The Hi bebeplus, model BT-100, electric breast pump is intended to be used by lactating women for expressing and collecting breast milk. It is intended for a single user.
Device Story
Electrically powered, software-controlled breast pump; intended for single-user, home-use by lactating women. System comprises motor unit, pump kit, and tubing. Operates via 12V DC adapter or rechargeable lithium battery. Provides single or double pumping options. Features basic massage/expression function and memory function to save operating sequences. Vacuum levels range 32-227 mmHg; cycling rates up to 63 cycles per minute. Includes back flow protection. User controls vacuum/cycle settings via interface with LCD display. Output is expressed breast milk collected in pump kit. Benefits include efficient milk expression and collection for breastfeeding mothers.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-5, -10), electrical safety (IEC 60601-1, -11), EMC (IEC 60601-1-2), and performance testing (cycle speed, suction pressure, battery life, operating life).
Technological Characteristics
Powered breast pump; PP funnel material. Microcontroller-based cycling control. Vacuum range 32-227 mmHg; cycle range 27-63 cpm. Power: 12V DC adapter or Li-Ion battery. Home-use environment. Biocompatibility per ISO 10993; Electrical safety per IEC 60601-1/1-11; EMC per IEC 60601-1-2.
Indications for Use
Indicated for lactating women for expressing and collecting breast milk. Intended for single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with a wing-like shape extending above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 1, 2016
Bistos Co., Ltd. % Dave Kim President Mtech Group 8310 Buffalo Speedway Houston, TX 77025
Re: K160274 Trade/Device Name: Hi Beber1us> model BT-100 Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: July 21, 2016 Received: July 22, 2016
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
For Division
Douglas Silverstein -S 2016.09.01 16:14:46 -04'00'
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160274
Device Name
Hi bebePlus , model BT-100
Indications for Use (Describe)
The Hi bebertus, model BT-100, electric breast pump is intended to be used by lactating women for expressing and collecting breast milk. It is intended for a single user.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
Z Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## K160274
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### Date 510k summary prepared: August 31, 2016
#### I. SUBMITTER
| Submitter's Name | Bistos Co., Ltd. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address | 7th Fl. A Bldg., Woolim Lions Valley 5-cha, 302<br>Galmachi-ro, Jungwon-gu, Seongnam-si, Gyeonggi-do,<br>Korea (Zip. 13201) |
| Submitter's Telephone | +82 (31) 7500340 |
| Contact person | Jenny Cho (jennycho@bistos.co.kr) / RA Manager |
| Official Correspondent | Dave Kim (davekim@mtech-inc.net) |
| Address | 8310 Buffalo Speedway, Houston, TX 77025 |
| Telephone | +713-467-2607 |
| Fax: | +713-583-8988 |
#### II. DEVICE
| Trade/proprietary Name | Hi bebeplus |
|------------------------|----------------------------|
| Model No. | BT-100 |
| Common or Usual Name | Electric Breast Pump |
| Regulation Name | Powered Breast Pump |
| Regulation Number | 21 CFR 884.5160 |
| Product Code | HGX: pump, breast, powered |
| Regulatory Class | Class II |
| Over the Counter Use | |
#### III. PREDICATE DEVICE
| Primary Manufacturer | Medela Ag |
|----------------------|-------------------------------------------|
| Device Name | Freestyle Deluxe, Freestyle Solution Set, |
| | Freestyle Basic, Freestyle Motor Warranty |
| 510(k) Number | K150499 |
| Regulation Name | Powered Breast Pump |
| Regulation Number | 21 CFR 884.5160 (Product Code: HGX) |
| Regulatory Class | Class II |
| Over the Counter Use | |
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
The Hi beberths, model BT-100, is an electrically powered, software-controlled breast pump intended to express and collect milk from the breast of lactating women. The breast pump system is comprised of a motor unit and pump kit including tubing..
The pump is intended to be used by a single user. The user has the option to pump breast milk from a single breast (single pumping) or from both breasts (double pumping). The Hi beberths, model BT-100 also includes back flow protection and it is powered by a 12V DC adaptor or rechargeable lithium battery.
The Hi beberths, model BT-100 has two operating functions: basic massage/expression function and memory function to save the operating sequence. The Hi beber™, model BT-100, is capable of providing vacuum levels from 32-227 mmHg with cycling rates up to 63 cycles per minute.
