← Product Code [KRD](/productcode/KRD) · K160219

# Hilal Embolization MicroCoils (K160219)

_Cook Incorporated · KRD · Aug 3, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160219

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** [KRD](/productcode/KRD.md)
- **Decision Date:** Aug 3, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Hilal Embolization MicroCoils™ are intended for arterial and venous embolization in the peripheral vasculature.

## Device Story

Hilal Embolization MicroCoils are platinum wire coils with nylon fibers; designed for arterial and venous embolization in peripheral vasculature. Device delivered via microcatheters (0.018-0.025 inch end hole diameter) using a loading cartridge and stylet. Available in straight, single-curl, and multiple-curl configurations; embolus lengths 0.5-6.0 cm; diameters 2-10 mm. Used by physicians in clinical settings to achieve vessel occlusion. Device functions as a mechanical embolization agent; provides physical obstruction to blood flow. MR Conditional status confirmed via testing.

## Clinical Evidence

Bench testing included wire tensile testing and coil uniaxial tensile strength testing. Animal testing (acute porcine model) evaluated straight and single-curl coils, demonstrating no arterial damage or coil migration. Performance, MRI, and biocompatibility testing leveraged from Nester® and Tornado® Embolization Coils (K153778).

## Technological Characteristics

Materials: platinum wire, nylon fibers. Configuration: straight, single-curl, multiple-curl. Delivery: pushable via microcatheter (0.018-0.025 inch). Dimensions: 0.5-6.0 cm length, 2-10 mm diameter. MR Conditional. Sterilization: not specified.

## Regulatory Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

## Predicate Devices

- Cook Retracta® Detachable Embolization Coil ([K123712](/device/K123712.md)/[K151676](/device/K151676.md))

## Reference Devices

- Nester® Embolization Coils ([K153778](/device/K153778.md))
- Tornado® Embolization Coils ([K153778](/device/K153778.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles forming the staff and a flowing ribbon representing the snakes.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 3, 2016

Cook Incorporated Mr. David Lehr Regulatory Affairs Specialist 750 Daniels Way, P.O. Box 489 Bloomington, Indiana 47402-0489

Re: K160219

Trade/Device Name: Hilal Embolization MicroCoils™ Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: June 23, 2016 Received: June 24, 2016

Dear Mr. David Lehr:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours

sincerely yours,

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160219

Device Name Hilal Embolization MicroCoils™

Indications for Use (Describe)

The Hilal Embolization MicroCoils™ are intended for arterial and venous embolization in the peripheral vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Cook Medical. The word "COOK" is in large, white, sans-serif letters on a red background. To the bottom of the word "COOK" is a red shape with the word "MEDICAL" in white, sans-serif letters. The registered trademark symbol is to the upper right of the word "COOK".

#### 2.0 510(k) SUMMARY

As required by 21 CFR §807.92 Date Prepared: August 1, 2016

#### I. SUBMITTER

| Submission:           | Traditional 510(k) Premarket Notification                     |
|-----------------------|---------------------------------------------------------------|
| Applicant:            | Cook Incorporated                                             |
| Contact:              | David Lehr, RAC                                               |
| Applicant Address:    | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 ext. 102309                                    |
| Contact Fax Number:   | (812) 332-0281                                                |

### II. DEVICE

| Trade Name:          | Hilal Embolization MicroCoilsTM              |
|----------------------|----------------------------------------------|
| Common Name:         | Vascular Embolization Device                 |
| Classification Name: | Device, Vascular, For Promoting Embolization |
| Regulation/Class:    | 21 CFR §870.3300/Class II                    |
| Product Code         | KRD                                          |

## III. PREDICATE DEVICE

The device subject of this submission is considered substantially equivalent to the predicate device, the Cook Retracta® Detachable Embolization Coil (K123712/K151676), which has never been subject to a design-related recall.

## IV. DEVICE DESCRIPTION

Hilal Embolization MicroCoils™ are manufactured from coiled platinum wire with equidistantly spaced nylon fibers and are available in straight, single-curl, and multiplecurl configurations. The Hilal coils are designed to be delivered by microcatheters with a minimum end hole diameter of 0.018 inch. The extended embolus lengths of the finished device range from 0.5 to 6.0 centimeters. The coiled embolus diameters range from 2 to 10 millimeters. The Hilal coils are loaded in a straight configuration into a loading cartridge. A loading stylet is also provided for loading this coil into the delivery catheter.

# V. INDICATIONS FOR USE

The Hilal Embolization MicroCoils™ are intended for arterial and venous embolization in the peripheral vasculature.

{4}------------------------------------------------

# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The subject device, the Hilal Embolization MicroCoils™, and the predicate device, the Retracta® Detachable Embolization Coil (K123712/151676), are substantially equivalent in that these devices have identical intended uses and similar technological characteristics. The predicate device is a combined embolization coil and delivery wire. Its embolization coil (made of platinum wire and nylon fibers) is delivered by a detachment mechanism. The subject device is a pushable embolization coil (also made of platinum wire and nylon fibers) delivered with a wire guide through a delivery catheter. The predicate coils (intended for delivery through catheters with an end hole size of 0.035 inch) have a helical shape and are available in lengths of 7 cm or 14 cm and diameters ranging from 4 to 20 mm. The subject Hilal coils (intended for delivery through catheters with end hole sizes of 0.018 through 0.025 inch) have straight, singlecurl, and multiple-curl shapes and are available in lengths ranging from 0.5 to 6 cm and in diameters ranging from 2 to 10 mm. Additionally, the subject device, like the predicate device, is labeled as MR Conditional. However, there are differences in the MR scanning conditions to reflect the testing performed on the subject device. Based on the comparison of the design, intended use, materials, fundamental technology, and principle of operation, the subject device is considered to be substantially equivalent to the currently marketed predicate device.

## VII. PERFORMANCE DATA

The subject device underwent the applicable testing listed below to ensure reliable design and performance under the testing parameters.

- . Wire Tensile Testing - Testing shows the raw wire has a peak load value greater than or equal to the minimum tensile strength requirements. The predetermined acceptance criteria were met.
- Coil Tensile Testing - Testing characterized the embolization coils' uniaxial tensile strength.
- . Animal Testing – An acute performance study was performed in a porcine model on straight and single curl Hilal coils, showing no evidence of arterial damage or coil migration. The predetermined acceptance criteria were met.

#### K160219

{5}------------------------------------------------

Cook Incorporated - Traditional 510(k) Hilal Embolization MicroCoilsTM August 1, 2016

Other testing was leveraged from the 510(k) submission for Nester® and Tornado® Embolization Coils (K153778). Performance (bench, MRI, and animal) and biocompatibility testing was conducted in accordance with applicable FDA guidance documents to confirm the reliable performance of critical device characteristics.

# VIII. CONCLUSIONS

The data included in this submission indicate that the subject device does not raise new questions of safety or effectiveness compared to the predicate device. This supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K160219](https://fda-staging.innolitics.com/device/K160219)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
