K160312 · Ilooda Co,., Ltd. · GEX · Oct 28, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160312
Device Name
FRAXIS DUO
Applicant
Ilooda Co,., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CO2 LASER Part: Fractional mode is indicated only for ablative skin resurfacing. Non-fractional mode is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery. HF electrosurgical Part : The FRAXIS DUO is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis
Device Story
Combination system featuring CO2 laser and HF electrosurgical modules; not used simultaneously. CO2 laser utilizes RF tube (10,600nm) with fractional or non-fractional handpieces for soft tissue ablation/coagulation. HF module uses bipolar handpiece with disposable micro-needle electrodes for electro-coagulation/hemostasis. Operated by physicians in clinical settings via LCD touch screen and footswitch. Laser energy delivered via articulated arm; HF energy delivered via needle array. Output parameters (power, pulse duration, depth) controlled by physician to optimize clinical effect. Benefits include precise tissue interaction for aesthetic and surgical applications.
Clinical Evidence
Bench testing only; no human clinical data. Performance supported by IEC 60601-1 (safety), IEC 60601-1-2 (EMC), EN 60825-1 (laser safety), and EN 60601-2-22. In vivo animal studies (micropig) evaluated histological ablation depth and thermal damage zones at various power/depth settings. Comparison with predicate device showed similar conical diamond-shaped tissue coagulation patterns in the dermis.
Technological Characteristics
Dual-modality system: CO2 laser (10,600nm, max 30W, air-cooled) and HF electrosurgical (2MHz, max 25W). Laser delivery via articulated arm; HF delivery via bipolar handpiece with disposable micro-needle electrodes (MTR-AC-25/64). User interface: LCD touch screen. Electrical: 100-240VAC. Biocompatibility per ISO 10993-1. Safety standards: IEC 60601-1, IEC 60601-1-2, EN 60825-1, EN 60601-2-22.
Indications for Use
Indicated for patients requiring ablative skin resurfacing (fractional mode), or incision, excision, ablation, vaporization, and coagulation of soft tissues (non-fractional mode) in dermatology, plastic surgery, ENT, gynecology, neurosurgery, dental, and oral surgery. Also indicated for electro-coagulation and hemostasis in dermatologic and general surgical procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 28, 2016
Ilooda Co., Ltd % Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway Houston, Texas 77025
Re: K160312
Trade/Device Name: Fraxis Duo Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX, GEI, OUH Dated: January 27, 2016 Received: February 5, 2016
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160312
Device Name FRAXIS DUO Instrument, surgical, powered, laser Micro-needle Fractional RF
Indications for Use (Describe)
### - CO2 LASER Part:
Fractional mode is indicated only for ablative skin resurfacing.
Non-fractional mode is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.
- HF electrosurgical Part : The FRAXIS DUO is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date 510k summary prepared: October 20, 2016
#### I. SUBMITTER
| Submitter's Name : | ILOODA CO LTD. |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's HQ Address: | 120, Jangan-ro 458 Beon-gil, Jangan-Gu,<br>Suwon-Si Gyeonggido, KOREA, 16200 |
| Submitter's Telephone: | +82 (031) 210 1622 |
| Contact person: | Yun-Jung HA (yjha@ilooda.com) / RD Manager |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
| Address: | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone: | +713-467-2607 |
| Fax: | +713-583-8988 |
| DEVICE | |
| Trade/proprietary name: | FRAXIS DUO |
| Common or Usual Name: | Instrument, surgical, powered, laser<br>Micro-needle Fractional RF |
| Regulation Name: | Laser surgical instrument for use in general and plastic surgery<br>and in dermatology.<br>Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number: | 21 CFR 878.4810 (Product Code: GEX)<br>21 CFR 878.4400 (Product Code: GEI, OUH) |
| Regulatory Class: | Class II |
| Prescription Use. | |
{4}------------------------------------------------
## PREDICATE DEVICE
| Primary Device Manufacturer: | Quanta System SpA |
|------------------------------|----------------------------------------------------------------------------------------|
| Device Name: | YOULASER C02 |
| 510(k) Number: | K123573 |
| Regulation Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Regulation Number: | 21 CFR 878.4810 (Product Code: GEX) |
| Regulatory Class: | Class II |
This predicate has not been subject to a design-related recall.
| Reference Device Manufacturer: | EndyMed Medical Ltd |
|-----------------------------------------------------------------|-------------------------------------------------------------|
| Device Name: | Intensif Applicator (Ilooda Secret) |
| 510(k) Number: | K130501 |
| Regulation Name: | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number: | 21 CFR 878.4400 (Product Code: OUH, GEI) |
| Regulatory Class: | Class II |
| This predicate has not been subject to a design-related recall. | |
#### DEVICE DESCRIPTION II.
FRAXIS DUO is a combination of two separate systems: CO2 laser and HF electrosurgical system.
