Dental Implant Abutment OKTAGON Bone Level RC, Dental Implant Abutment OKTAGON Bone Level NC
Applicant
Hager& Meisinger GmbH
Product Code
NHA · Dental
Decision Date
Nov 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Dental Implant Abutments Bone Level are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses. Abutments can be used in single tooth replacements and multiple tooth restorations. The Abutments are intended to be compatible to OKTAGON® Bone Level implants with diameters 3.3mm, 4.1mm and 4.8mm and with the lengths 8mm, 10mm, 12mm and 14mm. The Dental Implant System OKTAGON® Bone Level is intended for delayed loading, or for immediate loading when good primary stability is achieved and with appropriate occlusal load.
Device Story
Endosseous dental implant abutments designed to support prosthetic reconstructions (crowns, bridges, overdentures) in partially or fully edentulous patients. Device consists of angled abutments and alligator abutments; connects to OKTAGON® Bone Level implants via internal octagon/nut or conical connection secured by a screw. Used by dental professionals in clinical settings. Provides structural interface between endosseous implant and final prosthesis. Benefits include restoration of dental function and aesthetics. Mechanical stability achieved through titanium construction and screw-fixation.
Clinical Evidence
Bench testing only. Performance demonstrated via fatigue and breakage testing per ISO 14801. Biocompatibility evaluated per ISO 10993-1, ISO 7405, and ISO 10993-5. No clinical data presented.
Technological Characteristics
Materials: Titanium Grade 4 (abutments), Titanium Alloy TiAl6V4 (alligator abutments/screws). Standards: ASTM F136, ASTM F67, ISO 5832-2, ISO 5832-3. Connection: Internal octagon/nut or conical, screw-retained. Sterilization: Validated per ISO 17665-1/2 and ANSI/AAMI ST79. Mechanical testing: ISO 14801 fatigue testing.
Indications for Use
Indicated for partially or fully edentulous patients requiring prosthetic support via dental implants. Compatible with OKTAGON® Bone Level implants (diameters 3.3mm, 4.1mm, 4.8mm; lengths 8-14mm). Suitable for single tooth, multiple tooth, or total prostheses. Contraindicated for patients where immediate loading is not supported by primary stability.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Dental Endosseous Implant System OKTAGON® Bone Level (K143539)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 21, 2016
Hager & Meisinger GmbH Dr. Melanie May Regulatory Affairs Hansemannstrasse 10 Neuss. 41468 Germany
Re: K160132
Trade/Device Name: Endosseous Dental Implant Abutments OKTAGON® Bone Level Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: October 20, 2016 Received: October 20, 2016
Dear Dr. Melanie May:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K160132
#### Device Name
Endosseous Dental Implant Abutments OKTAGON® Bone Level
Indications for Use (Describe)
Indications for Use (Describe)
Dental Implant Abutments Bone Level are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.
Abutments can be used in single tooth replacements and multiple tooth restorations. The Abutments are intended to be compatible to OKTAGON® Bone Level implants with diameters 3.3mm, 4.1mm and with the lengths 8mm, 10mm, 12mm and 14mm.
The Dental Implant System OKTAGON® Bone Level is intended for delayed loading, or for immediate loading when good primary stability is achieved and with appropriate occlusal load.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
### 1. Applicant's Name and Address
Hager & Meisinger GmbH Hansemannstrasse 10 41468 Neuss, Germany Phone: (0049) 2131 2012-0 Fax: (0049) 2131 2012- 223 Contact Person: Dr. Melanie May Product approval and product validation (Regulatory Affairs)
### 2. Date prepared
Date prepared:
November 17th, 2016
Endosseous Dental Implant Abutments
#### 3. Name of the device
Trade Name:
| | OKTAGON® Bone Level |
|----------------------|-------------------------------------|
| Common Name: | Endosseous dental implant abutments |
| Classification Name: | Endosseous dental implant abutment |
| Product Code: | NHA |
| Regulation No: | 872.3630 |
| Class: | II |
| Panel: | Dental |
### 4. Predicate Devices:
| Predicate<br>Device | 510(k)<br>No. | Manufacturer | Trade Name |
|------------------------|---------------|--------------|------------------------------------------------------------|
| Primary<br>Predicate | K143539 | Meisinger | Dental Endosseous Implant<br>System OKTAGON® Bone<br>Level |
| Reference<br>Predicate | K132214 | Meisinger | Dental Abutment<br>OKTAGON® |
# 5. Device Description:
{4}------------------------------------------------
## 510(k) Summary
The Dental Implant System OKTAGON® Bone Level is an integrated system of endosseous dental implants which are designed to support prosthetic devices for partially of fully edentulous patients.
The devices covered in this submission are angled abutments and an addition to the currently available Dental Implants and Abutments OKTAGON® Bone Level.
The abutments are made of Titanium Grade 4, the alligator abutments and screws are made of Titanium Alloy (TiAl6V4); the connection to the implants is achieved by an internal octagon/nut construction and a metric thread.
