Align Anterior Ankle Fusion Plate

K160191 · Extremity Medical, LLC · HRS · Mar 25, 2016 · Orthopedic

Device Facts

Record IDK160191
Device NameAlign Anterior Ankle Fusion Plate
ApplicantExtremity Medical, LLC
Product CodeHRS · Orthopedic
Decision DateMar 25, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Align Anterior Ankle Fusion Plate is intended to facilitate arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints.

Device Story

Bone fixation device for ankle arthrodesis; consists of anatomical anterior tibia plate, locking/non-locking screws, and surgical instruments. Plate features central hole for angled 'home run' screw to provide axial compression across tibiotalocalcaneal (TTC) joints; 2-tab configuration includes anterior window for autologous bone grafting and fluoroscopic visualization of fusion. Used by orthopedic surgeons in clinical/OR settings to stabilize bone during fusion process; facilitates patient recovery by maintaining joint alignment and compression.

Clinical Evidence

No clinical testing was performed. Substantial equivalence is supported by bench testing, including pull-out and torque testing of screws, and engineering analysis.

Technological Characteristics

Metallic bone fixation appliance; includes anatomical anterior tibia plate and locking/non-locking screws. Features include central hole for axial compression screw and anterior window for bone grafting/fluoroscopic visualization. Mechanical properties verified via bench testing.

Indications for Use

Indicated for patients requiring arthrodesis of the ankle, specifically tibiotalocalcaneal and tibiotalar joints.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top of the seal in a circular fashion. The seal is black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 25, 2016 Extremity Medical, LLC Mr. Brian Smekal VP, Regulatory Affairs and Quality Assurance 300 Interpace Parkway, Suite 410 Parsippany, New Jersey 07054 Re: K160191 Trade/Device Name: Align Anterior Ankle Fusion Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 26, 2016 Received: January 27, 2016 Dear Mr. Smekal: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K160191 Device Name Align Anterior Ankle Fusion Plate Indications for Use (Describe) The Align Anterior Ankle Fusion Plate is intended to facilitate arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness: Align Anterior Ankle Fusion Plate | Submitter | Extremity Medical, LLC.<br>300 Interpace Parkway, Suite 410<br>Parsippany, NJ 07054 | |---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact<br>Person | Brian Smekal, MS, RAC<br>VP, Regulatory Affairs and Quality Assurance<br>Phone: (973) 588-8988<br>Email: bsmekal@extremitymedical.com | | Date Prepared | March 21, 2016 | | Trade Name | Align Anterior Ankle Fusion Plate | | Classification<br>Name and<br>Number | 21 CFR 888.3030 - Single/multiple component metallic bone fixation appliances and accessories; 21<br>CFR 888.3040 - Smooth or threaded metallic bone fixation fastener | | Product Code | HRS (plate, fixation, bone); HWC (screw, fixation, bone) | | Predicate<br>Devices | K081934 - Extremity Medical Compression Screw<br>K133691 – Extremilock Ankle Plating System<br>K132591 - Maxlock Extreme System<br>K121425 - Ortholoc 3Di Ankle Fusion Plating System and Ortholoc Bone Screws | | Device<br>Description | The Align Anterior Ankle Fusion Plate is a bone fixation device consisting of an anatomical, anterior<br>tibia plate and various locking and non-locking screws and a set of instruments used for implant site<br>preparation and delivery. The plate is available in three configurations, a slim plate and a 2-tab plate<br>(available in two widths). A central hole in the plate allows for angled placement of a "home run"<br>screw to obtain axial compression across the tibiotalarcalcaneal) (TTC) joints.<br>The 2-tab plate is designed with an anterior "window" to allow autologous grafting of the TT and<br>visualization of fusion under fluoroscopy. | | Indications for<br>use | The Align Anterior Ankle Fusion Plate is intended to facilitate arthrodesis of the ankle including<br>tibiotalocalcaneal and tibiotalar joints. | | Statement of<br>Technological<br>Comparison | The Align Anterior Ankle Fusion Plate and predicate devices are equivalent in terms of indications<br>for use, design, and material mechanical properties. | | Non-clinical<br>Testing | Bench testing including pull-out, torque on the screws and engineering analysis was used to<br>demonstrate that the mechanical strength of the subject device was substantially equivalent to the<br>predicate device. | | Clinical<br>Testing | No clinical testing was performed. | | Conclusion | The Align Anterior Ankle Fusion Plate is substantially equivalent to its predicate device. This<br>conclusion is based upon indications for use, principles of operation, design, and mechanical test<br>data. |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...