The Laser and High Frequency(=Radio frequency) can not be used simultaneously.
FRAXIS DUO's CO2 laser includes the system main body, differenthandpieces(Fractional type, Non-fractional type), footswitch and an LCD Touch screen control panel.
FRAXIS DUO's CO2 laser utilize a CO2 RF tube to generate a laser beam with a wavelength of 10,600nm and use different handpiece(Fractional type, Non-fractional type) for different indications for use. The physician can optimize the effect for different applications by controlling the power of the laser pulse and using a different handpiece.
The laser are supplied with different handpieces depending upon the device configuration.
FRAXIS DUO's High Frequency(=Radio Frequency) includes the system main body, a Bipolar handpiece equipped with disposable micro-needle electrodes, footswitch and an LCD touch screen control panel.
The HF energy is delivered using disposable micro-needle electrodes.
{5}------------------------------------------------
#### INDICATIONS FOR USE: III.
#### CO2 LASER Part: ।
Fractional mode is indicated only for ablative skin resurfacing.
Non-fractional mode is indicated for incision, excision, ablation, vaporization and coagulation of body soft tissues including intraoral tissues, in medical specialties including aesthetic (dermatology and plastic surgery), otorhinolaryngology (ENT), gynecology, neurosurgery, dental and oral surgery and genitourinary surgery.
#### HF electrosurgical part -
The FRAXIS DUO is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE IV. DEVICE
| | FRAXIS DUO | EndyMed Pro /<br>Intensif Applicator (Ilooda<br>Secret) (K130501) | YOULASER CO2<br>(k123573) | |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Manufacturer | Ilooda | EndyMed Medical Ltd | Quanta System SpA | |
| Intended use | - CO2 LASER Part:<br>Fractional mode is<br>indicated only for<br>ablative skin<br>resurfacing.<br>Non-fractional mode is<br>indicated for incision,<br>excision, ablation,<br>vaporization and<br>coagulation of body<br>soft tissues including<br>intraoral tissues, in<br>medical specialties<br>including aesthetic<br>(dermatology and<br>plastic surgery),<br>otorhinolaryngology<br>(ENT), gynecology,<br>neurosurgery, dental<br>and oral surgery and<br>genitourinary surgery. | | - CO2 LASER PART<br>YOULASER CO2 laser<br>when used in traditional, non<br>fractionated mode is<br>indicated for: incision,<br>excision, ablation,<br>vaporization and<br>coagulation of body soft<br>tissues including<br>intraoral tissues, in<br>medical specialties<br>including aesthetic<br>(dermatology and plastic<br>surgery),<br>otorhinolaryngology (ENT),<br>gynaecology,<br>neurosurgery, dental and<br>oral surgery and<br>genitourinary surgery.<br>YOULASER CO2laser<br>when used in<br>fractionated mode<br>(dot scanner) is indicated<br>only for ablative<br>skin resurfacing." | |
| | | FRAXIS DUO | EndyMed Pro /<br>Intensif Applicator (Ilooda<br>Secret) (K130501) | YOULASER CO2<br>(k123573) |
| | | -HF electrosurgical<br>Part :<br>The FRAXIS DUO is<br>intended for use in<br>dermatologic and<br>general surgical<br>procedures for electro-<br>coagulation and<br>hemostasis | The Intensif Applicator<br>is intended for use in<br>Dermatologic and<br>General Surgical<br>procedures for<br>electrocoagulation and<br>hemostasis. | |
| | Electrosurgical RF<br>applicator | Intensif Applicator<br>(Ilooda Secret)<br>(K130501) | Intensif Applicator<br>(Ilooda Secret) (K130501) | |
| | Output energy<br>type | CO2 / High frequency | High Frequency | CO2 |