The abutments are available in the following variations including the corresponding screws:
| Article<br>No. | Variation | Description | Material | | Connection to<br>Implant | Intended<br>Use |
|----------------|-----------|-----------------------------------------------------------------------------------------------|--------------------------------|--------------------------------|------------------------------------|---------------------------|
| | | | Abutment | Screw | | |
| 22025 | NC | Abutment, Angled, 18°,<br>Ø3.5mm, L 6.0 mm,<br>H1.5mm, cementable, incl.<br>basic screw 22030 | Titanium<br>Grade 4 | Titanium<br>Alloy<br>(TiAl6V4) | Octagonal,<br>fixating by<br>screw | For crowns<br>and bridges |
| 22026 | NC | Abutment, Angled, 18°,<br>Ø3.5mm, L 6.0 mm,<br>H3.0mm, cementable, 22030 | Titanium<br>Grade 4 | Titanium<br>Alloy<br>(TiAl6V4) | Octagonal,<br>fixating by<br>screw | For crowns<br>and bridges |
| 22030 | NC | Basic screw for NC abutment | Titanium<br>Alloy<br>(TiAl6V4) | - | | For use with<br>abutments |
| 22041 | NC | Alligator Abutment Angled,<br>15°, H3.0mm | Titanium<br>Alloy<br>(TiAl6V4) | - | Conical,<br>fixating by<br>screw | Overdenture<br>Prosthetic |
| 22042 | NC | Alligator Abutment Angled,<br>15°, H4.0mm | Titanium<br>Alloy<br>(TiAl6V4) | - | Conical,<br>fixating by<br>screw | Overdenture<br>Prosthetic |
| 22103 | RC | Abutment, Angled, 18°,<br>Ø5.0mm, L 8.0 mm,<br>H1.5mm, cementable, | Titanium<br>Grade 4 | Titanium<br>Alloy<br>(TiAl6V4) | Octagonal,<br>fixating by<br>screw | For crowns<br>and bridges |
| 22105 | RC | Abutment, Angled, 18°,<br>Ø5.0mm, L 8.0 mm,<br>H3.0mm, cementable, incl.<br>basic screw 22122 | Titanium<br>Grade 4 | Titanium<br>Alloy<br>(TiAl6V4) | Octagonal,<br>fixating by<br>screw | For crowns<br>and bridges |
{5}------------------------------------------------
## 510(k) Summary
| Article<br>No. | Variation | Description | Material | | Connection to<br>Implant | Intended<br>Use |
|----------------|-----------|-------------------------------------------|--------------------------------|-------|----------------------------------|---------------------------|
| | | | Abutment | Screw | | |
| 22122 | RC | Basic screw for RC abutment | Titanium<br>Alloy<br>(TiAl6V4) | - | - | For use with<br>abutments |
| 22129 | RC | Alligator Abutment Angled,<br>15°, H3.0mm | Titanium<br>Alloy<br>(TiAl6V4) | - | Conical,<br>fixating by<br>screw | Overdenture<br>Prosthetic |
| 22130 | RC | Alligator Abutment Angled,<br>15° H4.0mm | Titanium<br>Alloy<br>(TiAl6V4) | - | Conical,<br>fixating by<br>screw | Overdenture<br>Prosthetic |
| 22131 | RC/NC | Basic Screw for Alligator | Titanium<br>Alloy<br>(TiAl6V4) | - | - | Overdenture<br>Prosthetic |
### 6. Indications for use:
Dental Implant Abutments Bone Level are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.
Abutments can be used in single tooth replacements and multiple tooth restorations. The Abutments are intended to be compatible to OKTAGON® Bone Level implants with diameters 3.3mm, 4.1mm and 4.8mm and with the lengths 8mm, 10mm, 12mm and 14mm.
The Dental Implant System OKTAGON® Bone Level is intended for delayed loading, or for immediate loading when good primary stability is achieved and with appropriate occlusal load.
### 7. Performance tests and used standards
The following standards have been followed for the development, production, performance and safety testing of Dental Implant Abutment OKTAGON® Bone Level:
- Fatigue Testing according to ISO 14801
- Evaluation of Biocompatibility according to ISO 10993-1, ISO 7405 .
- Cytotoxicity Testing according to ISO 10993-5 .
{6}------------------------------------------------
## 510(k) Summary
- . Sterilization validation according to ISO 17665-1, ISO 17665-2 and ANSI/ AAMI ST 79
- Material Specification according to ASTM F136, ASTM F67, ISO 5832-. 2, ISO 5832-3
- Risk Management according to ISO 14971 ●
In order to demonstrate substantial equivalence performance tests (breakage and fatigue tests) have been conducted, fulfilling the requirements of ISO 14801 and the FDA's "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments".
The biocompatibility was evaluated according to ISO 10993-1. The evaluation shows, that the subject device is comparable in intended use, design and materials to the legally market predicate devices. There are no known negative biological effects of dental implants made of Titanium Grade 4. Thus the requirements of ISO 10993-1 are fulfilled.
