K160434 · Mri Interventions, Inc. · HAW · Mar 17, 2016 · Neurology
Device Facts
Record ID
K160434
Device Name
ClearPoint System
Applicant
Mri Interventions, Inc.
Product Code
HAW · Neurology
Decision Date
Mar 17, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The ClearPoint System is intended to provide stereotactic guidance and operation of instruments or devices during the planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners.
Device Story
ClearPoint System provides MRI-guided stereotactic navigation for neurosurgical procedures; includes workstation laptop with planning software, SMARTGrid planning grid, and SMARTFrame trajectory frame. Physician uses MRI images to plan trajectories; system provides real-time feedback via MRI-visible fiducial markers and targeting cannula. Hardware components (SMARTFrame, drill guides, accessories) are single-use and sterile. System used in MRI suite; physician operates workstation to guide instruments (biopsy needles, catheters, electrodes) to target. Output is visual trajectory guidance on workstation; assists in precise instrument placement. Benefits include accurate, real-time MRI-guided neurosurgical intervention.
Clinical Evidence
Bench testing only. Verification testing performed on 4.5mm and 6.0mm drill guides and Wharen Centering Guide. Results confirmed device functions as intended and maintains targeting accuracy of ±1.5mm at ≤125mm depth.
Technological Characteristics
Materials: Ultem, PEEK, titanium. Energy: MRI-compatible mechanical guidance. Connectivity: Workstation laptop. Sterilization: Ethylene oxide (EO) 10-6 SAL. Standards: ISO 10993-1 (biocompatibility), ISO 11135-1 (sterilization). Form factor: Stereotaxic frame with planning grid and hand controller.
Indications for Use
Indicated for patients undergoing neurological procedures requiring stereotactic guidance, including biopsies, catheter insertion, and electrode insertion, within 1.5T or 3.0T MRI environments.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three heads, each head facing forward. The eagle's body is formed by flowing lines, giving it a dynamic and modern appearance.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002
March 24, 2016
MRI Interventions. Inc. % John J. Smith, MD, JD Partner Hogan Lovells US LLP Columbia Square 555 Thirteenth Street, NW Washington, District of Columbia 20004
Re: K160434 Trade/Device Name: ClearPoint System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: II Product Code: HAW, ORR Dated: February 16, 2016 Received: February 16, 2016
Dear Dr. John Smith:
This letter corrects our substantially equivalent letter of March 17, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go
to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S 同心
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K160434
Device Name
#### ClearPoint System
Indications for Use (Describe)
The ClearPoint System is intended to provide stereotactic quidance and operation of instruments or devices during the planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
[ Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
{3}------------------------------------------------
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR Part 807.92 for the MRII ClearPoint System.
### 1. Company Making the Submission:
| Name of Owner: | MRI Interventions, Inc. |
|----------------|------------------------------|
| Address: | 5 Musick<br>Irvine, CA 92618 |
| Telephone: | 949-900-6833 |
| Fax: | 949-900-6834 |
| Contact: | Pete Piferi |
| E-mail: | ppiferi@mriinterventions.com |
| Date Prepared: | March 15, 2016 |
### 2. Device Name:
| Common Name: | Neurological Stereotaxic<br>Instrument |
|----------------------|----------------------------------------|
| Proprietary Name: | ClearPoint System |
| Classification Name: | Stereotaxic Instrument |
| Regulatory Class: | II |
| Regulation Number: | 21 C.F.R. § 882.4560 |
| Product Code: | HAW, ORR |
## 3. Predicate Device:
MRII ClearPoint System, K142505
## 4. Device Description:
The ClearPoint System is comprised of a workstation laptop with software, the SMARTGrid™ MRI-Guided Planning Grid, the SMARTFrame™ MRI-Guided Trajectory Frame, the SMARTFrame™ Accessory Kit and the SMARTFrame™ Hand Controller.
The SMARTGrid and associated Marking Tool are designed to assist the physician to precisely position the entry hole as called out in the trajectory planning software. The SMARTFrame is an Adjustable Trajectory Frame (ATF) that provides the quidance and fixation for neurosurgical tools. The MRI visible fluids of the Targeting Cannula along with the fiducial markers in the base of the frame allows for trajectory feedback when the physician views the MRI images, makes changes and confirms with subsequent MR images.
