K160525 · Ray Co., Ltd. · OAS · Jun 20, 2016 · Radiology
Device Facts
Record ID
K160525
Device Name
RCT700
Applicant
Ray Co., Ltd.
Product Code
OAS · Radiology
Decision Date
Jun 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
CBCT, panoramic x-ray imaging system with cephalostat, is an extra oral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dental maxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D Image is obtained using the standard narrow beam technique.
Device Story
Extraoral dental X-ray system (CBCT, panoramic, cephalometric) for imaging teeth, jaw, and oral structures. Uses rotating C-arm with X-ray generator and flat panel detector to capture 2D projections; PC software reconstructs data into 3D volumes or 2D panoramic/cephalometric images. Operated by dentists in clinical settings. System includes adjustable patient positioning (height) and console PC for image display. Provides diagnostic images for implantology, TMJ studies, and orthodontics. Benefits include high-resolution maxillofacial visualization for treatment planning.
Clinical Evidence
Bench testing only. Clinical images were provided for illustrative purposes to confirm system functionality but were not required to establish substantial equivalence. Two licensed practitioners verified clinical image quality as acceptable for diagnostic purposes.
Technological Characteristics
Extraoral X-ray system; Class II; Product Code OAS. Detectors: Indirect (CsI, GADOX) and Direct (CdTe) types. Connectivity: Giga-Ethernet. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-63, IEC 60601-1-2. DICOM 3.0 compatible. Wall or floor mount.
Indications for Use
Indicated for dental radiographic examination of teeth, jaw, and oral structures, including panoramic, implantology, TMJ studies, and cephalometry in patients over 5 years old. No restrictions on gender, weight, or health condition.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three intertwined figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the caduceus in a circular fashion. The seal is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 20, 2016
RAY Co., Ltd. c/o Jeff Rongero, Staff Engineer UL LLC 12 LABORATORY DRIVE RESEARCH TRIANGLE PARK, NC 27709
Re: K160525
Trade/Device Name: RCT700 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: OAS Dated: May 6, 2016 Received: May 10, 2016
Dear Jeff Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160525
Device Name RCT700
#### Indications for Use (Describe)
CBCT, panoramic x-ray imaging system with cephalostat, is an extra oral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dental maxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D Image is obtained using the standard narrow beam technique.
| Type of Use (Select one or both, as applicable) | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> <b> ☑ </b> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </div></td></tr><tr><td><div> <span> ☐ </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div></td></tr></table> | <div> <span> <b> ☑ </b> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> ☐ </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
| <div> <span> <b> ☑ </b> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | | |
| <div> <span> ☐ </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
| Date: | December 30, 2015 |
|----------------|------------------------------------------------------------------------------|
| APPLICANT | RAY Co.,Ltd |
| ADDRESS | #332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 443-823, Korea |
| Manufacturer | RAY Co.,Ltd<br>332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 443-823, Korea |
| | TEL : +82-31-605-1000<br>FAX : +82-2-6280-5534 |
| Contact Person | Changhwan Lee<br>e-mail : ch0406.lee@raymedical.co.kr |
#### Device Name
Trade/Proprietary Name: RCT700 Common Name: Dental panoramic/tomography and cephalometric x-ray system
#### Classification
Classification Name: Computed tomography x-ray system (21 CFR 892.1750) Class : Il Product code : OAS Panel : Radiology
#### Predicate device
| Parameter | Predicated Device-1 | Predicated Device-2 |
|--------------|---------------------|---------------------|
| Device Name | RAYSCAN α-Expert3D | RAYSCAN α-Expert3D |
| Manufacturer | RAY Co., Ltd | RAY Co., Ltd |
{4}------------------------------------------------
| | 510(K) Number | | |
|----------------------|-------------------------------------|-------------------------------------|--|
| | K122981<br>Traditional 510k | K142247<br>Traditional 510k | |
| Classification name | Computed tomography x-ray<br>system | Computed tomography x-ray<br>system | |
| Regulation number | 892.1750 | 892.1750 | |
| Primary product code | OAS | OAS | |
#### Description
System purpose RCT700 is 3D computed tomography for scanning hard tissues like bone and teeth. By rotating the c-arm which is embedded with high voltage generator all-in-one x-ray tube and a detector on each end, CBCT images of dental maxillofacial is attained by recombining data from the same level that are scanned from different angle.
