K160485 · STERIS Corporation · JOJ · Jun 1, 2016 · General Hospital
Device Facts
Record ID
K160485
Device Name
VERIFY VH2O2 Indicator Tape
Applicant
STERIS Corporation
Product Code
JOJ · General Hospital
Decision Date
Jun 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter. The tape may be used in the following sterilization cycles: - Non Lumen, Flexible and Lumen Cycles of the V-PRO 1, 1 Plus, maX and 60 Low Temperature Sterilizers - Standard Cycle of the STERRAD® 100S Sterilizer - Standard and Advanced Cycles of the STERRAD® NX Sterilizer - Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
Device Story
VERIFY™ VH2O2 Indicator Tape is a chemical process indicator used to secure wrapped medical items during hydrogen peroxide sterilization. The tape consists of coated crepe paper with diagonal lines of a chemical indicator. Upon exposure to hydrogen peroxide, the indicator undergoes a visual color change from pink to peach, yellow, or lighter, allowing clinicians to differentiate between processed and unprocessed packs. The device is used in clinical settings by healthcare staff. It provides a visual confirmation of exposure to the sterilization process, aiding in the management of sterile inventory and ensuring that only processed items are used in clinical procedures.
Clinical Evidence
Bench testing only. Performance demonstrated via ANSI/AAMI/ISO 11140-1:2014 and ANSI/AAMI/ISO 18472:2006 standards. Testing included simulated use in V-PRO and STERRAD sterilizers (pass/fail conditions), endpoint stability, light stability, tape adhesion (ASTM standards), biocompatibility (peroxide residual levels), off-set testing, and shelf-life evaluation. All tests passed.
Technological Characteristics
Substrate: Crepe paper. Indicator: Proprietary chemical indicator conforming to ANSI/AAMI/ISO 11140-1:2014 Type 1 requirements. Adhesive: Proprietary. Form factor: Tape. Free of lead, heavy metals, and natural rubber latex. Storage: 16-24°C, 30-60% RH.
Indications for Use
Indicated for use by healthcare providers to secure synthetic sterilization wraps and distinguish between processed and unprocessed units in specific low-temperature hydrogen peroxide sterilization cycles.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
STERRAD® Sealsure™ Chemical Indicator Tape (K111519)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a wing-like shape above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 1, 2016
Steris Corporation Anthony Piotrkowski Senior Manager, Regulatory Affairs 5960 Heisley Road Mentor, Ohio 44060
Re: K160485
Trade/Device Name: VERIFY™ VH2O2 Indicator Tape Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: April 29, 2016 Received: May 2, 2016
Dear Mr. Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160485
Device Name VERIFY™ VH2O2 Indicator Tape
Indications for Use (Describe)
The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, yellow or lighter.
The tape may be used in the following sterilization cycles:
- · Non Lumen, Flexible and Lumen Cycles of the V-PRO 1, 1 Plus, maX and 60 Low Temperature Sterilizers
- · Standard Cycle of the STERRAD® 100S Sterilizer
- · Standard and Advanced Cycles of the STERRAD® NX Sterilizer
- · Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX Sterilizer
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "STERIS" in a bold, serif typeface. A registered trademark symbol is located to the upper right of the final "S". The text is black against a white background.
Image /page/3/Picture/2 description: The image shows a logo with a series of horizontal, curved lines stacked on top of each other. The lines are all the same color, a light blue, and they create a wave-like pattern. The logo is simple and abstract, with no text or other elements present.