#### V. INDICATIONS FOR USE:
The Hi beberths, model BT-100, electric breast pump is intended to be used by lactating women for expressing and collecting breast milk. It is intended for a single user.
| | Hi bebeplus<br>BT-100 | Freestyle® Breast pump<br>(K150499) |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | | |
| Indication for use | The Hi bebeplus, model BT-100,<br>electric breast<br>pump is intended to be used by<br>lactating women for expressing<br>and collecting breast milk. It is<br>intended for a single user. | Freestyle® is a powered<br>breast pump to be used by<br>lactating women to express<br>and collect milk form their<br>breasts. The Freestyle® is<br>intended for a single user |
| Intended Use | Express and collect milk | Express and collect milk |
| Single User | | |
| Device | Yes | Yes |
| Environment of Use | Home | Home |
| Prescription<br>Designation | OTC | OTC |
| User Control | On-off switch Vacuum /Cycle<br>Up Vacuum /Cycle Down<br>Change Operation Mode | On-off switch Vacuum/Cycle<br>adjustment control |
| Visual<br>Indicator | LCD display | LCD display |
| Pumping options | Single or Double pumping | Single or Double pumping |
| Power Supply | Li-Ion battery or<br>AC adaptor provided<br>100V-240V, 50/60 Hz | Li-Ion battery or<br>AC adaptor provided |
| Adjustable<br>Vacuum<br>(expression) | 32 - 227 mmHg ±10% | 45-245 mmHg |
| Maximum Vacuum | 250 mmHg | 270 mmHg |
| Cycles per minute<br>(expression) | 27 - 63 ±5% | 49.8 - 81.6 |
| Suction<br>Settings | 10 | 9 |
| Adjustable<br>Suction Levels | Yes | Yes |
| Mode change | Yes | Yes |
| Cycling Control<br>Mechanism | Microcontroller | Microcontroller |
| Back Flow<br>Protection | Yes | Yes |
| Patient Contact<br>Material | Funnel made of PP | PP |
#### PREDICATE COMPARISON VI.
{5}------------------------------------------------
{6}------------------------------------------------
The Hi beberths, model BT-100 and the predicate device do not have identical indication for use statements; however, they have the same intended use (expressing and collecting milk from the breasts of lactating women).
In addition, there are differences in the device technological characteristics. The predicate device has 9 adjustable suction levels while the subject device has 10 adjustable suction levels. Also, the predicate device and subject device have different suction strengths and cycle speed ranges. These differences do not raise different questions of safety as compared to the predicate device. The suction levels both allow users to adjust the suction strength.
#### SUMMARY OF NON-CLINICAL TESTS VII.
Hi beber 100 BT-100 complies with voluntary standards for biocompatibility (in vitro cytotoxicity, irritation and sensitization testing), electrical safety, EMC testing, and use in the home healthcare environment. The following data were provided to support the substantial equivalence determination:
### Biocompatibility:
Testing was conducted in accordance with AAMI / ANSI / ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical Devices-- Part 5: Tests for In Vitro Cytotoxicity (1.929 Assay) and AAMI / ANSI / ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization. (Vaginal Irritation and Guinea Pig Maximization Sensitization)
### Electrical Safety:
Testing was conducted in accordance with AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) and IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance -Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
### Electromagnetic Compatibility:
Testing was conducted in accordance with IEC 60601-1-2:2007 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
### Software:
Software verification and validation testing as recommended in FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." (May 11, 2005)
{7}------------------------------------------------
FDA Guidance Document, Design Considerations for Devices Intended for Home Use, issued November 24, 2014 was also used to address home-use considerations.
Performance testing was conducted to show that the device meets its design requirements and performs as intended. The specifications were met for:
- Cycle Speed
- . Suction Pressure
- Battery Operating Time
- Battery Recharging Time
- . Operating life
These tests were conducted under conditions of single and double pumping mode and for the varying power sources (e.g., AC/DC power vs. battery power).
#### VIII. SUMMARY OF CLINICAL TESTS
Clinical testing was not required to demonstrate the substantial equivalence of the Hi bebertus, model BT-100 electric breast pump to its predicate device.
#### CONCLUSIONS IX.
Based on the information above, Hi bebert , model BT-100 electrical breast pump is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.