The HBOTC is intended for use as a nerve and muscle stimulator for patients wanting symptomatic relief and management of chronic pain as well as muscle conditioning and increasing local blood flow.
Device Story
HBOTC is a portable, battery-operated, two-channel electrical nerve and muscle stimulator for OTC use. Device inputs include user-selected frequency and intensity via buttons and dials. It transforms electrical energy into biphasic pulses (spike with exponential decay) delivered through self-adhesive electrodes. Output is displayed on an LCD. Used by patients in home settings for pain management and muscle conditioning. Healthcare providers do not operate the device; patients use it to manage pain or improve muscle performance. Benefits include non-pharmacological pain relief and muscle conditioning. Device includes timer, output jacks, and charging jack.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through software validation and compliance with IEC 60601-1, IEC 60601-1-11, and IEC 60601-2-10 standards.
Technological Characteristics
Portable, battery-operated (7.2V Ni-MH) electrical stimulator. ABS plastic housing. Two-channel output with galvanic isolation. Biphasic waveform (spike with exponential decay). Frequency range 1-70 Hz. Pulse width 5 ms @ 1k ohms. Maximum output voltage 100V @ 10k ohms. Includes LCD display, timer, and intensity controls. Software-controlled via microprocessor.
Indications for Use
Indicated for symptomatic relief and management of chronic pain; adjunctive treatment of post-surgical and post-traumatic pain; temporary relief of pain associated with sore/aching muscles in shoulder, waist, back, neck, arms, and legs due to exercise or normal activities. Also indicated for temporary increase of local blood circulation in healthy muscles and muscle conditioning to improve/facilitate muscle performance.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with a stylized eagle above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 15, 2016
HB Medical, Inc. % Robyn Scopis CEO Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, California 92831
Re: K160255
Trade/Device Name: HBOTC Muscle And Nerve Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: August 15, 2016 Received: August 16, 2016
Dear Robyn Scopis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J.Hoffmann -A
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160255
Device Name HBOTC Muscle and Nerve Stimulator
#### Indications for Use (Describe)
Transcutaneous Electrical Nerve Stimulation:
- 1. For symptomatic relief and management of chronic pain
- 2. For adjunctive treatment in the management of post-surgical and post-traumatic pain
3. For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise and normal household and work activities
Powered Muscle Stimulator
- 1. To temporarily increase local blood circulation in healthy muscles
- 2. For muscle conditioning by stimulating healthy muscle in order to improve or facilitate muscle performance
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY K160255 510(k) Owner HB Medical 5702 Bolsa Ave. Huntington Beach, CA 92649 Phone: (714) 330-7333 Facsimile: (714) 500-4092 Contact person Robyn Scopis Regulatory Specialists, Inc. 1801 Edgecliff Drive Fullerton, CA 92831 Phone: 949.262.0411 Fax: 949.552.2821 Email: robyn@regulatoryspecialists.com Preparation Date September 13, 2016 Primary Product Code: NUH Common Name Transcutaneous electrical nerve stimulator for pain relief Trade Name HBOTC Muscle and Nerve Stimulator stimulator, nerve, transcutaneous, Classification Name over-the-counter Regulation 882.5890 Class Class II Panel Neurology Secondary Product Code: NGX Common Name Powered muscle stimulator HBOTC Muscle and Nerve Stimulator Trade Name Classification Name stimulator, muscle, powered, for muscle conditioning Regulation 890.5850 Class Class II Panel Physical Medicine Primary Predicate K131910 Secondary Predicate K112485 Third Predicate K103738 Fourth Predicate K102051
{4}------------------------------------------------
## Description
The HBOTC is an OTC electrical nerve and muscle stimulator indicated for symptomatic relief and management of chronic pain as well as muscle conditioning and increasing local blood flow. The HBOTC consists of a portable battery-operated electrical stimulation device with two channels, two sets of lead wires, six packages of self-adhesive electrodes, and a battery charger. Each channel has a pair of buttons for selecting the desired frequency and a dial to control the intensity of the signal. The stimulator is also supplied with an output jack for each channel, a charging jack, timer buttons, and an LCD display.
# Intended Use
The HBOTC is intended for use as a nerve and muscle stimulator for patients wanting symptomatic relief and management of chronic pain as well as muscle conditioning and increasing local blood flow.
