Mineral Oil is used as an overlay for culture of embryos, oocytes, and sperm in assisted reproduction technology (ART) and micromanipulation procedures.
Device Story
Mineral Oil serves as an overlay for culture media in ART and micromanipulation procedures; prevents evaporation of small media volumes; maintains stable osmolality and pH for embryos, oocytes, and sperm. Device is a high-purity paraffin oil available in light and heavy density types; colorless, tasteless, clear, non-fluorescent. Used in clinical laboratory settings by embryologists or laboratory technicians. Product is filter-sterilized and supplied in light-resistant glass bottles. Does not directly contact gametes or embryos. Benefits include maintenance of optimal culture environment conditions, supporting embryo development.
Clinical Evidence
Bench testing only. Performance data includes: sterility (USP <71>), endotoxin (< 0.25 EU/mL), Mouse Embryo Assay (MEA >80% development at 96h), peroxide value (< 0.1 meq/kg), and readily carbonizable substance testing (USP <271>).
Technological Characteristics
High-purity paraffin oil; two types: Light (density 0.8200-0.8400 g/mL, viscosity 8.850-11.70 mm²/s) and Heavy (density 0.8500-0.8700 g/mL, viscosity 41.90-44.10 mm²/s). Filter sterilized. Packaged in light-resistant glass bottles. No electronic or software components.
Indications for Use
Indicated for use as an overlay for culture of embryos, oocytes, and sperm in assisted reproduction technology (ART) and micromanipulation procedures for patients undergoing fertility treatment.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of a bird-like figure, possibly representing an eagle or other national symbol. The image is in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 21, 2016
KITAZATO BioPharma Co., Ltd. % Diane Sudduth RA Consultant Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K160142 Trade/Device Name: Mineral Oil Regulation Number: 21 CFR§ 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: II Product Code: MOL Dated: January 20, 2016 Received: January 21, 2016
Dear Diane Sudduth,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160142
Device Name Mineral Oil
Indications for Use (Describe)
Mineral Oil is used as an overlay for culture of embryos, oocytes, and sperm in assisted reproduction technology (ART) and micromanipulation procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## MINERAL OIL
## K160142
#### 1. Submission Sponsor
KITAZATO Biopharma Co., Ltd. 81 Nakajima Fuji Shizuoka, 416-0907 Japan Phone number: +81-545-66-2202 Contact: Kyoko Izumi Title: Quality Assurance Manager
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (561) 305.5075 Office Phone: (512) 327.9997 Contact: Diana Sudduth, Senior Consultant, RA Email: project.management@emergogroup.com
#### 3. Date Prepared
April 12, 2016
#### 4. Device Identification
| Trade/Proprietary Name: | Mineral Oil |
|-------------------------|-----------------------------------------|
| Common/Usual Name: | Mineral Oil |
| Classification Name: | Reproductive media and supplements |
| Regulation Number: | 21 CFR 884.6180 |
| Product Code: | MQL, Reproductive media and supplements |
| Device Class: | Class II |
| Classification Panel: | Obstetrics/Gynecology |
#### 5. Legally Marketed Predicate Device(s)
Cook, Sydney IVF Culture Oil, K022002
#### 6. Device Description
{4}------------------------------------------------
Mineral Oil is used to cover medium during embryo, oocyte, and sperm culture in assisted reproduction technology (ART) and micro-manipulation procedures. Mineral Oil is recommended for use as an overlay for a small volume of medium to prevent evaporation stable osmolality and pH.
The material composition:
High Purity Paraffin Oil comes in two density types:
- Light type has a ratio between 0.8200 to 0.8400 g/mL (15°C) and a viscosity which is between 8.850 to 11.70mm²/s (37.78°C)
- Heavy type has a ratio between 0.8500 to 0.8700 g/mL (15°C) and a viscosity which is between 41.90 to 44.10mm²/s (37.78°C).
There are two types of products, Mineral Oil - Light and Mineral Oil - Heavy. Two different unit sizes (100mL and 50mL) are available for each type. Mineral oils are colorless, tasteless, clear oil fluids which do not produce fluorescence.
Both Mineral Oils (Heavy and Light) do not contact the ova or embryo during culture. Mineral Oil is filter sterilized and dispensed into sterile light-resistant glass bottles.
#### 7. Indication for Use Statement
Mineral Oil is used as an overlay for culture of embryos, oocytes, and sperm in assisted reproduction technology (ART) and micromanipulation procedures.
#### 8. Substantial Equivalence Discussion
The following Table 1 compares the Mineral Oil to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Kitazato BioPharma Co., Ltd. | Cook | Significant<br>Differences |
|----------------------|---------------------------------------|---------------------------------------|----------------------------|
| Trade Name | Mineral Oil | Sydney IVF Culture Oil | - |
| 510(k)<br>Number | K160142 | K022002 | - |
| Product Code | MQL | MQL | Same |
| Regulation<br>Number | 884.6180 | 884.6180 | |
| Regulation<br>Name | Reproductive media and<br>supplements | Reproductive media and<br>supplements | Same |
| | | | Table 1 - Comparison of Characteristics |
|--|--|--|-----------------------------------------|
| | | | |
{5}------------------------------------------------
| Clinical | | | |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Indications for<br>Use | Mineral Oil is used as an overlay<br>for culture of embryos, oocytes,<br>and sperm in assisted<br>reproduction technology (ART)<br>and micromanipulation<br>procedures. | The Sydney IVF Culture Oil is<br>intended for use as an oil<br>overlay for culture of gametes,<br>zygotes, or embryos in assisted<br>reproduction technology (ART)<br>and micromanupulation<br>procedures. | Similar |
| Patient<br>contact | This product does not contact<br>directly with gametes/embryos<br>as it is used as an overlay for<br>culturing in culture media. | This product does not<br>contact directly with<br>gametes/embryos as it is<br>used as an overlay for<br>culturing in culture media. | Same |
| Used in similar<br>population | The product is used for fertility<br>treatment of humans | The product is used for<br>fertility treatment of humans | Same |
| Technical | | | |
| MEA | ≥ 80%<br>(96h Blastocysts) | ≥ 80% | Same |
| Endotoxin | ≤ 0.25 EU/mL<br>(USP) | < 0.4 EU/mL | Similar |
| Sterilization<br>method | Aseptic filtration sterilization | Aseptic filtration sterilization | Same |
| Storage<br>method | 8-30 °C<br>Dark storage<br>Do not freeze | 2-8 °C<br>Do not freeze | Similar |
| Shelf life | 12 months | 8 weeks | Similar |
| Content | 100 mL<br>50 mL | 200 mL<br>50 mL | Similar |
| Composition | Mineral Oil | Mineral Oil | Same |
The technological characteristics of Mineral Oil are comparable to the predicate device. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
#### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of Mineral Oil and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, KITAZATO BIOPHARMA completed a number of non-clinical performance tests as described below:
{6}------------------------------------------------
- Appearance (color): clear, particle free -
- -Sterility: No microbial growth from sterility testing per USP <71>
- -Endotoxin Testing: Endotoxin values conform to the value < 0.25 EU/mL
- -MEA: >80% of 1-cell control embryos develop at 96 hours
- -Peroxide value (POV): < 0.1 meq/kg
- -Readily carbonizable substance test: no discoloration, color of sulfuric acid layer paler than control per USP <271>
- -Shelf-life testing
The Mineral Oil meets all the requirements for overall design, and sterilization results confirming that the design output meets the design inputs and specifications for the device.
#### 10. Statement of Substantial Equivalence
The results of the testing described above provide a reasonable assurance that Mineral Oil is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.