K160277 · GE Medical Systems Ultrasound and Primary Care Diagnostics · IYN · Mar 25, 2016 · Radiology
Device Facts
Record ID
K160277
Device Name
LOGIQ F8 Expert, LOGIQ F8, LOGIQ F6
Applicant
GE Medical Systems Ultrasound and Primary Care Diagnostics
Product Code
IYN · Radiology
Decision Date
Mar 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 are general purpose ultrasound imaging and analysis systems providing digital acquisition, processing and display capability, clinical applications including: Abdominal, Obstetrical, Gynecological, parts. Vascular/Peripheral Vascular, Transcranial. Small Pediatric, Musculoskeletal, Urological, Cardiac, Transvaginal, Transrectal, and Biopsy.
Device Story
The LOGIQ F8 Expert/F8/F6 are mobile, console-based diagnostic ultrasound systems. They acquire, process, and display digital ultrasound images for clinical diagnosis. The system includes a console, keyboard, touch panel, and 17-19 inch LCD display. It supports various transducers for imaging modes including B-mode, M-mode, PW/CW Doppler, Color Flow, Power Doppler, and 3D/4D imaging. Used in clinical settings by physicians and sonographers, the device provides real-time visualization and quantitative analysis (e.g., Auto-EF, Sonobiometry). Output is displayed on the integrated monitor for clinical decision-making, aiding in diagnosis across multiple anatomical regions. The system supports wireless connectivity and DICOM standards for data management.
Clinical Evidence
No clinical studies were required to support substantial equivalence. Evidence consists of bench testing, including acoustic output measurement, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in compliance with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2/UD 3, and ISO 10993-1.
Technological Characteristics
Mobile console-based ultrasound system (1600mm x 810mm x 720mm). Employs pulsed Doppler, echo imaging, and transducer technology. Connectivity via DICOM (NEMA PS 3.1-3.20). Biocompatible patient-contact materials per ISO 10993-1. Safety compliance: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37. Features include 3D/4D imaging, elastography, and coded pulse for harmonics.
Indications for Use
Indicated for ultrasound imaging, measurement, and analysis of the human body for abdominal, obstetrical, gynecological, small parts, vascular/peripheral vascular, transcranial, pediatric, musculoskeletal, urological, and cardiac applications, including transvaginal, transrectal, and biopsy procedures.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2016
GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC % Ms. Tracey Ortiz Regulatory Affairs Director 9900 W. Innovation Drive WAUWATOSA WI 53226
Re: K160277
Trade/Device Name: LOGIO F8 Expert / LOGIO F8 / LOGIO F6 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 29, 2016 Received: February 2, 2016
Dear Ms. Ortiz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are also blue. There are decorative swirls around the letters and the outer edge of the circle.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
Indications for Use
510(k) Number (if Known) -
Device Name LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6
Indications for Use (Describe)
The LOGIQ F8 Expert / LOGIQ F6 / LOGIQ F6 are general purpose ultrasound imaging and analysis systems providing digital acquisition, processing and display capability, clincal applications including: Abdominal, Obsterical, Gnall parts, Vascular/Peripheral Vascular, Transcranial, Pediatic, Musculoskeletal, Cardiac, Transvaginal, Transrectal, and Biopsy.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a blue circle. The circle is surrounded by a stylized pattern of swirling lines, also in blue, giving the impression of motion or energy.
### This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!"
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized, cursive font, enclosed within a circular border. The color of the logo is a light blue.
### Indications for Use Forms
The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 systems and for all of its probe/mode combinations. Combinations identified as "P" represents those previously cleared with another GE Ultrasound system.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The letters are white, creating a contrast against the blue background. The logo is simple and recognizable, representing the well-known multinational conglomerate.
### Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 Ultrasound System
| Clinical Application | B | M | Mode of Operation | | | | | | | | Other |
|------------------------------------|---|---|-------------------|----|-------|---------|-------|--------------------|---------------------|----------------|-------|
| | | | PW | CW | Color | Color M | Power | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | |
| <i>Anatomy/Region of Interest</i> | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | P | [5] |
| Abdominal[1] | P | P | P | N | P | | P | P | P | P | [5] |
| Pediatric | P | P | P | P | P | P | P | P | P | P | |
| Small Organ[2] | P | N | P | | P | | P | P | P | N | [6] |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | P | N | P | | P | | P | P | P | N | |
| Musculo-skeletal Conventional | P | N | P | | P | | P | P | P | N | |
| Musculo-skeletal Superficial | P | N | P | | P | | P | P | P | N | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| <i>Exam Type, Means of Access</i> | | | | | | | | | | | |
| Transcranial | P | P | P | N | P | N | P | P | P | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | N | N | N | | N | | N | N | N | | |
| Transvaginal | P | P | P | | P | | P | P | P | | |
| Intraoperative | | | | | | | | | | | |
| <i>Interventional Guidance</i> | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | P | P | P | P | P | P | P | P | P | P | [5] |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
* JCombined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
["] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a circular border. The letters and the border are in a light blue color, set against a white background.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with 4C-RS Transducer
| | Mode of Operation | | | | | | | | | | | |
|------------------------------------|-------------------|----|----|----|-------|---------|-------|-------------------|---------------------|----------------|-------|--|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
| Anatomy/Region of Interest | | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | P | | |
| Abdominal[1] | P | N | P | | P | | P | P | P | P | | |
| Pediatric | P | N | P | | P | | P | P | P | P | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | | |
| Peripheral Vascular | P2 | P2 | P2 | | P2 | | P2 | P2 | P2 | P2 | | |
| Musculo-skeletal Conventional | N | N | N | | N | | N | N | N | N | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | P | P | | | P | | P | P | P | P | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a circular border. The letters and the border are in a light blue color, set against a white background.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with 8C-RS Transducer
| Clinical Application | B | M | Mode of Operation<br>Doppler Modes | | | | | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse' | Other |
|----------------------------------|----|----|------------------------------------|----|-------|---------|-------|--------------------|---------------------|-----------------|-------|
| Anatomy/Region of Interest | | | PW | CW | Color | Color M | Power | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal1 | | | | | | | | | | | |
| Pediatric | P | P | P | | P | | P | P | P | P | |
| Small Organ2 | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac3 | P | P | P | | P | | P | P | P | P | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P3 | P3 | P3 | | P3 | | P3 | P3 | P3 | P3 | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage4 | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circular frame. The frame has four small, rounded protrusions evenly spaced around its circumference, giving it a slightly decorative appearance.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with 3Sc-RS Transducer
| | Mode of Operation | | | | | | | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|-----------------------------------|-------------------|---|----|----|-------|---------|-------|-------------------|---------------------|----------------|-------|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | | | | |
| <i>Anatomy/Region of Interest</i> | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | P | P | P | P3 | P | P | P | P | P | | |
| Pediatric | P | P | P | P3 | P | P | P | P | P | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | P | P | P | | P | P | P | P | P | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| <i>Exam Type, Means of Access</i> | | | | | | | | | | | |
| Transcranial | P | P | P | | P | P | P | P | P | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| <i>Interventional Guidance</i> | | | | | | | | | | | |
| Tissue Biopsy/Fluid | | | | | | | | | | | |
| Drainage[4] | P | P | P | | P | P | P | P | P | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[D] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circular frame. The frame has four small, rounded protrusions evenly spaced around its circumference, giving it a slightly decorative appearance.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with L6-12-RS Transducer
| | Mode of Operation | | | | | | | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|-----------------------------------|-------------------|---|----|----|-------|---------|-------|---------------------|---------------------|----------------|-------|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | | | | |
| <i>Anatomy/Region of Interest</i> | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | P | | P | | P | | P | P | P | P2 | |
| Small Organ[2] | P | | P | | P | | P | P | P | P2 | [6] |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | P | | P | | P | | P | P | P | P2 | |
| Musculo-skeletal Conventional | P | | P | | P | | P | P | P | P2 | |
| Musculo-skeletal Superficial | P | | P | | P | | P | P | P | P2 | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| <i>Exam Type, Means of Access</i> | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| <i>Interventional Guidance</i> | | | | | | | | | | | |
| Tissue Biopsy/Fluid | | | | | | | | | | | |
| Drainage[4] | P | | P | | P | | P | P | P | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circular frame. The frame has four small, rounded protrusions evenly spaced around its circumference, giving it a slightly decorative appearance.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with E8C-RS Transducer
| | Mode of Operation | | | | | | | | | | |
|------------------------------------|-------------------|----|---------------|----|-------|---------|-------|-------------------|---------------------|----------------|-------|
| Clinical Application | B | M | Doppler Modes | | | | | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| | | | PW | CW | Color | Color M | Power | | | | |
| <i>Anatomy/Region of Interest</i> | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | P | |
| Abdominal[1] | P3 | P3 | P3 | | P3 | | P3 | P3 | P3 | P3 | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| <i>Exam Type, Means of Access</i> | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | P3 | P3 | P3 | | P3 | | P3 | P3 | P3 | | |
| Transvaginal | P | P | P | | P | | P | P | P | | |
| Intraoperative | | | | | | | | | | | |
| <i>Interventional Guidance</i> | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | P | P | P | | P | | P | P | P | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P=previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. There are three curved lines emanating from the circle, giving the impression of movement or energy.
