K160245 · Visco Vision, Inc. · LPL · Oct 24, 2016 · Ophthalmic
Device Facts
Record ID
K160245
Device Name
Visco (olifilcon A) soft contact lens
Applicant
Visco Vision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Oct 24, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
The VISCO (olifilcon A) Toric Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters. The VISCO (olifilcon A) Multifocal Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity. Eye care practitioners may prescribe the lens for either single-use daily disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent the lens may be disinfected using a chemical disinfection system only.
Device Story
VISCO (olifilcon A) soft contact lenses; silicone hydrogel material; toric or multifocal configurations; UV-blocking monomer; light blue tint (reactive Blue19). Supplied sterile in buffered saline with Sodium Hyaluronate and Sodium Alginate. Prescribed by eye care practitioners for daily wear (single-use or frequent/planned replacement). Corrects ametropia (myopia/hyperopia) and presbyopia. Patients wear lenses to improve visual acuity. No active components; passive optical device.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including physiochemical studies per ISO 18369 and biocompatibility/toxicology assessments.
Indicated for aphakic and non-aphakic persons with non-diseased eyes requiring correction of myopia, hyperopia, astigmatism, or presbyopia. Powers range -20.00 to +20.00D; astigmatic correction -0.75 to -2.25D; add powers +0.75 to +2.75D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 24, 2016
Visco Vision Inc. % Ms. Jennifer Ting Manager Jens Medical Consulting Ltd. 6F. No. 39. Jixian Rd. Luzhou Dist. New Taipei City, TW 247
Re: K160245
Trade/Device Name: VISCO (olifilcon A) Soft Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: September 12, 2016 Received: September 14, 2016
Dear Ms. Ting:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160245
Device Name VISCO (olifilcon A) Soft Contact Lens
### Indications for Use (Describe)
The VISCO (olifilcon A) Toric Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.
The VISCO (olifilcon A) Multifocal Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.
Eye care practitioners may prescribe the lens for either single-use daily disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent the lens may be disinfected using a chemical disinfection system only.
Type of Use (Select one or both, as applicable)
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
Preparation Date: Sep. 12, 2016
#### Establishment Information: 1.1
| Name | Visco Vision Inc. |
|-----------|-------------------------------------------------------------|
| Address | No. 1, Xingye St., Guishan Dist., Taoyuan City 33341 Taiwan |
| Contact | Evan Huang |
| Phone No. | 886-3-3596868 |
| Fax No. | 886-3-3490202 |
#### 1.2 Contact Person:
| Company | Jens Medical Consulting Ltd. |
|----------|------------------------------|
| Name | Jennifer TING |
| Phone No | 886-2-82823192 |
| Fax No | 886-2-82867686 |
| e-mail: | jen.medical@msa.hinet.net |
#### Device Identification: 1.3
| Proprietary Name | VISCO (olifilcon A) Soft Contact Lens |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Soft (hydrophilic) Contact Lenses |
| Classification Name | Lenses, Soft Contact, Daily Wear<br>(21 CFR 886.5925, Product Code LPL)<br>Lenses, Soft Contact, Daily Wear (Disposable),<br>(21 CFR 886.5925, Product Code MVN) |
| Classification | II |
#### Legally Marketed Equivalent Device: 1.4
| Predicate Device Name | Biofinity (comfilcon A), K052560, CooperVision Inc. | |
|-----------------------|-----------------------------------------------------|--|
| | VISCO (olifilcon A), K141348, Visco Vision Inc. | |
| Product Code | LPL, MVN | |
#### 1.5 Device Description
The VISCO (olifilcon A) toric/multifocal soft contact lenses are manufactured in toric or multifocal configurations with UV blocker. The lens material, olifilcon A, is a silicon hydrogel with water content 47%. The VISCO Soft Contact Lens is light blue tinted with "reactive Blue19" for handling visibility purpose. A benzotriazole UV absorbing
{4}------------------------------------------------
monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280-315nm and less than 50% in the UVA range of 316-380mm. The lens is supplied in a sterile state, packaged in a buffered saline solution containing Sodium Hyaluronate and Sodium Alginate
#### 1.6 Indication for Use:
The VISCO (olifilcon A) Toric Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.
