K160139 · Freerider Corporation · ITI · Mar 20, 2017 · Physical Medicine
Device Facts
Record ID
K160139
Device Name
Luggie Chair
Applicant
Freerider Corporation
Product Code
ITI · Physical Medicine
Decision Date
Mar 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The FR-W04 (Luggie Chair) provides transportation for an elderly or disabled person. It can be used in a variety of indoor and outdoor settings.
Device Story
The FR-W04 (Luggie Chair) is a battery-powered, four-wheeled electric wheelchair for mobility-impaired individuals. It supports a single rider up to 360 lbs. The device features rear-wheel drive, regenerative electromechanical braking, and an adjustable seat. The user operates the chair via a controller mounted on the armrest, using a lever and speed knobs to manage direction and velocity. It is intended for use in indoor and outdoor environments. The device provides mobility, potentially increasing independence for the user. It is a non-clinical, bench-tested powered mobility aid.
Clinical Evidence
No clinical data. Bench testing only, including electromagnetic interference (IEC 6100-4-2/3/8, EN 55011/55022), ISO 7176 series (multiple sections), ISO 10993-1/5/10 (biocompatibility), and IEC 60601-1 (electrical safety).
Technological Characteristics
Battery-powered (24V, 10.5Ah Lithium) four-wheeled electric wheelchair. Features: rear-wheel drive, regenerative electromechanical brakes, Dynamic LiNX 40A controller, foldable seat/footrest. Dimensions: 37.44" L x 23.86" W. Max weight capacity: 360 lbs. Standards: ISO 7176, IEC 60601-1, ISO 10993, IEC 6100-4 series. Off-board AC-DC charger (100V-240V).
Indications for Use
Indicated for elderly or disabled persons requiring mobility assistance for transportation in indoor and outdoor settings. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human figures in profile, one behind the other. The figures are connected at the base and appear to be moving forward. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
March 20, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Freerider Corporation Michael Chen R&D Department and Assistant Vice President No. 22 Bengong 5th Road Kang-Shan District, TW
Re: K160139
Trade/Device Name: Luggie Chair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: March 7, 2017 Received: March 17, 2017
Dear Mr. Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Hoffmann -S
Carlos L. Peña. PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160139
Device Name Luggie Chair
Indications for Use (Describe)
The FR-W04 (Luggie Chair) provides transportation for an elderly or disabled person. It can be used in a variety of indoor and outdoor settings.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K160139
## 1. Contact Details
Applicant Name: Freerider Corporation
No. 22, Bengong 5th Road Kang-Shan Dist, Kaohsiung 820, Taiwan
| Contact Name: | Michael Chen |
|---------------|---------------------------------------|
| | Freerider Corporation |
| | No. 22, Bengong 5th Road |
| | Kang-Shan Dist, Kaohsiung 820, Taiwan |
| | Phone: 886-7-6223093 |
| | Fax: 886-7-6230373 |
Date Prepared: Nov 03, 2015
### 2. Device Name
Trade Name: Luggie Chair Common Name: Powered Wheelchair Classification Code: ITI Regulation Number: 21 CFR 890.3860
| 510(k) Number | Product Code | Trade Name | Manufacturer |
|---------------|--------------|-----------------------------------|-----------------|
| K033370 | ITI | Freerider FR 168-W<br>Power chair | Freerider Corp. |
| K151944 | INI | Freerider FR-L05<br>Luggie Super | Freerider Corp. |
#### 3. Legally Marketed Predicate Device(s)
### 4. Device Description
The FR-W04 (Luggie Chair) is a battery-powered, four-wheeled electric wheelchair intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings. The FR-W04 (Luggie Chair) is meant to be used by a single rider weighing up to 360 pounds. The wheelchair is rear-wheel drive and has electric, regenerative electromechanical brakes. The steering and user controls are provided on the armrest for ease of use by the rider. Steering is
{4}------------------------------------------------
controlled simply by turning the top controller in the desired direction. There is one lever, and speed knobs on the top controller to control movement speed of the wheel chair. The specification of control method is the same as the predicate device, Freerider FR168-W Power Chair (K033370).
The FR-W04 (Luggie Chair) has a controller and one lithium battery. There is also an off-board battery charger, which has also been previously cleared. It has an adjustable seat that has several height adjustments. The specification of battery charger and adjustable seat is the same as the predicate device, Freerider Luggie Super FR-L05 (K151944).
### 5. Intended Use/Indications for use
The FR-W04 (Luggie Chair) provides transportation for an elderly or disabled person. It can be used in a variety of indoor and outdoor settings.