Panoramic image scanning function for attaining image of whole teeth, and cephalometric scanning option for attaining cephalometric image are included. RCT700 options combination are as shown in the table below.
#### Indication for use
CBCT, panoramic x-ray imaging system with cephalostat, is an extra oral source x-ray system, which is intended for dental radiographic examination of the teeth, jaw, and oral structures, specifically for panoramic examinations and implantology and for TMJ studies and cephalometry, and it has the capability, using the CBVT technique, to generate dental maxillofacial 3D images. The device uses cone shaped x-ray beam projected on to a flat panel detector, and the examined volume image is reconstructed to be viewed in 3D viewing stations. 2D Image is obtained using the standard narrow beam technique
"CAUTION: Federal law restricts this device to sale by or on the order of a dentist."
#### Patient population
The patient population can be the possible person who can be taken X-ray diagnostic radiation exposure. There is no restriction for ethnic group, Gender, weight, health, or condition. We recommend patients for x-ray diagnostic radiation exposure to be over 5 years old.
{5}------------------------------------------------
#### Device functions
The purpose of this unit is diagnose clinical structures of a tooth and head areas mainly by using the characteristics of permeability from X-ray. The principle of formations are as following. The machine is made of X-ray generator and arms in which transfers X-ray signals to a sensor in 2D. Also, an object that has a magnification is required in a distance. Moreover, the unit has to be adjustable depending on height of a patient and PC system to reconstruct an image.
The arm parts are controlled for rotating and linear moving to synchronize between the sensor and X-ray generator to get the image of interests. The purpose of this mechanism is to provide the images in 2D or/and 3D as preferred to diagnose in a monitor. CBCT provides in 3D images as reconstructed and Panorama is to diagnose the structures in a panoramic view. Cephalometric allows for orthodontic treatment. These 3 functions could be in 1 system, Panorama with Cephalometric, or Panorama only system depending on the needs. To provide the features as mentioned above, digital transferring from permeated X-ray to absorbing to the sensor is essential and all the process are proceed in Detector. Detector transfers X-ray to light depending on the structure materials. Detector is separated into indirect method that the light is changed to digital signals on photodiode and direct method in which the light is directly transferred to digital signal. This unit is using both direct method depending on the interior structure materials.
#### Operating principle of the detector
X-ray data entered into the detector from an outside source is converted to visible ray by detector's Scintillator and the visible ray is transferred to CMOS ASIC and converted to an Electric signal to form an image signal.
Then the signal is changed to a Digital form and through the high speed LAN cable, the acquired image data is transmitted to the PC.
Acquired resulting image is displayed on the monitor by the PC Software.
#### Detector Options
RCT700 detector, depending on the modality, consists of SIX 650HD-E (Manufacturer: THALES) in case of the CT and one of three detectors in case of PANO, C10500D-43 (Manufacturer: Hamamatsu), XID-C15DP (Manufacturer: i3System) or SIX 650HD-E (Manufacturer: THALES).
Ceph is categorized by Scan Ceph or One shot Ceph according to the imaging methodology and the detector implemented in RCT700 Scan ceph is the XID-C24DS (Manufacturer: i3system) model whereas the detector used in the One shot Ceph is PaxScan 4336X (Manufacturer: Varian medical systems) and PaxScan 2530C (Manufacturer: Varian medical systems).
In particular, the detector used in SIX 650HD-E may also be used as a detector for PANO. Meaning the SIX 650HD-E detector is capable of being used as an independent detector in CT or the PANO modality or in CT and PANO Modality simultaneously.
