← Product Code [KGN](/productcode/KGN) · K160136

# Flowable Wound Matrix (K160136)

_Cook Biotech Incorporated · KGN · Sep 28, 2016 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160136

## Device Facts

- **Applicant:** Cook Biotech Incorporated
- **Product Code:** [KGN](/productcode/KGN.md)
- **Decision Date:** Sep 28, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

The Flowable Wound Matrix is intended for the management of wounds including: partial and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, 2nd degree burns, skin tears) and draining wounds.

## Device Story

Flowable Wound Matrix is a wound management device composed of particulate porcine Small Intestinal Submucosa (SIS) and fructose carrier. Supplied dry in a syringe; rehydrated with saline at point-of-care to form a gel-like consistency. Applied topically by clinician to wound site. Acts as a scaffold for cellular invasion and capillary growth; maintains supportive environment for wound healing. Intended for various wound types including ulcers, surgical, and trauma wounds. Benefits patient by facilitating tissue repair in complex or difficult-to-reach wound geometries.

## Clinical Evidence

Bench testing only. Evidence includes biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity), rehydration and deployment performance, package integrity, shelf life, and collagen characterization (Western Blot and Digital Scanning Calorimetry).

## Technological Characteristics

Particulate porcine Small Intestinal Submucosa (SIS) with fructose carrier. Dimensions: particles ≤1000 µm, 1.25cc volume. Rehydrated with saline via dual-syringe/luer connector system. Sterilized via E-beam. Single-use.

## Predicate Devices

- Oasis Wound Matrix ([K061711](/device/K061711.md))

## Reference Devices

- Integra Flowable Wound Matrix ([K072113](/device/K072113.md))
- Cook ECM Powder ([K152033](/device/K152033.md))
- MicroMatrix ([K153754](/device/K153754.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 28, 2016

Cook Biotech Incorporated Nick Wang, Ph.D. Regulatory Affairs Scientist 1425 Innovation Place West Lafayette, Indiana 47906

Re: K160136

Trade/Device Name: Flowable Wound Matrix Regulatory Class: Unclassified Product Code: KGN Dated: September 2, 2016 Received: September 6, 2016

Dear Dr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160136

Device Name Flowable Wound Matrix

Indications for Use (Describe) The Flowable Wound Matrix is intended for the management of wounds including:

- partial and full-thickness wounds,

- pressure ulcers,
- venous ulcers,
- diabetic ulcers,
- chronic vascular ulcers,
- tunneled/undermined wounds.
- surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence),
- trauma wounds (abrasions, lacerations, 2nd degree burns, skin tears),

- draining wounds.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

| Submitted By: | Perry W. Guinn<br>Vice President of Quality Assurance & Regulatory Affairs<br>Cook Biotech Incorporated<br>1425 Innovation Place<br>West Lafayette, IN 47906<br>(765) 497-3355<br>23 September 2016 |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

#### Name of Device:

| Trade/Proprietary Name: | Flowable Wound Matrix                                         |
|-------------------------|---------------------------------------------------------------|
| 510(k) Number:          | K160136                                                       |
| Common/Usual Name:      | Dressing, Wound, Collagen                                     |
| Classification Name:    | Dressing, Wound, Collagen                                     |
| Product Code:           | KGN                                                           |
| Device Class:           | The product code KGN is considered by FDA to be unclassified. |

#### Predicate Device:

Oasis Wound Matrix (K061711)

#### Reference Devices:

Integra Flowable Wound Matrix (K072113) Cook ECM Powder (K152033) MicroMatrix (K153754)

#### Intended Use:

The Flowable Wound Matrix is intended for the management of wounds including: partial and full-thickness wounds, pressure ulcers, venous ulcers, diabetic ulcers, chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence), trauma wounds (abrasions, lacerations, 2nd degree burns, skin tears) and draining wounds.

