K160272 · Endophys, Inc. · DYB · Mar 4, 2016 · Cardiovascular
Device Facts
Record ID
K160272
Device Name
Endophys Pressure Sensing Sheath Kit
Applicant
Endophys, Inc.
Product Code
DYB · Cardiovascular
Decision Date
Mar 4, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1340
Device Class
Class 2
Indications for Use
The Endophys Pressure Sensing Sheath Kit (Endophys Pressure Sensing Sheath, vessel dilator and guidewire) is intended to facilitate the introduction of diagnostic and interventional devices into the vasculature and to continuously measure blood pressure during the procedure when used with the Endophys Blood Pressure Monitor.
Device Story
The Endophys Pressure Sensing Sheath Kit consists of an introducer sheath with an integrated fiber optic pressure transducer, a vessel dilator, and a guidewire. Used in clinical settings by physicians to facilitate vascular access for diagnostic or interventional devices. The sheath connects to the Endophys Blood Pressure Monitor to provide continuous, high-resolution, instantaneous pulse wave hemodynamic data. The device assists in placement via a guidewire and dilator, which are removed once the sheath is positioned. The monitor displays blood pressure measurements, allowing the clinician to observe real-time hemodynamic status during procedures. This continuous monitoring aids clinical decision-making by providing immediate feedback on patient blood pressure, potentially improving patient safety during vascular interventions.
Clinical Evidence
Bench testing only. Performance data included luer verification, dimensional verification, accessories compatibility, radiopacity, tip deflection, kink resistance, sensor functionality, tip compression, liquid leakage, separation force, and corrosion resistance. The device met all specified criteria.
Technological Characteristics
Introducer sheath with integrated fiber optic pressure transducer. Includes vessel dilator and J-tipped guidewire. Features hemostasis valve, suture ring, and locking sheath hub. Radiopaque for imaging. Sterile (EO). Designed for connection to the Endophys Blood Pressure Monitor.
Indications for Use
Indicated for patients requiring vascular access for diagnostic or interventional procedures where continuous blood pressure monitoring is required.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human figures in profile, arranged in a row. The figures are connected by flowing lines, creating a sense of unity and continuity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2016
Endophys, Inc. Rebecca Pine Consultant Thanksgiving Tower, Suite 1930 1601 Elm Street Dallas, Texas 75201
Re: K160272
Trade/Device Name: Endophys Pressure Sensing Sheath Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB, DXO Dated: January 28, 2016 Received: February 2, 2016
Dear Rebecca Pine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 5. Indications for Use Statement
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Food and Drug Administration | | |
| Indications for Use | | Form Approved: OMB No. 0910-0120 |
| | | Expiration Date: January 31, 2017 |
| | | See PRA Statement below. |
| 510(k) Number (if known) | K160272 | |
| Device Name | Endophys Pressure Sensing Sheath Kit | |
| Indications for Use (Describe) | The Endophys Pressure Sensing Sheath Kit (Endophys Pressure Sensing Sheath, vessel dilator and guidewire) is intended to facilitate the introduction of diagnostic and interventional devices into the vasculature and to continuously measure blood pressure during the procedure when used with the Endophys Blood Pressure Monitor. | |
| Type of Use (Select one or both, as applicable) | | |
| <div><span style="font-family: Arial, sans-serif;"><b></b></span><span style="font-family: Symbol;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span><span style="font-family: Arial, sans-serif;"><b> </b></span></div> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 | PSC Publishing Service (301)-443-6740 | EF |
|----------------------|-------------|---------------------------------------|----|
|----------------------|-------------|---------------------------------------|----|
Page 1 of 1
PSC Polylishing Services (101) 445-6740
{3}------------------------------------------------
### 6. 510(k) Summarv
#### I. SUBMITTER
Endophys, Inc. Thanksgiving Tower, Suite 1930 1601 Elm Street Dallas, TX 75201
Contact person: Rebecca K Pine Phone: (760) 809-5178 Fax: (760) 290.3216 Date prepared: January 28, 2016
#### II. DEVICE
Name of the device: Endophys Pressure Sensing Sheath Kit Common of usual name: Introducer Catheter and accessories Classification name: Catheter Introducer Regulatory Class: 2 Product Code: DYB (classification); DXO (subsequent)
- III. PREDICATE DEVICE Endophys Pressure Sensing Sheath Kit (K141275) This predicate has not been subject to a design-related recall
No reference devices were used in this submission.
