20G x 1.25 NovaCath Secure IV Catheter System, 24G x 0.56 NovaCath Secure IV Catheter System
K160374 · Tangent Medical Technologies, Inc. · FOZ · Mar 11, 2016 · General Hospital
Device Facts
Record ID
K160374
Device Name
20G x 1.25 NovaCath Secure IV Catheter System, 24G x 0.56 NovaCath Secure IV Catheter System
Applicant
Tangent Medical Technologies, Inc.
Product Code
FOZ · General Hospital
Decision Date
Mar 11, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
The NovaCath™ Secure IV Catheter System is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature aids in the prevention of needlestick injuries. Upon catheter insertion, blood is contained within the device to aid in the prevention of blood exposure. The NovaCath™ device may be utilized in any patient population with consideration given to vascular anatomy and appropriateness of procedure. The 18, 20, and 22 gauge NovaCath™ devices are suitable for use with power injectors for a maximum of 300psi.
Device Story
NovaCath™ Secure IV Catheter System is an over-the-needle, radiopaque intravascular catheter with integrated extension tubing, secondary stabilization hub, female luer lock, and passive safety needle-shielding mechanism. Used in clinical settings for short-term peripheral vascular access; operated by healthcare professionals. Device functions as a closed system; self-sealing septum and luer lock contain blood after needle withdrawal to prevent exposure. Extension tubing clamp prevents blood loss during infusion set connection. Provides clinicians with reliable vascular access and sharps injury protection. Benefits include reduced risk of needlestick injuries and blood exposure during insertion and maintenance.
Clinical Evidence
Bench testing only. Verification testing included power injection flow rate and pressure testing (ISO 10555-1:2013) and sharps access testing in safe mode (ISO 23908:2011) for the new catheter models.
Technological Characteristics
Biocompatible polyurethane catheter; radiopaque; integrated extension tubing; passive safety needle-shielding mechanism; female luer lock; self-sealing septum. EO sterilized; single-use; non-pyrogenic; latex-free; DEHP-free. Complies with ISO 10555-1:2013 (flow/pressure) and ISO 23908:2011 (sharps protection).
Indications for Use
Indicated for patients requiring short-term (less than 30 days) vascular access for blood sampling, blood pressure monitoring, or fluid administration. Suitable for any patient population based on vascular anatomy and procedure appropriateness. 24G models are contraindicated for power injection.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a central emblem. The emblem consists of a stylized representation of three human profiles facing to the right, stacked one behind the other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2016
Mr. Greg Last Ouality Systems and Regulatory Affairs Manager Tangent Medical Technologies. Incorporated 8170 Jackson Road, Suite A Ann Arbor, Michigan 48103
Re: K160374
Trade/Device Name: NovaCath™ Secure IV Catheter System Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: FOZ Dated: February 8, 2016 Received: February 10, 2016
Dear Mr. Last:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K160374
Device Name NovaCathTM Secure IV Catheter System
### Indications for Use (Describe)
The NovaCath™ Secure IV Catheter System is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature aids in the prevention of needlestick injuries. Upon catheter insertion, blood is contained within the device to aid in the prevention of blood exposure. The NovaCath™ device may be utilized in any patient population with consideration given to vascular anatomy and appropriateness of procedure. The 18, 20, and 22 gauge NovaCath™ devices are suitable for use with power injectors for a maximum of 300psi.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Special 510(k): Device Modification 510(k) Summary Tangent Medical Technologies, Inc. NovaCath Secure IV Catheter System
| Submission Date: | February 8, 2016 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submission Owner<br>Information: | Tangent Medical Technologies, Inc.<br>8170 Jackson Rd. STE A<br>Ann Arbor, MI 48103 |
| Contact<br>Information: | Greg Last<br>Quality Systems and Regulatory Affairs Manager<br>Greg@tangentmedical.com<br>Tel: 734-527-4051<br>Fax: 734-253-2043 |
| Device Name: | NovaCath™ Secure IV Catheter System |
| Proprietary<br>Name: | NovaCath™ Secure IV Catheter System |
| Common Name: | Intravascular Catheter (short-term, less than 30 days, therapeutic) |
| Classification<br>Name: | Intravascular Catheter |
| Classification<br>Code: | FOZ |
| Regulation<br>Number: | 21 CFR §880.5200 |
| Predicate<br>Device: | NovaCath™ Secure IV Catheter System (K120839) |
| Device<br>Description: | The NovaCath™ Secure IV Catheter System is a short-term, therapeutic, IV catheter. The<br>NovaCath™ currently has four commercially available models. This special 510(k) adds<br>two additional models of different catheter lengths to the existing gauge size portfolio.<br>Two (2) additional models: |
| Gauge | 20G | 24G |
|-----------------|-------|-------|
| Catheter Length | 1.25" | 0.56" |
{4}------------------------------------------------
| Gauge | 18G | 20G | 22G | 24G |
|-----------------|-------|-------|-------|-------|
| Catheter Length | 1.25" | 1.00" | 1.00" | 0.75" |
Are being added to the existing four (4) models cleared in K120839:
The NovaCath™ Secure IV Catheter System is comprised of an over-the-needle, radiopaque intravascular catheter, integrated extension tubing with a secondary stabilization hub, a female luer lock and clamp, and a passive safety needle shielding mechanism. The design of the NovaCath™ Secure IV Catheter System is considered a closed system since it protects users from blood exposure during the catheter insertion procedure. Blood is contained within the device after needle withdrawl by a self-sealing septum and luer lock, which is provided with a one-time use vent plug. The clamp on the extension tubing is provided to eliminate blood exposure when the vent plug is replaced with an infusion set connection or other end cap.