{6}------------------------------------------------
| Device Designation | |
|----------------------------------------------------------|-------------------------------------------------------------------|
| Division | Description |
| CT + Pano | CT(SIX 650HD-E) + Pano(C10500D-43) |
| CT + Pano | CT(SIX 650HD-E) + Pano(XID-C15DP) |
| CT + Pano + Scan Ceph | CT(SIX 650HD-E) + Pano(C10500D-43) + Scan Ceph(XID-C24DS) |
| CT + Pano + Scan Ceph | CT(SIX 650HD-E) + Pano(XID-C15DP) + Scan Ceph(XID-C24DS) |
| CT + Pano + One shot Ceph(L) | CT(SIX 650HD-E) + Pano(C10500D-43) + One shot Ceph(PaxScan 4336X) |
| CT + Pano + One shot Ceph(L) | CT(SIX 650HD-E) + Pano(XID-C15DP) + One shot Ceph(PaxScan 4336X) |
| CT + Pano + One shot Ceph(S) | CT(SIX 650HD-E) + Pano(C10500D-43) + One shot Ceph(PaxScan 2530C) |
| CT + Pano + One shot Ceph(S) | CT(SIX 650HD-E) + Pano(XID-C15DP) + One shot Ceph(PaxScan 2530C) |
| CT | CT(SIX 650HD-E) |
| CT + Scan Ceph | CT(SIX 650HD-E) + Scan Ceph(XID-C24DS) |
| CT + One shot Ceph(L) | CT(SIX 650HD-E) + One shot Ceph(PaxScan 4336X) |
| CT + One shot Ceph(S) | CT(SIX 650HD-E) + One shot Ceph(PaxScan 2530C) |
| CT,Pano | CT,Pano(SIX 650HD-E) |
| CT,Pano + Scan Ceph | CT,Pano(SIX 650HD-E) + Scan Ceph(XID-C24DS) |
| CT,Pano + One shot Ceph(L) | CT,Pano(SIX 650HD-E) + One shot Ceph(PaxScan 4336X) |
| CT,Pano + One shot Ceph(S) | CT,Pano(SIX 650HD-E) + One shot Ceph(PaxScan 2530C) |
| Remark | |
| *Pano: C10500D-43(Hamamatsu) | |
| *Pano: XID-C15DP(i3System) | |
| *CT: SIX 650HD-E(THALES) | |
| *Scan Ceph: XID-C24DS(i3system) | |
| *One shot Ceph(S): PaxScan 2530C(Varian medical systems) | |
| *One shot Ceph(L): PaxScan 4336X(Varian medical systems) | |
RCT700 options combinations are as shown in the table below.
## Comparison with predicate device
#### Similarities:
The proposed device has the same intended for use and technical characteristic as the predicate device. A comparative analysis indicates that the following categories are similar: the mode of operation, 3D technology, performance specification, main features [such as X-ray tube, High voltage generator, CT detector, PANO detector and CEPH (Scan type, one-shot type) detector], Collimator operation type (Auto),
{7}------------------------------------------------
Class I with type B applied parts according to IEC 60601-1, focal size, DICOM 3.0 format compatible, network method, type of installation, and applicable standards. (See Table 1 for compares the characteristics between the predicate device)
#### Differences:
In case of RCT700, detector implemented in the device varies according to its modality; CT, PANO, Ceph and the corresponding detectors are described below.
In case of CT, Predicate devices RAYSCAN α-Expert 3D[K122981, Traditional] and RAYSCAN α-Expert3D[K142247, Traditional] included the C10900D model, however the Proposed device RCT700 includes the SIX 650HD-E model detector.
In case of PANO, Predicate devices RAYSCAN α-Expert 3D[K122981, Traditional] and RAYSCAN α-Expert3D[K142247, Traditional] uses the C10500D model whereas the Proposed device RCT700 is capable of incorporating one of 3 types of detector models; C10500D, SIX 650HD-E or XID-C15DP.
Cephalometry is classified into Scan type or the One shot type depending on the imaging method and in case of the Scan type, Predicate device included the XID-C24DS detector which is an equivalent model to that of RAYSCAN α-Expert3D[K142247, Traditional].