### Indications for Use
Transcutaneous Electrical Nerve Stimulation:
1. For symptomatic relief and management of chronic pain
2. For adjunctive treatment in the management of post-surgical and posttraumatic pain
3. For temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise and normal household and work activities
#### Powered Muscle Stimulator
1. To temporarily increase local blood circulation in healthy muscles
2. For muscle conditioning by stimulating healthy muscle in order to improve or facilitate muscle performance
# Comparison of Technological Characteristics to Predicate Devices
The predicates and the HBOTC were compared in the following areas and found to have minor different technological characteristics. The following differences have been determined to not impact the safety and effectiveness of the HBOTC:
{5}------------------------------------------------
| 510(k) Number | Subject Device | Primary<br>Predicate | Secondary<br>Predicate | Third Predicate | Fourth Predicate | Regulated Current or<br>Regulated Voltage | Regulated<br>Voltage | Regulated<br>Voltage | Regulated<br>Voltage | Regulated<br>Voltage | Regulated<br>Voltage |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------|---------------------------------------------------|
| | HBOTC<br>K160255 | MPP<br>K131910 | H4<br>K112485 | H4<br>K103738 | Pain Buddy<br>K102051 | Software/firmware/mi<br>croprocessor | Yes | Yes | Yes | Yes | Yes |
| Power Source | Ni-MH<br>rechargeable<br>battery (7.2 V) | Ni-MH<br>rechargeable<br>battery (7.2 V) | Ni-MH<br>rechargeable<br>battery (7.2 V) | Ni-MH<br>rechargeable<br>battery (7.2 V) | Two 1.5-Volt<br>AAA disposable<br>cells (3 V) | Automatic Overload<br>Trip | No | No | No | No | No |
| Intended Use | Transcutaneous<br>Electrical Nerve<br>Stimulation:<br>1. For<br>symptomatic<br>relief and<br>management of<br>chronic pain<br>2. For adjunctive<br>treatment in the<br>management of<br>post-surgical and<br>post-traumatic<br>pain<br>3. For temporary<br>relief of pain<br>associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, neck, upper<br>extremities<br>(arm), and lower<br>extremities (leg)<br>due to strain<br>from exercise and<br>normal<br>household and | The MPP is to be<br>used for<br>temporary relief<br>of pain<br>associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, neck,<br>upper<br>extremities<br>(arm), and lower<br>extremities (leg)<br>due to strain<br>from exercise<br>and normal<br>household and<br>work activities.<br>The MPP is<br>intended for<br>muscle<br>conditioning by<br>stimulating<br>muscle in order<br>to improve or<br>facilitate muscle<br>performance | The H-Wave H4<br>is indicated for<br>the treatment of<br>chronic pain,<br>acute pain, post-<br>surgical pain,<br>and temporary<br>pain. | 1. Relaxation of<br>muscle spasms;<br>2. Prevention or<br>retardation of<br>disuse atrophy;<br>3. Increasing<br>local blood<br>circulation;<br>4. Muscle re-<br>education;<br>5. Immediate<br>post-surgical<br>stimulation of<br>calf muscles to<br>prevent venous<br>thrombosis;<br>6. Maintaining<br>or increasing<br>range of motion. | TENS<br>stimulation is<br>used for<br>symptomatic<br>relief and<br>management of<br>chronic pain<br>and/or as an<br>adjunctive<br>treatment in the<br>management of<br>post-surgical<br>and post-<br>traumatic pain. | Automatic No-Load<br>Trip | Yes | Yes | Yes | Yes | Yes |
| | work activities | | | | | Automatic Shut Off? | No | No | No | No | No |
| | Powered Muscle<br>Stimulator<br>1. To temporarily<br>increase local<br>blood circulation<br>in healthy<br>muscles<br>2. For muscle<br>conditioning by<br>stimulating<br>healthy muscle in<br>order to improve<br>or facilitate<br>muscle<br>performance | | | | | Patient Override<br>Control | Yes | Yes | Yes | Yes | Yes |
| Line Current Isolation | Yes (battery<br>operated) | Yes (battery<br>operated) | Yes (battery<br>operated) | Yes (battery<br>operated) | Yes (battery<br>operated) | Indicator Display | | | | | |