Intended Use: The intended use of the device is identical to the predicate device.
> The NovaCath™ Secure IV Catheter System is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravascularly. The needle-shielding feature aids in the prevention of needlestick injuries. Upon catheter insertion, blood is contained within the device to aid in the prevention of blood exposure. The NovaCath™ device may be utilized in any patient population with consideration given to vascular anatomy and appropriateness of procedure. The 18, 20, and 22 gauge NovaCath™ devices are suitable for use with power injectors for a maximum of 300psi.
Comparison of The intended use of the device and the technological characteristics of the device are Technological unchanged from the predicate device. The two models being added to the NovaCath™ Characteristics: product line per this special 510(k) use the same technology and materials as the four commercially available models. The differences are in the needle length, catheter length, and housing sizes.
> The table below provides a comparison summary of the technological characteristics between the subject and predicate device.
{5}------------------------------------------------
| Substantial Equivalence Comparison | | |
|------------------------------------------------|----------------------------------------------|--------------------------------------------|
| Comparison Criteria | Subject Device<br>(NovaCath 201250 & 240560) | Predicate Device<br>(NovaCath K120839) |
| Same Intended Use | Yes | Yes |
| Prescription Device (Rx Only) | Yes | Yes |
| Biocompatible Polyurethane Catheter | Yes | Yes |
| Biocompatible Materials of Fabrication | Yes | Yes |
| Closed System | Yes | Yes |
| Built-In Extension from Main Assembly | Yes | Yes |
| Flashback Visualization | Yes | Yes |
| Sharps Injury Prevention Feature | Yes, Passive Integrated | Yes, Passive Integrated |
| Mechanical | Yes, Manual Needle<br>Retraction | Yes, Manual Needle Retraction |
| Intended Anatomical Location | Yes, Peripheral Vasculature | Yes, Peripheral Vasculature |
| Distal End Configuration | Yes, Meets ISO 10555-5 | Yes, Meets ISO 10555-5 |
| Proximal End Configuration | Yes, Female Luer Access Port | Yes, Femal Luer Access Port |
| Flow Rate (ml/min) | Yes, 20G: 40 ml/min<br>Yes, 24G: 14 ml/min | Yes, 20G: 46 ml/min<br>Yes, 24G: 14 ml/min |
| Catheter Stabilization | Yes, Primary and Secondary | Yes, Primary and Secondary |
| EO Sterilization Method | Yes | Yes |
| Packaged Sterile, Single Use | Yes | Yes |
| Non-Pyrogenic | Yes | Yes |
| Made without Latex Rubber | Yes | Yes |
| Made without DEHP | Yes | Yes |
| For Use with Power Injections up to<br>300 psi | Yes, 20G;<br>24G Contraindicated | Yes, 18G, 20G, 22G;<br>24G Contraindicated |
The device was developed and tested in accordance with released company procedures Summary of on design control (per 21 CFR 820.30) and risk analysis (per ISO 14971:2009). A Device declaration of conformity to design controls is located within the 510(k). Evaluation:
The risk analysis method used to assess the impact of the modifications was a Failure Modes and Effects Analysis (FMEA) per released procedures. The design verification tests completed are as follows:
| Modification | Test Performed | Consensus Standard |
|-----------------------------------------------------------------|-----------------------------------------------------------|--------------------|
| 20Gx1.25" Catheter | Power Injection Test for Flow<br>Rate and Device Pressure | ISO 10555-1:2013 |
| 20Gx1.25" Catheter | Flow Rate Test | ISO 10555-1:2013 |
| 24G x0.56" Catheter | Flow Rate Test | ISO 10555-1:2013 |
| 24G x0.56" Catheter<br>Modified Housing<br>("Slider") Component | Test Access to the Sharp in<br>Safe Mode | ISO 23908:2011 |
{6}------------------------------------------------
#### Conclusions: The NovaCath™ Secure IV Catheter System as described in this Special 510(k) has been found to perform in the substantial equivalent manner as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.