K160592 · Togo Medikit Co., Ltd. · FOZ · Aug 12, 2016 · General Hospital
Device Facts
Record ID
K160592
Device Name
SUPERCATH 6
Applicant
Togo Medikit Co., Ltd.
Product Code
FOZ · General Hospital
Decision Date
Aug 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi
Device Story
Intravascular catheter for venous/arterial access and fluid administration; single-use; short-term (<30 days). Features metallic introducer needle with side-notch for visual confirmation of vessel entry; built-in check valve in catheter hub reduces blood flashback during needle withdrawal; button-activated spring mechanism retracts needle into hub to prevent needlestick injuries. Includes slide component for catheter insertion assistance. Used by healthcare professionals in clinical settings. Output is vascular access; healthcare provider observes blood flashback in side-notch to confirm placement. Power injection capability (18-22G) allows use with contrast media injectors. Benefits include reduced needlestick risk and improved insertion control.
Clinical Evidence
Bench testing only. Testing included tensile strength (ISO 10555-1), air/liquid leakage, flow rate, power injection burst pressure, check valve leakage, and wing tensile strength. Biocompatibility testing included cytotoxicity, sensitization, systemic toxicity, irritation, pyrogenicity, implantation, genotoxicity, and hemocompatibility. No clinical data provided.
Indicated for patients requiring venous or arterial access for fluid administration. Suitable for short-term use (<30 days). 18-22 gauge catheters are indicated for use with power injectors at max 300 psi.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 12, 2016
TOGO MEDIKIT CO., LTD. c/o Izuni Maruo MIC International 4-1-17 Hongo Bunkyo-ku, Tokyo 113-0033 JAPAN
Re: K160592
Trade/Device Name: SUPERCATH 6 Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: II Product Code: FOZ Dated: July 13, 2016 Received: July 14, 2016
Dear Izumi Maruo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tina
Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160592
Device Name SUPERCATH 6
#### Indications for Use (Describe)
The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary-K160592
# a. Owner/Company name, address
TOGO MEDIKIT CO., LTD. 17148-6 Aza Kamekawa, Oaza Hichiya Hyuga City, Miyazaki, 883-0062, Japan
Daisuke Nagamizu Manager, Regulatory Affairs Section, Quality Assurance Department Phone: 011-81-982-53-8000 011-81-982-53-8008 Fax: Email: qc@togomedikit.co.jp
# b. Contact/Application Correspondent
Izumi Maruo Senior Consultant MIC International 4-1-17 Hongo, Bunkyo-ku Tokyo, 113-0033, Japan
Phone: 011-81-3-3818-8577 Fax: 011-81-3-3818-8573 maruo@mici.co.jp Email:
## c. Date prepared
February 29, 2016
# d. Name of device
| Trade Name: | SUPERCATH 6 |
|----------------------------|----------------------------------------------------------------------|
| Common Name: | Intravascular Catheter |
| Classification Name: | Catheter, Intravascular, Therapeutic<br>Short-term less than 30 days |
| Classification Regulation: | 21 CFR 880.5200 |
| Product Code: | FOZ |
| Classification Panel: | General Hospital |
{4}------------------------------------------------
## e. Predicate/reference devices
The SUPERCATH 6 is substantially equivalent to the following legally marketed devices:
| Predicate device | |
|------------------|-------------|
| 510(k): | K140419 |
| Trade name: | SUPERCATH 5 |
| Product code: | FOZ |
This submission uses a reference device for the power injection and changes in indications for use statement from the predicate device:
| Reference device | |
|------------------|---------------------------------------|
| 510(k): | K111236 |
| Trade name: | Introcan Safety® 3 Closed IV Catheter |
| Product code: | FOZ |
### f. Description of the device
The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days) and is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 - 22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.
The SUPERCATH 6 has technological characteristics as follows;
Its catheter hub has a built-in check valve which together with the healthcare professional's finger pressure on the blood vessel, assists to reduce blood flashback when the metallic introducer needle is withdrawn following blood vessel puncture. The built-in check valve is not intended to stop bleeding completely.
Pressing the button on the needle hub activates the coiled spring in the hub, retracting the metallic introducer needle into the needle hub.
The introducer needle of the SUPERCATH 6 has a side-notch to provide rapid visual confirmation of vessel entry. When the introducer needle is inserted into the vein, blood flows upward into side-notch and returns down along the inside of the catheter tube.