| Patient Leakage Current | | | | | | On/ Off Status | Yes | Yes | Yes | Yes | Yes |
| Normal Condition | 0 | 0 | 0 | 0 | 0 | Low Battery | Yes | Yes | Yes | Yes | Yes |
| Single fault condition | 0 | 0 | 0 | 0 | 0 | Voltage/ Current Level | Yes | Yes | Yes | Yes | Yes |
| Average DC current<br>through electrodes<br>when device is on but<br>no pulses are being<br>applied (μΑ) | 0 | 0 | 0 | 0 | 0 | Weight | 1.6 lb | 1.6 lb | 1.6 lb | 1.6 lb | 1.9 oz |
| Number of output<br>modes | 1 | 1 | 1 | 0<br>1 | 1 | Dimensions | 7" x 4.5" x 1.5" | 7" x 4.5" x 1.5" | 7" x 4.5" x 1.5" | 7" x 4.5" x 1.5" | 4.9" x 1.4" x 0.6" |
| Number of output<br>channels | 2 | 2 | 2 | 2 | 1 | Housing materials and<br>constructions | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing fastened<br>with screws | ABS plastic<br>housing<br>fastened with<br>screws |
| synchronous or<br>alternating | alternating | alternating | alternating | alternating | alternating | Waveform | biphasic | biphasic | biphasic | biphasic | biphasic |
| Method of Channel<br>Isolation | galvanic | galvanic | galvanic | galvanic | galvanic | Frequency | 1-70 Hz | 1-70 Hz | 1-70 Hz | 1-70 Hz | 8-80 Hz |
| Beat Frequency | N/A | N/A | N/A | N/A | N/A | | | | | | |
| Shape | spike with<br>exponential<br>decay | spike with<br>exponential<br>decay | spike with<br>exponential<br>decay | spike with<br>exponential<br>decay | Asymmetrical Bi-<br>Phasic<br>Rectangular | | | | | | |
| Maximum Output<br>Voltage @500Ω | 21V | 21V | 21V | 21V | 29V | | | | | | |
| Maximum Output<br>Voltage @2kΩ | 56v | 56v | 56v | 56v | 34V | | | | | | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Maximum Output<br>Voltage @10kΩ | 100V | 100V | 100V | 100V | 36V |
|----------------------------------|----------------------------------------------|----------------------------------------------|----------------------------------------------|----------------------------------------------|---------|
| Maximum Output<br>Current @500Ω | 42 mA | 42 mA | 42 mA | 42 mA | 60 mA |
| Maximum Output<br>Current @2kΩ | 28 mA | 28 mA | 28 mA | 28 mA | 17mA |
| Maximum Output<br>Current @10kΩ | 10 mA | 10 mA | 10 mA | 10 mA | 4mA |
| Pulse Width | 5 ms @ 1 k ohms | 5 ms @ 1 k ohms | 5 ms @ 1 k ohms | 5 ms @ 1 k ohms | 160 µs |
| Net Charge | o (equal positive<br>and negative<br>pulses) | o (equal positive<br>and negative<br>pulses) | o (equal positive<br>and negative<br>pulses) | o (equal positive<br>and negative<br>pulses) | unknown |
| Maximum Phase<br>Charge @500Ω | 16.8 µC | 16.8 µC | 16.8 µC | 16.8 µC | unknown |
| Maximum Current<br>Density @500Ω | 2 mA/cm² | 2 mA/cm² | 2 mA/cm² | 2 mA/cm² | unknown |
| Maximum Power<br>Density @500Ω | 0.042 W/cm² | 0.042 W/cm² | 0.042 W/cm² | 0.042 W/cm² | unknown |
| Burst Mode<br>Pulses per burst | N/A | N/A | N/A | N/A | unknown |
| Burst Mode<br>Bursts per second | N/A | N/A | N/A | N/A | unknown |
| ON Time (seconds) | N/A | N/A | N/A | N/A | unknown |
| OFF Time (seconds) | N/A | N/A | N/A | N/A | unknown |
{9}------------------------------------------------
The following non-clinical performance tests were conducted:
Software Validation
IEC 60601-1 (2012): Medical Electrical Equipment -Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-11 (2015): Medical electrical equipment – Part 1-11: General Requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
IEC 60601-2-10 (2012)/Amendment 1 2001, Medical electrical equipment - Part 2-10: Particular requirements for the safety of nerve and muscle stimulators.
Conclusions from non-clinical performance data
After performing non-clinical performance studies, the data shows that the HBOTC is substantially equivalent to the predicates as a powered muscle stimulator and transcutaneous electrical nerve stimulator for pain relief.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.