| | K160139 (subject device) | K033370 (predicate) | K151944 (predicate) |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General Device<br>Characteristics | | | |
| Manufacturer | Freerider | Freerider | Freerider |
| Model | FR-W04 (Luggie<br>Chair) | FR168W | Luggie Super |
| Indications for Use | The FR-W04<br>(Luggie Chair)<br>provides<br>transportation for an<br>elderly or disabled<br>person. It can be<br>used in a variety of<br>indoor and outdoor<br>settings. | The fr168w<br>powered<br>wheelchair<br>provides<br>transportation<br>for an elderly<br>or disabled<br>person. It can<br>be used in a<br>variety of<br>indoor and<br>outdoor<br>settings. | The FR-L05<br>(Luggie Super)<br>provides<br>transportation<br>for an elderly or<br>disabled person.<br>It can be used in<br>a variety of<br>indoor and<br>outdoor settings. |
| Framework | Folding seat and | | Folds for |
| | footrest | | transport |
| | Adjustable height | | |
| Overall Dimensions | | | |
| Length (in) | 37.44 | 32 | 39 |
| Width (in) | 23.86 | 22 | 20.7 |
| Height (in) | 30.11-33.11 | 40 (46 with<br>headrest) | 35.7 |
| Seat Dimensions | | | |
| Width (in) | 17 | | 16.9 |
| Depth (in) | 15.43 | | 13.2 |
| Height (in) | 8.98 | | 17.3 |
| Folded Dimensions (if<br>applicable) | | | |
| Length (in) | 31.73 | | 27.76 |
| Width (in) | 23.86 | | |
| Height (in) | 17 (w/o armrest) | | |
| Wheelchair Weight | | | |
| With batteries | 76.56 lbs. | 84.5 lbs. | 67.1 lbs. |
| Without batteries | 71.72 lbs. | 84.5 lbs. | 62.26 lbs. |
| Controller | Dynamic 40A | P&G VSI<br>Controller | Rhino R-series<br>50 |
| Motor (include output) | 2 | 2 | 1 |
| | 24V | 0.339hp | |
| Batteries | | | |
| Quantity | 1 | 2 | 1 |
| Type | Lithium<br>24V, 10.5Ah | 12V, U-1<br>sealed | Lithium<br>24V 10.5 Ah |
| Range per Charge | 8.19 mi (198 lbs.)<br>6.31 miles (360 lbs.) | 25 mi | 10.5 mi (198 lbs)<br>7.25 miles (360 lbs) |
| Charger | Off board<br>Switching Mode AC-DC<br>100V~240V 50/60Hz<br>Output :5A | Onboard standard, off board optional<br>4A<br>110V | Off board<br>Switching Mode AC-DC<br>100V~240V 50/60Hz<br>Output :5A |
| Actuator | Dynamic LiNX 40A module | | |
| Brake | Electronic regenerative, electromechanical | Electronic regenerative | Electronic regenerative, electromechanic al |
| Minimum braking distance | | | |
| Forward (in) | 73.14 | 44.8 | 75.9 |
| Reverse (in) | 30.87 | | 47.2 |
| Max speed | | | |
| Forward | 4 mph | 4 mph | 4 mph |
| Reverse | 2.4 mph | 3.8 mph | 2.4 mph |
| Rear Wheels (in) | 2 x 8 | 10 | 2 x 7 |
| Castors (in) | 2 x 7 | 8 | 2 x 6 |
| Anti-tip wheels | 2.55 in | 6 in | Yes |
| Max Weight Capacity | 360 lbs. | 300 lbs. | 360 lbs. |
| Curb Climbing ability | 0.98 in | | |
| Ground clearance | 1.22 in | | 2.13 in |
| Minimum Turning Radius | 30.7 in | 26 in | 41.3 in |
| Maximum Incline (°) | 6 | 5 | 6 |
| Footplates | Foldable | | |
| Armrest Type | Fold up | | |
| Warranty | Provided | | |
#### 6. Substantial Equivalence Comparison
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
The control method of the FR-W04 (Luggie chair) is substantially equivalent to Freerider FR168-W (K033370).
The device features of the FR-W04 (Luggie Chair) and the FR168-W (K033370) are similar. Both are electric wheelchairs that are battery operated and have automatic breaking systems. Operating mode and interface are very similar, the control lever is mounted on an armrest, and directional control is the same as FR168-W (K033370). The differences are as follows: The FR-W04 has an adjustable seat that has several height adjustments. The front and rear wheels are 7" and 8" respectively, smaller than that of the FR168-W (K033370).
The battery with charger and adjustable seat features on the FR-W04 (Luggie Chair) is substantially equivalent to that of the FR-L05 Luggie Super (K151944). The device features of the FR-W04 (Luggie Chair) and the FR-L05 Luggie Super (K151944) are similar. Both are battery operated and have automatic breaking systems. The charger and adjustable seat is the same between the two devices. The differences are as follows: The control lever of the FR-W04 is mounted on the armrest. The FR-W04 does not have a front handle bar and front chassis.
#### 7. Non-clinical Testing
Electromagnetic interference testing was conducted to IEC 6100-4-2, IEC 6100-4-3, IEC 6100-4-8, EN 12184:2009, EN 55011:2010, and EN55022. ISO 7176 testing to multiple sections; Section 1, Section 2, Section 4, Section 5, Section 6, Section 6, Section 7, Section 8, Section 9, Section 10, Section 11, Section 14, Section 14, Section 15, Section 16, and Section 21 was conducted, as well as ISO 10993-1, ISO 10993-5, ISO 10993-10 and IEC60601-1. Additional bench testing related to ground current leakage and summary matrix testing was also conducted. The FR-W04 (Luggie Chair) passed all testing.
## 8. Clinical Testing
No clinical testing is included in this submission.
### 9. Conclusion
The safety and effectiveness of the FR-W04 (Luggie Chair) was demonstrated by the
{8}------------------------------------------------
testing in compliance with national and international standards. The intended use, basic technology, and many features of the FR-W04 (Luggie Chair) are similar to the predicate device. No new issues of safety and effectiveness are raised by the differences between the FR-W04 (subject device), FR168-W (K033370), and FR-L05 Luggie Super (K151944). Therefore, the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.