The SUPERCATH 6 has a slide in order to assist insertion of the catheter. The user can choose either to push the catheter hub forward or push the slide forward to insert the catheter. When pushing the slide, the slide is stopped 12 mm forward by a stopper.
The SUPERCATH 6 is available in 14G, 16G, 18G, 20G, 22G and 24G.
## g. Indications for Use
The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is
{5}------------------------------------------------
designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 -22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi.
## h. Statement of substantial equivalence
The SUPERCATH 6 is the modified version of the SUPERCATH 5 (K140419). The SUPERCATH 6 is intended to be used with power injectors. The intended use of the SUPERCATH 6 and the SUPERCATH 5 (K140419) differ only in the use for power injection.
Regarding the reference device, the SUPERCATH 6 has a similar intended use to the Introcan Safety® 3 Closed IV Catheter (K111236). Both devices act as an intravascular catheter for short term use (less than 30 days) and the same gauge sizes are suitable for use with power injectors at a maximum pressure of 300 psi.
The SUPERCATH 5 (K140419) and the SUPERCATH 6 have the identical needlestick injury prevention feature, flashback visualization system, side-notch needle, and reduction of blood exposure system.
The comparison table for technological characteristics among the SUPERCATH 6, the SUPERCATH 5 (K140419), and the Introcan Safety® 3 Closed IV Catheter (K111236) is shown below.
| Item | SUPERCATH 6 | The predicate device; SUPERCATH (K140419) | The reference device; Introcan Safety® 3 Closed IV Catheter (K111236) |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The SUPERCATH 6 is intended to access a vein or artery and to administer fluids. The SUPERCATH 6 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. The 18 -22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi. | The SUPERCATH 5 is intended to access a vein or artery and to administer fluids. The SUPERCATH 5 is designed for single use and for short-term use (less than 30 days), is designed to minimize inadvertent needlesticks and to reduce accidental needlesticks. | Introcan Safety® 3 Closed IV Catheter is inserted into a patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure or administer fluids and blood intravascularly. The 18-22 gauge catheters may be used with power injectors at a maximum pressure of 300 psi with a luer lock connection only. |
| Needlestick injury prevention feature | The metallic introducer needle is | The metallic introducer needle is | The passive safety needle-shielding |
Comparison table for technological characteristics
{6}------------------------------------------------
| | retracted into the<br>needle hub when the<br>activation button on<br>the needle hub is<br>pressed. | retracted into the<br>needle hub when the<br>activation button on<br>the needle hub is<br>pressed. | mechanism of the<br>Introcan Safety® 3 is<br>located inside the<br>catheter hub. Upon<br>withdrawal of the<br>needle, the safety<br>shield engages as the<br>needle passes through<br>the catheter hub and<br>deploys automatically<br>to shield the needle<br>tip. |
|-------------------------------------------|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flashback<br>visualization | Yes | Yes | Yes |
| Reduction of<br>blood<br>exposure | Yes (Check valve*¹) | Yes (Check valve*¹) | Yes (Septum) |
| Side-notch needle*² | Yes | Yes | No |
| Radiopaque | Yes | Yes | Yes |
| Use<br>with<br>power<br>injectors | Yes (18-22 gauge) | No | Yes (18-22 gauge) |
| Sterilization | Ethylene oxide | Ethylene oxide | Ethylene oxide |
| Packaging | Tear-off pouch | Blister | Unknown |
| Multiple gauge sizes<br>and needle length | Yes | Yes | Yes |
Note:
*1: Check valve assists to reduce blood flashback when the metallic introducer needle is withdrawn following blood vessel puncture together with the healthcare professional's finger pressure.
*2: Side-notch needle is the introducer needle with a side-notch, to provide rapid visual confirmation of vessel entry. When the introducer needle is inserted into the vein, blood flows upward into the side-notch and returns down along the inside of the catheter tube.
The modifications to the SUPERCATH 6 from the SUPERCATH 5 (K140419) include following changes of components and raw materials:
- . All materials of the catheter hubs of the SUPERCATH 6 are different from those of the SUPERCATH 5 (K140419).
- The SUPERCATH 6 has the slide and the slide cap as new components
- Composition of the lubricating oil on the catheter of the SUPERCATH 6 is different from ● that of the SUPERCATH 5 (K140419).
