The Barone Jejunostomy Catheter Set is intended for patients greater than 12 years old requiring jejunal feeding and delivery of medication. The device can be placed via a surgically open or laparoscopic technique.
Device Story
The Barone Jejunostomy Catheter Set is a medical device used for enteral feeding and medication delivery in patients over 12 years old. The set includes a 10.2 Fr polyether-urethane jejunostomy catheter, a 5.5 Fr polyethylene stiffening cannula, a Coons wire guide, a dilator, entry access needles, and enteral connectors. The device is placed by a physician using either an open surgical or laparoscopic technique. The catheter features a tapered distal tip, six sideports, and a silicone suture wing for securement. The device functions as a conduit for delivering nutritional fluids or medications directly into the jejunum. It is designed to be MR Safe. The system includes specific design features to prevent misconnection with intravascular, hypodermic, or breathing system components, adhering to ISO 80369-1:2010 standards. The device benefits patients by providing a reliable, biocompatible access route for long-term enteral nutrition and medication administration.
Clinical Evidence
Bench testing only. Testing included tensile strength, fracture resistance, corrosion resistance, flex fatigue, liquid leakage, MRI compatibility (ASTM F2503), misconnection testing (ISO 80369-1:2010), torque testing, and biocompatibility (ISO 10993-1: cytotoxicity, sensitization, intracutaneous, systemic toxicity, subacute intraperitoneal toxicity, genotoxicity, and subcutaneous implant). All tests met predetermined acceptance criteria.
Technological Characteristics
Catheter: polyether-urethane, 10.2 Fr, 52 cm length, 6 sideports. Stiffener: 5.5 Fr polyethylene. Accessories: Coons wire guide, dilator, needles, enteral connectors. MR Safe per ASTM F2503. Biocompatibility per ISO 10993-1. Misconnection prevention per ISO 80369-1:2010. No software or electronic components.
Indications for Use
Indicated for patients >12 years old requiring jejunal feeding and medication delivery. Placed via open or laparoscopic surgery.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and dimension.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 18, 2017
Cook Incorporated Erum B. Nasir Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K160567
> Trade/Device Name: Barone Jejunostomy Catheter Set Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT, PIO, OCY Dated: April 20, 2017 Received: April 21, 2017
Dear Erum B. Nasir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160567
Device Name Barone Jejunostomy Catheter Set
Indications for Use (Describe)
The Barone Jejunostomy Catheter Set is intended for patients older than 12 years old requiring jejunal feeding and delivery of medication. The device can be placed via a surgically open or laparoscopic technique.
Type of Use (Select one or both, as applicable)
| ☑ Research and/or Educational Purposes |
|----------------------------------------|
| ☐ Commercial Purposes |
X | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The word "COOK" is in white, bold, sans-serif font and is placed on a red background. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font, also on a red background. The red background is a rectangle shape.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
## 510(k) SUMMARY
## Barone Jejunostomy Catheter Set 21 CFR §807.92 Date Prepared: May 17, 2017
### Submitted By:
Submission: Applicant: Contact: Applicant Address:
Traditional 510(k) Premarket Notification Cook Incorporated Erum Nasir Cook Incorporated 750 Daniels Way Bloomington, IN 47404 (812) 335-3575 x102607 (812) 332-0281
Contact Phone Number: Contact Fax Number:
### Device Information:
Trade Name: Common Name: Classification Name: Classification Regulations:
### Barone Jejunostomy Catheter Set
Tubes, Gastrointestinal (And Accessories) Gastrointestinal tube and accessories 21 CFR §876.5980, Product Codes KNT, PIO 21 CFR §876.1500, Product Codes OCY
Device Class/Classification Panel: Class II, Gastroenterology/Urology
## Predicate Devices:
- Primary Predicate: Compat® Jejunostomy Feeding Tube Kit (K991668) ●
- Additional Predicate: Flexiflo Laparoscopic Jejunostomy Kit (K925102) ●
## Device Description:
The Barone Jejunostomy Catheter Set is comprised of a jejunostomy catheter with a pre-loaded stiffening cannula, a Coons wire guide, a dilator, an enteral feeding adapter (funnel), two entry access needles, and a enteral transition connector. The Barone Jejunostomy Catheter is manufactured from polyether-urethane tubing. This catheter is a 10.2 Fr device with a proximal connector cap and a luer lock adapter, 6 distal sideports, a silicone suture wing, and a tapered distal tip; it has a length of 52 cm. The luer lock adapter is glued to the enteral feeding adapter (funnel) and thus is not functional for connecting to other devices. A 5.5 Fr polyethylene stiffener is also provided with the catheter and is matched for a transitional fit. The stiffener is designed with a pre-molded locking hub and a 0.035 inch diameter endhole.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font, stacked on top of the word "MEDICAL" in a smaller, white, sans-serif font. The text is set against a red background.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
### Indications for Use:
The Barone Jejunostomy Catheter Set is intended for patients greater than 12 years old requiring jejunal feeding and delivery of medication. The device can be placed via a surgically open or laparoscopic technique.
