K160259 · Zeltiq Aesthetics, Inc. · OOK · Mar 23, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160259
Device Name
ZELTIQ CoolSculpting System
Applicant
Zeltiq Aesthetics, Inc.
Product Code
OOK · General, Plastic Surgery
Decision Date
Mar 23, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4340
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K160259 · Mar 23, 2016
ZELTIQ CoolSculpting System
Zeltiq Aesthetics, Inc.
Published clinical literature
The sponsor conducted a literature review of 4,792 cryolipolysis treatments to demonstrate the safety and efficacy profile of the CoolSculpting System across multiple anatomical sites (abdomen, flanks, back, thighs, submental area, and banana roll).
Literature review; Cryolipolysis; Fat reduction; Safety profile
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Compilation of published clinical studies; Literature review of clinical studies
Individuals undergoing cryolipolysis for fat reduction; Sample Size: 4,792 treatments
Not applicable for this study
Fat layer reduction (ultrasound/caliper), circumferential reduction, volumetric reduction, blinded photo review, and incidence of adverse events
Indications for Use
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of bra fat, banana roll, submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in bra fat, back fat, banana roll, submental area, thigh, abdomen and flank. Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort. The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite. The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
Device Story
Portable thermoelectric cooling/heating system; applies controlled thermal energy to treatment sites. Comprised of control unit, detachable vacuum/surface applicators, and accessories (liners, gel/gelpads, cycle cards). Operated by clinicians in clinical settings. Facilitates cold-assisted lipolysis (fat breakdown) via controlled cooling; provides localized thermal therapy for pain/muscle spasms. Optional massage function improves local circulation and reduces cellulite appearance. Pretreatment skin wipes/gelpads ensure thermal contact. Clinicians use device to target visible fat bulges; output affects patient appearance and provides symptomatic relief for minor aches/pains. Benefits include non-surgical subcutaneous fat reduction and pain management.
Clinical Evidence
Literature review of 4,792 cryolipolysis treatments. Efficacy metrics: ultrasound fat layer reduction (mean 20.6%, 3.9 mm), caliper reduction (mean 22.3%), and independent photo review (89.7% correct). Circumferential reductions reported (0.9 cm to 6.8 cm). Safety profile: 0.13% incidence of serious adverse events (e.g., paradoxical hyperplasia, contour irregularity). Common side effects: erythema, bruising, numbness.
Technological Characteristics
Thermoelectric cooling/heating device. Features vacuum and non-vacuum surface applicators. System includes control unit, liners, gel/gelpads, cycle cards, foam borders, and securement system. Operates via controlled thermal transfer. No specific materials or software architecture details provided.
Indications for Use
Indicated for cold-assisted lipolysis of bra fat, back fat, banana roll, submental area, thigh, abdomen, and flank in individuals with BMI ≤ 30. Also indicated for localized thermal therapy to minimize pain post-trauma/surgery, relief of minor aches/pains/muscle spasms, and temporary improvement in local circulation/reduction in cellulite appearance.
Regulatory Classification
Identification
A contact cooling system for aesthetic use is a device that is a combination of a cooling pad associated with a vacuum or mechanical massager intended for the disruption of adipocyte cells intended for non-invasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special controls for this device is FDA's “Guidance for Industry and FDA Staff; Class II Special Controls Guidance Document: Contact Cooling System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with three heads in profile, representing the department's focus on people. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2016
Zeltiq Aesthetics, Inc. Shruti Jayakumar Regulatory Affairs Manager 4698 Willow Road Pleasanton, California 94588
Re: K160259
Trade/Device Name: Zeltiq Coolsculpting System Regulation Number: 21 CFR 878.4340 Regulation Name: Contact Cooling System For Aesthetic Use Regulatory Class: Class II Product Code: OOK Dated: January 29, 2016 Received: February 1, 2016
Dear Shruti Jayakumar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements
{1}------------------------------------------------
as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160259
Device Name CoolSculpting System
#### Indications for Use (Describe)
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of bra fat, banana roll, submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in bra fat, back fat, banana roll, submental area, thigh, abdomen and flank.
Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.
The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.
The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | ZELTIQ™ Aesthetics, Inc.<br>4698 Willow Road<br>Pleasanton, CA 94588 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Shruti Jayakumar<br>Regulatory Affairs Manager<br>ZELTIQ Aesthetics, Inc.<br>Phone: 925-474-2516<br>Fax: 925-474-8028 |
| DATE PREPARED: | March 14, 2016 |
| TRADE NAME: | ZELTIQ CoolSculpting System |
| COMMON NAME: | Skin Cooling Device |
| CLASSIFICATION NAME: | Contact Cooling System for Aesthetic Use |
| DEVICE CLASSIFICATION: | Class II, 21 CFR §878.4340 |
| PRODUCT CODE: | OOK |
DESCRIPTION OF THE DEVICE SUBJECT TO IDENTIFICATION
PREDICATE DEVICES: The ZELTIQ CoolSculpting System (K151179)
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The CoolSculpting System is a portable thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. The CoolSculpting System is comprised of a control unit, detachable vacuum and surface applicators and supplies such as liners, gel/gelpads, cycle cards, foam borders and securement system.