The ClearPoint System can be used with any MRI-compatible head fixation frame to immobilize the patient's head with respect to the scanner table, as well as with any imaging coil(s) that meet the physician's desired imaging quality. MRI Interventions also supplies an optional head fixation frame and imaging coil(s) that can be used with the ClearPoint System.
The ClearPoint Workstation includes the following:
- 1. ClearPoint Workstation Software (for trajectory planning and monitoring)
- 2. Laptop Computer
{4}------------------------------------------------
The hardware components of the ClearPoint System are the SMARTFrame and Accessories. They are all single use devices that are provided sterile. They include the following:
- 3. SMARTGrid Pack (interacts with the software to determine the desired location of the burr hole)
- a. Marking Grid
- b. Marking Tool
- 4. SMARTFrame Pack (SMARTFrame or SMARTFrame XG)
- a. SMARTFrame ("ATF") with Base
- b. Centering Tool
- c. Wharen Centering Guide
- d. Dock
- e. Device Lock (2 different diameters)
- f. Screwdriver
- Roll Lock Screw and Washer g.
- Rescue Screws (Extra Titanium Screws) 5.
- 6. Hand Controller (for use with the ATF) and Thumbwheel Extension
- 7. Accessory Pack
- a. Peel-away Sheath
- b. Stylet
- c. Lancet
- d. Depth Stop
- e. Ruler
- 8. Scalp Mount Base
- Guide Tube and Device Guide Packs (Guide Cannulas) 9.
- 10. SmartTip MRI Hand Drill and Drill Bit Kit
- 11. SmartTwist MRII Hand Drill and 4.5mm/6.0mm Drill Guide
- 12. MRI Neuro Procedure Drape, with Marker Pen and Cover
- 13. MR Camera Fiberscope Accessory Kit
- 14. SmartFrame MR Fiducial
Each of the above packs is sold separately. Each is intended to be used with the ClearPoint Workstation.
### 5. Indications for Use:
The ClearPoint System is intended to provide stereotactic guidance and operation of instruments or devices during the planning and operation of neurological procedures within the MRI environment and in conjunction with MR imaging. The ClearPoint System is intended as an integral part of procedures that have traditionally used stereotactic methodology. These procedures include biopsies, catheter and electrode insertion. The System is intended for use only with 1.5 and 3.0 Tesla MRI scanners.
### 6. Comparison of the Technological Characteristics of the Device with the Predicate Device:
Modifications to the predicate ClearPoint System are as follows:
- a) Addition of 4.5mm and 6.0mm Drill Guides that are compatible with the SmartFrame XG and can be used in place of the Targeting Cannula to facilitate the use of the cleared SmartTwist MRII Hand Drill (K151536) with the SmartFrame XG.
{5}------------------------------------------------
- b) Addition of the Wharen Centering Guide, an optional accessory designed to make the cleared Centering Tool (K100836) easier to use with the Scalp Mount Base. The use of this accessory is not required for proper use of the ClearPoint System, does not materially alter the clinical work flow, and does not change the intended use of the device. It can hold the Centering Tool in place and eliminate potential scalp movement during a) removing the lower pad of the Marking Grid, and b) during mounting of the Scalp Mount Base. Eliminating scalp movement during these steps will help ensure an accurate alignment of the SmartFrame to the entry point.
