Gold Anchor

K160209 · Naslund Medical AB · IYE · Jun 21, 2016 · Radiology

Device Facts

Record IDK160209
Device NameGold Anchor
ApplicantNaslund Medical AB
Product CodeIYE · Radiology
Decision DateJun 21, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2

Indications for Use

The Gold Anchor marker is indicated for use to radiographically mark soft tissue for future therapeutic procedures.

Device Story

Gold Anchor is a passive, single-use fiducial marker implanted into soft tissue to delineate tumors or lesions for radiographic visualization. The marker consists of a gold-iron alloy wire with cutouts, pre-loaded into 20G, 22G, or 25G needles. It is deployed into the target site via a stylet. The iron content (0.5% mass) enhances visibility on MRI. The device is used by clinicians to provide stable, identifiable landmarks for subsequent therapeutic procedures, such as radiation therapy. It is provided sterile via E-Beam radiation.

Clinical Evidence

Bench testing only. Performance data included MR safety testing per ASTM F2052-15 (magnetically induced displacement), ASTM F2182-11a (RF-induced heating), and ASTM F2119-07(2013) (MR image artifacts). Biocompatibility was evaluated via chemical characterization risk assessment in accordance with ISO 10993-17 and ISO 10993-18.

Technological Characteristics

Passive fiducial marker; gold-iron alloy (0.5% iron); wire-with-cutouts design; pre-loaded in 20G, 22G, or 25G needles; lubricated with Nusil Med-360 silicone oil; sterilized via E-Beam radiation. MR safety tested per ASTM F2052, F2182, and F2119.

Indications for Use

Indicated for radiographic marking of soft tissue to delineate tumors, lesions, or sites of interest for future therapeutic procedures in patients requiring such localization.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 21, 2016 NASLUND MEDICAL Ab Tomas Naslund Head of Supply Chain Vassvagen 21 Huddinge, 14139 SE Re: K160209 Trade/Device Name: Gold Anchor Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: January 25, 2016 Received: January 28, 2016 Dear Tomas Naslund: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Sincerely yours, Robert Ochs, Ph.D For Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K160209 Device Name Gold Anchor Indications for Use (Describe) The Gold Anchor marker is indicated for use to radiographically mark soft tissue for future therapeutic procedures. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K160209 # 5. 510(k) Summary Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Traditional 510(k) section 510(k) Summary of safety and effectiveness Information as required by section 807.92(c) #### I. SUBMITTER OF 510(k): | Company name: | Naslund Medical AB | |----------------|--------------------| | Address: | Vassvagen 21 | | | 141 39 Huddinge | | | Sweden | | Phone: | +46 732 620 717 | | Fax: | +46 850 900 381 | | Correspondent: | Tomas Naslund | #### II. DEVICE | Name of Device: | Gold Anchor™ | |-----------------------|-----------------------------| | Common or Usual Name: | Fiducial marker | | Classification Name: | Accelerator, Linear, Medica | | Regulatory Class: | II | | Product Code: | IYE | #### III. PREDICATE DEVICE The predicate devices are: - 1. Gold Anchor (K091645) - 2. VisiCoil Marker (K070305) #### IV. DEVICE DESCRIPTION The Gold Anchor™ Marker is a fiducial gold marker intended to be implanted within the body, either temporarily or permanently, to create identifying marks that can be seen on radiographic film or digital images. The marker is formed as a wire with cutouts and used to locate and delineate a tumor, lesion, or other site of interest. The marker comes pre-loaded in 25G, 22G and 20G needles delivered sterile and ready for use. Sterilization is achieved by E-Beam Radiation. This is a single-use device. The device is a passive implant. #### V. INDICATIONS FOR USE The Indications for Use statement for the subject device is identical to that of its predicate device Gold Anchor (K091645): - The Gold Anchor marker is indicated for use to radiographically mark soft tissue for future ● therapeutic procedures. {4}------------------------------------------------ ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE At a high level, the subject device and the predicate device Gold Anchor (K091645) are based on the following same technological characteristics: - Device delivered sterilized using Electron Beam Radiation. � - Device delivered pre-loaded in 25G or 22G needles. . - Use of a needle to reach the target tissue. � - Use of a stylet to deploy the fiducial gold marker to mark the soft tissue. � - Use of a material that lubricates the inside of the needle. . - . Use of a marker with a unique and patented design (wire with cutouts). The following technological differences exist between the subject device and the predicate device Gold Anchor (K091645): - The subject device is also available pre-loaded in a 20G needle. . - . The 25G needle of the predicate device was 120 mm long. The 25G needle of the subject device is 150 mm long. - � The material used to lubricate the inside of the subject device is Nusil Med-360 unrestricted silicone oil. - . The fiducial marker in the subject device contains a small amount of pure iron (0.5% of mass) mixed with the pure gold as an alloy to enhance the visibility of the marker on MRI. - . The predicate device only came with a Ø 0.28 mm marker. The subject device comes with either a Ø 0.40 mm marker or a Ø 0.28 mm marker. Comparison to the predicate device VisiCoil Marker (K070305): - The subject device is available pre-loaded in a 20G needle. The predicate device is avail-. able pre-loaded in thicker 17G. 18G, or 19G needles. ## VII. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination: Tests were conducted in accordance with FDA's guidance document, "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment Guidance for Industry and Food and Drug Administration Staff" and the following ASTM standards: - ASTM F2052-15 Standard Test Method for Measurement of Magnetically Induced Dis-. placement Force on Medical Devices in the Magnetic Resonance Environment. - ASTM F2182-11a Standard Test Method for Measurement of Radio Frequency Induced . Heating On or Near Passive Implants During Magnetic Resonance Imaging - . ASTM F2119-07(2013) Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants. Because of the minor amount of translational attraction (1-degree deflection angle) and no torque (qualitatively determined) at 3-Tesla, it was deemed unnecessary to conduct a quantitative evaluation of torque according to ASTM F2213-2006 (2011), Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment. A biocompatibility evaluation has been performed based on a chemical characterization risk assessment in accordance with ISO 10993-17 and ISO 10993-18.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...