The VISCO (olifilcon A) Multifocal Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.
Eye care practitioners may prescribe the lens for either single-use daily disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.
#### 1.7 Technological characteristics
The optical lens design specification:
| ● | Power Range | 20.00D to +20.00D |
|---|------------------------|-------------------------------------------|
| ● | Add Power (Multifocal) | Low (+0.75D ~ +1.25D) |
| | | Mid (+1.50D ~ +2.00D) |
| | | High (+2.25D ~ +2.75D) |
| ● | Cylinder Power (Toric) | -0.75D, -1.25D, -1.75D, -2.25D |
| | | Axis from 10° to 180° (in 10° increments) |
| ● | Diameter | 13.0 mm to 15.0 mm |
| ● | Center Thickness | 0.08mm @ -3.00D (Varies with Power) |
| ● | Base Curve | 8.0 mm to 9.2 mm |
The physical properties of the lenses are:
| • Refractive index: | 1.410 (hydrated) |
|---------------------|------------------|
|---------------------|------------------|
- Light transmittance: > 94%
{5}------------------------------------------------
- Water content: 47% by weight in normal saline 150 x 10-11 Oxygen permeability ●
- (cm²/sec)(ml O2/ml.mmHg) measured at 35°C
(intrinsic Dk-Colormetric method)
#### 1.8 Comparison table:
The characteristic comparison to predicate device is summarized in the following table.
| Similarities | | | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Item | Device | Predicate<br>(K052560) | predicate<br>(K141348) |
| Product Name | VISCO (olifilcon A) Toric<br>Soft Contact Lens<br>VISCO (olifilcon A)<br>Multifocal Soft Contact<br>Lens | BIOFINITY Toric<br>(comfilcon A) Soft Contact<br>Lens<br>BIOFINITY Multifocal<br>(comfilcon A) Soft Contact<br>Lens | Visco (olifilcon A) soft<br>contact lens |
| Manufacturer | Visco Vision Inc. | CooperVision Inc. | Visco Vision Inc. |
| Intended Use | Myopia, Hyperopia,<br>Prysbyopia | The same | Myopia, Hyperopia |
| Replacement<br>Schedule | Daily wear (monthly) or<br>Daily Disposable (Single<br>use) | daily wear (monthly) or<br>extended wear | Daily wear (monthly)<br>or Daily Disposable<br>(Single use) |
| USAN Name | olifilcon A | comfilcon A | olifilcon A |
| Material | Silicone Hydrogel | Silicone Hydrogel | Silicone Hydrogel |
| Lens Design | toric or multifocal | Spheric, aspheric, toric or<br>multifocal | spheric |
| Classification | Class II, | The same | The same |
| Type | Group I (low water,<br>nonionic) | The same | The same |
| Water Content | 47 % | 48 % | 47% |
| Oxygen<br>Permeability<br>(DK, 35°C ) | 150<br>(Fatt method) | 128<br>(Fatt method) | 150<br>(Fatt method) |
| Base Curve<br>Range | 8.0~9.2 | 8.0~9.5 | 8.0~9.2 |
| Diameter (mm) | 13.0~15.0 | 13.5~15.0 | 13.0~15.0 |
| Center Thickness | Varies with power (0.08<br>mm at -3.00D) | 0.05 mm – 0.50 mm | Varies with power<br>(0.08 mm at -3.00D) |
| Powers | -20.00D to +20.00D<br>in 0.25 steps | The same | -20.00D to +20.00D<br>in 0.25 steps |
| Add Power<br>(Multifocal) | Low (+0.75D ~ +1.25D)<br>Mid (+1.50D ~ +2.00D)<br>High (+2.25D ~ +2.75D) | +0.50D to +3.00D | NA |
| Cylinder Power<br>(Toric) | -0.75D, -1.25D, -1.75D,<br>-2.25D<br>Axis from 10° to 180° (in<br>10° increments) | -0.25D -5.00D<br>Axis from 0° to 180° | NA |
| Refractive Index | 1.410 | 1.40 | 1.410 |
| Light<br>Transmittance | 94% | >97% | 94% |
| Method of<br>Manufacture | Cast-Molded | The same | The same |
| Surface Treatment | No | No | No |
| Sterilization | steam | The same | The same |
| Packaging | Blister pack | The same | The same |
| Blue handling tint | Yes, reactive Blue19 | Yes, Phthalocyanine Blue | Yes, reactive Blue19 |
