← Product Code [NGX](/productcode/NGX) · K160323

# FlowKeepers (K160323)

_Dk Electronics, LLC · NGX · Apr 6, 2017 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160323

## Device Facts

- **Applicant:** Dk Electronics, LLC
- **Product Code:** [NGX](/productcode/NGX.md)
- **Decision Date:** Apr 6, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Powered Muscle Stimulator - To temporarily increase local blood circulation in healthy leg muscles - To stimulate healthy muscles in order to improve and facilitate muscle performance TENS - For temporary relief of pain associated with sore and aching muscles of the lower extremities (legs) due to strain from exercise or normal household and work activities

## Device Story

FlowKeepers® is a portable, battery-powered neuromuscular electrical stimulator (NMES) and TENS device. It delivers low-frequency (0.1 Hz), rectangular, bipolar, biphasic electrical pulses through the skin to leg muscles. Operated by the user, the device features a control unit with power buttons and LED indicators. It functions to stimulate healthy muscles for performance improvement and circulation, or to provide temporary pain relief. The device is intended for home use. Healthcare providers do not directly operate the device; users manage their own stimulation levels (1-15) and duration. The device includes an automatic shut-off feature (10 hours) and low-battery alerts. By inducing muscle contractions, it aims to alleviate muscle soreness and facilitate performance, providing a non-invasive, self-administered therapeutic option.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and specific requirements for nerve and muscle stimulators (IEC 60601-2-10) and home healthcare environments (IEC 60601-1-11). Results confirmed the device functioned as intended and met all safety standards.

## Technological Characteristics

Portable, battery-powered (3V Li-ion CR2032). Casing/body: ABS + PET; contacts: NBR. Waveform: Pulsed symmetrical biphasic and bipolar. Output: 1 channel, 15 intensity levels. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-1-4. Sterilization: Not specified. Connectivity: Standalone.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- Revitive IX® (OTC), Actegy Ltd. ([K143207](/device/K143207.md))
- Pennypad PP-904 Rapid Relief™ (OTC), HiVox Biotek, Inc. ([K140650](/device/K140650.md))

## Reference Devices

- [K160508](/device/K160508.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 6, 2017

DK Electronics, LLC Avery Huff Operations Mgr 413-B Elmwood Ave, Sharon Hill, Pennsylvania 19079

Re: K160323

Trade/Device Name: FlowKeepers® Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX, NUH Dated: February 22, 2017 Received: March 7, 2017

Dear Avery Huff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J. Hoffmann -S

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K160323

Device Name FlowKeepers®

Indications for Use (Describe)

Powered Muscle Stimulator - To temporarily increase local blood circulation in healthy leg muscles

- To stimulate healthy muscles in order to improve and facilitate muscle performance

#### TENS

- For temporary relief of pain associated with sore and aching muscles of the lower extremities (legs) due to strain from exercise or normal household and work activities

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary K160323

#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

| Submitter:         | DK Electronics, LLC                      |
|--------------------|------------------------------------------|
| Submitter Address: | 413-B Elmwood Ave, Sharon Hill, PA 19079 |
| Phone:             | 855-494-5158                             |
| Facsimile:         | 484-494-5163                             |
| Contact Person:    | Avery Huff                               |
| Email:             | avery_huff@yahoo.com                     |
| Date Prepared:     | 4/6/2017                                 |

#### Trade Name and Classification of Device

| Trade Name of Device: | FlowKeepers®                                                                                                                                  |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:          | Powered Muscle Stimulator<br>Neuromuscular electrical stimulator devices (NMES)                                                               |
| Classification:       | Powered Muscle Stimulator (21 CFR 890.5850)                                                                                                   |
| Product Codes:        | Primary - NGX (Stimulator, Muscle, Powered, For Muscle Conditioning)<br>Secondary - NUH (Stimulator, Nerve, Transcutaneous, Over-The-Counter) |
| Review Panel:         | Physical Medicine                                                                                                                             |

#### Predicate Devices

Revitive IX® (OTC), Actegy Ltd. (K143207) – NGX, NUH 21 CFR 890.5850 Pennypad PP-904 Rapid Relief™ (OTC), HiVox Biotek, Inc. (K140650) – NUH, 21 CFR 890.5890

#### Intended Use/ Indications for Use

Powered Muscle Stimulator

- · To temporarily increase local blood circulation in healthy leg muscles.
- · To stimulate healthy muscles in order to improve and facilitate muscle performance.

