: | Hygroscopic Laminaria Cervical Dilator (21 CFR<br>884.4260) |
| Regulatory Class: | II |
| Product Code: | PON |
#### III. Predicate Devices
Primary Predicate Device: Dilapan-S (K143447) manufactured by Medicem Technology s.r.o.
Reference Device: Cook Cervical Ripening Balloon (K131206) manufactured by Cook, Inc.
These predicate devices have not been subject to a design-related recall.
{4}------------------------------------------------
#### IV. Device Description
The Aqueduct 100 Cervical Dilator is a triple-balloon cervical dilation catheter which enables the simultaneous dilation of both sides of the cervical canal. The catheter consists of a 3-lumen shaft. One lumen inflates an anchoring balloon. A second lumen inflates the 2 cylindrical dilation balloons. The third lumen is for infusion of saline solution and also contains a fixed stiffening stylet to add rigidity to the catheter. The stylet is permanently sealed within the infusion lumen. In use, the catheter is inserted through the vagina and cervical canal and into the uterus where the distal spherical balloon is inflated. The catheter is then withdrawn until the spherical balloon comes into contact with the internal orifice of the uterus, which positions and anchors the catheter within the cervix. This aligns the two cylindrical balloons at the internal cervical os and external cervical os. The balloons are inflated with saline providing gradual mechanical dilation of the cervix.
After 3 minutes of dilation of the internal and external orifices of the uterus, a gentle, controlled injection of 1ml, and up to 2.5ml, of saline can be made through the catheter infusion lumen which exits between the two balloons (and inside the cervical canal). Evidence that cervical dilation is complete can be determined once droplets of the saline injection are observed entering through the external opening of the cervix. Optimal dilation of 8-9mm within the cervical canal is typically achieved following 5 minutes of balloon dilation. The entire procedure from catheter insertion to removal is completed in 6-7 minutes.
The Aqueduct 100 Cervical Dilator is packaged in a mylar/Tyvek pouch and EO sterilized to SAL 10-6.
#### V. Indications For Use
The Aqueduct 100 Cervical Dilator is intended to be used whenever cervical softening and dilation is desired. Some examples are: treatment of cervical stenosis, IUD placement and removal, Radium placement, drainage of uterine cavity, endometrial biopsy, uterine curettage, suction cannula aspiration, operative hysteroscopy.
#### VI. Comparison of Technological Characteristics
The design and function of the Aqueduct 100 Cervical Dilator is substantially equivalent to Dilapan-S (K143447) manufactured by Medicem Technology s.r.o. Both devices have the same intended use,
{5}------------------------------------------------
"whenever cervical softening and dilation is desired", and perform the comparable function "to dilate up to a 12mm diameter, in the case of Dilapan-S and 9mm diameter for the Aqueduct 100". The materials and mechanism of dilation are different as Dilapan-S is a hydrophilic polymer which relies on absorption of moisture to expand to its full diameter. The Aqueduct 100 is a polymer balloon that is inflated with a fixed volume of fluid to expand to its maximum diameter.
These differences do not raise different questions of safety and effectiveness
The Aqueduct 100 Cervical Dilatoris substantially equivalent in material, design and mechanism of operation to the Cook Cervical Ripening Balloon (K131206) manufactured by Cook, Inc. Both devices have multi-lumen polymer shafts, two polymer dilation balloons located near the tip of the catheter, and a stiffening stylet. The Aqueduct 100 Cervical Dilator includes a third balloon for positioning and anchoring the catheter within the cervical canal. The dilation balloons on the Cook product are made from an elastomeric material which increases in size as more fluid is injected, so that they do not have a fixed length or diameter. The Aqueduct 100 Cervical Dilator has nonelastic, fixed dimension balloons that expand up to 12mm diameter. The catheter sizes are 18 French for the Cook catheter, while the Aqueduct 100 Cervical Dilator is 10 French. The useable length of the Cook catheter is 40cm. The Aqueduct 100 Cervical Dilator useable length is 25cm. Inflation volumes are 40ml to 80ml for the Cook device and 7.5ml for the Aqueduct 100.
These differences do not raise different questions of safety and effectiveness
- VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
Biocompatibility tests were conducted on the Aqueduct 100 Cervical Dilator according to the requirements of ISO 10993-1 Biological Evaluation of Medical Devices Part-1: Evaluation and Testing. The following tests were conducted:
- . Cytotoxicity
- . Sensitization
- . Irritation
In Vitro Performance Testing
{6}------------------------------------------------
In vitro performance tests were performed on the Aqueduct 100 Cervical Dilator according to the requirements of ISO 10555-1, Sterile, Single Use Intravascular Catheters Part 1: General Requirements.
In-vitro performance testing conducted on the Aqueduct 100 Cervical Dilator included:
- Dimensional Verification
- Balloon Preparation, Deployment and Retraction
- Balloon Burst Pressure
- Balloon Fatigue
- . Balloon Inflation/Deflation Time
- Catheter Bond Strength
- Resistance to Stylet Penetration
Conclusion
Based upon these biocompatibility and in vitro performance tests, the Aqueduct 100 Cervical Dilator has been shown to be substantially equivalent to the currently marketed Dilapan-S (K143447) manufactured by Medicem Technology s.r.o. and Cook Cervical Ripening Balloon (K131206) manufactured by Cook, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.