In the One shot type, Predicate devices RAYSCAN α-Expert 3D[K122981, Traditional] included the SDX-4336CP model detector whereas RAYSCAN α-Expert3D[K142247, Traditional] and the Proposed device is capable of handling both the PaxScan 4336X and PaxScan 2530C detector models.
These 2 types of detectors are identical in technical characteristics, method and manufacturer and the only mentionable difference is that PaxScan 4336X detector is 42.4 (V) cm x 35.3 (H) cm in size and PaxScan 2530C detector is 24.9(V)cm x 30.2(H)cm.
Hereafter, the PaxScan 4336X mounted in the proposed device RCT700 One shot type will be referred to as Large size and PaxScan 2530C detector as the standard size.
Following is the comparison table listing the type of modality detectors mounted in Predicate devices RAYSCAN α-Expert 3D[K122981, Traditional] and RAYSCAN α-Expert3D[K142247, Traditional] and the optional detector types mounted in the proposed device RCT700.
| Division | Proposed Device | Predicate Device #1 | Predicate Device #2 | | |
|----------------------------------------------|--------------------------|--------------------------------|-------------------------|--------------------------------|---------------------|
| | RCT700 | RAYSCAN α-Expert3D | RAYSCAN α-Expert3D | | |
| | K160525,<br>Traditional | K122981,<br>Traditional | K142247,<br>Traditional | | |
| CT | Manufacturer | Thales | Hamamatsu | | |
| | Model | SiX 650HD-E | C10900D | Same as predicate<br>device #1 | |
| | Scintillator<br>Material | CsI (Indirect type) | CsI (Indirect type) | | |
| | Total pixel area | 144.0(W)x117.9(H)mm | 124.8(W)x124.8(H)mm | | |
| | Total pixel | 960x786 | 624x624 | | |
| | Pixel size | 150um | 200um | | |
| | | | | | |
| | Limiting<br>resolution | 3.3lp/mm | 2.5lp/mm | | |
| | Power<br>supply/Main | 5VDC / 7.5W | 5VDC / 1250mA | | |
| | MTF | 60% at 1LP/mm | 58% at 1LP/mm | | |
| | DQE | 0.45 at 1LP/mm | 0.22 at 1LP/mm | | |
| | VOXEL | 0.07~0.4mm | 0.28mm | | |
| | Manufacturer | | Hamamatsu | | |
| | Model | | C10500D | | |
| | Scintillator<br>Material | | CsI (Indirect type) | | |
| | Total pixel area | | 6.0(W)x1512(H)mm | | |
| PANO-1 | Total pixel | Same as predicate<br>device #1 | 60x1512 | Same as predicate<br>device #1 | |
| | Pixel size | | 100um | | |
| | Limiting<br>resolution | | 5.0lp/mm | | |
| | Power<br>supply/Main | | 5VDC / 1100mA | | |
| | MTF | | 70% at 1LP/mm | | |
| | DQE | | 0.5 at 1LP/mm | | |
| | Manufacturer | Thales | | | |
| | Model | SiX 650HD-E | | | |
| | Scintillator<br>Material | CsI (Indirect type) | | | |
| | Total pixel area | 144.0(W)x117.9(H)mm | | | |
| PANO-2 | Total pixel | N/A | | N/A | |
| | Pixel size | N/A | | | |
| | Limiting<br>resolution | 3.3lp/mm | | | |
| | Power<br>supply/Main | 5VDC / 7.5W | | | |
| | MTF | 60% at 1LP/mm | | | |
| | DQE | 0.45 at 1LP/mm | | | |
| PANO-3 | Manufacturer | i3System | | | |
| | Model | XID-C15DP | | | |
| | Scintillator<br>Material | CdTe(Direct type) | N/A | N/A | |
| | Total pixel area | 4.8(W)x150(H)mm | | | |
| | Total pixel | 48x1500 | | | |
| | Pixel size | 100um | | | |
| | Limiting resolution | 5.0lp/mm | | | |
| | Power supply/Main | 5VDC / 2A | | | |
| | MTF | 75% at 1LP/mm | | | |
| | DQE | 0.88 at 1LP/mm | | | |
| Ceph<br>(scan<br>type) | Manufacturer | | N/A | i3System | |