#### Device Description:

The Flowable Wound Matrix is a wound management device consisting of particulate Porcine Small Intestinal Submucosa (SIS) and fructose, a natural carrier. The device is an addition to the family of SIS-based wound management devices (Oasis Wound Matrix (K061711) and Cook ECM Powder (K152033)) already manufactured by Cook Biotech

{4}------------------------------------------------

Incorporated. The device is supplied dry, rehydrated with saline at the time of application, and delivered topically to the wound through a pre-supplied syringe. SIS, which composes the majority of the device, is the same base material as that of the predicate device Oasis Wound Matrix (K061711) and reference device Cook ECM Powder (K152033). In addition to SIS, the device also contains fructose, a carrier added only to facilitate the preparation and delivery of the device. Fructose is a sugar naturally found in the body and is readily metabolized. The Flowable Wound Matrix is meant to be employed by the user to manage wounds of the types outlined in the intended use of the device. The device achieves its intended use by providing a scaffold for cellular invasion and capillary growth, and maintaining a supportive environment for wound management.

### Comparison to Predicate Device:

The subject and predicate devices share the same fundamental technology; the main components of both devices are porcine SIS. The main difference between the subject and predicate is the device configuration and method of application. The Oasis Wound Matrix is supplied as a dry sheet of SIS and is applied topically by rehydrating and placing the sheet directly onto a wound. The Flowable Wound Matrix, in contrast, is particulate SIS supplied dry in a cylindrical shape. Upon rehydration with saline, the device acquires a gel-like consistency and can be applied directly to the wound site using the supplied syringe.

## Summary of Non-Clinical Tests:

The following testing was performed to demonstrate substantial equivalence:

- Biocompatibility Testing:
	- o Cytotoxicity
	- o Sensitization
	- Irritation/Intracutaneous Reactivity O
	- O Acute Systemic Toxicity
- Other Testing: ●
	- Rehydration and Deployment o
	- o Package Integrity
	- Shelf Life o
	- Collagen Characterization (Western Blot and Digital Scanning o Calorimetry)

{5}------------------------------------------------

### Substantial Equivalence:

Table 5-1 provides a comparison of the subject and predicate devices.

To provide further evidence of substantial equivalence, CBI used the 510(k) Decision-Making Flowchart from the FDA guidance document Evaluating Substantial Equivalence in Premarket Notifications [510(k)] (July 28, 2014) to compare and assess the intended use and the technological characteristics of the subject and predicate devices. Specifically, the intended uses of both devices are identical. In terms of technological characteristics, while the subject device differs from the predicate device in form (sheet vs. flowable), both devices share the same underlying technology, namely topical application of porcine SIS for wound management. In this 510(k), CBI submits biocompatibility, deployment, package and shelf-life testing as evidence to demonstrate the device can function as intended and performs comparably to the predicate device that is currently marketed for the same intended use. In conclusion, CBI believes the Flowable Wound Matrix is substantially equivalent to the Oasis Wound Matrix based on success in non-clinical testing and the identical nature of the fundamental technology and intended use.