# 510(k) Summary for VERIFY™ VH2O2 Indicator Tape
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (216) 354-2600 Fax No: (216) 639-4459
Anthony Piotrkowski Contact: Sr. Manager, Regulatory Affairs Telephone: (440) 392-7437 Fax No: (440) 357-9198 E-mail: tpiotrko@steris.com
> Submission Date: May 28, 2016
K160485
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
## TRADITIONAL PREMARKET NOTIFICATION [510(k)] VERIFY™ VH2O2 Indicator Tape
#### Device Name 1.
| Trade Name: | VERIFYTM <span style="border: 1px solid black;">VH2O2</span> Indicator Tape |
|----------------------|-----------------------------------------------------------------------------|
| Common/Usual Name: | Chemical Sterilization Process Indicator |
| Classification Name: | Sterilization Process Indicator<br>(21 CFR 880.2800, JOJ) |
#### 2. Predicate Device
STERRAD® Sealsure™ Chemical Indicator Tape (K111519)
#### Description of Device 3.
The VERIFY VH2O2 Indicator Tape consists of a special coated crepe paper with diagonal lines of a chemical indicator specific for hydrogen peroxide sterilization. It meets the performance specifications for a Type 1 process indicator as defined in ANSI/AAMI/ISO 11140-1:2014.
The VERIFY | VH2O2 Indicator Tape and its packaging are manufactured in the absence of lead, heavy metals, natural rubber latex and dry natural rubber.
#### 4. Intended Use/ Indications for Use
The VERIFY™ VH2O2 Indicator Tape is intended to secure items wrapped in synthetic wrap materials until used and to distinguish between processed and unprocessed units through a color change from the start color (pink) to peach, vellow or lighter.
The tape may be used in the following sterilization cycles:
- . Non Lumen, Flexible and Lumen Cycles of the V-PRO 1, 1 Plus, maX and 60 Low Temperature Sterilizers
- Standard Cycle of the STERRAD® 100S Sterilizer ●
- Standard and Advanced Cycles of the STERRAD® NX Sterilizer ●
- Express, Flex Scope, Duo and Standard Cycles of the STERRAD® 100 NX ● Sterilizer
{5}------------------------------------------------
## TRADITIONAL PREMARKET NOTIFICATION [510(k)] VERIFY™ VH2O2 Indicator Tape
| Proposed | Predicate (K111519) STERRAD<br>SEALSURE Chemical Indicator Tape | Comparison |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VERIFY VH2O2 Indicator Tape | | |
| The VERIFY VH2O2 Indicator Tape<br>is intended to secure items wrapped<br>in synthetic wrap materials until used<br>and to distinguish between processed<br>and unprocessed units through a<br>color change from the start color<br>(pink) to peach, yellow or lighter.<br>The tape may be used in the<br>following sterilization cycles:<br>• Non Lumen, Flexible and Lumen<br>Cycles of the V-PRO 1, 1 Plus,<br>maX and 60 Low Temperature<br>Sterilizers<br>• Standard Cycle of the<br>STERRAD® 100S Sterilizer<br>• Standard and Advanced Cycles of<br>the STERRAD® NX Sterilizer<br>• Express, Flex Scope, Duo and<br>Standard Cycles of the | STERRAD SEALSURE Chemical<br>Indicator Tape (PN 14202) is a process<br>indicator (ISO 11140-1:2205) intended<br>for use by healthcare providers to secure<br>non-woven sterilization packs and wraps<br>to be sterilized in the STERRAD<br>Sterilization Systems.<br>• Standard Cycle of STERRAD 100S<br>• Standard Cycle of STERRAD 50<br>• Standard Cycle of STERRAD 200<br>• Standard and Advanced Cycles of<br>STERRAD NX<br>• Standard, Flex, DUO and EXPRESS<br>Cycles of STERRAD 100NX<br>The color of the STERRAD SEALSURE<br>Chemical Indicator Tape changes from<br>red to gold (or lighter) when exposed to<br>hydrogen peroxide and is intended to<br>differentiate between processed and<br>unprocessed loads. | Both devices are<br>chemical indicator<br>tapes used to secure<br>packs for<br>sterilization in<br>hydrogen peroxide<br>sterilizers and to<br>indicate through a<br>color change that the<br>packs have been<br>exposed to hydrogen<br>peroxide. Testing is<br>included in this<br>submission to<br>demonstrate safe and<br>effective<br>performance of the<br>proposed device in<br>the indicated cycles. |
#### 5. Description of Safety and Substantial Equivalence
The VERIFY VH2O2 Indicator Tape has equivalent characteristics as compared to its predicate device, STERRAD® Sealsure™ Chemical Indicator Tape. Table 5.2 compares the proposed device's design to the predicate.