| | ClearPoint System<br>K142505 | ClearPoint System<br>Modified |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 882.4560 | 21 CFR 882.4560 |
| Product Code | HAW, ORR | HAW, ORR |
| Intended Use | The ClearPoint System is intended to<br>provide stereotactic guidance and<br>operation of instruments or devices<br>during the planning and operation of<br>neurological procedures within the<br>MRI environment and in conjunction<br>with MR imaging. The ClearPoint<br>System is intended as an integral part<br>of procedures that have traditionally<br>used stereotactic methodology.<br>These procedures include biopsies,<br>catheter and electrode insertion. The<br>System is intended for use only with<br>1.5 and 3.0 Tesla MRI scanners. | The ClearPoint System is intended to<br>provide stereotactic guidance and<br>operation of instruments or devices<br>during the planning and operation of<br>neurological procedures within the<br>MRI environment and in conjunction<br>with MR imaging. The ClearPoint<br>System is intended as an integral part<br>of procedures that have traditionally<br>used stereotactic methodology.<br>These procedures include biopsies,<br>catheter and electrode insertion. The<br>System is intended for use only with<br>1.5 and 3.0 Tesla MRI scanners. |
| Environment | MRI Suite | MRI Suite |
| Sterilization | EO 10-6 SAL | EO 10-6 SAL |
| SmartGrid Pack | MRI Planning Grid & Marking Tool | MRI Planning Grid & Marking Tool |
| SmartFrame Pack | SmartFrame XG, Scalp Mount Base,<br>Bone Screws, Scalp Mount Base,<br>Screws, Stand-Off Pins, Screwdriver,<br>Centering Tool, Dock and Lock, Roll<br>Lock Screw with Washer, Rescue<br>Screws (packaged separately) | SmartFrame XG, Scalp Mount Base,<br>Bone Screws, Scalp Mount Base,<br>Screws, Stand-Off Pins, Screwdriver,<br>Centering Tool, Wharen Centering<br>Guide (packaged and sold<br>separately), Dock and Lock, Roll Lock<br>Screw with Washer, Rescue Screws<br>(packaged separately) |
| Hand Controller | Thumbwheel Extension (Light Hand<br>Controller) | Thumbwheel Extension (Light Hand<br>Controller) |
| Accessory Kit | Peel-away Sheath, Lancet, Stylet,<br>Depth Stop, Ruler | Peel-away Sheath, Lancet, Stylet,<br>Depth Stop, Ruler |
| Drill Guides | Not included | 4.5mm & 6.0mm |
| Targeting<br>Cannula ID | 0.0825" | 0.0825" |
| Targeting<br>Cannula Material | Ultem and PEEK | Ultem and PEEK |
| Guide Tube /<br>Device Guide /<br>Drill Guide ID | 0.0938 & 0.141" | 0.0938, 0.141, 0.191, 0.250" |
| Guide Tube / | Ultem | Ultem |
| | ClearPoint System<br>K142505 | ClearPoint System<br>Modified |
| Device Guide /<br>Drill Guide<br>Material<br>Packaging | Sterile, Sealed Tray, Inside Sterile<br>Tyvek Pouch | Sterile, Sealed Tray, Inside Sterile<br>Tyvek Pouch<br>(Wharen Centering Guide and Drill<br>and Device Guides are Sterile in a<br>double Tyvek Pouch Without a Tray;<br>Wharen Centering Guide Packaging<br>includes PVC) |
| Targeting<br>Accuracy | $\pm$ 1.5mm @ ≤125mm | $\pm$ 1.5mm @ ≤125mm |
| Software | 1.5 | |
{6}------------------------------------------------
### 7. Performance Data:
Bench and phantom testing confirmed that the Wharen Centering Guide can be used to help hold the Centering Tool in place and maintain the selected entry point. Design verification was performed on the 4.5mm and 6.0mm Drill Guides with acceptable results. The tests demonstrated that the modified ClearPoint System functions as intended and is substantially equivalent to the legally marketed device.
The ClearPoint System complies with the following recognized consensus standards:
- AAMI/ANSI/ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing.
- · ANSI/AAMI/ISO 11135-1 Sterilization of health care products Ethylene oxide -Part 1: Requirements for development, validation, and routine control of sterilization process for medical devices.
### 8. Conclusions:
The modifications to the ClearPoint System were made in conformance with the company's design control procedures and the performance testing performed for the predicate ClearPoint System (K142505). The modified ClearPoint System has the same intended use and indications for use and similar technologies characteristics and principles of operation as the predicate ClearPoint System. The minor technological differences between the modified ClearPoint System and its predicate ClearPoint System raise no new issues of safety and effectiveness. Thus, the modifications are substantially equivalent to the previously cleared ClearPoint System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.