K160431 · Medical Compression Systems (Dbn) , Ltd. · JOW · Jul 13, 2016 · Cardiovascular
Device Facts
Record ID
K160431
Device Name
ActiveCare Home
Applicant
Medical Compression Systems (Dbn) , Ltd.
Product Code
JOW · Cardiovascular
Decision Date
Jul 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ActiveCare+SFT®and ActiveCare+DTx® Systems are portable, ambulatory, sequential, intermittent pneumatic compression devices (IPCDs) prescribed by health care professionals. The systems include a rechargeable battery powered option allowing patient mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised, such as during and after major orthopaedic surgery procedures e.g. total joint (hip and knee) arthroplasty. They are in the clinical setting or home environment and can be provided directly to the patient for home use. These devices are indicated for use in: - Preventing Deep Vein Thrombosis (DVT) - Diminishing post-operative pain and swelling - Reducing wound healing time - Patients at risk for Deep Vein Thrombosis (DVT) and related Pulmonary Embolism (PE) (Venous Thromboembolism (VTE)) - Treatment of venous stasis - Treatment and assistance in healing: Stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers - Enhancing blood circulation - Treatment of chronic venous insufficiency - Reducing edema The ActiveCare+SFT® and ActiveCare+DTx® Systems are intended to provide external compression in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow. In addition, the ActiveCare+DTx® System can detect hemodynamic changes in venous blood flow.
Device Story
ActiveCare@Home is a portable, battery-powered, sequential intermittent pneumatic compression device (IPCD) for clinical or home use. It simulates muscle contractions to enhance venous blood flow velocity in patients with venous impairment or reduced pulsatility, particularly post-orthopedic surgery. The device applies external compression synchronized with the patient's natural venous return profile. The system includes a Bluetooth module for data communication and an activity tracking feature that calculates a 'Patient's Activity Index' based on step counts. Output is displayed on an LCD screen, including battery life and activity data. A 2D barcode icon provides digital logistic information. Healthcare providers prescribe the device to improve circulation, prevent DVT/PE, and assist in healing venous/diabetic ulcers. The device allows patient mobility during treatment.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including risk analysis per ISO 14971:2012, software verification and validation, electromagnetic compatibility (EMC) per IEC 60601-1-2, and performance testing for the patient activity index and monitoring functionalities.
Technological Characteristics
Portable, battery-powered sequential IPCD. Features include a Bluetooth module for connectivity, LCD interface, and 2D barcode for logistics. Software is classified as 'Moderate' level of concern. Complies with IEC 60601-1-2 for EMC. Operates via pneumatic compression synchronized with patient venous flow.
Indications for Use
Indicated for patients at risk of DVT/PE (VTE) or experiencing venous impairment/reduced pulsatility, including post-orthopedic surgery. Used to prevent DVT, reduce post-op pain/swelling, accelerate wound healing, treat venous stasis, stasis dermatitis, venous/arterial/diabetic ulcers, chronic venous insufficiency, and edema.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 13, 2016
Medical Compression Systems (DBN) Ltd. % Orly Maor Regulatory Consultant 25 Sirkin Street Kfar Saba, 44421 Israel
Re: K160431
Trade/Device Name: ActiveCare(@)Home Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: May 23, 2016 Received: June 1, 2016
Dear Orly Maor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160431
Device Name ActiveCare@Home
The ActiveCare+SFT®and ActiveCare+DTx® Systems are portable, ambulatory, sequential, intermittent pneumatic compression devices (IPCDs) prescribed by health care professionals. The systems include a rechargeable battery powered option allowing patient mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised, such as during and after major orthopaedic surgery procedures e.g. total joint (hip and knee) arthroplasty. They are in the clinical setting or home environment and can be provided directly to the patient for home use.
These devices are indicated for use in:
- · Preventing Deep Vein Thrombosis (DVT)
- · Diminishing post-operative pain and swelling
- Reducing wound healing time
· Patients at risk for Deep Vein Thrombosis (DVT) and related Pulmonary Embolism (PE) (Venous Thromboembolism (VTE))
- · Treatment of venous stasis
- · Treatment and assistance in healing: Stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers
- · Enhancing blood circulation
- · Treatment of chronic venous insufficiency
- · Reducing edema
The ActiveCare+SFT® and ActiveCare+DTx® Systems are intended to provide external compression in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow. In addition, the ActiveCare+DTx® System can detect hemodynamic changes in venous blood flow.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medical Compression Systems (DBN) Ltd. The logo features a green stylized "M" next to the letters "CS" in black. Below the symbols is the text "Medical Compression Systems (DBN) Ltd" in black font.
# 510(k) SUMMARY Medical Compression Systems (DBN)
Active Care @ Home Device
## K160431
#### I. SUBMITTER
Medical Compression Systems (DBN) Ltd. 12 Ha'llan Street, PO Box 75, Or Akiva 30600, Israel Tel: +972 (4) 6266630 Fax: +972 (4) 6266640 E-mail: mcs@mcsmed.com
## Contact Person:
Orly Maor Company Regulatory Consultant 25 Sirkin Street Kfar Saba 44421, Israel Office; +972-9-745360 Cell: +972-52-8901368 Fax: +972-153-9-7453607 oram.ma@gmail.com
Date Prepared: May 23, 2016
#### II. DEVICE
Device:
Name of Device: Common or Usual Name: Classification: Product Code: Regulation No .: Class: Panel:
ActiveCare@Home ActiveCare@Home Sleeve, Limb, Compressible JOW 870.5800 II Cardiovascular
#### III. PREDICATE DEVICE
The predicate device is the ActiveCare@Home, cleared under K151377.