| Characteristic | Proposed | Predicate | Comparison |
|-----------------------|------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Substrate | Crepe Paper | Crepe Paper | Same |
| Indicator | Proprietary - conforms<br>with ISO 11140-1:2014<br>Type 1 requirements | Proprietary - conforms<br>with ISO 11140-1:2005<br>Class 1 requirements | Testing included in this<br>submission demonstrating<br>performance (See Table<br>5.3). |
| Adhesive | Proprietary | Proprietary | Testing included in this<br>submission demonstrating<br>performance (See Table<br>5.3). |
| Color Change | Pink to Peach/Yellow | Red to Gold | Similar start and endpoint<br>colors |
| Detection | hydrogen peroxide dose | hydrogen peroxide dose | Same |
| Shelf-life | 3 months to be<br>extended as real time<br>data allows to a<br>projected 24 months | 22 months | Testing included in this<br>submission to support<br>proposed device's shelf life<br>when stored under labeled<br>conditions (See Table 5.3). |
| Storage<br>Conditions | 16 - 24°C,<br>30 - 60% RH | 15 - 30°C,<br>Dry Conditions | |
Table 5.2 Device Comparison
{6}------------------------------------------------
### TRADITIONAL PREMARKET NOTIFICATION [510(k)] VERIFY™ VH2O2 Indicator Tape
The VERIFY VH2O2 Indicator Tape was tested in nonclinical testing. Performance of the chemical indicator, CI, was demonstrated by testing in an ANSI/AAMI/ISO 18472:2006-conforming resistometer according to the vapor-phased hydrogen peroxide, Type 1 chemical indicator requirements in ANSI/AAMI/ISO 11140-1:2014. CI performance was also tested in hydrogen peroxide sterilizers (V-Pro and STERRAD) with loads under simulated use conditions under pass and fail conditions. Endpoint stability was tested by evaluating the color of processed tape at various time points when stored at room temperature. Light stability was demonstrated by evaluating the tape after storage outside its packaging before exposure to peroxide sterilization. Tape adhesion was evaluated using ASTM adhesion standard and under simulated use conditions on wrapped loads processed through the sterilizer. Biocompatibility was demonstrated by testing peroxide residual levels of processed tape samples and Off-set testing per ANSI/AAMI/ISO 11140-1:2014. Shelf-life testing evaluated CI performance per ANSI/AAMI/ISO 11140-1:2014 and tape adhesion after various times of storage within the labeled storage conditions. Table 5.3 summarizes the testing performed to demonstrate substantial equivalence.
| Test | Result |
|-------------------------------------------------------------------------------------------------|--------|
| ANSI/AAMI/ISO 11140-1:2014 Vaporized Hydrogen Peroxide Process<br>Indicator Performance Testing | Pass |
| Simulated Use: Pass Performance Testing | Pass |
| Simulated Use: Fail Performance Testing | Pass |
| Endpoint Stability | Pass |
| Light Stability after Removal from Packaging | Pass |
| Adhesion Evaluation | Pass |
| Shelf Life Evaluation | Pass |
| Hydrogen Peroxide Residual Testing | Pass |
| Off-set (transference) Performance Testing | Pass |
| | | Table 5.3 Summary of Nonclinical Tests: | |
|--|--|-----------------------------------------|--|
|--|--|-----------------------------------------|--|
#### 6. Conclusion
The VERIFY | VH2O2 Indicator Tape is substantially equivalent to the predicate device, STERRAD® Sealsure™ Chemical Indicator Tape. Based on the performance testing provided in this submission and summarized above, the subject device is as safe and as effective as the legally marketed, claimed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.