### Diagnostic Ultrasound Indications for Use Form
### LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with RAB2-6-RS Transducer
| | Mode of Operation | | | | | | | | | | | |
|-------------------------------|-------------------|---|----|----|-------|---------|-------|---------------|--------------------|---------------------|-----------------|-------|
| Clinical Application | B | M | | | | | | Doppler Modes | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse' | Other |
| | | | PW | CW | Color | Color M | Power | | | | | |
| Anatomy/Region of Interest | | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | P | P | P | | P | | P | P | P | P | | [5] |
| Abdominal[1] | P | N | P | | P | | P | P | P | P | | [5] |
| Pediatric | | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Tissue Biopsy/Fluid | | | | | | | | | | | | |
| Drainage[4] | P | P | P | | P | P | P | P | P | P | | [5] |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P = previously cleared by FDA K151028; P=previously cleared by FDA K152758; P=previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric:
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Prescription Use (Per 21 CFR 801.109)
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a circular border. The letters and the border are in a light blue color, set against a white background.
### Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with 6S-RS Transducer
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|----|----|----|-------|---------|-------|---------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW | CW | Color | Color M | Power | Combined<br>Modes * | Harmonic<br>Imaging | Coded<br>Pulse | Other | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | p3 | p3 | p3 | p3 | p3 | p3 | p3 | p3 | p3 | | | |
| Small Organ [2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac [3] | p3 | p3 | p3 | p3 | p3 | p3 | p3 | p3 | p3 | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transcranial | N | N | N | N | N | N | N | N | N | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circular frame. The frame has four small, rounded protrusions evenly spaced around its circumference, giving it a slightly decorative appearance.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with L8-18i-RS Transducer
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|----|----|----|-------|---------|-------|--------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW | CW | Color | Color M | Power | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | P3 | P3 | P3 | | P3 | | P3 | P3 | N | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac [3] | | | | | | | | | | | |
| Peripheral Vascular | P3 | P3 | P3 | | P3 | | P3 | P3 | P3 | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | P3 | P3 | P3 | | P3 | | P3 | P3 | P3 | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage [4] | | | | | | | | | | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circular frame. The frame has four small, rounded protrusions evenly spaced around its circumference, giving it a slightly decorative appearance.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with E8Cs-RS Transducer
| | Mode of Operation | | | | | | | | | | |
|------------------------------------|-------------------|----|----|----|-------|---------|-------|----------------|------------------|-------------|-------|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined Modes | Harmonic Imaging | Coded Pulse | Other |
| Anatomy/Region of Interest | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | P4 | P4 | N | | P4 | | P4 | P4 | P4 | | |
| Abdominal[1] | P4 | P4 | N | | P4 | | P4 | P4 | P4 | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | N | N | N | | N | | N | N | N | | |
| Transvaginal | P4 | P4 | N | | P4 | | P4 | P4 | P4 | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid<br>Drainage[4] | P4 | P4 | N | | P4 | | P4 | P4 | P4 | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P =previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circular frame. The frame has four small, rounded protrusions evenly spaced around its circumference, giving it a slightly decorative appearance.