| Indication for Use | <b>Toric</b><br>The VISCO (olifilcon A)<br>Toric Soft Contact lenses<br>are indicated as daily wear<br>for the correction of<br>ametropia (myopia and<br>hyperopia) with<br>astigmatism in aphakic and<br>non-aphakic persons with<br>non-diseased eyes and<br>whose powers are from<br>-20.00 to +20.00 diopters<br>and astigmatic corrections<br>are from -0.75 to -2.25<br>diopters. | <b>Toric</b><br>Biofinity (comfilcon A)<br>toric soft contact lenses are<br>indicated for the correction<br>of ametropia (myopia or<br>hyperopia with<br>astigmatism) in aphakic<br>and non-aphakic persons<br>with non-diseased eyes in<br>powers from -20.00 to<br>+20.00 diopters and<br>astigmatic corrections<br>from -0.25 to -5.00<br>diopters. | NA |
| Multifocal | Multifocal | | |
| The VISCO (olifilcon A) | Biofinity (comfilcon A) | | |
| Multifocal Soft Contact | multifocal lenses are | | |
| lenses are indicated as | indicated for the correction | | |
| daily wear for the | of refractive ametropia | | |
| correction of ametropia | (myopia and hyperopia) | | |
| (myopia and hyperopia) | and emmetropia with | | |
| and emmetropia with | presbyopia in aphakic and | | |
| presbyopia in aphakic and | non-aphakic persons with | | |
| non-aphakic persons with | non-diseased eyes in | | |
| non-diseased eyes and | powers from -20.00 to | | |
| whose powers are from | +20.00 diopters and with | | |
| -20.00 to +20.00 diopters | add powers from +0.50 to | | |
| with add powers from | +3.00 diopters. The lenses | | |
| +0.75 to +2.75 diopters. | may be worn by persons | | |
| The lenses may be worn | who exhibit astigmatism of | | |
| by persons who exhibit | 2.00 diopters or less that | | |
| astigmatism of 1.00 | does not interfere with | | |
| diopters or less where the | visual acuity. | | |
| astigmatism does not | | | |
| interfere with visual acuity. | | | |
| Eye care practitioners may | Eye care practitioners may | Eye care practitioners<br>may prescribe the lens<br>for either single-use<br>disposable wear, or for<br>frequent/planned<br>replacement wear,<br>with cleaning,<br>disinfection, and<br>scheduled<br>replacement. When<br>prescribed for frequent<br>replacement, the lens<br>may be disinfected<br>using a chemical | |
| prescribe the lens for either | prescribe the lens for | | |
| single-use daily disposable | frequent replacement wear, | | |
| wear, or for | with cleaning, disinfecting | | |
| frequent/planned | and scheduled | | |
| replacement wear, with | replacements. | | |
| cleaning, disinfection, and | | | |
| scheduled replacement. | | | |
| When prescribed for | | | |
| frequent replacement, the | | | |
| lens may be disinfected | | | |
| using a chemical | | | |
| disinfection system only. | | disinfection system only. | |
| | | | |
| | | disinfection system | |
| | | only.The VISCO Soft | |
| | | (Hydrophilic) Contact | |
| | | Lenses help protect | |
| | | against transmission of | |
| | | harmful UV radiation | |
| | | to the cornea and into | |
| | | the eye. | |
| NA | Spherical and Aspherical | Spherical | |
| | Biofinity (comfilcon A) | The VISCO Soft | |
| | sphere and Asphere soft | (Hydrophilic) Contact | |
| | contact lenses are indicated | Lenses is indicated for | |
| | for the correction of | daily wear for the | |
| | ametropia (myopia and | correction of refractive | |
| | hyperopia) in aphakic and | ametropia (myopia and | |
| | non-aphakic persons with | hyperopia) in phakic | |
| | non-diseased eyes in | or aphakic persons | |
| | powers from -20.00 to | with non-diseased eyes | |
| | +20.00 diopters. The | who exhibit refractive…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.