| | Model | | | XID-C24DS | |
| | Scintillator<br>Material | | | CdTe (Direct type) | |
| | Total pixel area | | | 4.8(W)x240(H)mm | |
| | Total pixel | Same as predicate<br>device #2 | | 48x2400 | |
| | Pixel size | | | 100um | |
| | Limiting resolution | | | 5.0lp/mm | |
| | Power supply/Main | | | 5VDC / 2A | |
| | MTF | | | 75% at 1LP/mm | |
| | DQE | | | 0.88 at 1LP/mm | |
| | Manufacturer | | Samsung Mobile Display | Varian | |
| | Model | | SDX-4336CP | PaxScan 4336X | |
| | Scintillator<br>Material | | SDX-4336CP | GADOX (Indirect type) | |
| Ceph<br>(One<br>shot )-1,<br>Large<br>Size | Total pixel area | Same as predicate<br>device #2 | | 43.2 x 36.0 cm | 427(W)x356(H)mm |
| | Total pixel | | | 2880 x 2400 | 3072x2560 |
| | Pixel size | | | 150 um | 139um |
| | Limiting<br>resolution | | | 3.3 lp/mm | 3.6lp/mm |
| | Power supply/Main | | | 100-240VAC, 50/60Hz | 100-240VAC, 47-63Hz |
| | MTF | | | 45% at 1LP/mm | 54% at 1LP/mm |
| | DQE | | | 0.41 at 1P/mm | 0.2 at 1LP/mm |
| | Manufacturer | | N/A | Varian | |
| | Model | Same as predicate<br>device #2 | | PaxScan 2530C | |
| Ceph<br>(One<br>shot)-2,<br>Standard<br>Size | Scintillator<br>Material | | | GADOX (Indirect type) | |
| | Total pixel area | | | | |
| | Total pixel | | | | |
| | Pixel size | | | | 139um |
| | Limiting<br>resolution | | | | 3.6lp/mm |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| Power<br>supply/Main | | 100-240VAC, 47-63Hz |
|----------------------|--|---------------------|
| MTF | | 54% at 1LP/mm |
| DQE | | 0.2 at 1LP/mm |
Information for detector own 510(K) number or system in which it was cleared 510(K) number.
| Division | Model | Manufacturer | Own 510(K)<br>number | System in which it was cleared<br>510(k) number. |
|---------------|------------------|--------------|----------------------|--------------------------------------------------------------------------------------------------|
| CT | SiX 650HD-E | Thales | No | 1) System name: Unknown<br>2) Manufacturer: Unknown<br>3) 510(K) Number: Unknown |
| Pano-1 | C10500D | Hamamatsu | No | 1) System name: RAYSCAN α-Expert3D<br>2) Manufacturer: RAY CO., LTD<br>3) 510(K) Number: K122981 |
| Pano-2 | SiX 650HD-E | Thales | No | 1) System name: Unknown<br>2) Manufacturer: Unknown<br>3) 510(K) Number: Unknown |
| Pano-3 | XID-C15DP | i3System | No | 1) System name: PAPAYA plus<br>2) Manufacturer: Genoray<br>3) 510(K) Number: K130419 |
| Scan Ceph | XID-C24DS | i3System | No | 1) System name: RAYSCAN α-Expert3D<br>2) Manufacturer: RAY CO., LTD<br>3) 510(K) Number: K131695 |
| One shot Ceph | PaxScan<br>4336X | Varian | No | 1) System name: RAYSCAN α-Expert3D<br>2) Manufacturer: RAY CO., LTD<br>3) 510(K) Number: K142247 |
| One shot Ceph | PaxScan<br>2530C | Varian | No | 1) System name: RAYSCAN α-Expert3D<br>2) Manufacturer: RAY CO., LTD<br>3) 510(K) Number: K142247 |
{11}------------------------------------------------