{6}------------------------------------------------

|                             | Flowable Wound<br>Matrix (Subject<br>Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Oasis® Wound Matrix<br>(Predicate Device)                                                                                                                                                                                                                                                                                                                                                                                                                                          | Integra™ Flowable<br>Wound Matrix<br>(Reference Device)                                                                                                                                                                                                                                                                                                                                                                                                                                          | Cook ECM Powder<br>(Reference Device)                                                                                                                                                                                                                                                                                                                                                                                                                                           | MicroMatrix®<br>(Reference Device)                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                            | Flowable Wound<br>Matrix (Subject<br>Device)                                                                                                    | Oasis® Wound Matrix<br>(Predicate Device)    | Integra™ Flowable<br>Wound Matrix<br>(Reference Device)                                                                                         | Cook ECM Powder<br>(Reference Device) | MicroMatrix<br>(Reference Device)                                      |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|------------------------------------------------------------------------|
| 510(k)<br>number            | K160136                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | K061711                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | K072113                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | K152033                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K153754                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Dimensions                 | Particles of ≤1000 µm<br>in diameter<br>Volume of device is<br>1.25cc                                                                           | 2 cm x 2 cm to 20 cm<br>by 40 cm             | Collagen Particles of<br>200 – 2000 um in<br>diameter*<br>Volume of device is<br>3cc                                                            | Particles of ≤1000 µm<br>in diameter  | Two particle<br>distributions, particles<br>of <500 um and <1000<br>um |
| Indication<br>for Use       | The Flowable Wound<br>Matrix is intended for<br>the management of<br>wounds including:<br>partial and full-<br>thickness wounds,<br>pressure ulcers, venous<br>ulcers, diabetic ulcers,<br>chronic vascular ulcers,<br>tunneled/undermined<br>wounds, surgical<br>wounds (donor<br>sites/grafts, post-Moh's<br>surgery, post-laser<br>surgery, podiatric,<br>wound dehiscence),<br>trauma wounds<br>(abrasions, lacerations,<br>second-degree burns,<br>and skin tears),<br>draining wounds. | Oasis® Wound Matrix<br>is intended for the<br>management of wounds<br>including: partial and<br>full-thickness wounds,<br>pressure ulcers, venous<br>ulcers, diabetic ulcers,<br>chronic vascular ulcers,<br>tunneled/undermined<br>wounds, surgical<br>wounds (donor<br>sites/grafts, post-Moh's<br>surgery, post-laser<br>surgery, podiatric,<br>wound dehiscence),<br>trauma wounds<br>(abrasions, lacerations,<br>second-degree burns,<br>and skin tears),<br>draining wounds. | Integra™ Flowable<br>Wound Matrix is<br>intended for the<br>management of wounds<br>including: partial and<br>full-thickness wounds,<br>pressure ulcers, venous<br>ulcers, diabetic ulcers,<br>chronic vascular ulcers,<br>tunneled/undermined<br>wounds, surgical<br>wounds (donor<br>sites/grafts, post-Moh's<br>surgery, post-laser<br>surgery, podiatric,<br>wound dehiscence),<br>trauma wounds<br>(abrasions, lacerations,<br>second-degree burns,<br>and skin tears),<br>draining wounds. | Cook® ECM Powder is<br>intended for the<br>management of wounds<br>including: partial and<br>full-thickness wounds,<br>pressure ulcers, venous<br>ulcers, diabetic ulcers,<br>chronic vascular ulcers,<br>tunneled/undermined<br>wounds, surgical<br>wounds (donor<br>sites/grafts, post-Moh's<br>surgery, post-laser<br>surgery, podiatric,<br>wound dehiscence),<br>trauma wounds<br>(abrasions, lacerations,<br>second-degree burns,<br>and skin tears),<br>draining wounds. | MicroMatrix® is<br>intended for the<br>management of topical<br>wounds including:<br>partial and full-<br>thickness wounds,<br>pressure<br>ulcers, venous ulcers,<br>diabetic ulcers, chronic<br>vascular ulcers,<br>tunneled/undermined<br>wounds, surgical<br>wounds (donor sites/<br>grafts, post-Mohs<br>surgery, post-laser<br>surgery, podiatric,<br>wound dehiscence),<br>trauma wounds<br>(abrasions, lacerations,<br>second-degree burns,<br>and skin tears), and<br>draining wounds. | Technologic<br>al Features | Device is rehydrated to<br>a gel-like consistency<br>using saline in a dual<br>syringe/luer connector<br>system, and then<br>applied topically. | Sheet is rehydrated and<br>applied topically | Device is rehydrated to<br>a gel-like consistency<br>using saline in a dual<br>syringe/luer connector<br>system, and then<br>applied topically. | Powder, applied<br>directly topically | Particle, applied<br>directly topically                                |
| Collagen<br>Source          | Porcine small intestinal<br>submucosa                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Porcine small intestinal<br>submucosa                                                                                                                                                                                                                                                                                                                                                                                                                                              | Bovine tendon collagen                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Porcine small intestinal<br>submucosa                                                                                                                                                                                                                                                                                                                                                                                                                                           | Porcine urinary bladder<br>matrix                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Supplied<br>sterile?       | Yes                                                                                                                                             | Yes                                          | Yes                                                                                                                                             | Yes                                   | Yes                                                                    |
| Sterilization<br>method     | E-beam                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | E-beam                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                            |                                                                                                                                                 |                                              |                                                                                                                                                 |                                       |                                                                        |
| Intended for<br>single use? | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                            |                                                                                                                                                 |                                              |                                                                                                                                                 |                                       |                                                                        |

## Table 5-1. Substantial Equivalence Information

{7}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K160136](https://fda-staging.innolitics.com/device/K160136)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).