K160204 · Bontech, Inc. · MQB · Jun 9, 2016 · Radiology
Device Facts
Record ID
K160204
Device Name
BSD4343 Digital Flat Panel X-ray Detector
Applicant
Bontech, Inc.
Product Code
MQB · Radiology
Decision Date
Jun 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The BSD4343 detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general-purpose diagnostic procedures. It is not to be used for mammography.
Device Story
BSD4343 is a digital X-ray flat panel detector; intercepts X-ray photons using a Gadox:Tb scintillator; converts photons to visible light; illuminates an amorphous silicon (a-Si) TFT array to generate electrical signals; signals converted to digital values; images displayed on a monitor. Device integrates with an operating PC and X-ray generator; used in clinical settings for general radiography. Enables efficient diagnosis, information management, and network sharing of images. Benefits include digital conversion of X-ray images, replacing traditional film-based systems, and facilitating diagnostic workflows for healthcare professionals.
Clinical Evidence
Bench testing only. No clinical data was required. Performance was validated through non-clinical laboratory testing, including DQE, MTF, and NPS measurements, demonstrating resolution performance equivalent or superior to the predicate. Electrical safety and EMC testing conducted per IEC 60601-1 and IEC 60601-1-2.
Indicated for digital imaging in general radiographic systems for human anatomy; intended to replace film/screen-based systems for general-purpose diagnostic procedures. Not for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, with flowing lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2016
Bontech Inc. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K160204
Trade/Device Name: BSD4343 Digital Flat Panel X-ray Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: April 26, 2016 Received: April 26, 2016
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160204
Device Name BSD4343 Digital Flat Panel X-ray Detector
#### Indications for Use (Describe)
The BSD4343 detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general-purpose diagnostic procedures. It is not to be used for mammography.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 07
# 510(k) SUMMARY
{4}------------------------------------------------
## 1. Traditional 510(k) SUMMARY
This 510(k) summary is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.
Date 510K summary prepared : January 18, 2016
Submitter's Name, address, telephone number, a contact person:
| Submitter's Name : | BONTECH Inc. | |
|-------------------------|------------------------------------------------------------------------------------------------------------|--|
| Submitter's Address: | Youngtong-gu Youngtong-dong 980-3, Digital Empire D #1201,<br>Suwon city, Gyeonggi-do, South Korea 443-702 | |
| | Tel:+ 82-31-303-5254 / Fax: +82-31-303-5255 | |
| Contact person: | Mr. Kwang S Choi / RA Manager (cks@inhwagroup.com) | |
| Official Correspondent: | Dave Kim, MBA | |
| Address: | 8310 Buffalo Speedway, Houston, TX 77025 | |
| Telephone: | +713-467-2607 | |
| Fax: | +713-583-8988 | |
| Email: | davekim@mtech-inc.net | |
Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:
| Trade/proprietary name: | BSD4343 Digital Flat Panel X-ray Detector |
|-------------------------|-----------------------------------------------------|
| Device: | Solid State X-ay Imager (Flat Panel/Digital Imager) |
| Classification Name: | Stationary X-ray System |
| Regulation Number: | 21 CFR 892.1680 |
| Regulatory Class: | II |
| Product Code: | MQB |
| Predicate Device | |
| Manufacturer | Samsung Mobile Display Co., Ltd |
| Device | LLX240AB01 |
| 510(k) Number | K102587 |
| Decision Date | DEC 1, 2010 |
#### 2. Device Description
BSD4343 is a digital X-ray flat panel detector which intercepts x-ray photons and the scintillator (Gadox:Tb type) emits visible spectrum photons that illuminate an array of photo (a-SI)-detector that creates electrical signals. After the electrical signals are generated, it is converted to digital values, and the images will be displayed on the monitor. This device should be integrated with an operating PC and an X-Ray generator. It can digitalize x-ray images and transfer them for radiography diagnostics. Advanced digital image processing allows considerably efficient diagnosis, all kinds of information management, and sharing of image information on network.
{5}------------------------------------------------
## 3. Indications for Use
The BSD4343 detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic system in all general-purpose diagnostic procedures. It is not to be used for mammography.
#### 4. Summary of Design Control Risk management
The BSD4343 detector has been developed to meet the critical functional requirements and international safety standards. The risks and the hazardous impact of the device design were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the device design and production phase were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design and production were successfully mitigated and accepted.
## 5. Summary of the technological characteristics of the device compared to the predicate device:
The BSD4343 detector described in this 510(k) has similar indications for use and technical characteristics as the predicate device, LLX240AB01 digital flat panel X-ray detector manufactured by Samsung Mobile Display Co., Ltd.