{7}------------------------------------------------
## 510(k) Summary
### 8. Basis for substantial equivalence
The intended use for Dental Implant Abutment OKTAGON® Bone Level is identical to the named predicated devices.
| Manufacturer | Hager & Meisinger GmbH | Hager & Meisinger GmbH | Hager & Meisinger GmbH |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject device K160132 | Primary Predicate K143539 | Reference Predicate K132214 |
| Indications for Use | Dental Implant Abutments Bone Level are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.<br>Abutments can be used in single tooth replacements and multiple tooth restorations. The Abutments are intended to be compatible to OKTAGON® Bone Level implants with diameters 3.3mm, 4.1mm and 4.8mm and with the lengths 8mm, 10mm, 12mm and 14mm.<br>The Dental Implant System OKTAGON® Bone Level is intended for delayed loading, or for immediate loading when good primary stability is achieved and with appropriate occlusal load. | The implants are surgically placed in the maxillary and/ or mandibular arches to provide support for prosthetic restorations in edentulous or partially edentulous patients. The parts are intended to be used with OKTAGON® Bone Level abutments and prosthetic parts.<br>Dental Implant Abutments Bone Level are intended to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.<br>Abutments can be used in single tooth replacements and multiple tooth restorations. The Abutments are intended to be compatible to OKTAGON® Bone Level implants with diameters 3.3mm, 4.1mm and 4.8mm and with the lengths 8mm, 10mm, 12mm and 14mm.<br>The Dental Implant OKTAGON® system Bone Level is intended for delayed loading, or for immediate loading when good primary stability is achieved and with appropriate occlusal load. | Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions. Prosthetic applications can include individual crowns, bridges, partial or total prostheses.<br>Abutments can be used in single tooth replacements and multiple tooth restorations.<br>The Dental Implant Abutments OKTAGON® are intended to be compatible to OKTAGON® implants (Dental Implant OKTAGON®) with diameters 3.3mm, 4.1mm and 4.8mm in the variation Regular Platform, Wide Platform and Tapered Design with the lengths 8mm, 10mm, 12mm and 14mm. |
| Description | The OKTAGON® Dental Implant System is an integrated system of endosseous dental implants which are designed to support prosthetic devices for partially of fully edentulous patients. The devices covered in this submission are abutments in different versions including the | The OKTAGON® Dental Implant System is an integrated system of endosseous dental implants which are designed to support prosthetic devices for partially of fully edentulous patients. The devices covered in this submission are abutments in different versions including the | The OKTAGON® Dental Implant System is an integrated system of endosseous dental implants which are designed to support prosthetic devices for partially of fully edentulous patients. The devices covered in this submission are abutments. |
{8}------------------------------------------------
## 510(k) Summary
| Manufacturer | Hager & Meisinger GmbH | Hager & Meisinger GmbH | Hager & Meisinger GmbH |
|----------------------------------|-------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject device K160132 | Primary Predicate K143539 | Reference Predicate K132214 |
| | corresponding screws. | corresponding screws. | |
| Diameter, Length (L), Height (H) | Variation NC: | Variation NC: | Abutments: |
| | Angled abutments: | Abutments: | Ø2.5-4.8, L= 1.5 - 7.0 mm, |
| | Ø3.5, L= 6 mm, H = 1.5 mm and 3.0 mm | Ø3.5, L= 6 mm, H = 1.0 mm, 2.0, 3.0 and 3.5 mm | Alligator Abutments: |
| | Alligator abutments: | Alligator abutments: | H = 1.0, 1.5, 2.0, 3.0 and 4.0 mm |
| | H = 3.0 mm and 4.0 mm | H = 2.0, 3.0, 4.0 and 5.0 mm | |
| | Variation RC: | Variation RC: | |
| | Angled abutments: | Abutments: | |
| | Ø5.0 mm, L= 8.5 mm, H = 1.5 mm and 3.0 mm | Ø5.0 and 6.5, L= 8.5 mm, H = 1.0, 2.0 and 3.0 mm | |
| | Alligator abutments: | Alligator abutments: | |
| | H = 3.0 mm and 4.0 mm | H = 1.0, 2.0, 3.0, 4.0 and 5.0 mm | |
| Angulation | 18° for Abutments, 15° for angled Alligator | 0° for Abutments, 15° for angled Alligator | Max. 20° |
| Material | Titanium Grade 4 (abutments), Titanium Alloy (alligator abutments and screws) | Titanium Grade 4 (abutments), Titanium Alloy (alligator abutments and screws), synthetic material (plastic copings made of POM) | Titanium Grade 4 (abutments), Titanium Alloy (alligator abutments and screws), synthetic material (plastic copings made of POM) |
The intended use for Abutments is identical to the named predicated devices. The abutments have the same indications for use, material composition and the connection to implants is equivalent. In addition the principal design including measurements of abutments is identical to the previously cleared predicated devices.
Based on these observations, we conclude that the Dental Implant Abutment OKTAGON® Bone Level is substantially equivalent to the legally marketed predicate devices described.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.