The Indy OTW Vascular Retriever is intended to snare a foreign body and withdraw it to a peripheral vascular location.
Device Story
Indy OTW Vascular Retriever is a percutaneous retrieval device consisting of a 4-loop nitinol/platinum wire snare, an inner catheter, and an outer Flexor sheath. Used in peripheral vascular procedures; operated by physicians under fluoroscopic guidance. Device is introduced over a 0.035-inch wire guide. The snare is deployed to capture foreign bodies (e.g., wire guides, catheter fragments) and withdraw them to a peripheral location. Radiopaque markers on the sheath and catheter tip assist in visualization. Benefits include safe removal of misplaced or fractured intravascular devices, reducing potential for embolization or vessel damage.
Clinical Evidence
Bench testing only. Testing included biocompatibility (cytotoxicity, sensitization, reactivity, toxicity, hemocompatibility, pyrogenicity), tensile strength of joints, simulated use in anatomical models, liquid leak pressure, structural integrity, radiopacity under fluoroscopy, and immersion corrosion testing. All results met predetermined acceptance criteria.
Technological Characteristics
4-loop nitinol/platinum wire snare; nylon radiopaque inner catheter; Flexor sheath (nylon outer, embedded steel coil, PTFE inner liner). Compatible with 0.035-inch wire guides. Radiopaque markers for fluoroscopic visualization. Sterilized via EtO (SAL 10^-6).
Indications for Use
Indicated for snaring and withdrawing foreign bodies to a peripheral vascular location in patients requiring percutaneous retrieval of objects such as wire guides, coils, balloons, catheters, or filters.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged to suggest a sense of movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 2, 2016
Cook Incorporated Mr. Chad Schulenburg Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, IN 47404
Re: K160593
Trade/Device Name: Indy OTW Vascular Retriever Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: MMX Dated: July 29, 2016 Received: August 1, 2016
Dear Mr. Schulenburg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Brian D. Pullin -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160593
Device Name Indy OTW™ Vascular Retriever
Indications for Use (Describe)
The Indy OTW Vascular Retriever is intended to snare a foreign body and withdraw it to a peripheral vascular location.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The red background is a rectangle with a diagonal cut on the bottom right corner.
510(k) SUMMARY As required by 21 CFR §807.92 Date Prepared: July 29, 2016
#### I. SUBMITTER
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|-------------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Chad Schulenburg |
| Applicant Address: | Cook Incorporated |
| | 750 Daniels Way |
| | Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 ext. 104073 |
| Contact Fax Number: | (812) 332-0281 |
| | |
## II. DEVICE
Trade Name: Common Name: Classification Name: Regulation/Class: Product Code
Indy OTWTM Vascular Retriever Percutaneous Retrieval Device Embolectomy Catheter 21 CFR §870.5150/Class II MMX
# III. PREDICATE DEVICE
The device subject of this submission is considered substantially equivalent to the predicate device, the 4-Loop Vascular Retriever Snare (K112185).
## IV. DEVICE DESCRIPTION
The Indy OTW™ Vascular Retriever consists of a 4-loop nitinol wire snare attached to an inner catheter (which is attached proximally to a metal cannula) and includes an outer Flexor sheath. The device is compatible with a 0.035 inch wire guide for over-the-wire introduction. The Flexor sheath has a radiopaque band incorporated within the sheath material to identify the location of the sheath's distal tip.
## V. INDICATIONS FOR USE
The Indy OTWIM Vascular Retriever is intended to snare a foreign body and withdraw it to a peripheral vascular location.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol to the right of the word. Below the word "COOK" is the word "MEDICAL" in white. The background is red.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Indy OTWTM Vascular Retriever and the predicate device, the 4-Loop Vascular Retriever Snare (K112185), are substantially equivalent in that these devices have the same intended use and similar technological characteristics.