## 6. Substantial Equivalence
The BSD4343 detector and components conform to the FDA recognized standards as like the predicate device. Based on the recognized standard conformity evidences related to electro- mechanical-, software-, clinical-, and risk management, it is confirmed that BSD4343 is substantially equivalent to the predicate device.
| Characteristic | Proposed<br>BONTECH<br>BSD4343 | Proposed<br>Samsung Mobile Display<br>LLX240AB01 | Remark |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 510(k) number | - | K102587 | |
| Intended Use | The BSD4343 detector is<br>indicated for digital imaging<br>solution designed for general<br>radiographic system for human<br>anatomy. It is intended to replace<br>film or screen based radiographic<br>system in all general-purpose<br>diagnostic procedures. It is not to<br>be used for mammography. | LLX240AB01 Digital Flat Panel X-Ray<br>Detector is indicated for digital<br>imaging solution designed for<br>providing general radiographic<br>diagnosis of human anatomy<br>targeting both adult and children. It<br>is intended to replace film based<br>radiographic diagnostic systems and<br>provide a case diagnosis and<br>treatment planning for physicians<br>and other health care professionals.<br>Not to be used for mammography. | Same |
| Detector Type | Amorphous Silicon, TFT | Amorphous Silicon, TFT | Same |
{6}------------------------------------------------
| Scintillator | Gadolinium Oxysulfide | Gadolinium Oxysulfide | Same |
|----------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------|
| Imaging Area | 17 x 17 inches | 17 x 17 inches | Same |
| Pixel matrix | 3072 x 3072 (9.4 million) | 3072 x 3072 (9.4 million) | Same |
| Pixel pitch | 140 μm | 143 μm | Similarity |
| Resolution | 3.5 lp/mm | 3.5 lp/mm | Same |
| A/D conversion | 16 bit | 14 bit | Similarity |
| Grayscale | 16384 (14bit) | 16384 (14bit) | Same |
| Data output | RAW<br>*The RAW files are convertible<br>into DICOM 3.0 by console S/W | RAW<br>*The RAW files are convertible into<br>DICOM 3.0 by console S/W | Same |
| Dimensions | 460 x 460 x 15 mm | 500 x 496.6 x 45 mm | Similarity |
| Application | General Radiology system<br>Available with upright stand,<br>table, universal stand | General Radiology system<br>Available with upright stand, table,<br>universal stand | Same |
| picture | Image: BONTECH | Image: detector | Similarity |
When compared to the predicate device (K102587), the BSD4343 presented in this submission has the same:
- Intended Use
- Technological characteristics
- · Operating principle
- · Design features
- Performance
- Communication Method
There are no significant difference between BSD4343 and the predicate device that would adversely affect the use of the product. It is substantially equivalent to these devices in design, function, materials, operational principles and intended use.
## 7. Performance Testing/Data
To minimize electrical and mechanical hazards, BONTECH adheres to recognized and established industry practice, and all equipment complies with the relevant FDA and international standards. For example, electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) was performed and EMC testing was conducted in accordance with standard IEC 60601-1-2: 2007, EN 60601-1-2:2007
{7}------------------------------------------------
Through verification and validation activities, engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or identify new risks.
The instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
#### 8. Description of non-clinical tests.
#### - Non-clinical study
The non-clinical performance testing constrains that the main physical values for comparison of X-rav devices like DQE. MTF NPS are basically equivalent to the predicate device LLX240AB01 (K102587). The results show that BSD4343 offer similar or better resolution performance than LLX240AB01 at 0 ~ 3.5lp/mm spatial frequencies. Moreover, the ability of BSD4343 to utilize the input image signal are more efficient than LLX240AB01 at same patient exposure as shown in the detective quantum efficiency graph.
#### -Electrical safety and EMC
BSD4343 has been tested for electrical safety standard IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 *2007) + AM1 (2012) and electromagnetic compatibility IEC 60601-1-2: 2007
The software validation and verification testing was also performed. The results of nonclinical testing indicate that the BSD4343 detector is as safe and effective as the predicate device.
Compliance evidences were submitted for the following standards:
- IEC 60601-1: Test Report issued by 3rd party testing lab A
- A IEC 60601-1-2: Test Report issued by 3rd party testing lab
- A ISO 14971: Risk management file
#### 9. Description of clinical tests.
Images reviewed were not necessary to establish substantial equivalence based on the modifications to the device as evidenced in the laboratory performance data to show that the subject device operates as the indicated.
#### 10. Conclusion as to Substantial Equivalence
The BSD4343 detector is substantially equivalent to the predicate device LLX240AB01 (K102587). Both subject and predicate devices are same or very similar in the intended use, the design principle, the performance and the applicable standards. Some characteristics, for example, their appearance, pixel pitch and weight are different. However the compliance reports, performance demonstrations that these differences do not raise any new questions of safety and effectiveness. Therefore, BONTECH Inc. concludes the BSD4343 digital flat panel detector is substantially equivalent with the predicate device LLX240AB01 (K102587) of Samsung Mobile Display Co., Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.