# Statement of Substantial Equivalence
| Parameter | Proposed Device | Predicated Device | Predicated Device | | Ceph<br>(Optional) | Posterior/Anterior<br>Lateral<br>SMV<br>Carpus<br>Reverse Town's<br>Waters<br>Lateral wide | PA,AP<br>Lateral<br>SMV<br>Carpus<br>Reverse Town's<br>Waters | Posterior/Anterior<br>Lateral<br>SMV<br>Carpus<br>Reverse Town's<br>Waters<br>Lateral wide | | Touch monitor (panel) | Touch monitor (panel) | Touch monitor (panel) | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------------------------------------------------------------------------|---------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Manufacturer | RAY Co., Ltd. | RAY Co., Ltd. | RAY Co., Ltd. | Detector<br>Type | CT | SIX 650HD-E | C10900D | C10900D | Detector | Detector<br>- CT<br>SIX 650HD-E<br>- PANO<br>C10500D<br>SIX 650HD-E<br>XID-C15DP<br>- Ceph<br>XID-C24DS(Scan)<br>PaxScan 4336X(One shot, Large Size)<br>PaxScan 2530C(One shot, Standard<br>Size) | Detector<br>- CT<br>C10900D<br>- PANO<br>C10500D<br>- Ceph<br>SDX-4336CP (One shot) | Detector<br>- CT<br>C10900D<br>- PANO<br>C10500D<br>- Ceph<br>XID-C24DS(Scan)<br>PaxScan 4336X(One shot, Large Size)<br>PaxScan 2530C(One shot, Standard<br>Size) | |
| Device name | RCT700 | RAYSCAN α-Expert3D | RAYSCANα-Expert3D | | PANO | C10500D | C10500D | C10500D | Chinrest | Chinrest | Chinrest | Chinrest | |
| 510(K) Number | K160525<br>(Traditional 510K) | K122981<br>(Traditional 510K) | K142247<br>(Traditional 510K) | | | SIX 650HD-E | | | Head rest | Head rest | Head rest | Head rest | |
| Common<br>Name | Dental panoramic/tomography and<br>cephalometric x-ray system | Dental panoramic/tomography and<br>cephalometric x-ray system | Dental panoramic/tomography and<br>cephalometric x-ray system | | | XID-C15DP | | | Automatic Collimator | Automatic Collimator | Automatic Collimator | Automatic Collimator | |
| Indications<br>for use | RCT700, panoramic x-ray imaging<br>system with cephalostat, is an extraoral<br>source x-ray system, which is intended<br>for dental radiographic examination of<br>the teeth, jaw, and oral structures,<br>specifically for panoramic examinations<br>and implantology and for TMJ studies<br>and cephalometry, and it has the<br>capability, using the CBVT technique, to<br>generate dento-maxillo-facial 3D<br>images.<br>The device uses cone shaped x-ray<br>beam projected on to a flat panel<br>detector, and the examined volume<br>image is reconstructed to be viewed in<br>3D viewing stations. 2D Images are<br>obtained using the standard narrow<br>beam technique. | RAYSCAN α-Expert 3D, panoramic x-<br>ray imaging system with cephalostat, is<br>an extraoral source x-ray system, which<br>is intended for dental radiographic<br>examination of the teeth, jaw, and oral<br>structures, specifically for panoramic<br>examinations and implantology and for<br>TMJ studies and cephalometry, and it<br>has the capability, using the CBVT<br>technique, to generate dento-maxillo-<br>facial 3D images.<br>The device uses cone shaped x-ray<br>beam projected on to a flat panel<br>detector, and the examined volume<br>image is reconstructed to be viewed in<br>3D viewing stations. 2D Images are<br>obtained using the standard narrow<br>beam technique. | RAYSCAN α-Expert 3D panoramic x-<br>ray imaging system with cephalostat, is<br>an extraoral source x-ray system, which<br>is intended for dental radiographic<br>examination of the teeth, jaw, and oral<br>structures, specifically for panoramic<br>examinations and implantology and for<br>TMJ studies and cephalometry, and it<br>has the capability, using the CBVT<br>technique, to generate dento-maxillo-<br>facial 3D images.<br>The device uses cone shaped x-ray<br>beam projected on to a flat panel<br>detector, and the examined volume<br>image is reconstructed to be viewed in<br>3D viewing stations. 2D Images are<br>obtained using the standard narrow<br>beam technique. | | Ceph (Scan) | XID-C24DS | | XID-C24DS | Exposure switch | Exposure switch | Exposure switch | Exposure switch | |