| | PREDICATE DEVICE | SUBJECT DEVICE |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| | 4-Loop Vascular Retriever Snare<br>(K112185) | Indy OTW Vascular Retriever<br>(Subject of this Submission) |
| Regulation Number | 870.5150 | Identical |
| Product Code | MMX | Identical |
| Classification Name | Embolectomy Catheter | Identical |
| Class | II | Identical |
| Intended Use | Intended for use in the<br>cardiovascular system to manipulate<br>and retrieve foreign objects,<br>including, but not limited to, wire<br>guides, coils, balloons, catheters,<br>and filters. | Intended to snare a foreign body and<br>withdraw it to a peripheral vascular<br>location. |
| Model Number<br>Prefix | VRS | INDY |
| Snare Sub-Assembly | 5 Nitinol strands /<br>2 Tantalum strands | 5 Nitinol strands /<br>2 Platinum strands |
| Snare Loops | 4 | Identical |
| Inner Catheter | Radiopaque<br>Tetrafluoroethylene (TFE) | Nylon radiopaque tubing |
| Catheter<br>Length (cm) | 90 | 55, 100 |
| Sheath | Nylon | Flexor sheath:<br>Nylon outer layer,<br>Embedded steel coil,<br>PTFE inner liner |
| Outer Sheath<br>Size (Fr) | 10.0 | 8.0 |
| Connector Cap | Nylon | Identical |
#### Table 1: Comparison Table
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Cook Medical logo. The word "COOK" is in white, sans-serif, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif, uppercase letters on a darker red background. The logo is simple and modern.
| | PREDICATE DEVICE | SUBJECT DEVICE |
|-----------------------------------------------------|-------------------------------------------------|-------------------------------------------------------------|
| | 4-Loop Vascular Retriever Snare<br>(K112185) | Indy OTW Vascular Retriever<br>(Subject of this Submission) |
| Side Arm Assembly | Tuohy-Borst adapter material:<br>Polycarbonate | |
| | Male luer cap with strap material: | Identical |
| | High density polyethylene/polyvinyl<br>chloride | |
| Wire Guide<br>Compatibility<br>(diameter in inches) | N/A | 0.035 |
| Packaging | PETG tray<br>PET-PE / Tyvek pouch | Wire guide holder<br>PET-PE / Tyvek pouch |
| Sterilization Method | EtO | Identical |
| SAL | 10-6 | Identical |
| Shelf Life | 3 years | Identical |
The minor differences in dimensions, materials, and technology as compared to the predicate device were appropriately assessed and do not present any new questions of safety and/or effectiveness. The following devices were referenced to provide additional support: Flexor® Introducers (K142829), SeQure™ Snare System (K102484), Roadrunner® Uniglide® Hydrophilic Wire Guide (K110009), Approach® Pro ST Micro Wire Guides (K110163) and the Coda® Low Profile Balloon Catheter (K150970). Based on the comparison of the design, intended use, materials, fundamental technology, and principle of operation, the subject device is considered to be substantially equivalent to the currently marketed predicate device.
# VII. PERFORMANCE DATA
Biocompatibility testing and performance testing were conducted to ensure that the Indy OTWTM Vascular Retriever will not elicit an adverse biological response in patients and will reliably perform as intended.
- . Biocompatibility Testing - Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systematic Toxicity, Hemocompatibility, and Pyrogenicity tests were performed. All test results were acceptable and demonstrate that the device is biocompatible.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Cook Medical logo. The logo is white text on a red background. The word "COOK" is in a large, bold font, and the word "MEDICAL" is in a smaller font below it.
- 트 Tensile Strength Testing - Peak load values of joints between device components were in accordance with the predetermined acceptance criteria demonstrating that the joints have sufficient strength to remain intact during use.
- Simulated Use Testing The Indy OTW™ Vascular Retriever was capable of snaring and removing wire guides and catheter fragments or snaring and repositioning wire guides and catheters within an anatomical model in accordance with the instructions for use. The predetermined acceptance criteria were met.
- . Liquid Leak Pressure Testing –The sheath hub, Touhy-Borst, and catheter hub did not leak water while pressurized. The predetermined acceptance criterion was met.
- Structural Integrity Testing –The device maintained structural integrity under the conditions of the study. The predetermined acceptance criteria were met.
- 그 Radiopacity Testing - The catheter tip, sheath tip, and wire loops are visible under standard fluoroscopy. The predetermined acceptance criterion was met.
- Immersion Corrosion Testing – No evidence of corrosion was observed. The predetermined acceptance criterion was met.
# VIII. CONCLUSIONS
The results of the testing provide reasonable assurance that the subject device has been designed such that it will function as intended to snare a foreign body and withdraw it to a peripheral vascular location. The subject device does not raise new questions of safety or effectiveness as compared to the predicate device. This supports a determination of substantial equivalence of the subject device to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.