TENS

- · For temporary relief of pain associated with sore and aching muscles in the lower extremities (legs) due to strain from exercise, normal household duties and work activities.
#### Device Description

The FlowKeepers® is a portable, battery powered device and delivers electrical stimulation through the skin to the muscles. It provides low frequency (0.1 Hz) muscle stimulation with rectangular, bipolar, biphasic pulses, and operates in a range which balances between comfort and effective contractions. There is a control unit with power buttons and LED flasher. To see how FlowKeepers® technological characteristics are similar to the predicate device, see the comparison chart below under the Substantial Equivalence heading.

#### Performance Testing

Electromagnetic compatibility, electrical safety testing, and other performance testing were conducted in

{4}------------------------------------------------

accordance with the following standards

IEC 60601-1-8:2006 & A1:2012 Medical electrical equipment Part 1: General requirements for safety

IEC 60601-1-2 Edition 3: 2007-03 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance Collateral Standard: Electromagnetic compatibility- Requirements and tests

IEC 60601-1-11 Edition 1.0 2010-04

Medical electrical equipment Part 1-11: General Requirements for basic safety and essential performance Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

IEC 60601-1-4 Medical electrical equipment Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems

IEC 60601-2-10 Edition 2.0 2012-06 Medical electrical equipment Part 2: Particular requirements for the safety of nerve and muscle stimulators.

In all instances, the FlowKeepers® functioned as intended and the results observed were as expected.

#### Substantial Equivalence

The FlowKeepers® is substantially equivalent to other powered muscle stimulators such as Revitive IX® OTC (K143207). The FlowKeepers® has the same intended uses and similar indications, technological characteristics, and principles of operation as the Revitive IX® OTC. The difference in design and the minor technological differences between the FlowKeepers® and its predicate device Revitive IX® OTC raise no new issues of safety or effectiveness. Performance data demonstrate that the FlowKeepers® is as safe and effective as the predicate device. Thus, the FlowKeepers® is substantially equivalent.