#### IV. DEVICE DESCRIPTION
The ActiveCare Systems are portable, ambulatory, sequential, intermittent pneumatic compression devices (IPCDs) prescribed by healthcare professionals. The systems include a rechargeable battery powered option allowing patient
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for Medical Compression Systems (DBN) Ltd. The logo features a stylized green "M" next to the letters "CS" in black. Below the letters is the text "Medical Compression Systems (DBN) Ltd" in a smaller font size.
mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised such as during and after major orthopedic surgery procedures e.g. total joint(hip and knee) arthroplasty. They are intended for use in the clinical setting or home environment, and can be provided directly to the patient for home use.
#### V. INDICATIONS FOR USE
The ActiveCare Systems are portable, ambulatory, sequential, intermittent pneumatic compression devices (IPCDs) prescribed by healthcare professionals. The systems include a rechargeable battery powered option allowing patient mobility and ease of use. These devices simulate muscle contractions in order to treat or enhance blood flow velocity in individuals experiencing venous impairment or reduced pulsatility (dysfunction of the muscle pump) when blood flow may become challenged or compromised such as during and after major orthopedic surgery procedures e.g total joint(hip and knee) arthroplasty. They are intended for use in the clinical setting or home environment, and can be provided directly to the patient for home use.
These devices are indicated for use in:
- Preventing Deep Vein Thrombosis (DVT) ●
- . Diminishing post-operative pain and swelling
- Reducing wound healing time
- Patients at risk for deep vein thrombosis (DVT) and related pulmonary embolism (PE) (Venous Thromboembolism (VTE))
- Treatment of venous stasis
- Treatment and assistance in healing: Stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers
- Enhancing blood circulation
- Treatment of chronic venous insufficiency
- Reducing edema
The ActiveCare Systems are intended to provide external compression in synchrony with the specific patient's natural venous blood flow return profile in order to achieve a high pulsatile venous blood flow.
In addition, the ActiveCare+DTx System can detect hemodynamic changes in venous blood flow.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Medical Compression Systems (DBN) Ltd. The logo features a green, stylized "M" next to the letters "CS" in black. Below the letters is the text "Medical Compression" in a smaller font, followed by "Systems (DBN) Ltd".
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The modified ActiveCare@Home has the same intended use and indications, principles of operation, and technological characteristics as the cleared ActiveCare@Home. The minor differences do not raise any new questions of safety or effectiveness. Performance data demonstrates that the modified ActiveCare@Home is as safe and effective as the cleared ActiveCare@Home. Thus, the ActiveCare@Home is substantially equivalent to its predicate device.
The modified ActiveCare@Home is different from the cleared device in the following:
- 1. The modified ActiveCare@Home was equipped with hardware Bluetooth Module.
- 2. Two new Software Blocks were added to its original Software, which were designed to:
- · Establish a communication between the ActiveCare main processor and the Bluetooth module (Bluetooth Communication)
- · Collect and reprocess data already being measured by the device in order to detect the number of the patient's steps generated during the treatment period (hereinafter referred to in this document as "Patient's Activity Index").
- 3. Three minor modifications were added to the ActiveCare@Home device GUI:
- · Reduction of the battery icon size in order to generate space for displaying the Activity Index graph mentioned above.
- · Displaying under the battery icon an estimation of the remained battery life time (appears only when the ActiveCare@Home device is powered by the internal battery).
- · Adding a 2D barcode icon to the Archives screen, this will include digital information, in addition to the same data which is displayed on the LCD screen verbally - for logistic.
## VII. PERFORMANCE DATA
The following tests were performed and they are the same as those submitted in the original submissions:
- Risk analysis per ISO 14971:2012 -
- Software validation
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Medical Compression Systems (DBN) Ltd. The logo features a large, stylized letter 'M' in green, next to the letters 'CS' in black. Below the letters are the words "Medical Compression Systems (DBN) Ltd" in a smaller font.
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for
Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern. The level of concern is Moderate if a failure or latent design flaw could directly result in minor injury to the patient or operator. The level of concern is also "Moderate" if a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
#### Electromagnetic compatibility (EMC) -
Electromagnetic compatibility (EMC) per IEC 60601-1-2was conducted on the ActiveCare@Home device
### Performance Testing -Performance testing simulated use for the patient activity index, and monitoring functionalities
### VIII. CONCLUSIONS
The modified ActiveCare@Home device has the same intended use compared to the predicate device. The principal features of the device that were described show that the minor differences in device characteristics between the subject and predicate device do not raise any new questions of safety or effectiveness.
Performance data and software validation has been provided, establishing that the ActiveCare@Home device performs as intended and in a manner that is substantially equivalent to the predicate.
Therefore, the device may be found substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.