| Delivery system | | Laser Part<br>Articulated Arm with<br>Scanner, Normal(Smart<br>surgi) Handpiece | - | Articulated arm +<br>Normal or Scanner<br>handpiece |
| | | High frequency Part<br>Bipolar Handpiece | Bipolar Handpiece | |
| Operation mode | | Laser Part<br>Fractional mode<br>Scanner handpiece<br>Normal mode<br>(CW, Pulse, Single<br>Pulse,Ultra)<br>Normal handpiece | | Fractional mode<br>Surgical mode<br>(CW, Pulse) |
| | | -HF electrosurgical<br>Part :<br>Manual mode | Intensif mode | |
| | User interface | Color Touch Panel | Color Touch Screen | Color Touch Screen |
| | Electrical<br>Requirements | 100-240VAC, 50/60Hz,<br>5-3A | 100-230 VAC, 50-60Hz<br>1-3A, | 115 V ac; 50/60 Hz;<br>1000 VA or<br>230 V ac; 50/60 Hz;<br>1000 V A |
| Dimensions<br>(mm) | | Main unit :<br>410(W)x601(D)x1071(H)<br>Arm : Max 1,400(H)mm | Main Unit:<br>190 (W) x 300 (D) x<br>1150 (H) | Main unit :<br>480 (W) x 550(D) x<br>1200 (H)<br>Arm : Not known |
| | Weight<br>(Without arm) | 45kg | 33 kg | 55 Kg |
| | wavelength | 10,600nm | | |
| Spec.<br>_Laser | Medium of<br>transmission | CO2 Laser | - | CO2 Laser |
| | Aiming beam | Diode laser(Red),<br>Max 4mW, 655nm | - | Red Laser Diode,<br>3mW |
| | power | Max 30W | - | Max 30W |
| | Cooling | Air cooling | - | Air cooling |
| | Scanning<br>pattern | | - | Area, Shape, Spot<br>density |
| | FRAXIS DUO | EndyMed Pro /<br>Intensif Applicator (Ilooda<br>Secret) (K130501) | YOULASER CO2<br>(k123573) | |
| | (Fractional)<br>Image: [Fractional] | | | |
| | Pulse energy<br>(Fractional) | Max 150mJ | - | Up to 3J |
| | Pulse duration<br>(Fractional) | $20 \mu s \sim 5,000 \mu s$ | - | $50 \mu s - 20.000 \mu s$ |
| | Distance<br>(Fractional) | 1x1~20x20 mm | - | 18x18mm |
| | Spot size | Fractional : 100 ~ 120 $\mu m$<br>Non-fractional : Max<br>1.3mm | - | Fractional :200 $\mu m$<br>Non-fractional :<br>200 $\mu$ m, 400 $\mu$ m, 1000 $\mu$ m |
| | Pulse repetition<br>(Fractional) | 1,000Hz | | Up to 1kHz |
| | Number of<br>microbeams per<br>surface<br>area<br>(Fractional) | Max 289 spot/cm2 | | 100, 200 & 500 spot/cm2 |
| | Energy per<br>Microbeam<br>(Fractional) | 150mJ | | Fluence per microbeam<br>9.9 ÷84.8 J/cm2<br>(from Quanta USA<br>website) |
| | Total power<br>per surface<br>area<br>(Fractional) | Max.30W | | Max 30W |
| | Treatment<br>distance | 0.1mm - 2.0mm | | Not known |
| | Treatment time | 10-15min | | Not known |
| | Pulse rate<br>(Non-<br>fractional) | 1Hz - 1,000Hz | - | 2,000Hz |
| | Pulse width<br>(Non-<br>fractional) | Ultra : $20 \mu s \sim 5000 \mu s$ | | Surgical : up to 100 ms |
| Spec.<br>HF | High<br>frequency | 2MHz ± 10% | 1MHz ± 10% | |
| | Max power | Max 25W at 500Ω | Max 25W at 500Ω | |
| | Total power<br>delivered<br>per<br>treatment | Max.25W | Max.25W | |
| | Power of<br>per electrode<br>pin | Max.25W | Max.25W | |
| | Power density | 41W/cm² | 41W/cm² | |
| | Current density | 0.3A/cm² | 0.3A/cm² | |
| | RF duration | 50 ms ~ 950 ms | 50 ms ~ 950 ms | |
| | Treatment time | 10~15min<br>(recommended) | 10~15min<br>(recommended) | |
| | Needle<br>insert<br>depth | 0.5 ~ 3.5mm (0.1 step) | 0.5 ~ 3.5mm (0.1 step) | |
| | Intensity | 0 ~ 10 LEVEL | 0 ~ 10 LEVEL | |
| | | FRAXIS DUO | EndyMed Pro /<br>Intensif Applicator (Ilooda<br>Secret) (K130501) | YOULASER CO2<br>(k123573) |
| | Repetition | 0.2 / 0.5 / 1 / 2 sec /<br>Single | 0.2 / 0.5 / 1 / 2 sec /<br>Single | |
| | Connected<br>handpiece | Bipolar handpiece | Intensif handpiece<br>(Bipolar type) | |
| | Connected<br>electrodes | MTR-AC-25<br>MTR-AC-64 | Intensif 25 pin electrode…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.