K160257 · Cardiacassist, Inc. · DWF · Mar 4, 2016 · Cardiovascular
Device Facts
Record ID
K160257
Device Name
PROTEK Duo 31 Fr. Veno-Venous Cannula Set
Applicant
Cardiacassist, Inc.
Product Code
DWF · Cardiovascular
Decision Date
Mar 4, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4210
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.
Device Story
The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is a sterile, single-use device for extracorporeal life support. It consists of a dual-lumen, wire-reinforced cannula and a 15.5 Fr. introducer. The cannula features two concentric channels for simultaneous venous drainage and reinfusion. During insertion, the introducer is placed inside the inner lumen to facilitate advancement over a 0.038-inch guidewire into the internal jugular vein. The introducer includes a luer hub for contrast injection and a hemostasis cap to minimize blood loss. Once positioned, the cannula is connected to extracorporeal circulatory support equipment via barbed connectors. The device is used by clinicians in a hospital setting. It provides a single-cannula solution for blood access, potentially benefiting patients requiring higher blood flow rates than those supported by smaller-diameter cannulae.
Clinical Evidence
Bench testing only. Comparative hemolysis testing showed no difference between the subject device and the predicate over 6 hours. Pressure-flow testing confirmed the device supports higher flow rates than the smaller predicate. Tensile strength, pathway integrity, kink radius, and stiffness testing met all pre-defined acceptance criteria.
Technological Characteristics
Dual-lumen, wire-reinforced cannula body with concentric channels. Includes 15.5 Fr. introducer with luer hub. Materials and manufacturing methods are identical to the predicate. Radiopaque introducer for fluoroscopic visualization. Sterile, single-use device.
Indications for Use
Indicated for patients requiring extracorporeal life support procedures via internal jugular vein access for venous drainage and blood reinfusion.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 4, 2016
Cardiac Assist, Inc. Greg Johnson, PhD Director of Regulatory Affairs 240 Alpha Drive Pittsburgh, PA 15238
Re: K160257
Trade/Device Name: PROTEK Duo 31 Fr. Veno-Venous Cannula Set Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing Regulatory Class: Class II Product Code: DWF Dated: January 27, 2016 Received: February 1, 2016
Dear Greg Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICE<br>Food and Drug Administration | orm Approved: OMB No. 0910-0120<br>piration Date: January 31, 201 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Indications for Use | See PRA Statement below |
| 510(k) Number (if known)<br>K160257 | |
| PROTEK Duo 31 Fr. Veno-Venous Cannula Set<br>Device Name | |
| Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and<br>reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.<br>Indications for Use (Describe) The PROTEK Duo 31 Fr. ' | |
| | |
| | |
| | |
| | |
| | |
| | |
| or both, as applicable)<br>Type of Use (Select one | |
| 🇿 Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
| CONTINUE ON A SEPARATE PAGE IF NEEDED | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.<br>This section applies only to requirements of the Paperwork Reduction Act of 1995 | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and complete and review the collection of information. Send comments regarding this burden estimate or any o | |
| Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Staff<br>Office of Chief Information Officer<br>ood and Drug Administration<br>PRAStaff@fda.hhs.gov | |
| An agency may not conduct or sponsor, and a person is not respond to, a collection | |
| Page 1 of 1<br>FORM FDA 3881 (8/14) | PSC Publishing Services (301) 443-6740 |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for CardiacAssist, Inc. The logo features a circular graphic with arrows pointing in different directions, followed by the company name in a bold, blue font. The text "CardiacAssist, Inc." is written in a clear, sans-serif typeface, making it easily readable. The overall design is clean and professional, conveying a sense of reliability and innovation.
Date: 3/3/2016
## Applicant
CardiacAssist, Inc. 240 Alpha Drive Pittsburgh, PA 15238 Telephone: 412-963-7770 Fax: 412-963-0800
## Contact person
Greg Johnson, PhD Title: Director of Regulatory Affairs Phone: 412-963-7770 x266 e-mail: gjohnson@tandemheart.com
#### Device
| Trade/Proprietary Name: | PROTEK Duo 31 Fr. Veno-Venous Cannula Set |
|-------------------------|------------------------------------------------------------------------------------------------------|
| Common Name: | Veno-Venous Cannula and Introducer |
| Classification Name: | Cardiopulmonary bypass vascular catheter, cannula, or<br>tubing. (21 CFR 870.4210, Product Code DWF) |
## Predicate Device
PROTEK Duo 29 Fr. Veno-Venous Cannula (K140999)
#### Reference Device
Avalon Elite 31 Fr. Veno-Venous Cannula (K081820)
## Device Description
The PROTEK Duo 31 Fr. Veno-Venous Cannula Set consists of two components: a 31 Fr. Dual lumen Veno-Venous Cannula and a 15.5 Fr. Introducer. The Introducer is designed to accept a standard 0.038 inch guidewire. The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is intended as a single patient, single use, sterile device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CardiacAssist, Inc. The logo features a circular graphic with arrows pointing in different directions. The text "CardiacAssist, Inc." is written in a bold, sans-serif font to the right of the graphic.