## Comparison to Predicate Devices:
The Barone Jejunostomy Catheter Set and the predicate devices, the Compat® Jejunostomy Feeding Tube Kit (K991668) and the Flexiflo Laparoscopic Jejunostomy Kit (K925102), are substantially equivalent in that these devices have similar design and indications for use. Additionally, the proposed device has the same technological characteristics and methods of placement as those of the predicate devices. The differences between the subject device and the predicate devices include the materials, dimensions, number of sideports, retention mechanism, and included accessories.
### Performance Data:
The following tests were performed to demonstrate that the proposed Barone Jejunostomy Catheter Set met the applicable design and performance requirements and support a determination of substantial equivalence.
- . Tensile Testing (catheter, dilator, and enteral feeding adapter and enteral connector) -Testing demonstrated that the peak load value was greater than the predetermined acceptance criterion.
- . Tensile Testing (catheter-matched stiffener, wire guide) - Testing demonstrated that the peak load value was greater than or equal to the predetermined acceptance criterion.
- . Fracture Testing (wire guide) – Testing demonstrated that no signs of fracture were observed in the region of interest, according to the predetermined acceptance criterion.
- . Corrosion Resistance (wire guide) - Testing demonstrated that there was no sign of corrosion that could affect the functional performance of the wire guide. The predetermined acceptance criterion was met.
- Resistance to Damage by Flex (wire guide) Testing demonstrated that there were no . signs of defects or damage, including flaking or material loss, when subjected to repeated flexing. The predetermined acceptance criterion was met.
- . Liquid leakage (enteral feeding adapter and enteral connector) – Testing demonstrated that the connection did not leak. The predetermined acceptance criterion was met.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, block letters in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" written in white, block letters against a black background.
- MRI Testing MRI compatibility was assessed by evaluating magnetic field interactions ● (displacement force and torque), artifact, and RF-induced heating. The Barone Jejunostomy Catheter and the Enteral Feeding Adapter (funnel) are MR Safe according to ASTM F2503.
- Misconnection Testing Testing shows that the connector should not provide a secure . connection when forcefully assembled to any surface of the components of and should easily disengage from each connector of every other application category specified in ISO 80369-1:2010 or should a connection engage, then greater that 75% of the infusate should leak from the misconnection between the connector and the reference connector. Based on this testing, the potential for the misconnections in the following applications has been addressed in the labeling:
- o Intravascular or hypodermic applications (such as female luer lock fittings or male luer slip fittings).
- o Breathing systems and driving gases applications
- Torque Test (Catheter hub and enteral feeding adapter) Testing demonstrated that the ● peak load value was greater than the predetermined acceptance criterion.
- . Liquid leakage (Catheter and enteral feeding adapter) – Testing demonstrated that the connection did not leak. The predetermined acceptance criterion was met.
- Biocompatibility Testing Testing (i.e., cytotoxicity, sensitization, intracutaneous, . systemic toxicity, subacute intraperitoneal toxicity, genotoxicity, and subcutaneous implant) shows that the devices are biocompatible. In conformance with the applicable sections of ISO 10993-1, the predetermined acceptance criteria were met.
## Conclusion:
The results of these tests support a conclusion that the Barone Jejunostomy Catheter Set met the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate devices, the Compat® Jejunostomy Feeding Tube Kit (K991668) and the Flexiflo Laparoscopic Jejunostomy Kit (K925102).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.