- Formulation of the PVC of the SUPERCATH 6 is different from that of the SUPERCATH 5 (K140419)
- The SUPERCATH 6 does not have an attachment which the SUPERCATH 5 (K140419) has. ●
Following table is a brief comparison of main components and materials between the SUPERCATH
{7}------------------------------------------------
6 and the SUPECATH 5 (K140419);
| Component name | Materials | |
|------------------------|-----------------|-----------------------|
| | SUPERCATH 6 | SUPERCATH 5 (K140419) |
| Inner needle | | |
| -Introducer needle | Stainless steel | Stainless steel |
| -Outer and Inner cover | Polycarbonate | Polycarbonate |
| -Needle hub | Polycarbonate | Polycarbonate |
| -Attachment | - | Polypropylene |
| -Slide | Polycarbonate | - |
| -Slide cap | Polypropylene | - |
| Catheter | | |
| -Catheter tube | Polyurethane | Polyurethane |
| -Lubricating oil | Silicon | Silicon |
| -Catheter hub | Copolyester | Polycarbonate |
| -Wing | PVC | PVC |
Comparison table for components and materials
The attachment is used just during manufacturing process of the SUPERCATH 5 (K140419). The attachment is used in order to adjust the distance between the bevel of the introducer needle and the tip of the catheter. After the adjustment, the attachment is fixed and users cannot move the attachment. Although the SUPERCATH 6 does not have the attachment, the distance between the bevel of the introducer needle and the tip of the catheter is managed in our quality system. Therefore, the attachment does not affect substantial equivalence between the predicate device and the SUPERCATH 6.
Each gauge size of catheter hubs of the SUPERCATH 5 (K140419) and the SUPERCATH 6 is color-coded. However, all materials of the catheter hubs of the SUPERCATH 6 are different from those of the SUPERCATH 5 (K140419).
The SUPERCATH 6 has the slide and the slide cap as new components. User can choose to push the catheter hub forward or push the slide forward to insert the catheter. When pushing the slide, a stopper stops the slide 12 mm forward.
In order to evaluate any effects of the above changes on substantial equivalence, biocompatibility testing, sterilization, shelf-life testing, and performance testing were conducted. In conclusion, those testing demonstrated that the SUPERCATH 6 is substantially equivalent to the the SUPERCATH 5 (K140419).
#### Bench Testing i.
The following bench tests were performed to verify conformity to the recognized standards and demonstrate substantial equivalence to the SUPERCATH 5 (K140419).
# ● Tensile Strength for the Catheter
The tensile strength of the catheter meets acceptable minimum force until breakage when tested according to ISO 10555-1.
{8}------------------------------------------------
- Air and Liquid Leakage for the Hub Attachment
The catheter hub is impervious to air/liquid infiltration when subjected to positive pressure and aspiration when tested according to ISO 10555-1.
- · Flow rate
The flow rate through catheter meets allowable limits when tested according to ISO 10555-1.
- Power injection
The catheter burst pressure exceeds the peak pressure present in the catheter at maximum flow conditions when tested according to ISO10555-1
- Leakage at the Check Valve under pressure
The built-in check valve is impervious to liquid infiltration when subjected to positive pressure according to in-house standard.
- Tensile Strength for the Wing The tensile strength of the wing meets acceptable minimum force until breakage when tested according to in-house standard.
#### j. Biocompatibility Testing
Because the modifications to the SUPERCATH 6 from the SUPERCATH 5 (K140419) include changes of components and raw materials, we performed following biocompatibility tests;
- Cytotoxicity
- · Sensitization
- Systemic (acute) toxicity
- Irritation/ Intracutaneous reactivity
- · Pyrogenicity
- Implantation
- Subchronic and subacute systemic toxicity
- · Genotoxicity
- · Hemocompatibility
- Physicochemical tests
In the biocompatibility testing reports, no biocompatibility concern was raised.
# k. Simulated Clinical Use
The SUPERCATH 6 modified from the SUPERCATH 5 (K140419) has no new sharps injury prevention features. Therefore, this submission does not include the simulated clinical use testing report.
#### l. Conclusion
Based on the above discussion and enclosed sections regarding substantial equivalence to predicate device, TOGO MEDIKIT concludes that the SUPERCATH 6 is substantially equivalent to the SUPERCATH 5 (K140419).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.