{5}------------------------------------------------

## Comparison of FlowKeepers® with Revitive IX® (OTC) and
HiVox Biotek, Inc. Rapid ReliefTM Pennypad PP-904 (OTC)

|                                              | DK Electronics, LLC<br>FlowKeepers®                                                                                                                                                                                                                                                                                                                                                                           | Actegy Ltd.<br>Revitive IX®<br>(OTC)                                                                                                                                                                                                                                                                                                                                                                                    | HiVox Biotek, Inc.<br>Rapid Relief™<br>Pennypad PP-904 (OTC)                                                                                                                                                                                                                           |                                                               |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| 510(k) Number                                | K160323                                                                                                                                                                                                                                                                                                                                                                                                       | K143207                                                                                                                                                                                                                                                                                                                                                                                                                 | K140650                                                                                                                                                                                                                                                                                |                                                               |
| Product Codes:<br>Model                      | NGX, NUH<br>FlowKeepers®                                                                                                                                                                                                                                                                                                                                                                                      | NGX, NUH<br>RIX                                                                                                                                                                                                                                                                                                                                                                                                         | NUH<br>PP-904                                                                                                                                                                                                                                                                          |                                                               |
| Indications for Use (IFU)                    | Powered Muscle Stimulator:<br>-Temporarily increase local<br>blood circulation in healthy leg<br>muscles.<br>-To stimulate healthy muscles<br>in order to improve and<br>facilitate muscle performance.<br>TENS<br>-Temporary relief of pain<br>associated with sore and aching<br>muscles in the lower extremities<br>(legs) due to strain from<br>exercise, normal household<br>duties and work activities. | To temporarily increase local<br>blood circulation in healthy<br>leg muscles.<br><br>To stimulate healthy muscles<br>in order to improve and<br>facilitate muscle performance.<br>For temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back, upper<br>extremities (arms) and lower<br>extremities (legs) due to strain<br>from exercise or normal<br>household duties. | The Rapid Relief Pennypad<br>PP-904 is indicated for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the upper<br>and lower extremities (arm<br>and/or leg), and lower back<br>due to strain from exercise or<br>normal household and work<br>activities. |                                                               |
| Dimensions (mm) [WxHxD]                      | 140.6 mm x 67.14 mm x 10.5<br>mm                                                                                                                                                                                                                                                                                                                                                                              | Ø360 mm x 75 mm                                                                                                                                                                                                                                                                                                                                                                                                         | 113 mm x 70 mm x 9.7 mm                                                                                                                                                                                                                                                                |                                                               |
| Waveform                                     | Pulsed symmetrical biphasic and<br>bipolar                                                                                                                                                                                                                                                                                                                                                                    | Pulsed symmetrical biphasic,<br>rectangular and bipolar                                                                                                                                                                                                                                                                                                                                                                 | Symmetrical biphasic                                                                                                                                                                                                                                                                   |                                                               |
| Shape                                        | Rectangular                                                                                                                                                                                                                                                                                                                                                                                                   | Not Available                                                                                                                                                                                                                                                                                                                                                                                                           | Rectangular                                                                                                                                                                                                                                                                            |                                                               |
| Power Source(s)                              | DC 3 V Li Ion battery CR2032                                                                                                                                                                                                                                                                                                                                                                                  | DC 5V                                                                                                                                                                                                                                                                                                                                                                                                                   | DC 3 V Li Ion battery<br>CR2032                                                                                                                                                                                                                                                        |                                                               |
| Number of Output Modes                       | 1                                                                                                                                                                                                                                                                                                                                                                                                             | 1                                                                                                                                                                                                                                                                                                                                                                                                                       | 1                                                                                                                                                                                                                                                                                      |                                                               |
| Number of Output Channels                    | 1                                                                                                                                                                                                                                                                                                                                                                                                             | 2 (1 for Foot, 1 for body pads)                                                                                                                                                                                                                                                                                                                                                                                         | 1                                                                                                                                                                                                                                                                                      |                                                               |
| Automatic Shut Off                           | Auto Shut off at 10 hrs, low<br>battery                                                                                                                                                                                                                                                                                                                                                                       | Automatic Shut off, Override<br>Control                                                                                                                                                                                                                                                                                                                                                                                 | Automatic Shut off at 60<br>minutes, low battery                                                                                                                                                                                                                                       |                                                               |
| Automatic Overload Trip                      | No                                                                                                                                                                                                                                                                                                                                                                                                            | Not Available                                                                                                                                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                     |                                                               |
| Automatic No-Load Trip                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                           | Not Available                                                                                                                                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                     |                                                               |
| Patient Override Control                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                    |                                                               |
| Software/Firmware/Micropr<br>ocessor Control | Yes                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                    |                                                               |
| Weight                                       | 0.6 oz (with electrodes)<br>0.8 oz (with electrodes +<br>battery)                                                                                                                                                                                                                                                                                                                                             | 1725 g (not including PSU)                                                                                                                                                                                                                                                                                                                                                                                              | 13g (with battery)                                                                                                                                                                                                                                                                     |                                                               |
| Regulated Voltage Level                      | Yes                                                                                                                                                                                                                                                                                                                                                                                                           | Not Available                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                    |                                                               |
| Regulated Current Level                      | No                                                                                                                                                                                                                                                                                                                                                                                                            | Not Available                                                                                                                                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                     |                                                               |
| Timer Range (minutes)                        | 600 minutes ±10%                                                                                                                                                                                                                                                                                                                                                                                              | 1 to 60 min                                                                                                                                                                                                                                                                                                                                                                                                             | 15 minutes                                                                                                                                                                                                                                                                             |                                                               |