# Diagnostic Ultrasound Indications for Use Form LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 with BE9CS-RS Transducer
| | Mode of Operation | | | | | | | | | | |
|-------------------------------|-------------------|----|---------------------|----|-------|---------|-------|-------------------|---------------------|----------------|-------|
| Clinical Application | B | M | Doppler Modes<br>PW | CW | Color | Color M | Power | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Anatomy/Region of Interest | | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics | | | | | | | | | | | |
| Abdominal[1] | p5 | p5 | p5 | | p5 | | p5 | p5 | p5 | | |
| Pediatric | | | | | | | | | | | |
| Small Organ [2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Thoracic/Pleural | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transcranial | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | p5 | p5 | p5 | | p5 | | p5 | p5 | p5 | | |
| Transvaginal | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Interventional Guidance | | | | | | | | | | | |
| Tissue Biopsy/Fluid | | | | | | | | | | | |
| Drainage[4] | N | N | N | | N | | N | N | N | | |
| Vascular Access (IV, PICC) | | | | | | | | | | | |
| Nerve Block | | | | | | | | | | | |
Intended Use: Ultrasound imaging, measurement and analysis of the human body as follows:
N = new indication; P= previously cleared by FDA 133034; P2= previously cleared by FDA K141768;
P= previously cleared by FDA K151028; P=previously cleared by FDA K152758; P=previously cleared by FDA K141261
Notes: [1] Abdominal includes GYN and Urological/Prostate;
[2] Small Organ includes breast, testes, thyroid;
[3] Cardiac includes Adult and Pediatric;
[4] Interventional Guidance Tissue Biopsy is 2D biopsy guide
[5] 3D/4D imaging Mode
[6] Elastography imaging-Elasticity
[*]Combined modes are B/M, B/ Color M, B/PWD, B/Color/PWD, B/Power/PWD;
[*] Coded Pulse is for digitally encoded harmonics
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. There are swirling designs around the letters. The logo is simple and recognizable.
### 510(k) Summary
| | In accordance with 21 CFR 807.92 the following summary of information is provided: |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:<br>Submitter: | January 29, 2016<br>GE Medical Systems Ultrasound and Primary Care Diagnostics,<br>9900 Innovation Drive<br>Wauwatosa, WI 53226 |
| Primary Contact Person: | Tracey Ortiz<br>Regulatory Affairs Director<br>GE Healthcare<br>T:(262)676-6120<br>F:(414)918-8275 |
| Secondary Contact Person: | Jian Xie<br>Regulatory Affairs<br>GE Medical Systems (China) Co, Ltd.<br>T: +86 510 8527 8651<br>F: +86 510 8522 7347 |
| Device:<br>Trade Name:<br>Common/Usual Name:<br>Classification Names: | LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6<br>Ultrasound system<br>Class II |
| Product Code: | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Predicate Device(s): | LOGIQ F Series - K133034<br>LOGIQ S7 – K141261 Venue 50 - K152758<br>LOGIQ V3 / LOGIQ V5 -K141768<br>LOGIQ e-K151028 |
| Device Description: | The LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 are general<br>purpose diagnostic ultrasound systems which consists of a mobile<br>console approximately 1600mm high, 810mm in width and<br>720mm deep that provides digital acquisition, processing and<br>display capability. The user interface includes a computer<br>keyboard, 19 or 17 inch color LCD display and Touch panel.<br>LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 have the same<br>hardware and system features. |
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined in a stylized script, enclosed within a blue circle. There are decorative swirls around the outer edge of the circle, giving it a classic and recognizable appearance.
# GE Healthcare
510(k) Premarket Notification Submission
- Intended Use: The LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 are general purpose ultrasound imaging and analysis systems providing digital acquisition, processing and display capability, clinical applications including: Abdominal, Obstetrical, Gynecological, parts. Vascular/Peripheral Vascular, Transcranial. Small Pediatric, Musculoskeletal, Urological, Cardiac, Transvaginal, Transrectal, and Biopsy.
- Technology: The LOGIQ F8 Expert / LOGIQ F8 / LOGIQ F6 employ the same fundamental scientific technology as its predicate devices.
#### Determination of Comparison to Pr…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.