← Product Code [OAS](/productcode/OAS) · K160506

# Planmeca ProMax 3D Max, Planmeca Maximity (K160506)

_Planmeca Oy · OAS · May 20, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160506

## Device Facts

- **Applicant:** Planmeca Oy
- **Product Code:** [OAS](/productcode/OAS.md)
- **Decision Date:** May 20, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Planmeca ProMax 3D Max or Maximity is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital X-ray images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.

## Device Story

Planmeca ProMax 3D Max/Maximity uses cone beam computed tomography (CBCT) for 3D imaging and tomosynthesis for 2D imaging; captures cylindrical volume of data via several hundred sample images from different directions; data processed via separate 3D reconstruction PC; images viewed in Planmeca Romexis software on separate PC system; connected via Ethernet; operated by healthcare professionals in clinical settings; provides diagnostic support for dento-maxillo-facial, cervical spine, and ENT regions; aids clinical decision-making through visualization of anatomical structures.

## Clinical Evidence

Bench testing performed; clinical images of nose, sinuses, airways, middle ear, temporal bone, and vertebrae were acquired and reviewed by a specialist; images deemed to be of clinically usable diagnostic quality.

## Technological Characteristics

Cone beam computed tomography (CBCT) and tomosynthesis imaging; Ethernet connectivity; system consists of X-ray unit, 3D reconstruction PC, and Planmeca Romexis software workstation.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Planmeca ProMax 3D Max ([K093590](/device/K093590.md))
- Planmeca ProMax 3D Max ([K103659](/device/K103659.md))
- Carestream CS 9300

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 20, 2016

Planmeca Oy % Lars Moring Regulatory Affairs Manager Asentaiankatu 6 Helsinki, 00880 FINLAND

Re: K160506

Trade/Device Name: Planmeca ProMax 3D Max, Planmeca Maximity Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: OAS Dated: May 2, 2016 Received: May 5, 2016

Dear Lars Moring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

For

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration |  |
|-------------------------------------------------------------------------|--|
| <b>Indications for Use</b>                                              |  |
| Form Approved: OMB No. 0910-0120                                        |  |
| Expiration Date: January 31, 2017<br>See PRA Statement below.           |  |

| 510(k) Number (if known)       | K160506                                                                                                                                                                                                                                                                                                                               |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | Planmeca ProMax 3D Max<br>Planmeca Maximity                                                                                                                                                                                                                                                                                           |
| Indications for Use (Describe) | The Planmeca ProMax 3D Max or Maximity is a system intended to produce two-dimensional (2D) and three- dimensional (3D) digital X-ray images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients. |

| Type of Use (Select one or both, as applicable)                                                 |                                                                                     |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  
Food and Drug Administration  
Office of Chief Information Officer  
Paperwork Reduction Act (PRA) Staff  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

PSC Publishing Services (301) 443-6740 EF

a marka mana mana mana mana mana mara mara mara mara mara mara mara mara mar

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "PLANMECA" in white letters against a black background. The letters are bold and sans-serif. The word appears to be a logo or brand name.

# 510K) SUMMARY

# DATE

May 2, 2016

# PRODUCT, CLASSIFICATION NAME

Trade name: Planmeca ProMax 3D Max or Planmeca Maximity Common name: Computed tomography x-ray system Classification: OAS, Class II Regulation number: 892.1750

# MANUFACTURER

Planmeca Oy Asentajankatu 6 FI-00880 Helsinki, Finland Phone: +358 20 7795 500 Fax: +358 20 7795 396 Contact person: Lars Moring

## UNITED STATES SALES REPRESENTATIVE (U.S. DESIGNATED AGENT)

Planmeca USA Inc. 100 North Gary Avenue, Suite A Roselle, IL 60172 Phone: (630) 529 2300 Fax: (630) 529 1929 Contact person : Bob Pienkowski

#### INTENDED USE

The Planmeca ProMax 3D Max or Maximity is a system intended to produce two-dimensional (2D) and three-dimensional (3D) digital x-ray images of the dento-maxillo-facial, cervical spine and ENT (Ear, Nose, and Throat) regions at the direction of healthcare professionals as diagnostic support for pediatric and adult patients.

#### PRODUCT DESCRIPTION

The Planmeca ProMax 3D Max or Planmeca Maximity -x-ray unit uses cone beam computed tomography (CBCT) to produce three-dimensional (3D) images of the maxillofacial and ENT anatomies. Two dimensional (2D) images are produced with tomosynthesis method. In CBCT a cylindrical volume of data is captured in one imaging procedure. The data consists of several hundred sample images which are taken from different directions to cover a certain preprogrammed target area. These samples are used for 3D reconstruction (using a separate 3D reconstruction PC) that can be viewed in three dimensions in the Planmeca Romexis program (on a separate PC system). Ethernet computer networking technology is used.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "PLANMECA" in bold, white letters against a black background. The font is sans-serif and appears to be a custom or stylized typeface. The letters are closely spaced, creating a solid block of text. The overall impression is clean and modern.

# PREDICATE DEVICE

We consider this product modification to be similar in design, composition and function to the following devices introduced into commercial distribution after May 28, 1976:

Planmeca ProMax 3D Max # K093590 # K103659 Carestream CS 9300

# SUBSTANTIAL EQUIVALENCE

The product is principally just the same as in the previous 510(k) #K093590. The indications for use are expanded to include the ENT area.

The expanded intended use of the Planmeca ProMax 3D Max and the predicate device #K103659 is identical. The compared technical features for imaging technology, FOV, imaging parameters, resolution, and other basic characteristics are matching very closely, and the differences are so small that they do not have any effect on performance in practice. Both devices conform to given international performance standards.

The performance testing for ENT imaging applications was carried out taking clinical images for nose, sinuses, airways, middle ear, temporal bone and vertebrae. The images were reviewed by a specialist and were deemed to be of a clinically usable diagnostic quality.

# CONCLUSION

The comparison of characteristics supports substantial equivalence. Planmeca ProMax 3D Max is as safe and effective as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K160506](https://fda-staging.innolitics.com/device/K160506)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