- IV. DEVICE DESCRIPTION
The Endophys Pressure Sensing Sheath Kit is comprised of the Endophys Pressure Sensing Sheath, vessel dilator and guidewire. The Endophys Pressure Sensing Sheath and accessories are provided sterile (EO).
The Pressure Sensing Sheath ("PSS") is an introducer sheath with an integrated fiber optic pressure transducer. The PSS is provided with a dilator and a guidewire, which together make up the Endophys Pressure Sensing Sheath Kit ("PSS Kit"). The PSS Kit is intended to be used with the Endophys Blood Pressure Monitor ("BPM"), which connects to the PSS and displays the blood pressure measurements. Together with the BPM, the PSS is used to continuously monitor patient blood pressure during procedures requiring vascular access.
The PSS introducer sheath is used to assist vascular access in the same
{4}------------------------------------------------
manner as do standard catheter introducers. The PSS introducer sheath has an integrated pressure sensor that provides high resolution, continuous, and instantaneous pulse wave hemodynamics to monitor blood pressure. The PSS introducer sheath is intened only for connectin with the Endophys Blood Pressure Monitor.
The 6F Vessel Dilator is used to facilitate insertion of the PSS, and is packaged with the Endophys PSS Kit. Once the PSS has been properly positioned in the vessel, the Dilator is removed.
The guidewire is J tipped, 50cm (length), 0.89mm (diameter), and is used to assist in the placement of the Dilator/PSS Introducer Sheath through the skin and into the vasculature. Once the Dilator and Introducer are positioned over the Guidewire, the Guidewire is removed through the Dilator.
#### V. INDICATIONS FOR USE
The Endophys Pressure Sensing Sheath Kit (Endophys Pressure Sensing Sheath, vessel dilator and guidewire) is intended to facilitate the introduction of diagnostic and interventional devices into the vasculature and to continuously measure blood pressure during the procedure when used with the Endophys Blood Pressure Monitor.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Introducer catheters, by nature, are tubular structures intended to provide a pathway for therapeutic and/or diagnostic devices into the human vasculature. Specialized introducer catheter have been developed which incorporate sensors into these devices to allow active monitoring of blood pressure during interventional procedures. Blood pressure monitoring using a fiber optic transducer is the technological principle for both the subject and predicate devices. The technology requires use of a compatible blood pressure monitor to provide this data to the clinician.
At a high level, the subject and predicate devices are based on the following same technological elements:
- Compatibility with Endophys Blood Pressure Monitor .
- l Hemostasis valve- to prevent blood leakage during the procedure
- Suture ring- aid in placement
- Fiber optic pressure transducer I
- I Visibility under radiographic imaging
- I Locking sheath hub
{5}------------------------------------------------
The following technological differences exist between the subject and predicate devices:
- Modify sheath hub taper for ease of dilator insertion l
- Increase in sheath inside diameter from 0.82 to 0.89
- ' Incorporate locking dilator hub feature
- I Modify dilator/sheath dimensions to minimize sheath/dilator profile including a reduction in sheath shaft length and increased taper
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence.
- l Luer verification
- Dimensional verification ■
- Accessories compatibility
- I Radiopacity
- Tip deflection
- Kink resistance
- I Sensor functionality
- l Tip compression
- I Liquid leakage
- Separation force
- I Corrosion resistance
The modified Endophys Pressure Sensing Sheath met all specified criteria and did not raise new safety or performance questions. Based on the design verification performance the modified Endophys Pressure Sensing Sheath Kit was found to have the a safety and effectiveness profile that is similar to the predicate device.
#### VIII. CONCLUSIONS
The design testing performed as a part of the applied Design Controls to the modified Endophys Pressure Sensing Sheath Kit demonstrated that the performance of the modified device is equal to the legally marketed Endophys Pressure Sensing Sheath Kit.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.