| Indicator Display:<br>On/Off Status          | LED<br>Power On/Off, Level up/down                                                                                                                                                                                                                                                                                                                                                                            | Not available                                                                                                                                                                                                                                                                                                                                                                                                           | Power On                                                                                                                                                                                                                                                                               |                                                               |
| Low Battery                                  | Red LED                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                        |                                                               |
| Voltage/Current Level                        | Level 1-15                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                        |                                                               |
| Compliance with Voluntary<br>Standards       | UL 60601-1, EN60601-1-2,<br>EN 60601-1-11, EN60601-2-10<br>EN ISO 13485                                                                                                                                                                                                                                                                                                                                       | MDD (93/42EEC), EN<br>60601-1,<br>EN60601-1-2, EN 60601-1-<br>11,<br>EN60601-2-10                                                                                                                                                                                                                                                                                                                                       | EN 60601-1,EN60601-1-2,<br>EN60601-2-10<br>EN ISO 13485                                                                                                                                                                                                                                |                                                               |
| Compliance with 21 CFR<br>898*               | Yes                                                                                                                                                                                                                                                                                                                                                                                                           | Yes                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                    |                                                               |
| Housing Materials and<br>Construction        | Casing/body ABS + PET,<br>contacts NBR                                                                                                                                                                                                                                                                                                                                                                        | Casing/body ABS, footpads<br>NBR                                                                                                                                                                                                                                                                                                                                                                                        | Casing/body ABS Medical<br>Grade Silicon, contacts                                                                                                                                                                                                                                     |                                                               |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                               | DK Electronics, LLC<br>FlowKeepers®                                                                                                                                                                                                                                                                                                                                                                                     | Actegy Ltd.<br>Revitive IX® (OTC)                                                                                                                                                                                                                                                      | HiVox Biotek, Inc.<br>Rapid ReliefTM<br>Pennypad PP-904 (OTC) |
| Maximum<br>output voltage<br>(V) (±20%)      | @ 500 Ω<br>@ 2 kΩ<br>@ 10 kΩ                                                                                                                                                                                                                                                                                                                                                                                  | 166V<br>252V<br>290V                                                                                                                                                                                                                                                                                                                                                                                                    | 32V<br>Not Available<br>Not Available                                                                                                                                                                                                                                                  | 57.6V<br>89.6V<br>96.0V                                       |
| Maximum<br>output current<br>(±20%)          | @ 500 Ω<br>@ 2 kΩ<br>@ 10 kΩ                                                                                                                                                                                                                                                                                                                                                                                  | 332mA<br>126mA<br>29mA                                                                                                                                                                                                                                                                                                                                                                                                  | 10-14mA<br>Not Available<br>Not Available                                                                                                                                                                                                                                              | 115.2mA<br>44.8mA<br>9.6mA                                    |
| Pulse Duration (µs)                          |                                                                                                                                                                                                                                                                                                                                                                                                               | 150 µs +150 µs ±10%                                                                                                                                                                                                                                                                                                                                                                                                     | Not Available                                                                                                                                                                                                                                                                          | 200 µs                                                        |
| Frequency (Hz)                               |                                                                                                                                                                                                                                                                                                                                                                                                               | 100mHz ±10%                                                                                                                                                                                                                                                                                                                                                                                                             | 20-53Hz                                                                                                                                                                                                                                                                                | 2, 5, 40 Hz                                                   |
| Net Charge per Pulse (µC)                    |                                                                                                                                                                                                                                                                                                                                                                                                               | 0 @500Ω                                                                                                                                                                                                                                                                                                                                                                                                                 | Not Available                                                                                                                                                                                                                                                                          | 2304 @500Ω                                                    |
| Maximum Phase Charge (µC)                    |                                                                                                                                                                                                                                                                                                                                                                                                               | 2.25 @500Ω                                                                                                                                                                                                                                                                                                                                                                                                              | Not Available                                                                                                                                                                                                                                                                          | 23.04 @500Ω                                                   |
| Maximum Current Density<br>(mA/cm²)          |                                                                                                                                                                                                                                                                                                                                                                                                               | 0.789 mA@500Ω                                                                                                                                                                                                                                                                                                                                                                                                           | Not Available                                                                                                                                                                                                                                                                          | 2.828 @500Ω                                                   |
| Maximum Power Density (< 0.25 W/cm²)         |                                                                                                                                                                                                                                                                                                                                                                                                               | 0.00177 mW/cm²@ 500Ω                                                                                                                                                                                                                                                                                                                                                                                                    | Not Available                                                                                                                                                                                                                                                                          | 0.163 W/cm² @500Ω                                             |
| Burst Mode                                   |                                                                                                                                                                                                                                                                                                                                                                                                               | No                                                                                                                                                                                                                                                                                                                                                                                                                      | Not Available                                                                                                                                                                                                                                                                          | No                                                            |

{6}------------------------------------------------

#### DISCUSSION:

The predicate device, Revitive IX® (OTC) is similar to the subject device in the following ways:

1. It provides the same kind of muscle stimulation for the same indications

2. It delivers similar pulsed symmetric biphasic bipolar electric wave.

The main differences among the predicate devices are:

1. FlowKeepers® application is restricted to the calf muscles

2. FlowKeepers® is powered by DC delivered from a small Li+ coin battery and not by AC and a transformer.

3. FlowKeepers® pulse frequency is much smaller (0.2% -0.5%) compared to the predicate device, however, the difference does not impact the safety or effectiveness relative to other Powered Muscle Stimulators and TENS devices operating under these specifications

4. FlowKeepers® is a much smaller and lighter device compared to the predicate device (2% of weight)

5. FlowKeepers® parameters such as maximum output voltage and current, frequency, and pulse duration, although not exactly similar to these particular predicates, do not

adversely impact safety or effectiveness. FlowKeepers® parameters are similar to cleared device K160508.

#### CONCLUSION:

The subject device, FlowKeepers®, is as safe and effective as, and functions in a manner equivalent to the predicate device. The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device identified in the submission. Thus the subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K160323](https://fda-staging.innolitics.com/device/K160323)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com