Image /page/4/Figure/2 description: This image shows a medical device with several labeled components. The device features barbed connectors at one end and tantalum markers along its length. Insertion depth markers are visible, labeled with numbers 52, 50, and 48, and a wire-reinforced area is indicated in the middle section of the device.
Figure 1. PROTEK Duo 31 Fr. Veno-Venous Cannula
The PROTEK Duo 31 Fr. Veno-Venous Cannula (Figure 1) consists of two distinct lumens with a wire-reinforced cannula body. The inner lumen is located entirely within the outer lumen forming two concentric channels
The distal section (inner lumen/cannula) of the cannula body has six side holes near the distal tip opening. The proximal sections of each lumen are clear and not wire-reinforced to allow visualization of blood and to enable clamping to prevent blood flow during setup and removal of the cannula from the extracorporeal circulatory support equipment (see Figure 1). A non-vented barbed connector is affixed to both proximal ends (inner/distal and outer/proximal lumens) of the cannula and allow for connection of standard 3/8 inch blood circuit tubing for subsequent connection to extracorporeal circulatory support equipment.
The introducer (Figure 2) consists of a tube with a luer hub. The introducer fits inside the inner lumen of the cannula during insertion of the cannula/introducer assembly. The introducer is used to advance the cannula over a guidewire and facilitate cannula placement within the target vessel. The introducer has a luer hub at its proximal end to manage introducer insertion and removal from the cannula. The luer hub can also enable contrast injection, if necessary, to facilitate placement and final positioning of the cannula The hemostasis cap minimizes blood loss when the within the target vessel. cannula/introducer assembly is inserted into the target vessel. The introducer body is constructed from radiopaque material for visualization under fluoroscopy.
Image /page/4/Figure/7 description: The image shows a medical device, specifically an introducer needle. The device has a long, thin needle with a sharp tip on the right side of the image. On the left side of the image, there is an introducer hub and hemostasis cap, which are labeled with arrows pointing to them.
Figure 2: 15.5 Fr. Introducer
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for CardiacAssist, Inc. The logo features a stylized graphic to the left of the company name. The graphic consists of a circle with arrows pointing in different directions, suggesting circulation or assistance. The text "CardiacAssist, Inc." is written in a bold, sans-serif font, with the company name in a dark blue color.
#### Indications for Use
The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is intended for use as a single cannula for both venous drainage and reinfusion of blood via an internal jugular vein during extracorporeal life support procedures.
#### Comparison of Technological Characteristics
The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is identical to the predicate PROTEK Duo 29 Fr. Veno-Venous Cannula Set, with the exception that it is two French larger in diameter along the proximal length, 2.5 Fr larger in diameter along the distal portion of the device, and approximately 1.97 inches (5.0 cm) longer. It is designed for the same intended use as the PROTEK Duo 29 Fr. Veno-Venous Cannula Set, but in larger patients and/or those who require additional blood flow. All materials and methods of manufacture are identical.
#### Summary of Non-clinical Testing
Testing of the PROTEK Duo 31 Fr. Veno-Venous Cannula Set included comparative hemolysis, pressure-flow testing, tensile strength, pathway integrity, kink radius, and stiffness.
| Test | Method | Conclusion |
|--------------------------|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparative<br>Hemolysis | Side-by-side comparison with<br>predicate of bench top<br>hemolysis levels over 6 hours. | No difference in hemolysis levels<br>between test articles and predicate<br>controls. |
| Pressure-Flow | Measure pressure losses across<br>cannula at different flow rates. | Measured flow rates exceed the<br>smaller predicate at all levels of<br>pressure difference across the<br>cannula. Design specifications<br>were met. |
| Tensile Strength | Pull testing of both cannula and<br>introducer | Acceptance criteria were identical<br>to those of the predicate and were<br>met in all tests. |
| Pathway Integrity | Pressure testing | Acceptance criterion was identical<br>to that of the predicate and was met<br>in all tests. |
| Kink Radius | Flow rate reduction caused by<br>specified minimum bend radius<br>was measured | Acceptance criterion was identical<br>to that of the predicate and was met<br>in all tests. |
| Stiffness | Deflection testing to measure<br>force required to bend cannula<br>a specified distance | The cannula, the introducer, and the<br>cannula/introducer assembly met<br>acceptance criteria established to<br>ensure the cannula is stiff enough to<br>insert while remaining flexible<br>enough to avoid vessel injury. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the logo of CardiacAssist, Inc. The logo consists of a stylized graphic to the left of the company name. The graphic appears to be a circular design with internal elements, possibly representing a heart or circulatory system. The text "CardiacAssist, Inc." is written in a bold, sans-serif font, with the company name clearly emphasized.
## Conclusion
The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is made from the same materials, using the same manufacturing processes and sterilization techniques as the predicate device. Testing demonstrated that hemolysis, tensile strength, pathway integrity, and kink radius are identical to the predicate. Stiffness and pressure-flow properties differed as expected for the larger diameter cannula and met established acceptance criteria. The PROTEK Duo 31 Fr. Veno-Venous Cannula Set is determined to be substantially equivalent to the predicate PROTEK Duo 29 Fr. Veno-Venous Cannula Set.
{End of Section}
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.