### SUBSTANTIALLY EQUIVALENT TO:
The ZELTIQ CoolSculpting System is substantially equivalent to the ZELTIQ Dermal Cooling Device, also known as the ZELTIQ CoolSculpting System, which has been cleared for the indication of cold-assisted lipolysis of the flank (love handle), abdomen, thighs and submental area (K151179).
{4}------------------------------------------------
Clinical data from ZELTIQ and data from published scientific literature indicates that the CoolSculpting System has the same mechanism of action regardless of treatment site. The safety and efficacy profile remains the same.
### INDICATION FOR USE:
The CoolSculpting System is a skin cooling or heating device is indicated for cold-assisted lipolysis (breakdown of fat) of bra fat, back fat, banana roll, submental area, thigh, abdomen, and flank, or "love handles" in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in bra fat, banana roll, submental area, thigh, abdomen and flank.
Cooling with the device may also be used to minimize pain and thermal injury during laser and dermatological treatments and act as a local anesthetic for procedures that induce minor local discomfort.
The CoolSculpting System is also indicated for use to provide localized thermal therapy (hot or cold) to minimize pain post-trauma and post-surgery, and for temporary relief of minor aches, pains, and muscle spasms. The optional massage function can also be used for the temporary relief of minor muscle aches, pain, and spasm and for temporary improvement in local circulation and temporary reduction in the appearance of cellulite.
The ZELTIQ Pretreatment Skin Wipe and Gel/Gelpad facilitate thermal contact of the device with a patient's skin by mitigating minor variances in device-to-skin contact.
### TECHNICAL CHARACTERISTICS:
The CoolSculpting System is a thermoelectric cooling and heating device that applies controlled cooling or heating to a treatment site. This system features vacuum applicators of various sizes and non-vacuum surface applicators that are intended to provide clinicians with an additional option when treating a flat area of the body. The technological characteristics are the predicate devices. All share the same mechanism of cooling and heating for the same intended use. There is no change required to the device to accommodate additional treatment sites. The same as has been previously cleared.
### PERFORMANCE DATA:
A review of clinical publications revealed 4,792 cryolipolysis treatments during clinical studies. From these studies, we compiled the numbers of treatments in several anatomical areas: 1,695 treatments in the abdomen, 1,987 treatments in the flanks, 501 treatments in the back, 323 treatments in the inner thigh, 150 treatments in the lateral thigh, 3 treatments in the anterior thigh, 119 treatments in the submental area, and 14 treatments in the banana roll region.
Efficacy was measured by several techniques including ultrasound and caliper measurements, circumferential measurements, 3D quantification of volume reduction, and blinded, independent review of clinical photographs. Based on the compilation of data from these studies, the overall mean
{5}------------------------------------------------
ultrasound fat layer reduction ranged from 10.3 to 25.5% and 1.9 to 8.3 mm. Compiled mean caliper fat layer reduction ranged from 14.7 to 23.0%. Single studies showed mean 0.9 cm circumferential reduction in the inner thigh, 2.4 cm circumferential reduction in the flanks, 6.8 cm circumferential reduction in the abdomen, and 39.6 cm³ volumetric reduction in the flanks.
Based on the compilation of these various studies, the overall mean ultrasound fat layer thickness reduction was 20.6% and 3.9 mm. Compiled mean caliper fat layer reduction was 22.3%. The independent photo review was 89.7% correct, on average.
As shown by multiple clinical studies submitted for clearance to the agency, the summary of published data shows a similarly high safety and efficacy profile for the cryolipolysis procedure. Common procedural side effects include erythema, bruising, and numbness, which typically resolve within one month of treatment. Based on the literature review, 6 cases would be considered serious adverse events. These serious adverse events include three cases of paradoxical hyperplasia in the abdomen, one case of paradoxical hyperplasia in the abdomen, back, and flanks, one case of contour irregularity in the abdomen, and one case of contour irregularity in the flank. For 4,792 treatments in published studies, the incidence of serious adverse events is very low (0.13%). Given the fact that 76.8% of treatments were to the abdomen and flanks, this incidence rate shows no clear indication of treatment site specificity. The clinical publications indicate that cryolipolysis is a safe and effective non-surgical procedure for subcutaneous fat reduction.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.