| Mode of<br>Operation | Continuous operation with<br>intermittent, stated permissible<br>loading | Continuous operation with<br>intermittent, stated permissible<br>loading | Continuous operation with<br>intermittent, stated permissible<br>loading | | Ceph (One shot) | PaxScan 4336X | SDX-4336CP | PaxScan 4336X | Emergency stop switch | Emergency stop switch | Emergency stop switch | Emergency stop switch | |
| 3D technology | CBCT Cone beam<br>Computed Tomography | CBCT Cone beam<br>Computed Tomography | CBCT Cone beam<br>Computed Tomography | | | PaxScan 2530C | | PaxScan 2530C | Console PC set | Console PC set | Console PC set | Console PC set | |
| Performance<br>Specification | | 1) CBCT Computed tomography<br>2) Panoramic<br>3) Cephalometric(optional)<br>- One shot type<br>- Scan type | 1) CBCT Computed tomography<br>2) Panoramic<br>3) Cephalometric(optional)<br>- One shot type | 1) CBCT Computed tomography<br>2) Panoramic<br>3) Cephalometric(optional)<br>- One shot type<br>- Scan type | Exposure switch<br>Type | | "Deadman" Button type | "Deadman" Button type | "Deadman" Button type | Automatic<br>Collimator | CT exams<br>Panoramic exams<br>Cephalometric exams | CT exams<br>Panoramic exams<br>Cephalometric exams | CT exams<br>Panoramic exams<br>Cephalometric exams |
| Functional<br>Option | | Refer to Device Designation table in<br>description section of 510K summary. | Base<br>RAYSCAN α-3D :<br>CT+PANO<br>Option(CEPH)<br>RAYSCAN α-Multi 3D:<br>CT + PANO + One shot | Base<br>RAYSCAN α-3D :<br>CT+PANO<br>Option(CEPH)<br>RAYSCAN α-SM3D:<br>CT + PANO + SCAN CEPH<br>RAYSCAN α-M3DS:<br>CT + PANO + One shot(9.8 X 11.9<br>inch)<br>RAYSCAN α-M3DL:<br>CT + PANO + One shot(16.8 X 14.0<br>inch) | Main<br>Components | | Ceph Apparatus | Ceph Apparatus | Ceph Apparatus | Display Type | TFT LCD type(Normally black)<br>*1280x800 pixel | TFT LCD type(Normally black)<br>*1280x800 pixel | TFT LCD type(Normally black)<br>*1280x800 pixel |
| Exam<br>Mode | CT | Jaw(Standard)<br>Jaw-Fast<br>Surgical guide<br>Facial<br>Teeth<br>Endodontics<br>TMJ<br>Sinus<br>Airway | Standard<br>Surgical guide<br>Implant surgery<br>Endo treatment<br>TMJ<br>Sinus | Jaw(Standard)<br>Surgical guide<br>Implant surgery<br>Endo treatment<br>TMJ<br>Sinus | | | Vertical Carriage | Vertical Carriage | Vertical Carriage | Class | Class I with type B applied parts<br>according to IEC 60601-1 | Class I with type B applied parts<br>according to IEC 60601-1 | Class I with type B applied parts<br>according to IEC 60601-1 |
| | PANO | Standard(Normal)<br>TMJ<br>Sinus<br>Segmentation (Individual Tooth)<br>Bitewing<br>Orthogonal | Normal(Adult)<br>Pedodontics(Children)<br>TMJ<br>Sinus | Standard(Normal)<br>TMJ<br>Sinus<br>Segmentation (Individual Tooth)<br>Bitewing<br>Orthogonal | | | Rotator | Rotator | Rotator | Focal size | 0.5 | 0.5 | 0.5 |
| X-RAY Generator | X-RAY Generator | X-RAY Generator | | | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.